On the Model Validation of Active Implantable Medical Device for MRI Safety Assessment
The DOI matches its Crossref record; 35 of the 35 references it cites were found in a registry or an index.
Abstract
In this article, we discuss the model validation for active implantable medical devices (AIMDs) used for safety evaluations under magnetic resonance imaging (MRI) radio frequency (RF) coil emission. A mathematical derivation is presented to provide the guidance on selecting meaningful pathways for the model validation. Suggested validation pathways from current ISO 10974 are used as examples. It is shown that these standard pathways are: 1) inefficient since validations from several pathways are theoretically redundant and 2) incomplete or false since significantly different AIMD models can have identical validation outputs. Based on the developed guidance, two sets of pathways are proposed. It is demonstrated that for efficient and correct model validation, the electric-field distributions along validation pathways should be orthogonal to each other or at least has low correlations between each other. These guidelines can be implemented for future AIMD model validations in ISO10974.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“For example, for MR-induced heating (Section 2.3.1), a model of the power absorption of a new device can be tested by comparing against experiments using the same device in a saline-gel phantom [53].”
From Credibility assessment of in silico clinical trials for medical devices (2024) · cites this paper as [53]
Citations
Cites
Singular value decomposition and least squares solutions
Golub, G. H., & Reinsch, C. (1970) · Numerische Mathematik · doi:10.1007/bf02163027 · Cited by 2,929
Pukelsheim, F. (1994) · The American Statistician · doi:10.2307/2684253 · Cited by 1,242
Pukelsheim, F. (1994) · The American Statistician · doi:10.1080/00031305.1994.10476030 · Cited by 1,211 · cited 1×
“If the differences between the modeling and measurement results are within two-sigma variation, one can use one sigma as the uncertainty of the AIMD model [30].”
MethodTheory and Application of the Linear Model.
Nelder, J. A., & Graybill, F. A. (1977) · Journal of the Royal Statistical Society. Series A (General) · doi:10.2307/2344939 · Cited by 1,165
Theory and Application of the Linear Model.
Sen, P. K., & Graybill, F. A. (1978) · Journal of the American Statistical Association · doi:10.2307/2286558 · Cited by 961
ACR Guidance Document for Safe MR Practices: 2007
Kanal, E., Barkovich, A. J., Bell, C., Borgstede, J. P., Bradley, W. G., Froelich, J. W., Gilk, T., Gimbel, J. R., Gosbee, J., Kuhni-Kaminski, E., et al. (2007) · American Journal of Roentgenology · doi:10.2214/ajr.06.1616 · Cited by 787 · cited 1×
“However, patients with active implantable medical devices (AIMDs) are often prohibited from such procedures due to safety concerns [2].”
BackgroundThe Method of Moments in Electromagnetics
Gibson, W. C. (2007) · Chapman & Hall/CRC eBooks · doi:10.1201/9781420061468 · Cited by 692 · cited 1×
“This is similar to the use of a subdomain basis function in the method of moments to represent the solution space [35].”
MethodPark, S., Kamondetdacha, R., & Nyenhuis, J. A. (2007) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.21159 · Cited by 232 · cited 5×
“2957766 methods as described in [25], the transfer function (TF) method is widely used [23].”
BackgroundMethodInfluentialMRI and implanted medical devices: basic interactions with an emphasis on heating
Nyenhuis, J. A., Park, S.-M., Kamondetdacha, R., Amjad, A., Shellock, F. G., & Rezai, A. R. (2005) · IEEE Transactions on Device and Materials Reliability · doi:10.1109/tdmr.2005.859033 · Cited by 199
Complexity of MRI induced heating on metallic leads: Experimental measurements of 374 configurations
Mattei, E., Triventi, M., Calcagnini, G., Censi, F., Kainz, W., Mendoza, G., Bassen, H. I., & Bartolini, P. (2008) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-7-11 · Cited by 181
Self-citationNordbeck, P., Weiss, I., Ehses, P., Ritter, O., Warmuth, M., Fidler, F., Herold, V., Jakob, P. M., Ladd, M. E., Quick, H. H., et al. (2009) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21881 · Cited by 164
A Technique to Evaluate MRI-Induced Electric Fields at the Ends of Practical Implanted Lead
Feng, S., Qiang, R., Kainz, W., & Chen, J. (2014) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2014.2376523 · Cited by 140 · cited 3×
“The TF approach divides the safety evaluation into two parts: 1) the development and validation of the AIMD model under the RF field [23], [24] and 2) the evaluation of incident electric fields inside the human bodies without electrodes.”
MethodSelf-citationEffect of magnetic resonance imaging on DDD pacemakers
Erlebacher, J. A., Cahill, P. T., Pannizzo, F., & Knowles, R. (1986) · The American Journal of Cardiology · doi:10.1016/0002-9149(86)90768-x · Cited by 138 · cited 1×
“During MRI scans, the electromagnetic signals emitted by a radio frequency (RF) coil can penetrate into human bodies and interact strongly with AIMDs [3]–[15].”
BackgroundGolestanirad, L., KIRSCH, J. E., Bonmassar, G., Downs, S., Elahi, B., Martin, A. J., Iacono, M. I., Angelone, L. M., Keil, B. R., Wald, L. L., et al. (2018) · NeuroImage · doi:10.1016/j.neuroimage.2018.09.034 · Cited by 121 · cited 1×
“This kind of a folding pathway was observed in clinical settings [36].”
BackgroundGolestanirad, L., Angelone, L. M., KIRSCH, J. E., Downs, S., Keil, B. R., Bonmassar, G., & Wald, L. L. (2019) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2018.2885517 · Cited by 58
Liu, J., Zheng, J., Wang, Q., Kainz, W., & Chen, J. (2018) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2018.2851975 · Cited by 57 · cited 1×
“An alternative approach for the AIMD model validation is to use the transmission-line model described in [31].”
MethodSelf-citationComplete Loss of ICD Programmability After Magnetic Resonance Imaging
Fiek, M., REMP, T., Reithmann, C., & Steinbeck, G. (2004) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2004.00573.x · Cited by 52 · cited 1×
“age or the RF-induced heating, numerical and experimental investigations were performed [14]–[24].”
MethodSohn, S., DelaBarre, L. J., Gopinath, A., & Vaughan, J. T. (2014) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2014.2331621 · Cited by 47 · cited 4×
“During MRI scans, the electromagnetic signals emitted by a radio frequency (RF) coil can penetrate into human bodies and interact strongly with AIMDs [3]–[15].”
BackgroundInfluentialMattei, E., Calcagnini, G., Censi, F., Triventi, M., & Bartolini, P. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2043102 · Cited by 43
Evaluation of MRI RF electromagnetic field induced heating near leads of cochlear implants
Zeng, Q. Y., Wang, Q., Zheng, J., Kainz, W., & Chen, J. (2018) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aacbf2 · Cited by 34 · cited 3×
“The induced voltage near the IPG can be evaluated by [26], [27]”
BackgroundMethodSelf-citationTanaka, R., Yumoto, T., Shiba, N., Okawa, M., Yasuhara, T., Ichikawa, T., Tokunaga, K., Date, I., & Ujike, Y. (2012) · Journal of neurosurgery · doi:10.3171/2012.9.jns12738 · Cited by 26 · cited 1×
“The induced voltage between the device electrode and the IPG can lead to device damage [3], while the localized energy near the device tip may cause instantaneous tissue damage [16]–[20].”
BackgroundAn MRI Compatible RF MEMs Controlled Wireless Power Transfer System
Byron, K., Winkler, S. A., Robb, F. J. L., Vasanawala, S. S., Pauly, J. M., & Scott, G. (2019) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2019.2902554 · Cited by 25
Kawamura, T., Saito, K., Kikuchi, S., Takahashi, M., & Ito, K. (2009) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2009.2029712 · Cited by 24
Zeng, Q. Y., Liu, J., Angelone, L. M., Lloyd, T. D. R., Wedan, S., Chen, J., & Kainz, W. (2018) · IEEE Transactions on Electromagnetic Compatibility · doi:10.1109/temc.2018.2862249 · Cited by 24 · cited 3×
“The induced voltage near the IPG can be evaluated by [26], [27]”
BackgroundMethodSelf-citationChen, Z., Solbach, K., Erni, D., & Rennings, A. (2016) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2016.2631169 · Cited by 23
A Novel Implant Electrode Design Safe in the RF Field of MRI Scanners
McCabe, S., & Scott, J. B. (2017) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2017.2669977 · Cited by 21 · cited 1×
“The induced voltage between the device electrode and the IPG can lead to device damage [3], while the localized energy near the device tip may cause instantaneous tissue damage [16]–[20].”
BackgroundElabyad, I. A., Herrmann, T., Bruns, C., Bernarding, J., & Erni, D. (2017) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2017.2708707 · Cited by 20
Synthesis and Design of Programmable Subwavelength Coil Array for Near-Field Manipulation
Gao, F., Zhang, F., Wakatsuchi, H., & Sievenpiper, D. F. (2015) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2015.2450711 · Cited by 19
Fetus MRI at 7 T: ${B}_{1}$ Shimming Strategy and SAR Safety Implications
Xin, X., Huang, Q., Gao, Y., Li, B., Xu, Y., & Chen, W. (2013) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2013.2247053 · Cited by 16
Liu, J., Zheng, J., Wang, Q., Kainz, W., & Chen, J. (2020) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2020.2978598 · Cited by 16 · cited 1×
“An alternative approach for the AIMD model validation is to use the transmission-line model described in [31].”
MethodSelf-citationElabyad, I. A., Terekhov, M., Stefanescu, M. R., Lohr, D., Fischer, M., & Schreiber, W. (2019) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2019.2913636 · Cited by 15
Kordzadeh, A., & Zanche, N. D. (2017) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2017.2668406 · Cited by 9
Simplified Transfer Function Assessment of Implantable Leads for MRI Safety Evaluations
Feng, S., Wang, Q., Wen, C., Zheng, J., Kainz, W., & Chen, J. (2018) · IEEE Transactions on Electromagnetic Compatibility · doi:10.1109/temc.2018.2867814 · Cited by 9 · cited 1×
“We also developed a smoothed spurious AIMD model (Tf3) by filtering out those unrealistic high-frequency variations, as shown in [33].”
MethodSelf-citationExcitation and RF Field Control of a Human-Size 10.5-T MRI System
Bluem, P., de Moortele, P. V., Adriany, G., & Popovic, Z. B. (2018) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2018.2884405 · Cited by 8
Gandji, N. P., Lee, G., Semouchkin, G. B., Semouchkina, E., Neuberger, T., & Lanagan, M. T. (2018) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2018.2874266 · Cited by 6
Research works printed but not linked: 2
- hadamard, 1893, Résolution d’une question relative aux determinants, Bulletin des Sciences Mathématiques, 17, 240
- 2019, Magnetic resonance imaging (MRI) exams
Cited by
Credibility assessment of in silico clinical trials for medical devices
Pathmanathan, P., Aycock, K., Badal, A., Bighamian, R., Bodner, J., Craven, B. A., & Niederer, S. (2024) · PLOS Computational Biology · doi:10.1371/journal.pcbi.1012289 · Cited by 62
“For example, for MR-induced heating (Section 2.3.1), a model of the power absorption of a new device can be tested by comparing against experiments using the same device in a saline-gel phantom [53].”
InfluentialWang, Y., Guo, R., Hu, W., Zheng, J., Wang, Q., Jiang, J., Kurpad, K. N., Kaula, N. F., Long, S. A., Chen, J., et al. (2021) · Magnetic Resonance in Medicine · doi:10.1002/mrm.28967 · Cited by 27
Self-citationA technique for the reduction of RF‐induced heating of active implantable medical devices during MRI
Wang, Y., Zheng, J., Guo, R., Wang, Q., Kainz, W., Long, S. A., & Chen, J. (2021) · Magnetic Resonance in Medicine · doi:10.1002/mrm.28953 · Cited by 17
Self-citationDeveloping AIMD Models Using Orthogonal Pathways for MRI Safety Assessment
Zheng, J., Wang, Z., Wang, Q., Hu, S., Gu, Z., Kainz, W., & Chen, J. (2020) · IEEE Transactions on Electromagnetic Compatibility · doi:10.1109/temc.2020.2997236 · Cited by 15
“4, a small error in the AIMD development is expected [15].”
BackgroundMethodSelf-citationWang, Y., Zheng, J., Wang, Q., Chen, J., & Kainz, W. (2020) · IEEE Transactions on Electromagnetic Compatibility · doi:10.1109/temc.2020.3039488 · Cited by 7
“Ideally, the incident E fields of these exposures should be orthogonal to each other and generate a relatively high (two to three times higher than the background heating) lead-tip heating [20], [21].”
BackgroundSelf-citation
Metadata
- Authors
- Wang, Z., Zheng, J., Wang, Y., Kainz, W., & Chen, J.
- Affiliations
- University of Houston (US)
- United States Food and Drug Administration (US)
- Venue
- IEEE Transactions on Microwave Theory and Techniques
- Published
- 2020-06
- Volume
- 68
- Issue
- 6
- Pages
- 2234-2242
- ISSN
- 0018-9480 (ISSN-L)
- 1557-9670
- OpenAlex
- W2997094537
- Cited by
- 22
- References
- 34
- Citations by year
- 2025: 5 · 2024: 4 · 2023: 2 · 2022: 3 · 2021: 3 · 2020: 5
- FWCI
- 1.83
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- Locations
- IEEE Transactions on Microwave Theory and Techniques (publishedVersion)
- Funders
- National Science Foundation (1440107)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Wireless Power Transfer Systems
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Biophysics › Electromagnetic Fields and Biological Effects
- Keywords
- active implantable medical devices, MRI safety, model validation, radiofrequency coils, electric field distribution
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.