Propagation of spontaneously actuated pulsive vibration in human heart wall and in vivo viscoelasticity estimation
Open accessThe DOI matches its Crossref record; 29 of the 29 references it cites were found in a registry or an index.
Abstract
Though myocardial viscoelasticity is essential in the evaluation of heart diastolic properties, it has never been noninvasively measured in vivo. By the ultrasonic measurement of the myocardial motion, we have already found that some pulsive waves are spontaneously excited by aortic-valve closure (AVC) at end-systole (T0). These waves may serve as an ideal source of the intrinsic heart sound caused by AVC. In this study, using a sparse sector scan, in which the beam directions are restricted to about 16, the pulsive waves were measured almost simultaneously at about 160 points set along the heart wall at a sufficiently high frame rate. The consecutive spatial phase distributions, obtained by the Fourier transform of the measured waves, clearly revealed wave propagation along the heart wall for the first time. The propagation time of the wave along the heart wall is very small (namely, several milliseconds) and cannot be measured by conventional equipment. Based on this phenomenon, we developed a means to measure the myocardial viscoelasticity in vivo. In this measurement, the phase velocity of the wave is determined for each frequency component. By comparing the dispersion of the phase velocity with the theoretical one of the Lamb wave (the plate flexural wave), which propagates along the viscoelastic plate (heart wall) immersed in blood, the instantaneous viscoelasticity is determined noninvasively. This is the first report of such noninvasive determination. In in vivo experiments applied to five healthy subjects, propagation of the pulsive wave was clearly visible in all subjects. For the 60-Hz component, the typical propagation speed rapidly decreased from 5 m/s just before the time of AVC (t = T0 - 8 ms) to 3 m/s at t = T0 + 10 ms. In the experiments, it was possible to determine the viscosity more precisely than the elasticity. The typical value of elasticity was about 24-30 kPa and did not change around the time of AVC. The typical transient values of viscosity decreased rapidly from 400 Pa x s at t = T0 - 8 ms to 70 Pa x s at t = T0 + 10 ms. The measured shear elasticity and viscosity in this study are comparable to those obtained for the human tissues using audio frequency in in vitro experiments reported in the literature.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Acoustic Fields and Waves in Solids
Auld, B. A. (1973) · Medical Entomology and Zoology · Cited by 5,827
Human heart: tagging with MR imaging--a method for noninvasive assessment of myocardial motion.
Zerhouni, E. A., Parish, D. M., Rogers, W. J., Yang, A., & Shapiro, E. P. (1988) · Radiology · doi:10.1148/radiology.169.1.3420283 · Cited by 1,430 · cited 4×
“This method offers potential for in vivo myocardial tissue characterization of diastolic properties, which cannot be obtained by conventional echocardiography, TDI, CT, or MRI.”
InfluentialHandbook of Physiology—The Cardiovascular System
Guyton, A. C. (1985) · Critical Care Medicine · doi:10.1097/00003246-198505000-00025 · Cited by 1,147
Real-Time Strain Rate Imaging of the Left Ventricle by Ultrasound
Heimdal, A., Støylen, A., Torp, H., & Skjærpe, T. (1998) · Journal of the American Society of Echocardiography · doi:10.1016/s0894-7317(98)70151-8 · Cited by 850
Alterations with Age in the Viscoelastic Properties of Human Arterial Walls
Learoyd, B. M., & Taylor, M. G. (1966) · Circulation Research · doi:10.1161/01.res.18.3.278 · Cited by 681 · cited 2×
“In [42], the artery was subjected to pressure oscillations, and μ1 and μ2 were measured from the human carotid artery to the femoral artery.”
Color Doppler Myocardial Imaging: A New Technique for the Assessment of Myocardial function
Sutherland, G. R., Stewart, M. J., Groundstroem, K. W. E., Moran, C. M., Fleming, A., Guell-Peris, F. J., Riemersma, R. A., Fenn, L. N., Fox, K. A. A., & McDicken, W. N. (1994) · Journal of the American Society of Echocardiography · doi:10.1016/s0894-7317(14)80001-1 · Cited by 617 · cited 1×
“The tissue Doppler imaging (TDI) technique [16]–[21]— modified ultrasound two-dimensional (2-D) color flow mapping—enables determination of motion distribution of the myocardium in real time.”
Strain and strain rate imaging: a new clinical approach to quantifying regional myocardial function
Sutherland, G. R., Di Salvo, G., Claus, P., D'hooge, J., & Bijnens, B. (2004) · Journal of the American Society of Echocardiography · doi:10.1016/j.echo.2004.03.027 · Cited by 606 · cited 2×
“Even in current measurement, however, the sampling frequency of the motion of the heart wall is low (at most 200 Hz [21]), that is, the sampling period is 5 ms, which is too long to detect the propagation time of the wave.”
Miyatake, K., Yamagishi, M., Tanaka, N., Uematsu, M., Yamazaki, N., Mine, Y., Sano, A., & Hirama, M. (1995) · Journal of the American College of Cardiology · doi:10.1016/0735-1097(94)00421-l · Cited by 495
Ultrasonic imaging of internal vibration of soft tissue under forced vibration
Yamakoshi, Y., Sato, J., & Sato, T. (1990) · IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control · doi:10.1109/58.46969 · Cited by 474 · cited 2×
“As shown in Table I, the order of magnitude of the phase velocity values is similar to those measured recently using low-frequency, shear-wave vibrations reported in the literature for human quadriceps muscle [2], for human bicep muscle [5], for pig leg/shoulder muscles [1], and for beef muscle [4].”
Shear elasticity probe for soft tissues with 1-D transient elastography
Sandrin, L., Tanter, M., Gennisson, J.-L., Catheline, S., & Fink, M. (2002) · IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control · doi:10.1109/58.996561 · Cited by 427 · cited 3×
“According to [2] and [5], the shear wave velocity and shear elasticity of the human muscle increase as a load is applied in in vivo experiments.”
A solution to diffraction biases in sonoelasticity: The acoustic impulse technique
Catheline, S., Wu, F., & Fink, M. (1999) · The Journal of the Acoustical Society of America · doi:10.1121/1.426907 · Cited by 282 · cited 2×
“3 m/s Beef muscle (2-ms pulsed excitation) [4]”
Transcutaneous measurement and spectrum analysis of heart wall vibrations
Kanai, H., Sato, M., Koiwa, Y., & Chubachi, N. (1996) · IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control · doi:10.1109/58.535480 · Cited by 277 · cited 2×
“To measure the original vibrations of the heart sounds (which are audible by stethoscope) using ultrasound from the chest wall, we have already developed a method to transcutaneously measure the heart-wall vibrations as a waveform at one point or multiple points preset along an ultrasonic beam in the heart wall [22],…”
MethodSelf-citationA model for longitudinal and shear wave propagation in viscoelastic media
Szabo, T. L., & Wu, J. (2000) · The Journal of the Acoustical Society of America · doi:10.1121/1.428630 · Cited by 274 · cited 1×
“In [50], experimental results on soft tissue have been discussed in the framework of the linear theory of viscoelasticity relating stress and strain on the basis of the Voigt, Maxwell, and Kelvin models.”
Oestreicher, H. L. (1951) · The Journal of the Acoustical Society of America · doi:10.1121/1.1906828 · Cited by 257 · cited 3×
“In [32], for the low frequency of 500–5,000 Hz, the shear parameters were measured as μ = 2.”
InfluentialMeasurement of Viscoelastic Properties of Arteries in the Living Dog
GOW, B. S., & TAYLOR, M. G. (1968) · Circulation Research · doi:10.1161/01.res.23.1.111 · Cited by 209 · cited 2×
“About 2 Hz 290–1,460 kPa 2,800–19,000 Pa·s Canine descending thoracic [43]”
Sonoelastic determination of human skeletal muscle elasticity
Levinson, S. F., Shinagawa, M., & Sato, T. (1995) · Journal of Biomechanics · doi:10.1016/0021-9290(94)00173-2 · Cited by 203 · cited 2×
“According to [2] and [5], the shear wave velocity and shear elasticity of the human muscle increase as a load is applied in in vivo experiments.”
Noninvasive evaluation of local myocardial thickening and its color-coded imaging
Kanai, H., Hasegawa, H., Chubachi, N., et al. (1997) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/58.655190 · Cited by 193 · cited 2×
“To measure the original vibrations of the heart sounds (which are audible by stethoscope) using ultrasound from the chest wall, we have already developed a method to transcutaneously measure the heart-wall vibrations as a waveform at one point or multiple points preset along an ultrasonic beam in the heart wall [22],…”
MethodSelf-citationDoppler tissue imaging: myocardial wall motion velocities in normal subjects
Palka, P., Lange, A., Fleming, A., et al. (1995) · Journal of the American Society of Echocardiography · doi:10.1016/s0894-7317(05)80380-3 · Cited by 175
Ultrasonic shear wave properties of soft tissues and tissuelike materials
Madsen, E. L., Sathoff, H. J., & Zagzebski, J. A. (1983) · The Journal of the Acoustical Society of America · doi:10.1121/1.390158 · Cited by 171 · cited 1×
“In [37], using a similar method, μ1 and μ2 were 124 kPa and 0.”
Kulke, M., Fujita-Becker, S., Rostkova, E., Neagoe, C., Labeit, D., Manstein, D. J., Gautel, M., & Linke, W. A. (2001) · Circulation Research · doi:10.1161/hh2201.099453 · Cited by 170 · cited 1×
“Not only the myocardial elasticity but also the viscosity depend on the interaction between the giant muscle protein titin and actin filaments in the sarcomere during both diastolic stretch and systolic shortening [52].”
Diffraction field of a low frequency vibrator in soft tissues using transient elastography
Catheline, S., Thomas, J.-L., Wu, F., & Fink, M. A. (1999) · IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control · doi:10.1109/58.775668 · Cited by 167
Viscoelastic properties of pig kidney in shear, experimental results and modelling
Nasseri, S., Bilston, L. E., & Phan-Thien, N. (2002) · Rheologica Acta · doi:10.1007/s003970200017 · Cited by 140 · cited 5×
“In [44], using a standard strain-controlled rheometer, the shear elasticity and viscosity of fresh pig kidney were 1–7.”
InfluentialThomas, S. P., Kucera, J. P., Bircher-Lehmann, L., Rudy, Y., Saffitz, J. E., & Kléber, A. G. (2003) · Circulation Research · doi:10.1161/01.res.0000074916.41221.ea · Cited by 134 · cited 1×
“3 m/s [8], [9] and 2–3 m/s [10], respectively.”
Loeffler, L., & SAGAWA, K. (1975) · Circulation Research · doi:10.1161/01.res.36.4.498 · Cited by 55 · cited 3×
“Many researchers have measured the relaxation time in the step response of the soft tissue [34], [53], in which the relaxation time can be compared with the viscosity using appropriate modeling.”
Adler, R. S., Rubin, J. M., Bland, P. H., & Carson, P. L. (1990) · Ultrasound in Medicine & Biology · doi:10.1016/0301-5629(90)90021-4 · Cited by 47 · cited 1×
“Analysis of the motion-mode (M-mode) image [14], [15] is a candidate for transthoracic measurement of the minute and rapid heart-wall motion.”
Dynamic mechanical measurements during the life to death transition in animal tissues1
Fitzgerald, E. R. (1975) · Biorheology · doi:10.3233/bir-1975-12611 · Cited by 39 · cited 1×
“Such behavior is influenced by collagen and its interactions with other components of the tissue [44], by water content [45], and by time after death [51].”
Acoustic shear wave displacement measurement using ultrasound
Dutt, V., Kinnick, R. R., & Greenleaf, J. F. (2002) · Proceedings/Proceedings - IEEE Ultrasonics Symposium · doi:10.1109/ultsym.1996.584202 · Cited by 25
Onset of pulsatile waves in the heart walls at end-systole
Kanai, H., Yonechi, S., Susukida, I., et al. (2000) · Ultrasonics · doi:10.1016/s0041-624x(99)00117-1 · Cited by 19 · cited 1×
“A steep notch pulse then occurs exactly at T0, which coincides with the beginning of the IR period [27].”
Self-citationRegional ischemia increases sensitivity of left ventricular relaxation to volume in pigs
Solomon, S. B., & Glantz, S. A. (1999) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1999.276.6.h1994 · Cited by 10
Cited by
Ultrafast imaging in biomedical ultrasound
Tanter, M., & Fink, M. (2014) · IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control · doi:10.1109/tuffc.2014.2882 · Cited by 1,151
An Overview of Elastography-An Emerging Branch of Medical Imaging
Sarvazyan, A., J. Hall, T., W. Urban, M., Fatemi, M., R. Aglyamov, S., & S. Garra, B. (2011) · Current Medical Imaging Reviews · doi:10.2174/157340511798038684 · Cited by 428
Ultrasound Elastography: The New Frontier in Direct Measurement of Muscle Stiffness
Brandenburg, J. E., Eby, S. F., Song, P., Zhao, H., Brault, J. S., Chen, S., & An, K.-N. (2014) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/j.apmr.2014.07.007 · Cited by 327
Ultrafast Cardiac Ultrasound Imaging
Cikes, M., Tong, L., Sutherland, G. R., & D’hooge, J. (2014) · JACC: Cardiovascular Imaging · doi:10.1016/j.jcmg.2014.06.004 · Cited by 220
“Kanai (29), 2005 Coarse sampling Shear wave imaging 5 healthy volunteers Myocardial viscoelastic properties (Pulsive wave propagation) Phased array; 450 Hz”
BackgroundInfluentialCouade, M., Pernot, M., Messas, E., Bel, A., Ba, M., Hagege, A., Fink, M., & Tanter, M. (2011) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2010.2076829 · Cited by 190
“wavelengths (larger than the wall thickness) and have been shown to be guided by the wall and to behave as Lamb waves [6].”
Metadata
- Authors
- Kanai, H.
- Venue
- IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control
- Published
- 2005-11
- Volume
- 52
- Issue
- 11
- Pages
- 1931-1942
- ISSN
- 0885-3010 (ISSN-L)
- OpenAlex
- W2105448221
- Cited by
- 213
- References
- 29
- FWCI
- 4.29
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.