Numerical Analysis of Blood Damage Potential of the HeartMate II and HeartWare HVAD Rotary Blood Pumps
The DOI matches its Crossref record; 22 of the 22 references it cites were found in a registry or an index.
Abstract
Implantable left ventricular assist devices (LVAD s) became the therapy of choice in treating end‐stage heart failure. Although survival improved substantially and is similar in currently clinically implanted LVAD s HeartMate II (HM II) and H eart W are HVAD, complications related to blood trauma are frequently observed. The aim of this study was to compare these two pumps regarding their potential blood trauma employing computational fluid dynamics. High‐resolution structured grids were generated for the pumps. N ewtonian flow was calculated, solving R eynolds‐averaged N avier–S tokes equations with a sliding mesh approach and a k‐ω shear stress transport turbulence model for the operating point of 4.5 L/min and 80 mm Hg. The pumps were compared in terms of volumes subjected to certain viscous shear stress thresholds, below which no trauma was assumed (von W illebrand factor cleavage: 9 Pa, platelet activation: 50 Pa, and hemolysis: 150 Pa), and associated residence times. Additionally, a hemolysis index was calculated based on a E ulerian transport approach. Twenty‐two percent of larger volumes above 9 Pa were observed in the HVAD; above 50 Pa and 150 Pa the differences between the two pumps were marginal. Residence times were higher in the HVAD for all thresholds. The hemolysis index was almost equal for the HM II and HVAD. Besides the gap regions in both pumps, the inlet regions of the rotor and diffuser blades have a high hemolysis production in the HM II, whereas in the HVAD, the volute tongue is an additional site for hemolysis production. Thus, in this study, the comparison of the HM II and the HVAD using numerical methods indicated an overall similar tendency to blood trauma in both pumps. However, influences of turbulent shear stresses were not considered and effects of the pivot bearing in the HM II were not taken into account. Further in vitro investigations are required.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Two-equation eddy-viscosity turbulence models for engineering applications
Menter, F. R. (1994) · AIAA Journal · doi:10.2514/3.12149 · Cited by 21,593
ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure 2012
Authors/Task Force Members, McMurray, J. J. V., Adamopoulos, S., Anker, S. D., Auricchio, A., Böhm, M., Dickstein, K., Falk, V., Filippatos, G., Fonseca, C., et al. (2012) · European Heart Journal · doi:10.1093/eurheartj/ehs104 · Cited by 6,879
Extended Mechanical Circulatory Support With a Continuous-Flow Rotary Left Ventricular Assist Device
Pagani, F. D., Miller, L. W., Russell, S. D., Aaronson, K. D., John, R., Boyle, A. J., Conte, J. V., Bogaev, R. C., MacGillivray, T. E., Naka, Y., et al. (2009) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2009.03.055 · Cited by 869
Aaronson, K. D., Slaughter, M. S., Miller, L. W., McGee, E. C., Cotts, W., Acker, M. A., Jessup, M. L., Gregorič, I. D., Loyalka, P., Frazier, O. H., et al. (2012) · Circulation · doi:10.1161/circulationaha.111.058412 · Cited by 688
Red Blood Cell Damage by Shear Stress
Leverett, L. B., Hellums, J. D., Alfrey, C. P., & Lynch, E. C. (1972) · Biophysical Journal · doi:10.1016/s0006-3495(72)86085-5 · Cited by 682
Giersiepen, M., Wurzinger, L. J., Opitz, R., & Reul, H. (1990) · The International Journal of Artificial Organs · doi:10.1177/039139889001300507 · Cited by 594
Starling, R. C., Naka, Y., Boyle, A. J., Gonzalez-Stawinski, G., John, R., Jorde, U., Russell, S. D., Conte, J. V., Aaronson, K. D., McGee, E. C., et al. (2011) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2010.10.062 · Cited by 452
Shear stress enhances the proteolysis of von Willebrand factor in normal plasma
Tsai, H., Sussman, II, & Nagel, R. (1994) · Blood · doi:10.1182/blood.v83.8.2171.2171 · Cited by 422
Acquired von Willebrand Syndrome in Continuous-Flow Ventricular Assist Device Recipients
Crow, S. S., Chen, D., Milano, C. A., Thomas, W., Joyce, L. D., Piacentino, V., Sharma, R., Wu, J. R., Arepally, G. M., Bowles, D. E., et al. (2010) · The Annals of Thoracic Surgery · doi:10.1016/j.athoracsur.2010.04.099 · Cited by 388
Fraser, K. H., Zhang, T., Taskin, M. E., Griffith, B. P., & Wu, Z. J. (2012) · Journal of Biomechanical Engineering · doi:10.1115/1.4007092 · Cited by 384
Slaughter, M. S., Pagani, F. D., McGee, E. C., Birks, E. J., Cotts, W. G., Gregoric, I., Howard Frazier, O., Icenogle, T., Najjar, S. S., Boyce, S. W., et al. (2013) · The Journal of Heart and Lung Transplantation · doi:10.1016/j.healun.2013.04.004 · Cited by 349
Geisen, U., Heilmann, C., Beyersdorf, F., Benk, C., Berchtold-Herz, M., Schlensak, C., Budde, U., & Zieger, B. (2008) · European Journal of Cardio-Thoracic Surgery · doi:10.1016/j.ejcts.2007.12.047 · Cited by 331
Human red blood cell hemolysis in a turbulent shear flow: Contribution of Reynolds shear stresses
Sallam, A. M., & Hwang, N. H. C. (1984) · Biorheology · doi:10.3233/bir-1984-21605 · Cited by 245
A review of computational fluid dynamics analysis of blood pumps
Behbahani, M., Behr, M., Hormes, M., Steinseifer, U., Arora, D., Coronado, O. M., & Pasquali, M. (2009) · European Journal of Applied Mathematics · doi:10.1017/s0956792509007839 · Cited by 163
Platelet and Coagulation Parameters Following Millisecond Exposure to Laminar Shear Stress
Wurzinger, L. J., Opitz, R., Blasberg, P., & Schmid-Schönbein, H. (1985) · Thrombosis and Haemostasis · doi:10.1055/s-0038-1657744 · Cited by 135
Platelet Dysfunction in Outpatients With Left Ventricular Assist Devices
Steinlechner, B., Dworschak, M., Birkenberg, B., Duris, M., Zeidler, P., Fischer, H., Milošević, L., Wieselthaler, G. M., Wolner, E., Quehenberger, P., et al. (2008) · The Annals of Thoracic Surgery · doi:10.1016/j.athoracsur.2008.10.027 · Cited by 134
Birschmann, I., Dittrich, M. T., Eller, T., Wiegmann, B., Reininger, A. J., Budde, U., & Strüber, M. (2013) · The Journal of Heart and Lung Transplantation · doi:10.1016/j.healun.2013.11.010 · Cited by 114
A Validated Computational Fluid Dynamics Model to Estimate Hemolysis in a Rotary Blood Pump
Arvand, A., Hormes, M., & Reul, H. (2005) · Artificial Organs · doi:10.1111/j.1525-1594.2005.29089.x · Cited by 108
The effect of shear stress on protein conformation
Stasio, E. D., & CRISTOFARO, R. D. (2010) · Biophysical Chemistry · doi:10.1016/j.bpc.2010.07.002 · Cited by 102
Taskin, M. E., Fraser, K., Zhang, T., Gellman, B. N., Fleischli, A., Dasse, K. A., Griffith, B. P., & Wu, Z. J. (2010) · Artificial Organs · doi:10.1111/j.1525-1594.2010.01017.x · Cited by 97
Chiu, W.-C., Slepian, M. J., & Bluestein, D. (2014) · ASAIO Journal · doi:10.1097/mat.0000000000000034 · Cited by 66
Current Status of Left Ventricular Assist Device Technology
Molina, E., & Boyce, S. W. (2013) · Seminars in Thoracic and Cardiovascular Surgery · doi:10.1053/j.semtcvs.2013.02.002 · Cited by 43
Metadata
- Authors
- Thamsen, B., Blümel, B., Schaller, J., Paschereit, C. O., Affeld, K., Goubergrits, L., & Kertzscher, U.
- Venue
- Artificial Organs
- Publisher
- Wiley
- Published
- 2015-08
- Volume
- 39
- Issue
- 8
- Pages
- 651-659
- ISSN
- 0160-564X (ISSN-L)
- 1525-1594
- OpenAlex
- W2113003448
- Cited by
- 191
- References
- 22
- FWCI
- 7.86
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.