Occlusion Detection Time in Insulin Pumps at Two Different Basal Rates
Open accessAbstract
Background: The detection of insulin infusion set (IIS) occlusions is an important feature of insulin pumps with regard to patient safety. However, there are no requirements for a time limit until an alarm has to be triggered after an occlusion occurred. The standard IEC 60601-2-24 is applicable for insulin pumps and describes test settings and procedures to determine occlusion detection time (ODT). Methods: In this study, ODT of six different insulin pump models with different IIS (in total 10 different insulin pump systems) was tested for two basal rates (1.0 U/h and 0.1 U/h). Results: Differences were seen between the tested pump systems. At a basal rate of 1.0 U/h all insulin pump systems showed an acceptable ODT of less than 5 hours. However, at a basal rate of 0.1 U/h, as often used in children, the median ODT ranged from approximately 4 hours to more than 40 hours. With the lower basal rate, median ODT was longer than 6-8 hours for 9 of the 10 systems. Conclusions: Insulin pump users should not blindly rely on occlusion alarms but perform regular glucose monitoring and manufacturers should develop mechanisms that allow an earlier detection at low basal rates.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Occlusion alarms can take time to trigger depending on the pump's infusion rate, with delays up to 40 hours post-failure.
“Furthermore, occlusion alarms can take time to trigger depending on the pump's infusion rate, with delays up to 40 hours post-failure.17”
From Development of a Real-time Force-based Algorithm for Infusion Failure Detection (2024) · cites this paper as [17]
The Results report that occlusion detection took substantially longer at 0.1 U/h than at 1.0 U/h, supporting the infusion-rate dependence. But “up to 40 hours” understates the observed delays: the longest median was 41 hours 21 minutes, and an individual alarm took 65 hours 40 minutes. (LLM, conf=0.87)
Experimental — accuracy not yet measuredCurrent insulin pump systems exhibit substantial variability in occlusion detection times, with delayed alarms-especially at lower basal rates-heightening the risk of unnoticed infusion failures and prolonged hyperglycemic excursions.
“As demonstrated by Freckmann et al, current insulin pump systems exhibit substantial variability in occlusion detection times, with delayed alarms—especially at lower basal rates—heightening the risk of unnoticed infusion failures and prolonged hyperglycemic excursions.23”
From Clinical utility of the twiist Automated Insulin Delivery System: first study with twiist (2026) · cites this paper as [23]
The prototype's occlusion detection time was similar or better (distinctly shorter) than that of commercially available pumps tested in comparable settings.
“Our study results suggest that the tested pump prototype provides delivery accuracy and occlusion detection that is similar or even better compared to all commercially available pumps tested with a similar experimental approach.”
From Accuracy Evaluation of a Novel Reusable Patch Pump Prototype (2023) · cites this paper as [4]
Experimental testing was based on IEC 60601-2-24 and on procedures published previously.
“Experimental testing was based on IEC 60601-2-24 and on procedures published previously. 3-5”
From Accuracy Evaluation of a Novel Reusable Patch Pump Prototype (2023) · cites this paper as [4]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667 · see citing works
Hyperglycemic crises in adult patients with diabetes
Kitabchi, A. E., Umpierrez, G. E., Miles, J. M., & Fisher, J. N. (2009) · Diabetes Care · doi:10.2337/dc09-9032 · Cited by 2,062 · see citing works
Attia, N. A., Jones, T. W., Holcombe, J. H., & Tamborlane, W. V. (1998) · Diabetes Care · doi:10.2337/diacare.21.5.817 · Cited by 98 · see citing works
Hanas, R., Lindgren, F., & Lindblad, B. (2009) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2008.00441.x · Cited by 93 · see citing works
Hypoglycemia and ketoacidosis with insulin pump therapy in children and adolescents
Hanås, R., & Ludvigsson, J. (2006) · Pediatric Diabetes · doi:10.1111/j.1399-543x.2006.00169.x · Cited by 91 · see citing works
Insulin Infusion Set Use: European Perspectives and Recommendations
Deiss, D., Adolfsson, P., Alkemade-van Zomeren, M., Bolli, G. B., Charpentier, G., Cobelli, C., Danne, T., Girelli, A., Mueller, H., Verderese, C. A., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.07281.sf · Cited by 62 · see citing works
Borot, S., Franc, S., Cristante, J., Penfornis, A., Benhamou, P.-Y., Guerci, B., Hanaire, H., Renard, E., Reznik, Y., Simon, C., et al. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814543946 · Cited by 47 · see citing works
Insulin Pump Occlusions: For Patients Who Have Been Around the (Infusion) Block
Klonoff, D. C., Freckmann, G., & Heinemann, L. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817700545 · Cited by 41 · cited 1× · see citing works
“In addition, manufacturers should develop IIS less prone to occlusions and mechanisms that allow for a timely detection of occlusions.(14) Abbreviations”
BackgroundSelf-citationReichel, A., Rietzsch, H., Köhler, H., Pfützner, A. H., Gudat, U., & Schulze, J. (2009) · Experimental and Clinical Endocrinology & Diabetes · doi:10.1055/s-0029-1211971 · Cited by 33 · see citing works
Gibney, M., Xue, Z., Swinney, M., Bialonczyk, D., & Hirsch, L. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0342 · Cited by 32 · see citing works
van Bon, A. C., Dragt, D., & DeVries, J. H. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0259 · Cited by 25 · see citing works
Basal Rate Subcutaneous Insulin Infusion: Absorption Kinetics and Relation to Local Blood Flow
Hildebrandt, P. R., & Birch, K. (1988) · Diabetic Medicine · doi:10.1111/j.1464-5491.1988.tb01023.x · Cited by 20 · see citing works
Flagged first, then by citations
Research works printed but not linked: 2
- [2]2. IEC 60601-2-24:2012. Medical electrical equipment—part 2-24: particular requirements for the basic safety and essential performance of infusion pumps and controllers. 2012. [Google Scholar]
- [3]3. IEC 60601-2-24:1998. Medical electrical equipment—part 2-24: particular requirements for the safety of infusion pumps and controllers. 1998. [Google Scholar]
Cited by
Sorted by citations
Bode, B. W., Garg, S. K., Norwood, P. C., Morales, C., Hardy, T. A., Liu, R., & Ignaut, D. A. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.0224 · Cited by 34 · see citing works
Regittnig, W., Tschaikner, M., Tuca, A., Simic, A., Feiel, J., Schaller‐Ammann, R., Licht, A. H., Jungklaus, M., & Pieber, T. R. (2022) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.14594 · Cited by 27 · see citing works
Waldenmaier, D., Zschornack, E., Buhr, A., Pleus, S., Haug, C., & Freckmann, G. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0445 · Cited by 21 · see citing works
Self-citationEvaluation of the Accuracy of Current Tubeless Pumps for Continuous Subcutaneous Insulin Infusion
Ziegler, R., Oliver, N., Waldenmaier, D., Mende, J., Haug, C., & Freckmann, G. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.0525 · Cited by 11 · see citing works
Self-citationHong, K. Y., Lee, E. K., Kim, Y., Choi, D.-C., & Min, J. J. (2023) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-023-00973-y · Cited by 7 · see citing works
Modelling and simulation of occlusions in insulin pumps
Formo, M. W., Stavdahl, O., & Fougner, A. L. (2021) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc46164.2021.9630219 · Cited by 5 · see citing works
Development of a Real-time Force-based Algorithm for Infusion Failure Detection
Blanco, L. E., Wilcox, J. H., Hughes, M. S., & Lal, R. A. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241247530 · Cited by 2 · see citing works
Modeling and Simulation of Insulin Pump Set Failures Based on the Power Flow Method
Wang, W., Guo, D., Quan, L., & Liu, S. (2025) · IEEE Access · doi:10.1109/access.2025.3554261 · Cited by 2 · see citing works
Chappel, E. (2023) · Global Journal of Engineering and Technology Advances · doi:10.30574/gjeta.2023.16.3.0181 · Cited by 1 · see citing works
Accuracy Evaluation of a Novel Reusable Patch Pump Prototype
Mende, J., Baumstark, A., Waldenmaier, D., Amirouche, A., Barraud, A., Fridez, P., Haug, C., & Freckmann, G. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221097997 · Cited by 1 · see citing works
Self-citationKetone-Based Alert System for Insulin Pump Failures
Aiello, E. M., Laffel, L. M., Patti, M., & Doyle, F. J. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231209339 · Cited by 1 · see citing works
Ibrahim, M., Beneyto, A., Contreras, I., & Vehı́, J. (2023) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2023.10.441 · Cited by 1 · see citing works
Mende, J., Eichenlaub, M., Waldenmaier, D., Haug, C., Freckmann, G., & Ziegler, R. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211070815 · Cited by 1 · see citing works
Self-citationClinical utility of the twiist Automated Insulin Delivery System: first study with twiist
Ruiz, A., Mcdonough, A., Snow, K., Bode, B., & Mitri, J. (2026) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2026-006307
Chappel, E. (2024) · Encyclopedia · doi:10.3390/encyclopedia4020052
Insulin Delivery Hardware: Pumps and Pens
Lal, R., & Leelarathna, L. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2025.8803.rl
Contemporary methods and problems of glucose monitoring in diabetes
Friedecky, B., & Kratochvíla, J. (2018) · Klinická biochemie a metabolismus · doi:10.61568/kbm.2018.023
Kontinuierliche Glukose- und Ketonmessung: Perspektiven für modernes Diabetesmanagement
Kapellen, T. M., Brinkmann, C. K., Freckmann, G., Gölz, S., Tan, S., Seufert, J., Siegmund, T., Ziegler, R., & Szendrödi, J. (2026) · Diabetologie und Stoffwechsel · doi:10.1055/a-2896-0531
Self-citation
- Authors
- Freckmann, G., Kamecke, U., Waldenmaier, D., Haug, C., & Ziegler, R.
- Affiliations
- Universität Ulm (DE)
- Diabetes Clinic for Children and Adolescents, Muenster, Germany
- Corresponding author
- Freckmann, Guido
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2017-12-29
- Volume
- 12
- Issue
- 3
- Pages
- 608-613
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC6154229
- OpenAlex
- W2781380700
- Cited by
- 18
- References
- 12
- Citations by year
- 2026: 2 · 2025: 2 · 2024: 2 · 2023: 4 · 2022: 2 · 2021: 2 · 2020: 3 · 2018: 1
- FWCI
- 0.87
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
- Health Sciences › Health Professions › Emergency Medical Services › Patient Safety and Medication Errors
- Keywords
- insulin pump, patient safety, children, glucose monitoring
- MeSH terms
- Equipment Failure, Humans, Hypoglycemic Agents, Insulin, Insulin Infusion Systems
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review