Fully Implantable Multi-Channel Measurement System for Acquisition of Muscle Activity
Abstract
This paper presents intramuscular electromyogram (EMG) signals obtained with a fully implantable measurement system that were recorded during goal directed arm movements. In a first implantation thin film electrodes were epimysially implanted on the deltoideus of a rhesus macaque and the encapsulation process was monitored by impedance measurements. Increase of impedance reached a constant level after four weeks indicating a complete encapsulation of electrodes. EMG recorded with these electrodes yielded a signal-to-noise ratio of about 80 dB at 200 Hz. The EMG recorded during goal-directed arm movements showed a high similarity to movements in the same direction and at the same time presented clear differences between different movement directions in time domain. Six classifiers and seven time and frequency domain features were investigated with the aim of discriminating the direction of arm movement from EMG signals. Reliable recognition of arm movements was achieved for a subset of the movements under investigation only. A second implantation of the whole measurement system for nine weeks demonstrated simple handling during surgery and good biotolerance in the animals.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“In contrast to surface EMG, implantable EMG has many advantages, such as higher signal quality, less movement artifacts, and the ability to record small and deep muscle activity.[364]”
From Flexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics (2022) · cites this paper as [364]
In the citing paper
“In contrast, surface electrodes generally are more flexible,[364,369] whose morphology in the limited space between this epimysium and the skin is either 1D or 2D thin electrode layers[370,371] or very tiny electrodes (millimeter or even sub-millimeter level).[369,372]”
From Flexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics (2022) · cites this paper as [364]
Citations
Cites
Welch, P. (1967) · IEEE Transactions on Audio and Electroacoustics · doi:10.1109/tau.1967.1161901 · Cited by 12,277 · cited 1×
“For calculation of signal-to-noise ratio (SNR) power spectral density of the recorded EMG signals was calculated using Welch’s method [40].”
MethodA new strategy for multifunction myoelectric control
Hudgins, B., Parker, P., & Scott, R. N. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.204774 · Cited by 2,149 · cited 1×
“[43] classified four different contraction patterns using six features calculated on 40 ms long frames of EMG achieving 91.”
MethodMyoelectric signal processing for control of powered limb prostheses
Parker, P., Englehart, K., & Hudgins, B. (2006) · Journal of Electromyography and Kinesiology · doi:10.1016/j.jelekin.2006.08.006 · Cited by 651 · cited 2×
“The observed SNR of 80 dB at 200 Hz is a considerable improvement compared to the SNR < 10 of surface EMG [6].”
BackgroundMethodControl of Multifunctional Prosthetic Hands by Processing the Electromyographic Signal
Zecca, M., Micera, S., Carrozza, M. C., & Dario, P. (2002) · Critical Reviews? in Biomedical Engineering · doi:10.1615/critrevbiomedeng.v30.i456.80 · Cited by 606 · cited 1×
“Classification of movement direction was based on seven features in time and frequency domain [41]–[44].”
BackgroundStudy of stability of time-domain features for electromyographic pattern recognition
Tkach, D., Huang, H., & Kuiken, T. A. (2010) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-7-21 · Cited by 582 · cited 3×
“Investigations on prosthesis control [42]–[44] evaluated two channels of sEMG and focused on classification of clearly distinguishable contraction patterns that were intentionally generated by human subjects.”
BackgroundControl of Multifunctional Prosthetic Hands by Processing the Electromyographic Signal
Zecca, M., Micera, S., Carrozza, M. C., & Dario, P. (2017) · Critical Reviews in Biomedical Engineering · doi:10.1615/critrevbiomedeng.v45.i1-6.150 · Cited by 570
A Comparison of Surface and Intramuscular Myoelectric Signal Classification
Hargrove, L. J., Englehart, K., & Hudgins, B. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.889192 · Cited by 518 · cited 2×
“Further drawbacks originate from sensitivity to changes in skin conductivity and pickup of external electromagnetic noise [14].”
BackgroundEMG feature evaluation for movement control of upper extremity prostheses
Zardoshti-Kermani, M., Wheeler, B. C., Badie, K., & Hashemi, R. M. (1995) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.481972 · Cited by 477 · cited 3×
“A threshold between 50 mV and 100 mV was reported in the literature [42].”
BackgroundBioimpedance and Bioelectricity Basics
(2008) · Elsevier eBooks · doi:10.1016/b978-0-12-374004-5.x0001-3 · Cited by 462 · cited 1×
“They normally have higher polarization impedance [8], although recently developed electrodes yield comparable performance during long-term EMG measurements [9].”
BackgroundSmith, L. H., Hargrove, L. J., Lock, B. A., & Kuiken, T. A. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2010.2100828 · Cited by 443 · cited 1×
“classification performance [42], [47] the short window length”
Background
Research works printed but not linked: 8
- abu-saleh, 2011, Ein ASIC in 130 nm-Technologie für die Aufnahme von EMG-Signalen zur Ansteuerung einer Prothese, Anal Proc
- abu-saleh, 2010, Ein implantierbares System zur Aufnahme von EMG-Signalen zur Ansteuerung einer Prothese, In Proc BMT
- lewis, 0, Performance of implanted multi-site EMG recording electrodes: In vivo impedance measurements and spectral analysis, Proc of IFESS 2010, 133
- cardona, 2011, Inductive energy transmission system and real-time data link for intelligent implants, In Proc BMT, 56
- lewis, 2012, Acquisition of muscle activity with a fully implantable multi-channel measurement system I2MTC, Proc IEEE I2MTC, 996
- dietl, 2010, Verbundprojekt MyoPlant, In Proc BMT
- willison, 1963, A method for measuring motor unit activity in human muscle, J Physiol, 168, 35
- rafolt, 1997, An implantable multichannel telemetry system, Proc XIV Int Symp Biotelemetry, 277
Cited by
Flexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics
Heng, W., Solomon, S., & Gao, W. (2022) · Advanced Materials · doi:10.1002/adma.202107902 · Cited by 609
Toward higher-performance bionic limbs for wider clinical use
Farina, D., Vujaklija, I., Brånemark, R., Bull, A. M. J., Dietl, H., Graimann, B., Hargrove, L. J., Hoffmann, K.-P., Huang, H., Ingvarsson, T., et al. (2021) · Nature Biomedical Engineering · doi:10.1038/s41551-021-00732-x · Cited by 263
“94 ; MIRA, ref.”
Self-citationHand Gesture Recognition Using Three-Dimensional Electrical Impedance Tomography
Jiang, D., Wu, Y., & Demosthenous, A. (2020) · IEEE Transactions on Circuits & Systems II Express Briefs · doi:10.1109/tcsii.2020.3006430 · Cited by 80
“3006430 electrodes for intramuscular [3], epimysial [4] and intraneural recording [5] can achieve in-depth signal acquisition at higher resolution, but their implementation is limited due to their invasiveness.”
BackgroundApplication of Min-Max Normalization on Subject-Invariant EMG Pattern Recognition
Islam, M. J., Ahmad, S., Haque, F., Reaz, M. B. I., Bhuiyan, M. A. S., & Islam, M. R. (2022) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2022.3220286 · Cited by 73
“So, the EMG signal reflects the functional state of hand movements [1], [2].”
BackgroundControl of a Robotic Hand Using a Tongue Control System—A Prosthesis Application
Johansen, D., Cipriani, C., Popović, D. B., & Struijk, L. N. S. A. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2517742 · Cited by 75
“is the electromyogram (EMG) signal; several studies have been carried out using surface EMG [1]–[9] and intramuscular [10], [11] recordings.”
Background
Metadata
- Authors
- Lewis, S., Russold, M., Dietl, H., Ruff, R., Audi, J. M. C., Hoffmann, K.-P., Abu-Saleh, L., Schroeder, D., Krautschneider, W. H., Westendorff, S., Gail, A., Meiners, T., & Kaniusas, E.
- Affiliations
- Otto Bock Healthcare Products, Vienna, Austria
- Ottobock (Austria) (AT)
- Fraunhofer Institute for Biomedical Engineering (DE)
- Fraunhofer Institute for Biomedical Engineering (IBMT), Sankt Ingbert, Germany
- Hamburg University of Technology, Hamburg, Germany
- Hamburg University of Technology (DE)
- University of Göttingen (DE)
- Werner-Wicker-Klinik (DE)
- Vienna University of Technology, Vienna, Austria
- Venue
- IEEE Transactions on Instrumentation and Measurement
- Published
- 2013-07
- Volume
- 62
- Issue
- 7
- Pages
- 1972-1981
- ISSN
- 0018-9456 (ISSN-L)
- 1557-9662
- OpenAlex
- W2052543545
- Cited by
- 52
- References
- 41
- Citations by year
- 2026: 2 · 2025: 2 · 2024: 6 · 2023: 3 · 2022: 7 · 2021: 7 · 2020: 3 · 2019: 4 · 2018: 2 · 2017: 3 · 2016: 5 · 2015: 4 · 2014: 2 · 2013: 2
- FWCI
- 2.39
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- IEEE Transactions on Instrumentation and Measurement (publishedVersion)
- Fraunhofer-Publica (Fraunhofer-Gesellschaft) (submittedVersion)
- GRO (submittedVersion)
- tub.dok (Hamburg University of Technology) (submittedVersion)
- Funders
- Medizinische Universität Wien
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Keywords
- intramuscular electromyography, thin film electrodes, electrode encapsulation, impedance monitoring, rhesus monkeys, deltoid muscle
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Fully Implantable Multi-Channel Measurement System for Acquisition of Muscle Activity. https://app.cochat.ai/papers/10.1109/tim.2013.2253992
- The paper
- Lewis, S., Russold, M., Dietl, H., Ruff, R., Audi, J. M. C., Hoffmann, K.-P., Abu-Saleh, L., Schroeder, D., Krautschneider, W. H., Westendorff, S., Gail, A., Meiners, T., & Kaniusas, E. (2013). Fully Implantable Multi-Channel Measurement System for Acquisition of Muscle Activity. IEEE Transactions on Instrumentation and Measurement. https://doi.org/10.1109/tim.2013.2253992
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.