Real-Time Performance Assessment of High-Order Tremor Estimators Used in a Wearable Tremor Suppression Device
Open accessAbstract
The side effects and complications of traditional treatments for treating pathological tremor have led to a growing research interest in wearable tremor suppression devices (WTSDs) as an alternative approach. Similar to how the human brain coordinates the function of the human system, a tremor estimator determines how a WTSD functions. Although many tremor estimation algorithms have been developed and validated, whether they can be implemented on a cost-effective embedded system has not been studied; furthermore, their effectiveness on tremor signals with multiple harmonics has not been investigated. Therefore, in this study, four tremor estimators were implemented, evaluated, and compared: Weighted-frequency Fourier Linear Combiner (WFLC), WFLC-based Kalman Filter (WFLC-KF), Band-limited Multiple FLC, and enhanced High-order WFLC-KF (eHWFLC-KF). This study aimed to evaluate the performance of each algorithm on a bench-top tremor suppression system with 18 recorded tremor motion datasets; and compare the performance of each estimator. The experimental evaluation showed that the eHWFLC-KF-based WTSD achieved the best performance when suppressing tremor with an average of 89.3% reduction in tremor power, and an average error when tracking voluntary motion of 6.6°/s. Statistical analysis indicated that the eHWFLC-KF-based WTSD is able to reduce the power of tremor better than the WFLC and WFLC-KF, and the BMFLC-based WTSD is better than the WFLC. The performance when tracking voluntary motion is similar among all systems. This study has proven the feasibility of implementing various tremor estimators in a cost-effective embedded system, and provided a real-time performance assessment of four tremor estimators.
Overview
CoChat Summary
TL;DR
This paper implements and compares four tremor estimators (WFLC, WFLC-KF, BMFLC, eHWFLC-KF) in real time on a wearable tremor suppression device using a bench-top mechatronic simulator and 18 recorded parkinsonian tremor datasets. The eHWFLC-KF-based system achieved the best tremor suppression, with an average 89.3% reduction in tremor power, while voluntary-motion tracking was similar across systems.
Key findings
- The eHWFLC-KF-based WTSD achieved the highest average tremor power suppression ratios: 91.1% for resting tremor, 91% for postural tremor, and 85.7% for kinetic tremor; its average error when tracking voluntary motion was 6.6°/s.
- Statistical analysis showed eHWFLC-KF suppressed tremor significantly better than WFLC and WFLC-KF; BMFLC was better than WFLC; no significant difference was found between eHWFLC-KF and BMFLC for postural and kinetic tremor, and more data were said to be required to find a significant difference with BMFLC.
- No significant difference was found in ETVM or correlation coefficient between eHWFLC-KF and any other algorithm; all mean ETVM values were under 10°/s, and all mean correlation coefficients were greater than 0.9, though BMFLC’s correlation coefficient was significantly lower than WFLC and WFLC-KF.
- Computational complexity was O(m²) time and O(m) space for WFLC and BMFLC, and O(m³) time and O(m²) space for WFLC-KF and eHWFLC-KF; the paper notes Coppersmith-Winograd could reduce KF-based time complexity to O(m²·³⁷⁵).
Methods
- Four tremor estimators were implemented on a cost-effective embedded system within a wearable tremor suppression glove and tested on a bench-top mechatronic tremor simulator reproducing 18 motion datasets from 18 individuals with parkinsonian tremor; the datasets included resting, postural, and kinetic/voluntary motion subsets, with a total MDS-UPDRS tremor rating of 8.72 ± 2.47.
- Angular velocity from an IMU was sampled at 50 Hz and used as the control variable; post-trial processing in MATLAB used Butterworth bandpass filters (3–18 Hz for tremor, 0.1–2 Hz for voluntary motion).
- Metrics included tremor power suppression ratio (TPSR), error when tracking voluntary motion (ETVM), correlation coefficient, time delay, and computational complexity; statistical comparisons used Shapiro-Wilk, MANOVA or Kruskal-Wallis with Bonferroni correction, α = 0.05, and 80% power in SPSS.
Limitations
- The paper states an inherent limitation is the use of recorded tremor motion on a single-axis bench-top simulator, which gives higher repeatability than human trials but may not fully resemble actual tremor with multiple degrees of freedom; the simulator also lacks mass-spring components and cannot accurately simulate human joint impedance.
- The paper states that algorithm parameters were chosen from the literature and were not optimized to minimize tremor in the current experimental setup, so parameter optimization is needed for future WTSDs.
- The paper states that the relationship between algorithm performance and computational power was not studied, and that deciding whether overall performance exceeds other algorithms in humans may require functional tests and qualitative feedback; it also notes time delay is an inherent drawback of tremor estimators.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures











Verification
Claims the paper makes
2 Valid · 9 Not valid · 18 Not checked · 61 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Furthermore, research on using the KF and its derivative algorithms [21], [31], [32] has shown improved accuracy in tremor estimation compared to gradient descent methods.”
What the sources say
- Not checked — no full text of the sourcecites [21] [21] Y. Zhou et al. , 'Design and validation of a high-order weightedfrequency Fourier linear combiner-based Kalman filter for parkinsonian tremor estimation,' in Annual International Conference of the IEEE Engineering in Medicine and Biology Society , Orlando, FL, USA, August 16-20, 2016, pp. 5893-5896. - source
- Partly supportedcites [31] [31] S. Shahtalebi et al. , 'Wake: Wavelet decomposition coupled with adaptive Kalman filtering for pathological tremor extraction,' Biomedical Signal Processing and Control , vol. 48, pp. 179-188, 2019. - sourceExperimental — accuracy not yet measured
The Abstract and Section 5 describe WAKE as an adaptive Kalman-filtering method that improves tremor-estimation accuracy over its counterparts; Section 4 compares it with BMFLC and E-BMFLC. Our copy does not identify those comparators as gradient-descent methods, so it does not establish the specific comparison claimed. (LLM, conf=0.78)
- Partly supportedcites [32] [32] L. Yao et al. , 'Improved detection of parkinsonian resting tremor with feature engineering and Kalman filtering,' Clinical Neurophysiology , vol. 131, no. 1, pp. 274-284, 2020. - sourceExperimental — accuracy not yet measured
The Discussion (Features, classifiers, and Kalman filter; page unavailable) says Kalman filtering improved tremor detection, and the Results report a 17.0% gain in specificity. The paper does not report a comparison with gradient descent methods or establish improved tremor-estimation accuracy relative to them. (LLM, conf=0.87)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The experimental evaluation showed that the eHWFLC-KF-based WTSD achieved the best performance when suppressing tremor with an average of 89.3% reduction in tremor power, and an average error when tracking voluntary motion of 6.6°/s.”
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The experimental evaluation showed that the eHWFLC-KF-based WTSD achieved the best performance when suppressing tremor with an average of 89.3% reduction in tremor power, and an average error when tracking voluntary motion of 6.6°/s.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper states — it cites no source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The performance when tracking voluntary motion is similar among all systems.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Statistical analysis indicated that the eHWFLC-KF-based WTSD is able to reduce the power of tremor better than the WFLC and WFLC-KF, and the BMFLC-based WTSD is better than the WFLC.”
hedgedstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
43 more claims
Create a free account to see all 90 claims and every source verdict.
- Probably not a substantial claim. 22
- It may state two separate claims as one. 25
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ V. CONCLUSIONS
“the computational complexity of the eHWFLCKF is higher than the other algorithms.”
§ TABLE II
“| WFLC-KF | O ( m 3 ) | O ( m 2 ) | | eHWFLC-KF | O ( m 3 ) | O ( m 2 ) |”
Citations
Cites
Matrix multiplication via arithmetic progressions
Coppersmith, D., & Winograd, S. (1990) · Journal of Symbolic Computation · doi:10.1016/s0747-7171(08)80013-2 · Cited by 2,248
Adaptive Filtering: Algorithms and Practical Implementation
Diniz, P. S. R. (2008) · doi:10.1007/978-0-387-68606-6 · Cited by 1,172
Radhakrishnan, S., Kolippakkam, D., & Mathura, V. S. (2008) · Bioinformatics: A Concept-Based Introduction · doi:10.1007/978-0-387-84870-9_3 · Cited by 842
Adaptive cancelling of physiological tremor for improved precision in microsurgery
Riviere, C. N., Rader, R. S., & Thakor, N. V. (1998) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.686791 · Cited by 260 · cited 3×
“The WFLC algorithm [18] is an adaptive extension of the FLC with the ability to track different frequencies.”
MethodReal-Time Epileptic Seizure Prediction Using AR Models and Support Vector Machines
Chisci, L., Mavino, A., Perferi, G., Sciandrone, M., Anile, C., Colicchio, G., & Fuggetta, F. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2038990 · Cited by 255
Deep Brain Stimulation and Tremor
Lyons, K. E., & Pahwa, R. (2008) · Neurotherapeutics · doi:10.1016/j.nurt.2008.01.004 · Cited by 143
Diagnosis and Treatment of Common Forms of Tremor
Puschmann, A., & Wszolek, Z. (2011) · Seminars in Neurology · doi:10.1055/s-0031-1271312 · Cited by 138
Online tremor suppression using electromyography and low-level electrical stimulation
Dosen, S., Muceli, S., Dideriksen, J. L., Romero, J. P., Rocon, E., Pons, J., & Farina, D. (2015) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2014.2328296 · Cited by 120
A practical introduction to data structures and algorithm analysis
(1996) · Cited by 112
An adaptive estimation of periodic signals using a Fourier linear combiner
Vaz, C., Kong, X., & Thakor, N. (1994) · IEEE Transactions on Signal Processing · doi:10.1109/78.258116 · Cited by 108 · cited 2×
“tiple Fourier Linear Combiners (FLCs) [23].”
Method
Research works printed but not linked: 1
- [37][37] Y. Zhou, 'A wearable mechatronic ddevice for hand tremor monitoring and suppression: development and evaluation,' Ph.D. dissertation, School of Biomedical Engineering, Western University, 2019. -
Cited by
Shagan, A., Chase-Markopoulou, C., Walter, J. R., Sellhorn-Timm, J., Shao, Y., Nadler, T., Benson, A., Wochner, I., Rumley, E. H., Wurster, I., et al. (2025) · Device · doi:10.1016/j.device.2025.100719 · Cited by 8
Rasheed, A., Kim, J., Ang, W. T., & Veluvolu, K. C. (2025) · ISA Transactions · doi:10.1016/j.isatra.2024.12.040 · Cited by 6
A Tremor-Suppressing Strategy Based on the Equivalent-Input-Disturbance Approach
Xie, M., She, J., Liu, Z., Yang, Z., & Sato, D. (2024) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2024.3375911 · Cited by 6
Paik, P., Zhou, X., & Atashzar, S. F. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3575999 · Cited by 3
Designs of Upper Limb Tremor Suppression Orthoses: Efficacy and Wearer's Comfort
ChiGan, M., Chen, M., & Jing, M. (2024) · Journal of Medical Devices · doi:10.1115/1.4066968 · Cited by 2
Metadata
- Authors
- Zhou, Y., Habibollahi, Z., Ibrahim, A., Jenkins, M. E., Naish, M. D., & Trejos, A. L.
- Affiliations
- Western University (CA)
- Venue
- IEEE Transactions on Neural Systems and Rehabilitation Engineering
- Published
- 2022
- Volume
- 30
- Pages
- 2856-2865
- ISSN
- 1534-4320 (ISSN-L)
- 1558-0210
- OpenAlex
- W4301390986
- Cited by
- 8
- References
- 42
- Citations by year
- 2026: 1 · 2025: 5 · 2024: 2
- FWCI
- 0.90
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- IEEE Transactions on Neural Systems and Rehabilitation Engineering (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
- Scholarship@Western (Western University) (submittedVersion)
- Funders
- Natural Sciences and Engineering Research Council (NSERC) of Canada and the Canadian Institutes of Health Research (CIHR) through a Collaborative Health Research Projects (396234)
- Canadian Foundation for Innovation
- Ontario Research Fund
- Topics
- Life Sciences › Neuroscience › Cognitive Neuroscience › Motor Control and Adaptation
- Health Sciences › Medicine › Clinical Neurology › Deep Brain Stimulation and Movement Disorders
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Keywords
- pathological tremor, Kalman filter, embedded system implementation, real-time performance, multiple harmonics
- MeSH terms
- Wearable Electronic Devices, Algorithms, Fourier Analysis, Humans, Motion, Tremor
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Real-Time Performance Assessment of High-Order Tremor Estimators Used in a Wearable Tremor Suppression Device. https://app.cochat.ai/papers/10.1109/tnsre.2022.3211450
- The paper
- Zhou, Y., Habibollahi, Z., Ibrahim, A., Jenkins, M. E., Naish, M. D., & Trejos, A. L. (2022). Real-Time Performance Assessment of High-Order Tremor Estimators Used in a Wearable Tremor Suppression Device. IEEE Transactions on Neural Systems and Rehabilitation Engineering. https://doi.org/10.1109/tnsre.2022.3211450
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.