On the Model Validation of Active Implantable Medical Device for MRI Safety Assessment
The DOI matches its Crossref record; 35 of the 35 references it cites were found in a registry or an index.
Abstract
In this article, we discuss the model validation for active implantable medical devices (AIMDs) used for safety evaluations under magnetic resonance imaging (MRI) radio frequency (RF) coil emission. A mathematical derivation is presented to provide the guidance on selecting meaningful pathways for the model validation. Suggested validation pathways from current ISO 10974 are used as examples. It is shown that these standard pathways are: 1) inefficient since validations from several pathways are theoretically redundant and 2) incomplete or false since significantly different AIMD models can have identical validation outputs. Based on the developed guidance, two sets of pathways are proposed. It is demonstrated that for efficient and correct model validation, the electric-field distributions along validation pathways should be orthogonal to each other or at least has low correlations between each other. These guidelines can be implemented for future AIMD model validations in ISO10974.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“For example, for MR-induced heating (Section 2.3.1), a model of the power absorption of a new device can be tested by comparing against experiments using the same device in a saline-gel phantom [53].”
From Credibility assessment of in silico clinical trials for medical devices (2024) · cites this paper as [53]
Citations
Cites
Singular value decomposition and least squares solutions
Golub, G. H., & Reinsch, C. (1970) · Numerische Mathematik · doi:10.1007/bf02163027 · Cited by 2,929
Pukelsheim, F. (1994) · The American Statistician · doi:10.2307/2684253 · Cited by 1,242
Pukelsheim, F. (1994) · The American Statistician · doi:10.1080/00031305.1994.10476030 · Cited by 1,211 · cited 1×
“If the differences between the modeling and measurement results are within two-sigma variation, one can use one sigma as the uncertainty of the AIMD model [30].”
MethodTheory and Application of the Linear Model.
Nelder, J. A., & Graybill, F. A. (1977) · Journal of the Royal Statistical Society. Series A (General) · doi:10.2307/2344939 · Cited by 1,165
Theory and Application of the Linear Model.
Sen, P. K., & Graybill, F. A. (1978) · Journal of the American Statistical Association · doi:10.2307/2286558 · Cited by 961
ACR Guidance Document for Safe MR Practices: 2007
Kanal, E., Barkovich, A. J., Bell, C., Borgstede, J. P., Bradley, W. G., Froelich, J. W., Gilk, T., Gimbel, J. R., Gosbee, J., Kuhni-Kaminski, E., et al. (2007) · American Journal of Roentgenology · doi:10.2214/ajr.06.1616 · Cited by 787 · cited 1×
“However, patients with active implantable medical devices (AIMDs) are often prohibited from such procedures due to safety concerns [2].”
BackgroundThe Method of Moments in Electromagnetics
Gibson, W. C. (2007) · Chapman & Hall/CRC eBooks · doi:10.1201/9781420061468 · Cited by 692 · cited 1×
“This is similar to the use of a subdomain basis function in the method of moments to represent the solution space [35].”
MethodPark, S., Kamondetdacha, R., & Nyenhuis, J. A. (2007) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.21159 · Cited by 232 · cited 5×
“2957766 methods as described in [25], the transfer function (TF) method is widely used [23].”
BackgroundMethodInfluentialMRI and implanted medical devices: basic interactions with an emphasis on heating
Nyenhuis, J. A., Park, S.-M., Kamondetdacha, R., Amjad, A., Shellock, F. G., & Rezai, A. R. (2005) · IEEE Transactions on Device and Materials Reliability · doi:10.1109/tdmr.2005.859033 · Cited by 199
Complexity of MRI induced heating on metallic leads: Experimental measurements of 374 configurations
Mattei, E., Triventi, M., Calcagnini, G., Censi, F., Kainz, W., Mendoza, G., Bassen, H. I., & Bartolini, P. (2008) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-7-11 · Cited by 181
Self-citation
Research works printed but not linked: 2
- hadamard, 1893, Résolution d’une question relative aux determinants, Bulletin des Sciences Mathématiques, 17, 240
- 2019, Magnetic resonance imaging (MRI) exams
Cited by
Credibility assessment of in silico clinical trials for medical devices
Pathmanathan, P., Aycock, K., Badal, A., Bighamian, R., Bodner, J., Craven, B. A., & Niederer, S. (2024) · PLOS Computational Biology · doi:10.1371/journal.pcbi.1012289 · Cited by 62
“For example, for MR-induced heating (Section 2.3.1), a model of the power absorption of a new device can be tested by comparing against experiments using the same device in a saline-gel phantom [53].”
InfluentialWang, Y., Guo, R., Hu, W., Zheng, J., Wang, Q., Jiang, J., Kurpad, K. N., Kaula, N. F., Long, S. A., Chen, J., et al. (2021) · Magnetic Resonance in Medicine · doi:10.1002/mrm.28967 · Cited by 27
Self-citationA technique for the reduction of RF‐induced heating of active implantable medical devices during MRI
Wang, Y., Zheng, J., Guo, R., Wang, Q., Kainz, W., Long, S. A., & Chen, J. (2021) · Magnetic Resonance in Medicine · doi:10.1002/mrm.28953 · Cited by 17
Self-citationDeveloping AIMD Models Using Orthogonal Pathways for MRI Safety Assessment
Zheng, J., Wang, Z., Wang, Q., Hu, S., Gu, Z., Kainz, W., & Chen, J. (2020) · IEEE Transactions on Electromagnetic Compatibility · doi:10.1109/temc.2020.2997236 · Cited by 15
“4, a small error in the AIMD development is expected [15].”
BackgroundMethodSelf-citationWang, Y., Zheng, J., Wang, Q., Chen, J., & Kainz, W. (2020) · IEEE Transactions on Electromagnetic Compatibility · doi:10.1109/temc.2020.3039488 · Cited by 7
“Ideally, the incident E fields of these exposures should be orthogonal to each other and generate a relatively high (two to three times higher than the background heating) lead-tip heating [20], [21].”
BackgroundSelf-citationHu, W., Guo, R., Wang, Q., Zheng, J., Tsang, J., Kainz, W., Long, S. A., & Chen, J. (2023) · Magnetic Resonance in Medicine · doi:10.1002/mrm.29650 · Cited by 7
Self-citationAkter, M. K., Guo, R., Islam, M. Z., Zheng, J., Kainz, W., Long, S. A., & Chen, J. (2024) · IEEE Transactions on Electromagnetic Compatibility · doi:10.1109/temc.2024.3403508 · Cited by 5
“This poses a significant risk of immediate tissue damage due to local temperature elevation induced by RF energy deposition [12], [13], [14], [15].”
BackgroundSelf-citationA survey on state of the inter body radio communication channel: Performance and solutions
Mezher, M., Alabbas, A. R., & Ilyas, M. (2022) · Transactions on Emerging Telecommunications Technologies · doi:10.1002/ett.4697 · Cited by 6
“70 The researchers demonstrated a Slotted Triangular Flared (STF) patch applied on a Teflon substrate.”
BackgroundPrediction of Active Implantable Medical Device Electromagnetic Models Using a Neural Network
Chang, J., Lan, Q., Guo, R., Zheng, J., Chen, J., & Kainz, W. (2021) · Digest - IEEE Antennas and Propagation Society. International Symposium · doi:10.1109/aps/ursi47566.2021.9704511 · Cited by 4
Self-citationA Novel Automated System for Evaluating AIMD Safety in MRI RF Fields
Zuo, Z., Shen, A., Ebrahimi, F., Wang, Q., Zheng, J., Jeong, H., Kumar, A., V, & Chen, J. (2025) · IEEE Transactions on Electromagnetic Compatibility · doi:10.1109/temc.2025.3556975 · Cited by 3
“To quantify the difference of tangential electric fields along these trajectories, SVD is used to capture the eigenvalues of tangential electric field matrices for three pathways designed in this study [16], [27].”
InfluentialSelf-citationAnalysis of a Flexible Dual-Channel Octagonal Coil System for UHF MRI
Curreli, N., Lodi, M. B., Melis, A., Puddu, C., Casu, S., Fanti, A., Djuric, N. M., Retico, A., & Mazzarella, G. (2022) · IEEE Access · doi:10.1109/access.2022.3209676 · Cited by 4
“Therefore, it is of great importance to avoid inhomogeneity in the spatial distribution of the transmitted RF fields which could lead to high induced currents and produce large phaseshifted contributions to the B1-field, with consequent critical problems in patient safety [2], [38], [39], [40].”
BackgroundIslam, M. Z., Guo, R., Akter, M. K., Zheng, J., Kainz, W., Long, S. A., & Chen, J. (2024) · Magnetic Resonance in Medicine · doi:10.1002/mrm.30389 · Cited by 2
Self-citation2020 Index IEEE Transactions on Microwave Theory and Techniques Vol. 68
(2020) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2021.3049615 · Cited by 2
Temperature Rise Mechanism of Pacemaker Induced by a Time-Varying Gradient Magnetic Field
Li, Y.-H., Mo, C., Li, J., Yang, P., Wang, J. M., Yu, H., Hu, S., Zhang, P.-M., & Liu, X. (2024) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2024.3485457 · Cited by 1
“However, when patients with AIMDs are scanned, the magnetic field of the MRI can create safety issues by potentially heating the metallic components of the implants [3].”
BackgroundRapid MRI RF-Induced Heating Prediction for Shoulder Prosthesis with Artificial Neural Networks
Liu, Y., Zhang, P., Li, Y.-H., & Guo, R. (2025) · Lecture notes in electrical engineering · doi:10.1007/978-981-96-4067-6_12 · Cited by 1
Wang, Z., Zuo, Z., Song, S., Zheng, J., Wang, Q., Wang, Y., Kainz, W., & Chen, J. (2025) · IEEE Transactions on Electromagnetic Compatibility · doi:10.1109/temc.2025.3566298 · Cited by 1
“Because low lead-tip heating decreases the overall TF T validation accuracy and linearly dependent electric fields result in redundant validation data leading to incomplete validation [12].”
Self-citationUllah, S., Hameed, S., Bhusal, B., Bonmassar, G., & Golestanirad, L. (2025) · IEEE International Symposium on Electromagnetic Compatibility · doi:10.1109/emcsipi52291.2025.11170258 · Cited by 1
Targeted Advertising for the Broadcasting Industry in DVB Markets
Pettazzi, A. (2020) · SMPTE Motion Imaging Journal · doi:10.5594/jmi.2020.3001858 · Cited by 1
The Independence of AIMD Lead Transfer Functions on the Trajectory Shapes
Zuo, Z., Wang, Q., Xu, C. Z., Zheng, J., Kumar, A., V, Kainz, W., & Chen, J. (2023) · URSI General Assembly and Scientific Symposium · doi:10.23919/ursigass57860.2023.10265600 · Cited by 1
Self-citationErratum to “On the Model Validation of Active Implantable Medical Device for MRI Safety Assessment”
Wang, Z., Zheng, J., Wang, Y., Kainz, W., & Chen, J. (2020) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2020.2978595 · Cited by 1
Self-citationLead-Specific Transfer Function Validation Strategy for MRI RF Heating Evaluation
Guo, R., Wang, Z., Song, S., Zheng, J., Wang, Q., & Chen, J. (2022) · Digest - IEEE Antennas and Propagation Society. International Symposium · doi:10.1109/ap-s/usnc-ursi47032.2022.9886193
“To understand the RF-induced heating near tip electrodes, standard method in ISO/TS 10974:2018 [5] that uses the lead transfer function (TFT) in combination with the electric (E)-fields tangential to the lead pathway are used [7], [8].”
MethodSelf-citationYang, L., Akter, M. K., Guo, R., Zheng, J., Steckner, M., Kainz, W., & Chen, J. (2025) · IEEE Transactions on Electromagnetic Compatibility · doi:10.1109/temc.2025.3586966
“The transfer function was validated experimentally using phantom measurements with independent lead pathways [26].”
Self-citation
Metadata
- Authors
- Wang, Z., Zheng, J., Wang, Y., Kainz, W., & Chen, J.
- Affiliations
- University of Houston (US)
- United States Food and Drug Administration (US)
- Venue
- IEEE Transactions on Microwave Theory and Techniques
- Published
- 2020-06
- Volume
- 68
- Issue
- 6
- Pages
- 2234-2242
- ISSN
- 0018-9480 (ISSN-L)
- 1557-9670
- OpenAlex
- W2997094537
- Cited by
- 22
- References
- 34
- Citations by year
- 2025: 5 · 2024: 4 · 2023: 2 · 2022: 3 · 2021: 3 · 2020: 5
- FWCI
- 1.83
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- Locations
- IEEE Transactions on Microwave Theory and Techniques (publishedVersion)
- Funders
- National Science Foundation (1440107)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Wireless Power Transfer Systems
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Biophysics › Electromagnetic Fields and Biological Effects
- Keywords
- active implantable medical devices, MRI safety, model validation, radiofrequency coils, electric field distribution
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.