Comparison of electrical velocimetry and thermodilution techniques for the measurement of cardiac output
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
Aim: To compare a new method of non‐invasive determination of cardiac output based on electrical velocimetry (EV‐CO) with invasive thermodilution methods. Methods: Fifty critically ill patients were enrolled into the study. EV‐CO was compared with cardiac output measured by a pulmonary artery catheter (PA‐CO) in one group (n= 25) and by a femoral artery catheter (PiCCO‐CO) in a second group (n= 25), by simultaneous measurements. Standard electrocardiography electrodes were used for non‐invasive measurements, and EV‐CO was calculated using the Bernstein–Osypka equation. The invasive measurements of PA‐CO and PiCCO‐CO were made by the injection of iced 0.9% saline and the recording of thermodilution curves. Results: The precision values of EV‐CO, PA‐CO and PiCCO‐CO measurements were ± 0.46 [95% confidence interval (95% CI), ± 0.06], ± 0.57 (95% CI, ± 0.09) and ± 0.48 l/min (95% CI, ± 0.08 l/min), respectively. The mean differences between EV‐CO and PA‐CO or PiCCO‐CO were –0.05 ± 0.71 and 0.22 ± 0.78 l/min, respectively. The lower and upper limits of agreement for the comparison of EV‐CO with PA‐CO were −1.47 and 1.37 l/min (95% CI, ± 0.25 l/min), respectively. In the comparison of EV‐CO and PiCCO‐CO, lower and upper limits of –1.34 and 1.78 l/min (95% CI, ± 0.27 l/min) were found. The percentage errors between EV‐CO and PA‐CO or PiCCO‐CO were 26.5% and 26.4%, respectively. Conclusions: The values of cardiac output were statistically comparable between the groups. Therefore, electrical velocimetry is a suitable method to evaluate haemodynamic variables with clinically acceptable accuracy.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Cited by
Peyton, P. J., & Chong, S. W. (2010) · Anesthesiology · doi:10.1097/aln.0b013e3181ee3130 · Cited by 456
Use of Central Venous Oxygen Saturation to Guide Therapy
Walley, K. R. (2011) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.201010-1584ci · Cited by 242
“To fulfill the perceived clinical need for similar measurements a number of new technologies have emerged, including echocardiography (6), ultrasonic cardiac output monitors (7), arterial pulse pressure analysis techniques (8, 9), impedance techniques (10, 11), alternative dye dilution techniques (12), and so on, to p…”
BackgroundNoori, S., Drabu, B., Soleymani, S., & Seri, I. (2012) · Archives of Disease in Childhood - Fetal and Neonatal Edition · doi:10.1136/fetalneonatal-2011-301090 · Cited by 146
Non-Invasive Cardiac Output Measurement in Low and Very Low Birth Weight: A Method Comparison
Grollmuss, O., & Gonzalez, P. (2014) · Frontiers in Pediatrics · doi:10.3389/fped.2014.00016 · Cited by 52
Narula, J., Chauhan, S., Ramakrishnan, S., & Gupta, S. K. (2017) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1053/j.jvca.2016.12.009 · Cited by 39
New technologies in pediatric anesthesia
Holtby, H., Skowno, J. J., Kor, D. J., Flick, R. P., & Uezono, S. (2012) · Pediatric Anesthesia · doi:10.1111/pan.12007 · Cited by 35
Awadhare, P., Patel, R., McCallin, T., Mainali, K., Jackson, K., Starke, H., & Bhalala, U. (2022) · Frontiers in Pediatrics · doi:10.3389/fped.2022.857106 · Cited by 9
Sumbel, L., Nagaraju, L., Ogbeifun, H., Agarwal, A., & Bhalala, U. (2022) · Progress in Pediatric Cardiology · doi:10.1016/j.ppedcard.2022.101551 · Cited by 4
Metadata
- Authors
- Zoremba, N., Bickenbach, J., Krauss, B., Rossaint, R., Kuhlen, R., & Schälte, G.
- Venue
- Acta Anaesthesiologica Scandinavica
- Publisher
- Wiley
- Published
- 2007-10-18
- Volume
- 51
- Issue
- 10
- Pages
- 1314-1319
- ISSN
- 0001-5172 (ISSN-L)
- 1399-6576
- OpenAlex
- W2149297659
- Cited by
- 105
- References
- 1
- FWCI
- 3.67
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.