Initial Measurements with the PETsys TOFPET2 ASIC Evaluation Kit and a Characterization of the ASIC TDC
Open accessThe DOI matches its Crossref record; 48 of the 48 references it cites were found in a registry or an index.
Abstract
For a first characterization, we used the two KETEK-PM3325-WB SiPMs each equipped with a 3 mm x 3 mm x 5 mm LYSO scintillation crystal provided with the PETsys TOFPET2 ASIC evaluation kit. We changed the lower of two discriminator thresholds (D_T1) in the timing branch from vth_t1 = 5-30. The overvoltage was varied in a range of 1.25-7.25 V. The ambient temperature was kept at 16 °C. For all measurements, we performed an energy calibration including a correction for saturation. We evaluated the energy resolution, the coincidence resolving time (CRT) and the coincidence rate. At an overvoltage of 6 V, we obtained an energy resolution of about 10% full width at half maximum (FWHM), a CRT of approximately 210 ps FWHM and 400 ps full width at tenth maximum, the coincidence rate showed only small variations of about 5%. To investigate the influence of the ambient temperature, it was varied between 12 °C and 20 °C. At 12 °C and an overvoltage of 6.5 V, a CRT of approx. 195 ps FWHM and an energy resolution of about 9.5% FWHM could be measured. Observed satellite peaks in the time difference spectra were investigated in more detail. We could show that the location of the satellite peaks is correlated with a programmable delay element in the trigger circuit.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures


![Fig. 3 Sketch of the TOFPET2 ASIC trigger scheme used in the timing branch. Adapted from [40].](https://figures.cochat.ai/paper-figures/02ae9b94dbc343e6ba137b0bc5e1ffea/initial-measurements-with-the-petsys-tofpet2-asic-evaluation-kit-and-a-figure-3.png)










Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Rinne, J. O., Brooks, D. J., Rossor, M. N., Fox, N. C., Bullock, R., Klunk, W. E., Mathis, C. A., Blennow, K., Barakos, J., Okello, A. A., et al. (2010) · The Lancet Neurology · doi:10.1016/s1474-4422(10)70043-0 · Cited by 701
Bar‐Shalom, R., Yefremov, N., Guralnik, L., Gaitini, D. E., Frenkel, A., Kuten, A., Altman, H., Keidar, Z., & Israel, O. (2003) · PubMed · Cited by 594
Ryan, C. G., Clayton, E., Griffin, W. L., Sie, S. H., & Cousens, D. R. (1988) · Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms · doi:10.1016/0168-583x(88)90063-8 · Cited by 510 · cited 1×
“The 511 keV and 1275 keV peaks in the energy value histogram (Figure 4a) were identified by calculating and subtracting the background [39]–[41] and performing a peak search using a Markov chain algorithm [42].”
MethodWeber, W. A., Petersen, V., Schmidt, B., Tyndale-Hines, L., Link, T., Peschel, C., & Schwaiger, M. (2003) · Journal of Clinical Oncology · doi:10.1200/jco.2003.12.004 · Cited by 493
Brain PET in the Diagnosis of Alzheimer’s Disease
Marcus, C., Mena, E., & Subramaniam, R. M. (2014) · Clinical Nuclear Medicine · doi:10.1097/rlu.0000000000000547 · Cited by 382
Recent developments in time-of-flight PET
Vandenberghe, S., Mikhaylova, E., D’Hoe, E., Mollet, P., & Karp, J. S. (2016) · EJNMMI Physics · doi:10.1186/s40658-016-0138-3 · Cited by 368 · cited 1×
“[coincidence resolving time (CRT)] with a timing resolution in the order of a few hundred picoseconds [1], [2].”
BackgroundTowards picosecond resolution with single-photon avalanche diodes
Cova, S., Longoni, A., & Andreoni, A. (1981) · Review of Scientific Instruments · doi:10.1063/1.1136594 · Cited by 219 · cited 1×
“signal contains timing information in the order of a few tens of picoseconds [10], [12]–[15].”
BackgroundRecent advances and future advances in time-of-flight PET
Moses, W. W. (2007) · Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment · doi:10.1016/j.nima.2007.06.038 · Cited by 205 · cited 3×
“Furthermore, SiPMs can be operated in a magnetic field why they are the photosensor of choice for applications that integrate a PET into an MRI [1]–[3].”
BackgroundMethodAdvances in time-of-flight PET
Surti, S., & Karp, J. S. (2016) · Physica Medica · doi:10.1016/j.ejmp.2015.12.007 · Cited by 173 · cited 12×
“[coincidence resolving time (CRT)] with a timing resolution in the order of a few hundred picoseconds [1], [2].”
BackgroundMethodInfluentialGundacker, S., Auffray, E., Pauwels, K., & Lecoq, P. (2016) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/7/2802 · Cited by 173 · cited 1×
“/MeV), short decay time (∼40 ns), and fast rise time (∼70 ps), which result in excellent CRT values [1], [10], [11].”
Background
Research works printed but not linked: 9
- 37. PETsys TOF ASIC Evaluation Kit (Flyer), V21 Edition, PETsys Electron. SA, Porto Salvo, Portugal, Nov. 2017. [Online]. Available: http://www.petsys.com/
- 28. TOFPET ASIC V1-Short Data Sheet (Revision 1.2), PETsys Electron. SA, Porto Salvo, Portugal, 2014.
- 54. G. F. Knoll, Radiation Detection and Measurement. Hoboken, NJ, USA: Wiley, 2010.
- 40. TOFPET 2 SiPM Readout ASIC, PETsys Electron. SA, Porto Salvo, Portugal, 2018.
- 12. Product Data Sheet SiPM-Silicon Photomultiplier PM3315- WB/PM3325-WB, KETEK GmbH, Munich, Germany, Nov. 2016. [Online]. Available: https://www.ketek.net/
- 49. TOFPET2 ASIC-Evaluation Report, PETsys Electron. SA, Porto Salvo, Portugal, 2017. [Online]. Available: http://www.petsys.com/
- 42. TOFPET2 ASIC Evaluation Kit-Software User Guide V2018.04, PETsys Electron. SA, Porto Salvo, Portugal, 2018.
- 41. TOFPET2 ASIC Evaluation Kit-Hardware User Guide (V1.2), PETsys Electron. SA, Porto Salvo, Portugal, 2018.
- 43. Personal Communication, PETsys Electron. SA, Porto Salvo, Portugal, 2018.
Cited by
Liang, S., Wu, H., Zhen, L., Hua, Q., Garg, S., Kaddoum, G., Hassan, M. M., & Yu, K. (2022) · IEEE Transactions on Intelligent Transportation Systems · doi:10.1109/tits.2022.3158253 · Cited by 367
“At present, devices that can implement edge computing include SoC, FPGA, Application Specific Integrated Circuit(ASIC) [36], Central Processing Units (CPU) and GPU.”
MethodSND@LHC: the scattering and neutrino detector at the LHC
Acampora, G., Ahdida, C., Albanese, R., Albrecht, C., Alexandrov, A., Andreini, M., Anokhina, A., Asada, T., Auberson, N., Baldanza, C., et al. (2024) · Journal of Instrumentation · doi:10.1088/1748-0221/19/05/p05067 · Cited by 105
Sensors for Positron Emission Tomography Applications
Jiang, W., Chalich, Y., & Deen, M. J. (2019) · Sensors · doi:10.3390/s19225019 · Cited by 112
Park, H., Yi, M., & Lee, J. S. (2022) · Biomedical Engineering Letters · doi:10.1007/s13534-022-00234-y · Cited by 70
Evaluation of the PETsys TOFPET2 ASIC in multi-channel coincidence experiments
Nadig, V., Schug, D., Weissler, B., & Schulz, V. (2021) · EJNMMI Physics · doi:10.1186/s40658-021-00370-x · Cited by 65
“These single events were further processed using an analysis software developed at our institute [52].”
BackgroundMethodResultInfluentialSelf-citation
Metadata
- Authors
- Schug, D., Nadig, V., Weissler, B., Gebhardt, P., & Schulz, V.
- Affiliations
- RWTH Aachen University (DE)
- Venue
- IEEE Transactions on Radiation and Plasma Medical Sciences
- Published
- 2019-07
- Volume
- 3
- Issue
- 4
- Pages
- 444-453
- ISSN
- 2469-7303 (ISSN-L)
- 2469-7311
- OpenAlex
- W2885357483
- Cited by
- 56
- References
- 48
- Citations by year
- 2026: 1 · 2025: 5 · 2024: 9 · 2023: 11 · 2022: 7 · 2021: 13 · 2020: 4 · 2019: 6
- FWCI
- 5.27
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- IEEE Transactions on Radiation and Plasma Medical Sciences (publishedVersion, cc-by)
- arXiv (Cornell University) (submittedVersion)
- RWTH Publications (RWTH Aachen) (submittedVersion, other-oa)
- arXiv (Cornell University)
- Funders
- Deutsche Forschungsgemeinschaft (SCHU 2973/2-1)
- Horizon 2020 Framework Programme (667211)
- Topics
- Physical Sciences › Physics and Astronomy › Radiation › Radiation Detection and Scintillator Technologies
- Physical Sciences › Physics and Astronomy › Nuclear and High Energy Physics › Particle Detector Development and Performance
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Keywords
- time-of-flight PET, LYSO scintillator, coincidence time resolution, energy resolution, coincidence frequency, overvoltage, energy calibration, ambient temperature dependence, satellite peaks, trigger circuit
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Initial Measurements with the PETsys TOFPET2 ASIC Evaluation Kit and a Characterization of the ASIC TDC. https://app.cochat.ai/papers/10.1109/trpms.2018.2884564
- The paper
- Schug, D., Nadig, V., Weissler, B., Gebhardt, P., & Schulz, V. (2019). Initial Measurements with the PETsys TOFPET2 ASIC Evaluation Kit and a Characterization of the ASIC TDC. IEEE Transactions on Radiation and Plasma Medical Sciences. https://doi.org/10.1109/trpms.2018.2884564
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.