Real-Time Performance Assessment of High-Order Tremor Estimators Used in a Wearable Tremor Suppression Device

Open access
Zhou, Yue, Habibollahi, Zahra, Ibrahim, Anas, Jenkins, Mary E., Naish, Michael D., Trejos, Ana Luisa · 2022 · IEEE Transactions on Neural Systems and Rehabilitation Engineering

Abstract

The side effects and complications of traditional treatments for treating pathological tremor have led to a growing research interest in wearable tremor suppression devices (WTSDs) as an alternative approach. Similar to how the human brain coordinates the function of the human system, a tremor estimator determines how a WTSD functions. Although many tremor estimation algorithms have been developed and validated, whether they can be implemented on a cost-effective embedded system has not been studied; furthermore, their effectiveness on tremor signals with multiple harmonics has not been investigated. Therefore, in this study, four tremor estimators were implemented, evaluated, and compared: Weighted-frequency Fourier Linear Combiner (WFLC), WFLC-based Kalman Filter (WFLC-KF), Band-limited Multiple FLC, and enhanced High-order WFLC-KF (eHWFLC-KF). This study aimed to evaluate the performance of each algorithm on a bench-top tremor suppression system with 18 recorded tremor motion datasets; and compare the performance of each estimator. The experimental evaluation showed that the eHWFLC-KF-based WTSD achieved the best performance when suppressing tremor with an average of 89.3% reduction in tremor power, and an average error when tracking voluntary motion of 6.6°/s. Statistical analysis indicated that the eHWFLC-KF-based WTSD is able to reduce the power of tremor better than the WFLC and WFLC-KF, and the BMFLC-based WTSD is better than the WFLC. The performance when tracking voluntary motion is similar among all systems. This study has proven the feasibility of implementing various tremor estimators in a cost-effective embedded system, and provided a real-time performance assessment of four tremor estimators.

Overview

CoChat Summary

TL;DR

This paper implements and compares four tremor estimators (WFLC, WFLC-KF, BMFLC, eHWFLC-KF) in real time on a wearable tremor suppression device using a bench-top mechatronic simulator and 18 recorded parkinsonian tremor datasets. The eHWFLC-KF-based system achieved the best tremor suppression, with an average 89.3% reduction in tremor power, while voluntary-motion tracking was similar across systems.

Key findings

  • The eHWFLC-KF-based WTSD achieved the highest average tremor power suppression ratios: 91.1% for resting tremor, 91% for postural tremor, and 85.7% for kinetic tremor; its average error when tracking voluntary motion was 6.6°/s.
  • Statistical analysis showed eHWFLC-KF suppressed tremor significantly better than WFLC and WFLC-KF; BMFLC was better than WFLC; no significant difference was found between eHWFLC-KF and BMFLC for postural and kinetic tremor, and more data were said to be required to find a significant difference with BMFLC.
  • No significant difference was found in ETVM or correlation coefficient between eHWFLC-KF and any other algorithm; all mean ETVM values were under 10°/s, and all mean correlation coefficients were greater than 0.9, though BMFLC’s correlation coefficient was significantly lower than WFLC and WFLC-KF.
  • Computational complexity was O(m²) time and O(m) space for WFLC and BMFLC, and O(m³) time and O(m²) space for WFLC-KF and eHWFLC-KF; the paper notes Coppersmith-Winograd could reduce KF-based time complexity to O(m²·³⁷⁵).

Methods

  • Four tremor estimators were implemented on a cost-effective embedded system within a wearable tremor suppression glove and tested on a bench-top mechatronic tremor simulator reproducing 18 motion datasets from 18 individuals with parkinsonian tremor; the datasets included resting, postural, and kinetic/voluntary motion subsets, with a total MDS-UPDRS tremor rating of 8.72 ± 2.47.
  • Angular velocity from an IMU was sampled at 50 Hz and used as the control variable; post-trial processing in MATLAB used Butterworth bandpass filters (3–18 Hz for tremor, 0.1–2 Hz for voluntary motion).
  • Metrics included tremor power suppression ratio (TPSR), error when tracking voluntary motion (ETVM), correlation coefficient, time delay, and computational complexity; statistical comparisons used Shapiro-Wilk, MANOVA or Kruskal-Wallis with Bonferroni correction, α = 0.05, and 80% power in SPSS.

Limitations

  • The paper states an inherent limitation is the use of recorded tremor motion on a single-axis bench-top simulator, which gives higher repeatability than human trials but may not fully resemble actual tremor with multiple degrees of freedom; the simulator also lacks mass-spring components and cannot accurately simulate human joint impedance.
  • The paper states that algorithm parameters were chosen from the literature and were not optimized to minimize tremor in the current experimental setup, so parameter optimization is needed for future WTSDs.
  • The paper states that the relationship between algorithm performance and computational power was not studied, and that deciding whether overall performance exceeds other algorithms in humans may require functional tests and qualitative feedback; it also notes time delay is an inherent drawback of tremor estimators.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1109/tnsre.2022.3211450

Cites

40 of 40 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Fig. 1 A prototype of a WTSD developed for testing on individuals with pathological tremor. The key components used for this study consist of a 3D printed glove frame (#1) and a brushless DC motor (#2). (a) Top view. (b) Bottom view. (c) Side view.
Fig. 1 A prototype of a WTSD developed for testing on individuals with pathological tremor. The key components used for this study consist of a 3D printed glove frame (#1) and a brushless DC motor (#2). (a) Top view. (b) Bottom view. (c) Side view.CC-BY · From the PDF
Fig. 2 Bench-top experimental setup for evaluating four different tremor estimators. (a) Top view of the experimental setup: The bench-top mechatronic tremor simulator consists of a 3D printed hand model (#10), a brushless DC motor (#8), and a motor controller (#7). The WTSD consists of a 3D printed glove frame (#1), a brushless DC motor (#2), a motor controller (#4), a communication converter (#5), a microcontroller (#6), and an inertial measurement unit (#9). A J-Link debug probe (#3) is connected to the microcontroller for code debugging and downloading. (b) Exploded CAD view of the bench-top mechatronic tremor simulator.
Fig. 2 Bench-top experimental setup for evaluating four different tremor estimators. (a) Top view of the experimental setup: The bench-top mechatronic tremor simulator consists of a 3D printed hand model (#10), a brushless DC motor (#8), and a motor controller (#7). The WTSD consists of a 3D printed glove frame (#1), a brushless DC motor (#2), a motor controller (#4), a communication converter (#5), a microcontroller (#6), and an inertial measurement unit (#9). A J-Link debug probe (#3) is connected to the microcontroller for code debugging and downloading. (b) Exploded CAD view of the bench-top mechatronic tremor simulator.CC-BY · From the PDF
Fig. 3 Control flow diagram of the bench-top mechatronic tremor simulator and the WTSD.
Fig. 3 Control flow diagram of the bench-top mechatronic tremor simulator and the WTSD.CC-BY · From the PDF
Fig. 4 Time series and frequency distribution of one original resting tremor motion collected from one individual with parkinsonian tremor (red) and the suppressed motion outputs of the WTSD (blue) using different algorithms. 'a', 'b', 'c', and 'd' represent the motion collected from the WTSD using the WFLC, the BMFLC, the WFLC-KF, and the eHWFLC-KF, respectively.
Fig. 4 Time series and frequency distribution of one original resting tremor motion collected from one individual with parkinsonian tremor (red) and the suppressed motion outputs of the WTSD (blue) using different algorithms. 'a', 'b', 'c', and 'd' represent the motion collected from the WTSD using the WFLC, the BMFLC, the WFLC-KF, and the eHWFLC-KF, respectively.CC-BY · From the PDF
Fig. 5 Distribution of the TPSR of resting tremor (top), postural tremor (middle), and kinetic tremor (bottom). A statistically significant difference ( p < .05) is labeled by a '*', and a statistically highly significant difference ( p < .001) is labeled by a '**'.
Fig. 5 Distribution of the TPSR of resting tremor (top), postural tremor (middle), and kinetic tremor (bottom). A statistically significant difference ( p < .05) is labeled by a '*', and a statistically highly significant difference ( p < .001) is labeled by a '**'.CC-BY · From the PDF
Fig. 6 Tremor suppression ratio achieved from the systems with different tremor estimators. The x axis shows the common logarithm of the power of the original tremor. The y axis shows the TPSR. Different markers indicate the results collected from different systems. Each solid line presents the first-order polynomial fit of the results from each system.
Fig. 6 Tremor suppression ratio achieved from the systems with different tremor estimators. The x axis shows the common logarithm of the power of the original tremor. The y axis shows the TPSR. Different markers indicate the results collected from different systems. Each solid line presents the first-order polynomial fit of the results from each system.CC-BY · From the PDF
Fig. 7 Estimated marginal means of the TPSR that was divided into two classes by the median value of the power of the original tremor. The Low Tremor Power class includes the data below the median value, and the High Tremor Power class includes the data above the median value. A statistically significant difference ( p < .05) is indicated by a '*'.
Fig. 7 Estimated marginal means of the TPSR that was divided into two classes by the median value of the power of the original tremor. The Low Tremor Power class includes the data below the median value, and the High Tremor Power class includes the data above the median value. A statistically significant difference ( p < .05) is indicated by a '*'.CC-BY · From the PDF
Fig. 8 Time series of the ETVM collected from one sample dataset tested with different algorithms. 'a', 'b', 'c', and 'd' represent the motion collected from the WTSD using the WFLC, the BMFLC, the WFLC-KF, and the eHWFLC-KF, respectively.
Fig. 8 Time series of the ETVM collected from one sample dataset tested with different algorithms. 'a', 'b', 'c', and 'd' represent the motion collected from the WTSD using the WFLC, the BMFLC, the WFLC-KF, and the eHWFLC-KF, respectively.CC-BY · From the PDF
Fig. 9 Distribution of the ETVM of different algorithms. No significant difference was found between each pair of algorithms.
Fig. 9 Distribution of the ETVM of different algorithms. No significant difference was found between each pair of algorithms.CC-BY · From the PDF
Fig. 10 Distribution of the correlation coefficient between the estimated voluntary motion and the original voluntary motion. A significant difference ( p < .05) is indicated by a '*'.
Fig. 10 Distribution of the correlation coefficient between the estimated voluntary motion and the original voluntary motion. A significant difference ( p < .05) is indicated by a '*'.CC-BY · From the PDF
Fig. 11 Distribution of the time delay between measured voluntary motion and the original voluntary motion. No significant difference was found between each pair of algorithms.
Fig. 11 Distribution of the time delay between measured voluntary motion and the original voluntary motion. No significant difference was found between each pair of algorithms.CC-BY · From the PDF

Verification

Claims the paper makes

2 Valid · 9 Not valid · 18 Not checked · 61 Own results and interpretations

43 more claims

Create a free account to see all 90 claims and every source verdict.

See all claims — free account
  • Probably not a substantial claim. 22
  • It may state two separate claims as one. 25

Create a free account to see these claims.

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

Each contradiction between two passages listed here was confirmed by an independent check.

1 of 1

Citations

Cites

Show all 42 references

Research works printed but not linked: 1

  1. [37][37] Y. Zhou, 'A wearable mechatronic ddevice for hand tremor monitoring and suppression: development and evaluation,' Ph.D. dissertation, School of Biomedical Engineering, Western University, 2019. -

Cited by

See all 8

Metadata

Authors
Zhou, Y., Habibollahi, Z., Ibrahim, A., Jenkins, M. E., Naish, M. D., & Trejos, A. L.
Affiliations
Western University (CA)
Venue
IEEE Transactions on Neural Systems and Rehabilitation Engineering
Published
2022
Volume
30
Pages
2856-2865
ISSN
  • 1534-4320 (ISSN-L)
  • 1558-0210
OpenAlex
W4301390986
Cited by
8
References
42
Citations by year
2026: 1 · 2025: 5 · 2024: 2
FWCI
0.90
Type
journal article
Language
en
Open access
Locations
  • IEEE Transactions on Neural Systems and Rehabilitation Engineering (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
  • Scholarship@Western (Western University) (submittedVersion)
Funders
  • Natural Sciences and Engineering Research Council (NSERC) of Canada and the Canadian Institutes of Health Research (CIHR) through a Collaborative Health Research Projects (396234)
  • Canadian Foundation for Innovation
  • Ontario Research Fund
Topics
  • Life Sciences › Neuroscience › Cognitive Neuroscience › Motor Control and Adaptation
  • Health Sciences › Medicine › Clinical Neurology › Deep Brain Stimulation and Movement Disorders
  • Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
Keywords
pathological tremor, Kalman filter, embedded system implementation, real-time performance, multiple harmonics
MeSH terms
Wearable Electronic Devices, Algorithms, Fourier Analysis, Humans, Motion, Tremor

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Cited sources are checked with a language model’s help (gpt-6-sol).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Real-Time Performance Assessment of High-Order Tremor Estimators Used in a Wearable Tremor Suppression Device. https://app.cochat.ai/papers/10.1109/tnsre.2022.3211450
The paper
Zhou, Y., Habibollahi, Z., Ibrahim, A., Jenkins, M. E., Naish, M. D., & Trejos, A. L. (2022). Real-Time Performance Assessment of High-Order Tremor Estimators Used in a Wearable Tremor Suppression Device. IEEE Transactions on Neural Systems and Rehabilitation Engineering. https://doi.org/10.1109/tnsre.2022.3211450

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.