A Systematic Literature Review of the Factors that Influence the Accuracy of Consumer Wearable Health Device Data
Open accessThe claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Red LEDs are more commonly used in medical settings, are not absorbed as well by the skin allowing light to transmit deeper and detect multiple biological measures, but have a lower signal-to-noise ratio and are more susceptible to motion artifacts.
“In contrast, red LEDs, which are more commonly used in medical settings, are not absorbed as well by the skin allowing the light to transmit deeper and allowing for detection of multiple biological measures, but it has a lower signal-to-noise ratio and is more susceptible to motion artifacts23.”
From Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research (2020) · cites this paper as [23]
Motion artifacts and level of physical activity are two related factors identified as influencing wearable heart rate accuracy.
“two related factors identified as influencing wearable HR accuracy are motion artifacts23 and level of physical activity46.”
From Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research (2020) · cites this paper as [23]
Wearables that use red light will likely be more effective on darker skin tones because red light is not absorbed by melanin, which may allow for more accurate heart rate measurements.
“wearables that use red light will likely be more effective on darker skin tones as it is not absorbed by melanin, which may allow for more accurate HR measurements23.”
From Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research (2020) · cites this paper as [23]
Wearables that use green PPG or a combination of green and red PPG are likely to perform better during movement and physical activity.
“indicating that wearables that use green PPG or a combination of green and red PPG are likely to perform better during movement and physical activity23.”
From Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research (2020) · cites this paper as [23]
Wearable manufacturers use proprietary algorithms to translate PPG signals to heart rate measurements, which may influence accuracy.
“Wearable manufacturers use proprietary algorithms to translate PPG signals to HR measurements, which may influence accuracy23.”
From Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research (2020) · cites this paper as [23]
PPG green light is less susceptible to motion artifacts, whereas PPG red light is more susceptible to motion artifacts.
“one strength of PPG green light is that it is less susceptible to motion artifacts, whereas PPG red light is more susceptible to motion artifacts”
From Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research (2020) · cites this paper as [23]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967 · see citing works
Migueles, J. H., Cadenas‐Sánchez, C., Ekelund, U., Nyström, C. D., Mora-González, J., Löf, M., Labayen, I., Ruiz, J. R., & Ortega, F. B. (2017) · Sports Medicine · doi:10.1007/s40279-017-0716-0 · Cited by 1,879 · see citing works
Wearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912 · see citing works
Haghi, M., Thurow, K., & Stoll, R. (2017) · Healthcare Informatics Research · doi:10.4258/hir.2017.23.1.4 · Cited by 712 · see citing works
Zero-Velocity Detection—An Algorithm Evaluation
Skog, I., Händel, P., Nilsson, J.-O., & Rantakokko, J. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2060723 · Cited by 666 · see citing works
Shcherbina, A., Mattsson, C., Waggott, D., Salisbury, H., Christle, J., Hastie, T., Wheeler, M., & Ashley, E. (2017) · Journal of Personalized Medicine · doi:10.3390/jpm7020003 · Cited by 625 · see citing works
The Current State of Physical Activity Assessment Tools
Ainsworth, B. E., Cahalin, L. P., Buman, M. P., & Ross, R. (2014) · Progress in Cardiovascular Diseases · doi:10.1016/j.pcad.2014.10.005 · Cited by 441 · see citing works
Kakria, P., Tripathi, N. K., & Kitipawang, P. (2015) · International Journal of Telemedicine and Applications · doi:10.1155/2015/373474 · Cited by 410 · see citing works
Assessing Physical Activity Using Wearable Monitors: Measures of Physical Activity
BUTTE, N. F., EKELUND, U., & WESTERTERP, K. R. (2012) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0b013e3182399c0e · Cited by 359 · see citing works
Accuracy of Heart Rate Watches: Implications for Weight Management
Wallen, M. P., Gomersall, S. R., Keating, S. E., Wisløff, U., & Coombes, J. S. (2016) · PLOS ONE · doi:10.1371/journal.pone.0154420 · Cited by 336 · see citing works
Xie, J., Wen, D., Liang, L., Jia, Y., Gao, L., & Lei, J. (2018) · JMIR mHealth and uHealth · doi:10.2196/mhealth.9754 · Cited by 310 · see citing works
Nelson, B. W., & Allen, N. B. (2019) · JMIR mHealth and uHealth · doi:10.2196/10828 · Cited by 310 · see citing works
Behavior Change Techniques Present in Wearable Activity Trackers: A Critical Analysis
Mercer, K., Li, M., Giangregorio, L., Burns, C., & Grindrod, K. (2016) · JMIR mHealth and uHealth · doi:10.2196/mhealth.4461 · Cited by 304 · see citing works
Twenty-four Hours of Sleep, Sedentary Behavior, and Physical Activity with Nine Wearable Devices
Rosenberger, M. E., Buman, M. P., Haskell, W. L., McConnell, M. V., & Carstensen, L. L. (2015) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000000778 · Cited by 298 · see citing works
A Motion-Tolerant Adaptive Algorithm for Wearable Photoplethysmographic Biosensors
Yousefi, R., Nourani, M., Ostadabbas, S., & Panahi, I. M. S. (2013) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2013.2264358 · Cited by 229 · see citing works
Validity of Consumer-Based Physical Activity Monitors for Specific Activity Types
Nelson, M. B., Kaminsky, L. A., Dickin, D. C., & Montoye, A. H. K. (2016) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000000933 · Cited by 206 · see citing works
WORLDWIDE SURVEY OF FITNESS TRENDS FOR 2019
Thompson, W. R. (2018) · ACSMʼs Health & Fitness Journal · doi:10.1249/fit.0000000000000438 · Cited by 193 · see citing works
Influence of skin type and wavelength on light wave reflectance
Fallow, B. A., Tarumi, T., & Tanaka, H. (2013) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-013-9436-7 · Cited by 162 · see citing works
Alternatives to polysomnography (PSG): a validation of wrist actigraphy and a partial-PSG system
Kosmadopoulos, A., Sargent, C., Darwent, D., Zhou, X., & Roach, G. D. (2014) · Behavior Research Methods · doi:10.3758/s13428-013-0438-7 · Cited by 160 · see citing works
Straiton, N., Alharbi, M., Bauman, A., Neubeck, L., Gullick, J., Bhindi, R., & Gallagher, R. (2018) · Maturitas · doi:10.1016/j.maturitas.2018.03.016 · Cited by 147 · see citing works
Chowdhury, S. S., Hyder, R., Hafiz, M. S. B., & Haque, M. A. (2016) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2016.2632201 · Cited by 102 · see citing works
Smart technologies toward sleep monitoring at home
Park, K. S., & Choi, S. H. (2019) · Biomedical Engineering Letters · doi:10.1007/s13534-018-0091-2 · Cited by 95 · see citing works
A Real-time PPG Quality Assessment Approach for Healthcare Internet-of-Things
Naeini, E. K., Azimi, I., Rahmani, A. M., Liljeberg, P., & Dutt, N. D. (2019) · Procedia Computer Science · doi:10.1016/j.procs.2019.04.074 · Cited by 90 · see citing works
Promises and Challenges in the Use of Consumer-Grade Devices for Sleep Monitoring
Roomkham, S., Lovell, D. R., Cheung, J., & Perrin, D. (2018) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2018.2811735 · Cited by 84 · see citing works
Simpson, L. A., Eng, J. J., Klassen, T. D., Lim, S. B., Louie, D. R., Parappilly, B., Sakakibara, B. M., & Zbogar, D. (2015) · Journal of Rehabilitation Medicine · doi:10.2340/16501977-1993 · Cited by 79 · see citing works
Smart Attacks against Intelligent Wearables in People-Centric Internet of Things
Liu, J., & Sun, W. Y. (2016) · IEEE Communications Magazine · doi:10.1109/mcom.2016.1600553cm · Cited by 78 · see citing works
Healthcare algorithms by wearable inertial sensors: a survey
Ao, B., Gaoli, F., Wang, Y., et al. (2015) · China Communications · doi:10.1109/cc.2015.7114054 · Cited by 77 · see citing works
Matsumura, K., Rolfe, P. J., Lee, J., & Yamakoshi, T. (2014) · PLoS ONE · doi:10.1371/journal.pone.0091205 · Cited by 72 · see citing works
E-technology and Pedometer Walking Program to Increase Physical Activity at Work
Faghri, P. D., Omokaro, C., Parker, C., Nichols, E., Gustavesen, S., & Blozie, E. (2008) · The Journal of Primary Prevention · doi:10.1007/s10935-007-0121-9 · Cited by 71 · see citing works
Fitness Trackers: Fit for Health but Unfit for Security and Privacy
Fereidooni, H., Frassetto, T., Miettinen, M., Sadeghi, A., & Conti, M. (2017) · IEEE/ACM Conference on Connected Health: Applications, Systems and Engineering Technologies · doi:10.1109/chase.2017.54 · Cited by 56 · see citing works
Validity of wrist-worn consumer products to measure heart rate and energy expenditure
Thiebaud, R. S., Funk, M. D., Patton, J. C., Massey, B. L., Shay, T. E., Schmidt, M. G., & Giovannitti, N. (2018) · Digital Health · doi:10.1177/2055207618770322 · Cited by 56 · see citing works
A practical guidance for assessments of sedentary behavior at work: A PEROSH initiative
Holtermann, A., Schellewald, V., Mathiassen, S. E., Gupta, N., Pinder, A., Punakallio, A., Veiersted, K. B., Weber, B., Takala, E., Draicchio, F., et al. (2017) · Applied Ergonomics · doi:10.1016/j.apergo.2017.03.012 · Cited by 52 · see citing works
Photoplethysmogram at green light: Where does chaos arise from?
Sviridova, N., Zhao, T., Aihara, K., Nakamura, K., & Nakano, A. (2018) · Chaos Solitons & Fractals · doi:10.1016/j.chaos.2018.09.016 · Cited by 52 · see citing works
Comparative analysis of actigraphy performance in healthy young subjects
Bellone, G., Plano, S. A., Cardinali, D. P., Chada, D. P., Vigo, D. E., & Golombek, D. A. (2016) · Sleep Science · doi:10.1016/j.slsci.2016.05.004 · Cited by 47 · see citing works
Consumer sleep monitors: is there a baby in the bathwater?
Bianchi, M. T., Russo, K., & Goparaju, B. (2015) · Nature and Science of Sleep · doi:10.2147/nss.s94182 · Cited by 43 · see citing works
Thorup, C. B., Andreasen, J. J., Sørensen, E. E., Grønkjær, M., Dinesen, B., & Hansen, J. (2017) · BMJ Open · doi:10.1136/bmjopen-2016-011742 · Cited by 43 · see citing works
Perceived barriers to walking for physical activity.
Dunton, G. F., & Schneider, M. L. (2006) · PubMed Central · Cited by 41 · see citing works
WearDrive: fast and energy-efficient storage for wearables
Huang, J., Badam, A., Chandra, R., & Nightingale, E. B. (2015) · Cited by 41 · see citing works
Yang, D., Cheng, Y., Zhu, J., Xue, D., Abt, G., Ye, H., & Peng, Y. (2018) · IEEE Access · doi:10.1109/access.2018.2805223 · Cited by 40 · see citing works
Lee, S. M., Shin, H., & Hahm, C. (2016) · International Conference on Advanced Communication Technology · doi:10.1109/icact.2016.7423470 · Cited by 38 · see citing works
Tang, S. K. D., Goh, Y. Y. S., Wong, M. L. D., & Lew, Y. L. (2016) · International Conference on Intelligent and Advanced Systems · doi:10.1109/icias.2016.7824118 · Cited by 34 · see citing works
Energy efficient heart rate sensing using a painted electrode ECG wearable
Krachunov, S., Beach, C., Casson, A. J., Pope, J., Fafoutis, X., Piechocki, R. J., & Craddock, I. J. (2017) · Global Internet of Things Summit (GIoTS) · doi:10.1109/giots.2017.8016260 · Cited by 26 · see citing works
Lang, M. (2018) · Chapman University Digital Commons (Chapman University) · Cited by 21 · see citing works
Movement noise cancellation in PPG signals
Ban, D., & Kwon, S. (2016) · International Conference on Consumer Electronics · doi:10.1109/icce.2016.7430517 · Cited by 12 · see citing works
Lee, S., Shin, H., & Hahm, C. (2016) · International Conference on Advanced Communication Technology · doi:10.1109/icact.2016.7423469 · Cited by 12 · see citing works
Kalantarian, H., Lee, S. I., Mishra, A. K., et al. (2013) · IEEE International Conference on Pervasive Computing and Communications Workshops (PerCom Workshops) · doi:10.1109/percomw.2013.6529578 · Cited by 11 · see citing works
Circuits and systems for energy efficient smart wearables
Sharma, A., Pande, T., Aroul, A. L. P., Soundarapandian, K., & Lee, W. (2016) · Technical digest - International Electron Devices Meeting/Technical digest · doi:10.1109/iedm.2016.7838359 · Cited by 5 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Guidelines for wrist-worn consumer wearable assessment of heart rate in biobehavioral research
Nelson, B. W., Low, C. A., Jacobson, N., Areán, P., Torous, J., & Allen, N. B. (2020) · npj Digital Medicine · doi:10.1038/s41746-020-0297-4 · Cited by 241 · see citing works
- Authors
- Mahloko, L., & Adebesin, F.
- Affiliations
- University of Pretoria (ZA)
- Corresponding author
- Adebesin, Funmi
- ORCID
- Adebesin, Funmi
- Venue
- Lecture Notes in Computer Science
- Publisher
- Springer International Publishing
- Published
- 2020
- Volume
- LNCS-12067
- Issue
- Part II
- Pages
- 96-107
- ISSN
- 0302-9743 (ISSN-L)
- 1611-3349
- PMC
- PMC7134235
- OpenAlex
- W3016016217
- Cited by
- 23
- References
- 47
- Citations by year
- 2026: 1 · 2025: 1 · 2024: 1 · 2023: 4 · 2022: 8 · 2021: 5 · 2020: 3
- FWCI
- 9.08
- Type
- conference paper
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- Locations
- Lecture notes in computer science (publishedVersion)
- PubMed Central (publishedVersion)
- HAL (submittedVersion, cc-by)
- UpSpace Institutional Repository (University of Pretoria) (submittedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Context-Aware Activity Recognition Systems
- Health Sciences › Health Professions › General Health Professions › Mobile Health and mHealth Applications
- Keywords
- data accuracy, heart rate measurement, physical activity monitoring, sleep monitoring, fitness tracking, sensor types, tracking algorithm, energy consumption, processing capacity, developing countries
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review