Insulin induces a progressive increase in the resistance of subcutaneous tissue to fluid flow: Implications for insulin pump therapy
Open accessAbstract
Aim: To determine the effect of insulin on the resistance of subcutaneous tissue to the flow of infusion fluids. Materials and methods: Thirty subjects with type 1 diabetes wore two Accu‐Chek Spirit Combo insulin pumps with Accu‐Chek FlexLink infusion sets (Roche Diabetes Care, Mannheim, Germany) for 7 days. One pump was filled with insulin aspart (Novo Nordisk, Bagsvaerd, Denmark) and used for continuous subcutaneous insulin infusion (CSII). The other pump was filled with insulin diluting medium (IDM; Novo Nordisk) and used to deliver IDM subcutaneously at rates identical to those employed for CSII. Both infusion sites were assessed daily by measuring the pressure required to infuse various bolus amounts of IDM. Results: On day 1, maximum pressure (P max ) and tissue flow resistance (TFR; calculated from measured pressure profiles) were similar for both infusion sites ( P > 0.20). During the subsequent study days, the P max and TFR values observed at the IDM infusion site remained at levels comparable to those seen on day 1 ( P > 0.13). However, at the site of CSII, P max and TFR progressively increased with CSII duration. By the end of day 7, P max and TFR reached 25.8 */2.11 kPa (geometric mean */geometric standard deviation) and 8.64 */3.48 kPa*s/μL, respectively, representing a remarkable 3.5‐ and 20.6‐fold increase relative to the respective P max and TFR values observed on day 1 ( P < 0.001). Conclusion: Our results suggest that insulin induces a progressive increase in the resistance of subcutaneous tissue to the introduction of fluid; this has important implications for the future design of insulin pumps and infusion sets.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
SIND may occur as frequently as once per month and is detected more frequently by chemical measurement of (or by symptoms of) hyperglycemia or DKA than by pump alerts or alarms.
“Survey data and pump downloads reveal that SIND may occur as frequently as once per month and is detected more frequently by chemical measurement of (or by symptoms of) hyperglycemia or DKA than by pump alerts or alarms [22], [23].”
From Performance of a Continuous Subcutaneous Insulin Infusion (CSII) Pump With Acoustic Volume and Flow Sensing in Simulated High-Consequence Situations (2024) · cites this paper as [23]
The Discussion (page unavailable) says infusion-site leakage often goes unnoticed until hyperglycaemia is detected because current pumps lack leakage-detection alarms. It does not establish that SIND occurs once per month, report survey or pump-download findings, or compare detection by DKA symptoms with pump alerts across SIND events. (LLM, conf=0.83)
Experimental — accuracy not yet measuredCurrent occlusion alarm systems cannot detect other types of infusion failures such as leakages, dislodged cannulas, and silent occlusions.
“However, these detection thresholds are not standardized across manufacturers and cannot detect other types of infusion failures such as leakages, dislodged cannulas, and “silent” (ie, no alert) occlusions.15,16”
From Development of a Real-time Force-based Algorithm for Infusion Failure Detection (2024) · cites this paper as [15]
The Discussion (page unavailable) says current pumps have no leakage-detection alarms and that most cannula malplacements may go unrecognized for hours. It also says some malplacements are eventually recognized by high-pressure alarms, so an absolute inability to detect them overstates the source. The paper does not address “silent” occlusions. (LLM, conf=0.86)
Experimental — accuracy not yet measuredOcclusion detection thresholds are not standardized across manufacturers.
“However, these detection thresholds are not standardized across manufacturers and cannot detect other types of infusion failures such as leakages, dislodged cannulas, and “silent” (ie, no alert) occlusions.15,16”
From Development of a Real-time Force-based Algorithm for Infusion Failure Detection (2024) · cites this paper as [15]
Testing of retrofitted pressure sensors in CSII pumps suggests that the incorporation of more precise fluid monitoring could improve infusion failure detection.
“Alternatively, the testing of retrofitted pressure sensors in CSII pumps suggests that the incorporation of more precise fluid monitoring could improve the infusion failure detection.15”
From Development of a Real-time Force-based Algorithm for Infusion Failure Detection (2024) · cites this paper as [15]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Log-normal distributions across the sciences: keys and clues
LIMPERT, E., STAHEL, W. A., & ABBT, M. (2001) · BioScience · doi:10.1641/0006-3568(2001)051[0341:lndats]2.0.co;2 · Cited by 2,621 · see citing works
Insulin-pump therapy for type 1 diabetes mellitus
Pickup, J. C. (2012) · New England Journal of Medicine · doi:10.1056/nejmct1113948 · Cited by 270 · see citing works
Nonmetabolic Complications of Continuous Subcutaneous Insulin Infusion: A Patient Survey
Pickup, J. C., Yemane, N., Brackenridge, A., & Pender, S. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0192 · Cited by 123 · see citing works
Randomized Trial of Infusion Set Function: Steel Versus Teflon
Patel, P. J., Benasi, K., Ferrari, G., Evans, M. G., Shanmugham, S., Wilson, D. M., & Buckingham, B. A. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0119 · Cited by 90 · see citing works
Occlusion Detection Time in Insulin Pumps at Two Different Basal Rates
Freckmann, G., Kamecke, U., Waldenmaier, D., Haug, C., & Ziegler, R. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817750404 · Cited by 18 · see citing works
Flagged first, then by citations
Research works printed but not linked: 27
- [1]Beck RW, Bergenstal RM, Laffel LM, Pickup JC. Advances in technology for management of type 1 diabetes. Lancet. 2019; 394: 1265‐ 1273. 31533908
- [2]Tauschmann M, Hovorka R. Technology in the management of type 1 diabetes mellitus ‐ current status and future prospects. Nat Rev Endocrinol. 2018; 14: 464‐ 475. 29946127
- [4]Philip BK, Philip JH. Characterization of flow in intravenous catheters. IEEE Trans Biomed Eng. 1986; 33: 529‐ 531. 3710511
- [5]Philip JH, Philip BK. Simplified approach to assessing intravenous flow characteristics in the therapeutic infusion range. IEEE Trans Biomed Eng. 1988; 35: 1093‐ 1094. 3220504
- [6]Levick JR. Flow through interstitium and other fibrous matrices. Q J Exp Physiol. 1987; 72: 409‐ 437. 3321140
- [7]Swartz MA, Fleury ME. Interstitial flow and its effects in soft tissues. Annu Rev Biomed Eng. 2007; 9: 229‐ 256. 17459001
- [8]Richter WF, Bhansali SG, Morris ME. Mechanistic determinants of biotherapeutics absorption following SC administration. AAPS J. 2012; 14: 559‐ 570. 22619041 PMC3385825
- [9]Søeborg T, Rasmussen CH, Mosekilde E, Colding‐Jørgensen M. Bioavailability and variability of biphasic insulin mixtures. Eur J Pharm Sci. 2012; 46: 198‐ 208. 21703346
- [11]Højbjerre L, Skov‐Jensen C, Kaastrup P, Pedersen PE, Stallknecht B. Effect of steel and teflon infusion catheters on subcutaneous adipose tissue blood flow and infusion counter pressure in humans. Diabetes Technol Ther. 2009; 15: 301‐ 306. 19425878
- [12]Patte C, Pleus S, Wiegel C, et al. Effect of infusion rate and indwelling time on tissue resistance pressure in small‐volume subcutaneous infusion like in continuous subcutaneous insulin infusion. Diabetes Technol Ther. 2013; 15: 289‐ 294. 23427865
- [16]Greenwood JA, Sandomire MM. Sample size required for estimating the standard deviation as a per cent of its true value. J Am Stat Assoc. 1950; 45: 257‐ 260.
- [17]Burnett RW. Accurate estimation of standard deviations for quantitative methods used in clinical chemistry. Clin Chem. 1975; 21: 1935‐ 1938. 1192586
- [18]Brange J. Galenics of Insulin. Berlin, Heidelberg, New York: Springer‐Verlag; 1987.
- [19]Hrynyk M, Neufeld RJ. Insulin and wound healing. Burns. 2014; 40: 1433‐ 1446. 24810536
- [20]Wang J, Xu J. Effects of topical insulin on wound healing – a review of animal and human evidences. Diabetes Metab Syndr Obes. 2020; 13: 719‐ 727. 32214835 PMC7078652
- [21]Swinney MR, Cox AL, Hawkins ED, et al. Insulin, not the preservative m‐cresol, instigates loss of infusion site patency over extended durations of CSII in diabetic swine. J Pharm Sci. 2020; 110: 1418‐ 1426. 33321138
- [22]Karlin AW, Ly TT, Pyle L, et al. Duration of infusion set survival in lipohypertrophy versus nonlipohypertrophied tissue in patients with type 1 diabetes. Diabetes Technol Ther. 2016; 18: 429‐ 435. 27227290 PMC4931738
- [23]Hirsch L, Byron K, Gibney M. Intramuscular risk at insulin injection sites – measurement of the distance from skin to muscle and rationale for shorter‐length needles for subcutaneous insulin therapy. Diabetes Technol Ther. 2014; 16: 867‐ 873. 25329935
- [24]Renard E, Guerci B, Leguerrier AM, Boizel R. Lower rate of initial failures and reduced occurrence of adverse events with a new catheter model for continuous subcutaneous insulin infusion: prospective, two‐period, observational, multicenter study. Diabetes Technol Ther. 2010; 12: 769‐ 773. 20809682
- [27]Welsh JB, Vargas S, Williams G, Moberg S. Designing the modern pump – engineering aspects of continuous subcutaneous insulin infusion software. Diabetes Technol Ther. 2010; 12(Suppl 1): S37‐ S42. 20515305
- [34]Zhang JY, Shang T, Chattaraj S, et al. Advances in insulin pump infusion sets symposium report. J Diabetes Sci Technol. 2021; 15: 705‐ 709. 33703930 PMC8120061
- [35]Melberg SG, Havelund S, Villumsen J, Brange J. Insulin compatibility with polymer materials used in external pump infusion systems. Diabetic Med. 1988; 5: 243‐ 247. 2967145
- [36]Slominski AT, Zmijewski MA. Glucocorticoids inhibit wound healing – novel mechanism of action. J Invest Dermatol. 2017; 137: 1012‐ 1014. 28411834 PMC5412951
- [37]Lal RA, Hsu L, Zhang J, et al. Longevity of the novel ConvaTec infusion set with lantern technology. Diabetes Obes Metab. 2021; 23: 1973‐ 1977. 33822472 PMC8720264
- [38]Waldenmaier D, Zschornack E, Buhr A, Pleus S, Haug C, Freckmann G. A prospective study of insulin infusion set use for up to 7 days – early replacement reasons and impact on glycemic control. Diabetes Technol Ther. 2020; 22: 734‐ 741. 32167382
- [39]Odeh RE, Chou YM, Owen DB. The precision for coverages and sample size requirements for normal tolerance intervals. Commun Stat‐Simul C. 1987; 16: 969‐ 985.
- [40]Grunberger G, Bhargava A, Ly T, et al. Human regular U‐500 insulin via continuous subcutaneous insulin infusion versus multiple daily injections in adults with type 2 diabetes – the VIVID study. Diabetes Obes Metab. 2020; 22: 434‐ 441. 31865633 PMC7065168
Cited by
Sorted by citations
Butterfield, R. D., & Sims, N. M. (2024) · IEEE Open Journal of Engineering in Medicine and Biology · doi:10.1109/ojemb.2024.3408092 · Cited by 6 · see citing works
Development of a Real-time Force-based Algorithm for Infusion Failure Detection
Blanco, L. E., Wilcox, J. H., Hughes, M. S., & Lal, R. A. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241247530 · Cited by 2 · see citing works
- Authors
- Regittnig, W., Tschaikner, M., Tuca, A., Simic, A., Feiel, J., Schaller‐Ammann, R., Licht, A. H., Jungklaus, M., & Pieber, T. R.
- Affiliations
- Medical University of Graz (AT)
- Joanneum Research (AT)
- Roche Diabetes Care GmbH (Germany) (DE)
- Corresponding author
- Regittnig, Werner
- Venue
- Diabetes, Obesity and Metabolism
- Publisher
- Wiley
- Published
- 2021-11-17
- Volume
- 24
- Issue
- 3
- Pages
- 455-464
- ISSN
- 1462-8902 (ISSN-L)
- 1463-1326
- PMC
- PMC9299465
- OpenAlex
- W3209858923
- Cited by
- 27
- References
- 5
- Citations by year
- 2025: 2 · 2024: 7 · 2023: 2 · 2022: 15 · 2021: 1
- FWCI
- 4.67
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Diabetes Obesity and Metabolism (publishedVersion, cc-by-nc-nd)
- PubMed Central (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- Funders
- Medizinische Universität Graz
- Roche
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
- Health Sciences › Medicine › Dermatology › Skin Diseases and Diabetes
- Keywords
- insulin, insulin pump therapy, insulin aspart, infusion sets, type 1 diabetes, infusion pressure
- MeSH terms
- Diabetes Mellitus, Type 1, Humans, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Subcutaneous Tissue
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review