Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology
Open accessAbstract
Citation information : McAlinden C, Khadka J & Pesudovs K. Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology. Ophthalmic Physiol Opt 2011, 31 , 330–338. doi: 10.1111/j.1475‐1313.2011.00851.x The ever‐expanding choice of ocular metrology and imaging equipment has driven research into the validity of their measurements. Consequently, studies of the agreement between two instruments or clinical tests have proliferated in the ophthalmic literature. It is important that researchers apply the appropriate statistical tests in agreement studies. Correlation coefficients are hazardous and should be avoided. The ‘limits of agreement’ method originally proposed by Altman and Bland in 1983 is the statistical procedure of choice. Its step‐by‐step use and practical considerations in relation to optometry and ophthalmology are detailed in addition to sample size considerations and statistical approaches to precision (repeatability or reproducibility) estimates.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“This was consistent with recommendations from McAlinden et al that sample sizes for agreement studies should include at least 100 subjects(McAlinden et al., 2011).”
From Acceptability and Concordance of Self- Versus Clinician- Sampling for HPV Testing among Rural South Indian Women (2021) · cites this paper as McAlinden et al., 2011
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A Coefficient of Agreement for Nominal Scales
Cohen, J. (1960) · Educational and Psychological Measurement · doi:10.1177/001316446002000104 · Cited by 42,398
Intraclass correlations: Uses in assessing rater reliability.
Shrout, P. E., & Fleiss, J. L. (1979) · Psychological Bulletin · doi:10.1037/0033-2909.86.2.420 · Cited by 24,532
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619
A concordance correlation coefficient to evaluate reproducibility.
Lin, L. I.-K. (1989) · Biometrics · doi:10.2307/2532051 · Cited by 8,785
Scheffé, H. (1960) · Soil Science · doi:10.1097/00010694-196006000-00016 · Cited by 5,625
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Comprehensive Textbook of Psychiatry
Schwab, J. J. (1968) · Psychosomatics · doi:10.1016/s0033-3182(68)71813-2 · Cited by 1,291
Measurement reliability and agreement in psychiatry
Shrout, P. E. (1998) · Statistical Methods in Medical Research · doi:10.1177/096228029800700306 · Cited by 1,211
Clinical biostatistics: LIV. The biostatistics of concordance
Kramer, M. S., & Feinstein, A. R. (1981) · Clinical Pharmacology & Therapeutics · doi:10.1038/clpt.1981.18 · Cited by 1,087
Improving the Reliability of the Pelli-Robson Contrast Sensitivity Test
Bullimore, M. A., Elliott, D. B., & Bailey, I. L. (1991) · Noninvasive Assessment of the Visual System · doi:10.1364/navs.1991.md12 · Cited by 257
Cited by
Sorted by citations
Statistical guidelines for the analysis of data obtained from one or both eyes
Armstrong, R. A. (2012) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12009 · Cited by 538
Goligher, E. C., Laghi, F. A., Detsky, M. E., Farias, P., Murray, A., Brace, D., Brochard, L. J., Sebastien-Bolz, S., Rubenfeld, G. D., Kavanagh, B. P., et al. (2015) · Intensive Care Medicine · doi:10.1007/s00134-015-3687-3 · Cited by 451
Goligher, E. C., Laghi, F., Detsky, M. E., Farias, P., Murray, A., Brace, D., Brochard, L. J., Bolz, S.-S., Rubenfeld, G. D., Kavanagh, B. P., et al. (2015) · Intensive Care Medicine · doi:10.1007/s00134-015-3724-2 · Cited by 315
Sample Size for Assessing Agreement between Two Methods of Measurement by Bland−Altman Method
Lu, M., Zhong, W.-H., Liu, Y., Miao, H., Li, Y., & Ji, M. (2016) · The International Journal of Biostatistics · doi:10.1515/ijb-2015-0039 · Cited by 307
Comparison of subjective grading and objective assessment in meibography
Pult, H., & Riede-Pult, B. H. (2012) · Contact Lens and Anterior Eye · doi:10.1016/j.clae.2012.10.074 · Cited by 277
Precision (repeatability and reproducibility) studies and sample-size calculation
McAlinden, C., Khadka, J., & Pesudovs, K. (2015) · Journal of Cataract & Refractive Surgery · doi:10.1016/j.jcrs.2015.06.029 · Cited by 248
Self-citationReporting Standards for a Bland–Altman Agreement Analysis: A Review of Methodological Reviews
Gerke, O. (2020) · Diagnostics · doi:10.3390/diagnostics10050334 · Cited by 246
McAlinden, C., Khadka, J., & Pesudovs, K. (2011) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.10-7093 · Cited by 236
Self-citationCorneal Deformation Measurement Using Scheimpflug Noncontact Tonometry
Hon, Y., & Lam, A. K. C. (2012) · Optometry and Vision Science · doi:10.1097/opx.0b013e318279eb87 · Cited by 235
Exact Parametric Confidence Intervals for Bland‐Altman Limits of Agreement
Carkeet, A. (2015) · Optometry and Vision Science · doi:10.1097/opx.0000000000000513 · Cited by 216
Should Pearson's correlation coefficient be avoided?
Armstrong, R. A. (2019) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12636 · Cited by 203
Tian, L., Qu, J., zhang, X.-Y., & Sun, X. (2016) · Journal of Ophthalmology · doi:10.1155/2016/8013621 · Cited by 188
Jiang, Y., Yuan, Q., Lv, W., Xi, S., Huang, W., Sun, Z., Chen, H., Zhao, L., Liu, W., Hu, Y., et al. (2018) · Theranostics · doi:10.7150/thno.28018 · Cited by 125
Sánchez-Cano, A. I., Orduna, E., Segura, F. J., FUENTE, C. L. D. L., Cuenca, N., Abecia, E., & Pinilla, I. (2014) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2014.05.035 · Cited by 114
Ctori, I., & Huntjens, B. (2015) · PLoS ONE · doi:10.1371/journal.pone.0129005 · Cited by 112
The effect of topical atropine on the choroidal thickness of healthy children
Zhang, Z., Zhou, Y., Xie, Z., Chen, T., Gu, Y., Lu, S., & Wu, Z. (2016) · Scientific Reports · doi:10.1038/srep34936 · Cited by 111
The Effect of Cycloplegia on the Lenstar and the IOLMaster Biometry
Huang, J., McAlinden, C., Su, B., Pesudovs, K., Feng, Y., Hua, Y., Yang, F., Pan, C., Zhou, H., & Wang, Q. (2012) · Optometry and Vision Science · doi:10.1097/opx.0b013e3182772f4f · Cited by 104
Self-citationAberrations and Topography in Normal, Keratoconus‐Suspect, and Keratoconic Eyes
Gordon‐Shaag, A., Millodot, M., Ifrah, R., & Shneor, E. (2012) · Optometry and Vision Science · doi:10.1097/opx.0b013e318249d727 · Cited by 101
Repeatability of the Pentacam HR in Various Grades of Keratoconus
Kreps, E. O., Jiménez‐García, M., Issarti, I., Claerhout, I., Koppen, C., & Rozema, J. J. (2020) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2020.06.013 · Cited by 92
Hernández‐Camarena, J. C., Chirinos‐Saldaña, P., Navas, A., Ramírez-Miranda, A., Mota, A. D. L., Jiménez-Corona, A., & Graue-Hernández, E. O. (2014) · Journal of Refractive Surgery · doi:10.3928/1081597x-20140815-02 · Cited by 90
VISUAL FUNCTION MEASURES IN EARLY AND INTERMEDIATE AGE-RELATED MACULAR DEGENERATION
Chandramohan, A., Stinnett, S. S., Petrowski, J. T., Schuman, S. G., Toth, C. A., Cousins, S. W., & Lad, E. M. (2016) · Retina · doi:10.1097/iae.0000000000001002 · Cited by 89
Huang, J., Chen, H., Li, Y., Chen, Z., Gao, R., Yu, J., Zhao, Y., Lu, W., McAlinden, C., & Wang, Q. (2019) · Journal of Refractive Surgery · doi:10.3928/1081597x-20190109-01 · Cited by 89
Self-citationLiao, H., Zhang, Z., Chen, J., Liao, M., Xu, L., Wu, Z., Yuan, K., Song, B., & Zeng, Y. (2019) · Annals of Surgical Oncology · doi:10.1245/s10434-019-07815-9 · Cited by 90
Selective laser trabeculoplasty (SLT) vs other treatment modalities for glaucoma: systematic review
McAlinden, C. (2013) · Eye · doi:10.1038/eye.2013.267 · Cited by 87
Self-citationChen, W., McAlinden, C., Pesudovs, K., Wang, Q., Lü, F., Feng, Y., Chen, J., & Huang, J. (2012) · Journal of Cataract & Refractive Surgery · doi:10.1016/j.jcrs.2012.04.031 · Cited by 83
Self-citationPreliminary Report on a Novel Virtual Reality Perimeter Compared With Standard Automated Perimetry
Razeghinejad, R., Gonzalez-Garcia, A., Myers, J. S., & Katz, L. J. (2020) · Journal of Glaucoma · doi:10.1097/ijg.0000000000001670 · Cited by 81
Evaluation of a new rebound tonometer for self-measurement of intraocular pressure
Dabasia, P. L., Lawrenson, J. G., & Murdoch, I. (2015) · British Journal of Ophthalmology · doi:10.1136/bjophthalmol-2015-307674 · Cited by 81
Reproducibility of Optical Coherence Tomography Measurements in Children
Altemir, I., Pueyo, V., Elía, N., Polo, V., Larrosa, J. M., & Oros, D. (2012) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2012.06.012 · Cited by 76
Rozema, J. J., Wouters, K., Mathysen, D. G. P., & Tassignon, M. (2014) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2014.08.014 · Cited by 76
McAlinden, C., Wang, Q., Gao, R., Zhao, W., Yu, A., Li, Y., Guo, Y., & Huang, J. (2016) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2016.09.019 · Cited by 75
Self-citationAxial Length Measurement Failure Rates with the IOLMaster and Lenstar LS 900 in Eyes with Cataract
McAlinden, C., Wang, Q., Pesudovs, K., Yang, X., Bao, F., Yu, A., Lin, S., Feng, Y., & Huang, J. (2015) · PLoS ONE · doi:10.1371/journal.pone.0128929 · Cited by 74
Self-citationSchiano‐Lomoriello, D., Hoffer, K. J., Abicca, I., & Savini, G. (2021) · Scientific Reports · doi:10.1038/s41598-020-79674-4 · Cited by 69
The development and evaluation of the new Ocular Surface Disease Index-6
Pult, H., & Wolffsohn, J. S. (2019) · The Ocular Surface · doi:10.1016/j.jtos.2019.08.008 · Cited by 67
Sridhar, K., & Charles, A. L. (2019) · Food Chemistry · doi:10.1016/j.foodchem.2019.02.110 · Cited by 65
Rosenfield, M., Hue, J. E., Huang, R., & Bababekova, Y. (2011) · Ophthalmic and Physiological Optics · doi:10.1111/j.1475-1313.2011.00887.x · Cited by 64
Measurement of Intraocular Pressure by Patients With Glaucoma
Pronin, S., Brown, L., Megaw, R. D., & Tatham, A. J. (2017) · JAMA Ophthalmology · doi:10.1001/jamaophthalmol.2017.3151 · Cited by 61
Sebbag, L., Kass, P. H., & Maggs, D. J. (2015) · Journal of the American Veterinary Medical Association · doi:10.2460/javma.246.4.426 · Cited by 60
Zhao, J., Chen, Z., Zhou, Z., Ding, L., & Zhou, X. (2012) · Clinical and Experimental Optometry · doi:10.1111/j.1444-0938.2012.00793.x · Cited by 59
Wilson, S., Ctori, I., Shah, R., Suttle, C., & Conway, M. L. (2022) · Ophthalmic and Physiological Optics · doi:10.1111/opo.13022 · Cited by 61
The Quality of Vision Questionnaire
McAlinden, C., Skiadaresi, E., Gatinel, D., Cabot, F. A., Huang, J., & Pesudovs, K. (2013) · Optometry and Vision Science · doi:10.1097/opx.0b013e3182993856 · Cited by 58
Self-citationSubjective Quality of Vision Before and After Cataract Surgery
Skiadaresi, E., McAlinden, C., Pesudovs, K., Polizzi, S., Khadka, J., & Ravalico, G. (2012) · Archives of Ophthalmology · doi:10.1001/archophthalmol.2012.1603 · Cited by 57
Self-citationZhang, Z., Huang, X., Meng, X., Chen, T., Gu, Y., Wu, Y., & Wu, Z. (2017) · Scientific Reports · doi:10.1038/s41598-017-08174-9 · Cited by 57
Blue‐blocking Filters and Digital Eyestrain
Palavets, T., & Rosenfield, M. (2018) · Optometry and Vision Science · doi:10.1097/opx.0000000000001318 · Cited by 57
Masmali, A. M., Alrabiah, S. A., Alharbi, A., El-Hiti, G. A.-R., & Almubrad, T. (2014) · Eye & Contact Lens Science & Clinical Practice · doi:10.1097/icl.0000000000000002 · Cited by 56
Xu, Z., Hua, Y., Qiu, W., Li, G., & Wu, Q. E. (2018) · BMC Ophthalmology · doi:10.1186/s12886-018-0683-8 · Cited by 55
Carkeet, A., & Goh, Y. T. (2016) · Statistical Methods in Medical Research · doi:10.1177/0962280216665419 · Cited by 54
Evaluation of 6 biometers based on different optical technologies
Montés‐Micó, R. (2021) · Journal of Cataract & Refractive Surgery · doi:10.1097/j.jcrs.0000000000000690 · Cited by 53
Comparison of Horizontal Corneal Diameter Measurements Using the Orbscan IIz and Pentacam HR Systems
Salouti, R., Nowroozzadeh, M. H., Zamani, M., Ghoreyshi, M., & Khodaman, A. R. (2013) · Cornea · doi:10.1097/ico.0b013e3182a40786 · Cited by 52
Hua, Y., Qiu, W., Xiao, Q., & Wu, Q. E. (2018) · PLoS ONE · doi:10.1371/journal.pone.0193023 · Cited by 51
Comparing accommodative function between the dominant and non-dominant eye
Momeni‐Moghaddam, H., McAlinden, C., Azimi, A., Sobhani, M., & Skiadaresi, E. (2013) · Graefe s Archive for Clinical and Experimental Ophthalmology · doi:10.1007/s00417-013-2480-7 · Cited by 50
Self-citationFerrer‐Blasco, T., Domínguez‐Vicent, A., Esteve-Taboada, J. J., Aloy, M. Á., Adsuara, J. E., & Montés‐Micó, R. (2016) · Graefe s Archive for Clinical and Experimental Ophthalmology · doi:10.1007/s00417-016-3555-z · Cited by 50
Agreement of ocular biometric measurements in young healthy eyes between IOLMaster 700 and OA-2000
Liao, X., Peng, Y., Liu, B., Tan, Q., & Lan, C. (2020) · Scientific Reports · doi:10.1038/s41598-020-59919-y · Cited by 48
Bolzanni, H., Oriá, A. P., Raposo, A. C., & Sebbag, L. (2020) · Veterinary Ophthalmology · doi:10.1111/vop.12751 · Cited by 46
Contrast sensitivity and higher-order aberrations in Keratoconus subjects
Shneor, E., Pinero, D. P., & Doron, R. (2021) · Scientific Reports · doi:10.1038/s41598-021-92396-5 · Cited by 46
López‐Miguel, A., Martínez-Almeída, L., González-García, M. J., Coco-Martín, M. B., Sobrado‐Calvo, P., & Maldonado, M. J. (2012) · Journal of Cataract & Refractive Surgery · doi:10.1016/j.jcrs.2012.08.061 · Cited by 46
Valenzuela, E. D., Welsh, S., & Dubín, A. (2014) · Revista Brasileira de Terapia Intensiva · doi:10.5935/0103-507x.20140038 · Cited by 45
Feng, Y., Reinstein, D. Z., Nitter, T., Archer, T. J., McAlinden, C., Chen, X., Bertelsen, G., Utheim, T. P., & Stojanović, A. (2022) · Journal of Refractive Surgery · doi:10.3928/1081597x-20220414-01 · Cited by 44
Self-citationSelf-measurement with Icare HOME tonometer, patients’ feasibility and acceptability
Cvenkel, B., Velkovska, M. A., & Jordanova, V. D. (2019) · European Journal of Ophthalmology · doi:10.1177/1120672118823124 · Cited by 44
A double-blind test of blue-blocking filters on symptoms of digital eye strain
Rosenfield, M., Li, R. T., & Kirsch, N. T. (2020) · Work · doi:10.3233/wor-203086 · Cited by 44
Translation, Cultural Adaptation, and Rasch Analysis of the Visual Function (VF-14) Questionnaire
Khadka, J., Huang, J., Mollazadegan, K., Gao, R., Chen, H., Zhang, S., Wang, Q., & Pesudovs, K. (2014) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.14-14017 · Cited by 43
Self-citationDomínguez‐Vicent, A., Pérez‐Vives, C., Ferrer‐Blasco, T., García‐Lázaro, S., & Montés‐Micó, R. (2016) · International Journal of Ophthalmology · doi:10.18240/ijo.2016.07.20 · Cited by 43
Comparison of Two Main Orthokeratology Lens Designs in Efficacy and Safety for Myopia Control
Lu, W., Ning, R., Diao, K., Ding, Y., Chen, R., Zhou, L., Lian, Y., McAlinden, C., Sanders, F., Xia, F., et al. (2022) · Frontiers in Medicine · doi:10.3389/fmed.2022.798314 · Cited by 42
Self-citationAssessment of Tear Film Quality among Smokers Using Tear Ferning Patterns
Masmali, A. M., Al-Shehri, A., Alanazi, S. A., Abusharaha, A., Fagehi, R. A., & El-Hiti, G. A.-R. (2016) · Journal of Ophthalmology · doi:10.1155/2016/8154315 · Cited by 42
Assessment of subjective refraction with a clinical adaptive optics visual simulator
Hervella, L., Villegas, E. A., Prieto, P. M., & Artal, P. (2018) · Journal of Cataract & Refractive Surgery · doi:10.1016/j.jcrs.2018.08.022 · Cited by 42
An eye on the Shih Tzu dog: Ophthalmic examination findings and ocular surface diagnostics
Sebbag, L., Silva, A. P. S. M., Santos, Á. P. B., Raposo, A. C., & Oriá, A. P. (2022) · Veterinary Ophthalmology · doi:10.1111/vop.13022 · Cited by 42
Evaluation of the precision of contrast sensitivity function assessment on a tablet device
Dörr, M., Lesmes, L. A., Elze, T., Wang, H., Lu, Z., & Bex, P. J. (2017) · Scientific Reports · doi:10.1038/srep46706 · Cited by 41
Koefoed, V. F., Baste, V., Roumes, C., & Høvding, G. (2014) · Acta Ophthalmologica · doi:10.1111/aos.12487 · Cited by 39
Proptosis—Correlation and Agreement between Hertel Exophthalmometry and Computed Tomography
Ramli, N. M., Kala, S., Samsudin, A., Rahmat, K., & Abidin, Z. Z. (2015) · Orbit · doi:10.3109/01676830.2015.1057291 · Cited by 38
Agreement of white-to-white measurements with swept-source OCT, Scheimpflug and color LED devices
Tañá‐Rivero, P., Aguilar-Córcoles, S., Rodríguez‐Prats, J. L., Montés‐Micó, R., & Ruíz-Mesa, R. (2020) · International Ophthalmology · doi:10.1007/s10792-020-01552-9 · Cited by 38
Vaishampayan, V., Kapoor, A., & Gumfekar, S. P. (2023) · The Canadian Journal of Chemical Engineering · doi:10.1002/cjce.24915 · Cited by 37
Measurement properties of continuous text reading performance tests
Brussee, T., van Nispen, R. M. A., & van Rens, G. H. M. B. (2014) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12158 · Cited by 37
León, A., Medrano, S. M., & Rosenfield, M. (2012) · Ophthalmic and Physiological Optics · doi:10.1111/j.1475-1313.2012.00891.x · Cited by 37
Investigation of Ocular Tear Ferning in Controlled and Uncontrolled Diabetic Subjects
Masmali, A. M., Maeni, Y. A., El-Hiti, G. A.-R., Murphy, P. J., & Almubrad, T. (2017) · Eye & Contact Lens Science & Clinical Practice · doi:10.1097/icl.0000000000000419 · Cited by 37
Repeatability and Diurnal Variation of Tear Ferning Test
Masmali, A. M., Albahlal, J. M., El-Hiti, G. A.-R., Akhtar, S., Purslow, C., Murphy, P. J., & Almubrad, T. (2015) · Eye & Contact Lens Science & Clinical Practice · doi:10.1097/icl.0000000000000116 · Cited by 35
Enders, C., Baeurle, F., Lang, G. E., Dreyhaupt, J., Lang, G. K., Loidl, M., & Werner, J. U. (2019) · Ophthalmologica · doi:10.1159/000499114 · Cited by 35
A descriptive study of samples sizes used in agreement studies published in the PubMed repository
Han, O., Tan, H. W., Julious, S. A., Sutton, L., Jacques, R., Lee, E. C., Lewis, J., & Walters, S. J. (2022) · BMC Medical Research Methodology · doi:10.1186/s12874-022-01723-5 · Cited by 34
Repeatability and agreement of ocular biometry measurements: Aladdin versus Lenstar
McAlinden, C., Gao, R., Yu, A., Wang, X., Yang, J., Yu, Y., Chen, H., Wang, Q., & Huang, J. (2017) · British Journal of Ophthalmology · doi:10.1136/bjophthalmol-2016-309365 · Cited by 34
Self-citationMasmali, A. M., Alanazi, S. A., Alotaibi, A. G., Fagehi, R. A., Abusharaha, A., & El-Hiti, G. A.-R. (2019) · Clinical ophthalmology · doi:10.2147/opth.s201127 · Cited by 33
Ruíz-Mesa, R., Aguilar-Córcoles, S., Montés‐Micó, R., & Tañá‐Rivero, P. (2020) · Expert Review of Medical Devices · doi:10.1080/17434440.2020.1772050 · Cited by 33
Wang, W., Miao, Y., Savini, G., McAlinden, C., Chen, H., Hu, Q., Wang, Q., & Huang, J. (2017) · Scientific Reports · doi:10.1038/s41598-017-13800-7 · Cited by 33
Self-citationComparing Autorefractors for Measurement of Accommodation
Aldaba, M., López, S. G., Vilaseca, M., Pujol, J., & Arjona, M. (2015) · Optometry and Vision Science · doi:10.1097/opx.0000000000000685 · Cited by 32
Repeatability of ocular aberration measurements in patients with keratoconus
Jinabhai, A. N., Radhakrishnan, H., & O’Donnell, C. (2011) · Ophthalmic and Physiological Optics · doi:10.1111/j.1475-1313.2011.00868.x · Cited by 32
Repeatability and reproducibility of a new optical biometer in normal and keratoconic eyes
Yağcı, R., Güler, E., Kulak, A. E., Erdoğan, B. D., Balcı, M., & Hepşen, İ. F. (2014) · Journal of Cataract & Refractive Surgery · doi:10.1016/j.jcrs.2014.04.039 · Cited by 32
Madhivanan, P., Nishimura, H., Ravi, K., Pope, B., Coudray, M., Arun, A., Krupp, K., Jayakrishna, P., & Srinivas, V. (2021) · Asian Pacific Journal of Cancer Prevention · doi:10.31557/apjcp.2021.22.3.971 · Cited by 31
“This was consistent with recommendations from McAlinden et al that sample sizes for agreement studies should include at least 100 subjects(McAlinden et al., 2011).”
MethodViswanathan, D., Kumar, N. L., Males, J. J., & Graham, S. L. (2014) · Acta Ophthalmologica · doi:10.1111/aos.12622 · Cited by 31
Asgari, S., Hashemi, H., Jafarzadehpur, E., Mohamadi, A., Rezvan, F., & Fotouhi, A. (2016) · International Ophthalmology · doi:10.1007/s10792-016-0189-4 · Cited by 31
A comparison of plusoptiX A12 measurements with cycloplegic refraction
Fogel-Levin, M., Doron, R., Wygnanski-Jaffe, T., Ancri, O., & Ben-Zion, I. (2016) · Journal of American Association for Pediatric Ophthalmology and Strabismus · doi:10.1016/j.jaapos.2016.04.006 · Cited by 31
Esteve-Taboada, J. J., Domínguez‐Vicent, A., Monsálvez-Romín, D., Águila‐Carrasco, A. J. D., & Montés‐Micó, R. (2017) · Graefe s Archive for Clinical and Experimental Ophthalmology · doi:10.1007/s00417-017-3663-4 · Cited by 31
Herrera, L., NAVARRO, I. P., Sánchez-Cano, A. I., Pérez-García, D., Remón, L., Almenara, C., Caramello, C., Cristóbal, J. Á., & Pinilla, I. (2015) · Retina · doi:10.1097/iae.0000000000000636 · Cited by 30
Validation of refraction and anterior segment parameters by a new multi-diagnostic platform (VX120)
Gordon‐Shaag, A., Pinero, D. P., Kahloun, C., Markov, D., Parnes, T., Gantz, L., & Shneor, E. (2018) · Journal of Optometry · doi:10.1016/j.optom.2017.12.003 · Cited by 30
Phung, L., Gregori, N. Z., Ortiz, A. G., Shi, W., & Schiffman, J. (2015) · Retina · doi:10.1097/iae.0000000000000818 · Cited by 30
The Italian Catquest-9SF cataract questionnaire: translation, validation and application
Skiadaresi, E., Ravalico, G., Polizzi, S., Lundström, M., González‐Andrades, M., & McAlinden, C. (2016) · Eye and Vision · doi:10.1186/s40662-016-0043-9 · Cited by 30
Self-citationThe interaction between homatropine and optical blur on choroidal thickness
Sander, B. P., Collins, M. J., & Read, S. A. (2018) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12450 · Cited by 30
van Doorn, L. L. A., Evans, B. J. W., Edgar, D. F., & Fortuin, M. F. (2013) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12101 · Cited by 29
Tunset, A., Kjær, P., Chreiteh, S. S., & Jensen, T. S. (2013) · Chiropractic & Manual Therapies · doi:10.1186/2045-709x-21-26 · Cited by 29
McAlinden, C., Schwiegerling, J. T., Khadka, J., & Pesudovs, K. (2020) · Journal of Cataract & Refractive Surgery · doi:10.1097/j.jcrs.0000000000000084 · Cited by 28
Self-citationOrthokeratology and Contact Lens Quality of Life Questionnaire (OCL-QoL)
McAlinden, C., & Lipson, M. J. (2017) · Eye & Contact Lens Science & Clinical Practice · doi:10.1097/icl.0000000000000451 · Cited by 27
Self-citationNarayanan, D., Cheng, H., Tang, R. A., & Frishman, L. J. (2014) · Documenta Ophthalmologica · doi:10.1007/s10633-014-9467-5 · Cited by 27
Tajbakhsh, Z., Salouti, R., Nowroozzadeh, M. H., Mohammad, A. A., Tabatabaee, S., & Zamani, M. (2012) · Ophthalmic and Physiological Optics · doi:10.1111/j.1475-1313.2012.00942.x · Cited by 27
Rasch analysis of three dry eye questionnaires and correlates with objective clinical tests
McAlinden, C., Gao, R., Wang, Q., Zhu, S., Yang, J., Yu, A., Bron, A. J., & Huang, J. (2017) · The Ocular Surface · doi:10.1016/j.jtos.2017.01.005 · Cited by 27
Self-citation
Metadata
- Authors
- McAlinden, C., Khadka, J., & Pesudovs, K.
- Affiliations
- Flinders University (AU)
- Corresponding author
- McAlinden, Colm
- Venue
- Ophthalmic and Physiological Optics
- Published
- 2011-05-26
- Volume
- 31
- Issue
- 4
- Pages
- 330-338
- ISSN
- 0275-5408 (ISSN-L)
- 1475-1313
- OpenAlex
- W2126376404
- Cited by
- 436
- References
- 49
- Citations by year
- 2026: 22 · 2025: 24 · 2024: 33 · 2023: 30 · 2022: 37 · 2021: 42 · 2020: 47 · 2019: 33 · 2018: 22 · 2017: 28 · 2016: 27 · 2015: 27 · 2014: 20 · 2013: 15 · 2012: 22
- FWCI
- 5.01
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Ophthalmic and Physiological Optics (publishedVersion)
- PubMed (publishedVersion)
- Adelaide Research & Scholarship (AR&S) (University of Adelaide) (submittedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Health Sciences › Medicine › Ophthalmology › Ophthalmology and Visual Impairment Studies
- Keywords
- repeatability, reproducibility, limits of agreement, correlation coefficient, ophthalmic instruments, sample size, measurement precision, optometry
- MeSH terms
- Humans, Ophthalmology, Optometry, Vision, Ocular, Reproducibility of Results, Qualitative Research
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review