Occlusion Detection Time in Insulin Pumps at Two Different Basal Rates

Open access
Freckmann, Guido, Kamecke, Ulrike, Waldenmaier, Delia, Haug, Cornelia, Ziegler, Ralph·2018·Journal of Diabetes Science and Technology

Abstract

Background: The detection of insulin infusion set (IIS) occlusions is an important feature of insulin pumps with regard to patient safety. However, there are no requirements for a time limit until an alarm has to be triggered after an occlusion occurred. The standard IEC 60601-2-24 is applicable for insulin pumps and describes test settings and procedures to determine occlusion detection time (ODT). Methods: In this study, ODT of six different insulin pump models with different IIS (in total 10 different insulin pump systems) was tested for two basal rates (1.0 U/h and 0.1 U/h). Results: Differences were seen between the tested pump systems. At a basal rate of 1.0 U/h all insulin pump systems showed an acceptable ODT of less than 5 hours. However, at a basal rate of 0.1 U/h, as often used in children, the median ODT ranged from approximately 4 hours to more than 40 hours. With the lower basal rate, median ODT was longer than 6-8 hours for 9 of the 10 systems. Conclusions: Insulin pump users should not blindly rely on occlusion alarms but perform regular glucose monitoring and manufacturers should develop mechanisms that allow an earlier detection at low basal rates.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1177/1932296817750404

Cites

12 of 12 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Occlusion alarms can take time to trigger depending on the pump's infusion rate, with delays up to 40 hours post-failure.

    “Furthermore, occlusion alarms can take time to trigger depending on the pump's infusion rate, with delays up to 40 hours post-failure.17”

    From Development of a Real-time Force-based Algorithm for Infusion Failure Detection (2024) · cites this paper as [17]

    The Results report that occlusion detection took substantially longer at 0.1 U/h than at 1.0 U/h, supporting the infusion-rate dependence. But “up to 40 hours” understates the observed delays: the longest median was 41 hours 21 minutes, and an individual alarm took 65 hours 40 minutes. (LLM, conf=0.87)

    Experimental — accuracy not yet measured
  • Current insulin pump systems exhibit substantial variability in occlusion detection times, with delayed alarms-especially at lower basal rates-heightening the risk of unnoticed infusion failures and prolonged hyperglycemic excursions.

    “As demonstrated by Freckmann et al, current insulin pump systems exhibit substantial variability in occlusion detection times, with delayed alarms—especially at lower basal rates—heightening the risk of unnoticed infusion failures and prolonged hyperglycemic excursions.23”
  • The prototype's occlusion detection time was similar or better (distinctly shorter) than that of commercially available pumps tested in comparable settings.

    “Our study results suggest that the tested pump prototype provides delivery accuracy and occlusion detection that is similar or even better compared to all commercially available pumps tested with a similar experimental approach.”
  • Experimental testing was based on IEC 60601-2-24 and on procedures published previously.

    “Experimental testing was based on IEC 60601-2-24 and on procedures published previously. 3-5”

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 2

  1. [2]2. IEC 60601-2-24:2012. Medical electrical equipment—part 2-24: particular requirements for the basic safety and essential performance of infusion pumps and controllers. 2012. [Google Scholar]
  2. [3]3. IEC 60601-2-24:1998. Medical electrical equipment—part 2-24: particular requirements for the safety of infusion pumps and controllers. 1998. [Google Scholar]

Cited by

Sorted by citations

Authors
Freckmann, G., Kamecke, U., Waldenmaier, D., Haug, C., & Ziegler, R.
Affiliations
  • Universität Ulm (DE)
  • Diabetes Clinic for Children and Adolescents, Muenster, Germany
Corresponding author
Freckmann, Guido
Venue
Journal of Diabetes Science and Technology
Published
2017-12-29
Volume
12
Issue
3
Pages
608-613
ISSN
1932-2968 (ISSN-L)
OpenAlex
W2781380700
Cited by
18
References
12
Citations by year
2026: 2 · 2025: 2 · 2024: 2 · 2023: 4 · 2022: 2 · 2021: 2 · 2020: 3 · 2018: 1
FWCI
0.87
Type
journal article
Language
en
Open access
Locations
  • Journal of Diabetes Science and Technology (publishedVersion)
  • Europe PMC (PubMed Central) (publishedVersion)
  • PubMed Central (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
  • Health Sciences › Health Professions › Emergency Medical Services › Patient Safety and Medication Errors
Keywords
insulin pump, patient safety, children, glucose monitoring
MeSH terms
Equipment Failure, Humans, Hypoglycemic Agents, Insulin, Insulin Infusion Systems

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review