Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology
Open accessAbstract
Citation information : McAlinden C, Khadka J & Pesudovs K. Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology. Ophthalmic Physiol Opt 2011, 31 , 330–338. doi: 10.1111/j.1475‐1313.2011.00851.x The ever‐expanding choice of ocular metrology and imaging equipment has driven research into the validity of their measurements. Consequently, studies of the agreement between two instruments or clinical tests have proliferated in the ophthalmic literature. It is important that researchers apply the appropriate statistical tests in agreement studies. Correlation coefficients are hazardous and should be avoided. The ‘limits of agreement’ method originally proposed by Altman and Bland in 1983 is the statistical procedure of choice. Its step‐by‐step use and practical considerations in relation to optometry and ophthalmology are detailed in addition to sample size considerations and statistical approaches to precision (repeatability or reproducibility) estimates.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“This was consistent with recommendations from McAlinden et al that sample sizes for agreement studies should include at least 100 subjects(McAlinden et al., 2011).”
From Acceptability and Concordance of Self- Versus Clinician- Sampling for HPV Testing among Rural South Indian Women (2021) · cites this paper as McAlinden et al., 2011
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A Coefficient of Agreement for Nominal Scales
Cohen, J. (1960) · Educational and Psychological Measurement · doi:10.1177/001316446002000104 · Cited by 42,398
Intraclass correlations: Uses in assessing rater reliability.
Shrout, P. E., & Fleiss, J. L. (1979) · Psychological Bulletin · doi:10.1037/0033-2909.86.2.420 · Cited by 24,532
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619
A concordance correlation coefficient to evaluate reproducibility.
Lin, L. I.-K. (1989) · Biometrics · doi:10.2307/2532051 · Cited by 8,785
Scheffé, H. (1960) · Soil Science · doi:10.1097/00010694-196006000-00016 · Cited by 5,625
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Comprehensive Textbook of Psychiatry
Schwab, J. J. (1968) · Psychosomatics · doi:10.1016/s0033-3182(68)71813-2 · Cited by 1,291
Measurement reliability and agreement in psychiatry
Shrout, P. E. (1998) · Statistical Methods in Medical Research · doi:10.1177/096228029800700306 · Cited by 1,211
Clinical biostatistics: LIV. The biostatistics of concordance
Kramer, M. S., & Feinstein, A. R. (1981) · Clinical Pharmacology & Therapeutics · doi:10.1038/clpt.1981.18 · Cited by 1,087
Improving the Reliability of the Pelli-Robson Contrast Sensitivity Test
Bullimore, M. A., Elliott, D. B., & Bailey, I. L. (1991) · Noninvasive Assessment of the Visual System · doi:10.1364/navs.1991.md12 · Cited by 257
Cited by
Sorted by citations
Items 101–200 of 433
Calvo-Sanz, J. A., Portero-Benito, A., & Puente, A. A. (2018) · Graefe s Archive for Clinical and Experimental Ophthalmology · doi:10.1007/s00417-018-3909-9 · Cited by 27
Özçalışkan, Ş., & Yenerel, N. M. (2019) · Clinical and Experimental Optometry · doi:10.1111/cxo.12888 · Cited by 26
Angle-to-angle and spur-to-spur distance analysis with high-resolution optical coherence tomography
Montés‐Micó, R., Tañá‐Rivero, P., Aguilar-Córcoles, S., Ruiz-Santos, M., Rodríguez-Carrillo, M. D., & Ruíz-Mesa, R. (2020) · Eye and Vision · doi:10.1186/s40662-020-00208-0 · Cited by 26
Repeatability of whole-cornea measurements using a new swept-source optical coherence tomographer
Tañá‐Rivero, P., Aguilar-Córcoles, S., Ruíz-Mesa, R., & Montés‐Micó, R. (2020) · European Journal of Ophthalmology · doi:10.1177/1120672120944022 · Cited by 26
Huang, J., Zhao, Y., Savini, G., Yu, G., Yu, J., Chen, Z., Tu, R., & Zhao, Y. (2020) · Journal of Ophthalmology · doi:10.1155/2020/8946364 · Cited by 25
Xie, X., Corradetti, G., Song, A. E., Pardeshi, A. A., Sultan, W., Lee, J. Y., Yu, F., Zhang, L., Chen, S., Chopra, V., et al. (2020) · PLoS ONE · doi:10.1371/journal.pone.0240110 · Cited by 25
Item Banking Enables Stand‐Alone Measurement of Driving Ability
Khadka, J., Fenwick, E., Lamoureux, E. L., & Pesudovs, K. (2016) · Optometry and Vision Science · doi:10.1097/opx.0000000000000958 · Cited by 25
Self-citationCambridge Colour Test: reproducibility in normal trichromats
Fernandes, T. P., dos Santos, N. A., & Paramei, G. V. (2020) · Journal of the Optical Society of America A · doi:10.1364/josaa.380306 · Cited by 24
Assessment of corneal biomechanics, tonometry and pachymetry with the Corvis ST in myopia
Wang, X., McAlinden, C., Zhang, H., Yan, J. R., Wang, D., Wei, W., & Mi, S. (2021) · Scientific Reports · doi:10.1038/s41598-020-80915-9 · Cited by 24
Self-citationTañá-Sanz, P., Ruiz-Santos, M., Rodríguez-Carrillo, M. D., Aguilar-Córcoles, S., Montés‐Micó, R., & Tañá‐Rivero, P. (2021) · Expert Review of Medical Devices · doi:10.1080/17434440.2021.1905518 · Cited by 24
Qin, M., Yuan, Y., Wang, Y., Li, P., Chen, W., Wang, Y., Yang, M., Wu, J., Ji, M., Luo, J., et al. (2022) · BMC Ophthalmology · doi:10.1186/s12886-021-02116-w · Cited by 24
Redondo, B., Vera, J., Luque‐Casado, A., García‐Ramos, A., & Jiménez, R. (2019) · Vision Research · doi:10.1016/j.visres.2019.07.001 · Cited by 27
Forecasting Progressive Trends in Keratoconus by Means of a Time Delay Neural Network
Jiménez‐García, M., Issarti, I., Kreps, E. O., Dhubhghaill, S. N., Koppen, C., Varssano, D., Rozema, J., & on behalf of The REDCAKE Study Group (2021) · Journal of Clinical Medicine · doi:10.3390/jcm10153238 · Cited by 23
Femtosecond Laser–assisted LASIK Improves Quality of Life
Meidani, A., Tzavara, C. K., Dimitrakaki, C., Pesudovs, K., & Tountas, Y. (2012) · Journal of Refractive Surgery · doi:10.3928/1081597x-20120403-01 · Cited by 23
Self-citationIntra‐ and inter‐ examiner repeatability of cycloplegic retinoscopy among young children
McCullough, S. J., Doyle, L. A., & Saunders, K. J. (2016) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12341 · Cited by 22
Venkataraman, A. P., Brautaset, R., & Domínguez‐Vicent, A. (2022) · PLoS ONE · doi:10.1371/journal.pone.0278269 · Cited by 22
The effect of pupillary dilatation on IOL power measurement by using the IOLMaster
Khambhiphant, B., Chatbunchachai, N., & Pongpirul, K. (2015) · International Ophthalmology · doi:10.1007/s10792-015-0063-9 · Cited by 22
Agreement between a partial coherence interferometer and 2 manual keratometers
Bullimore, M. A., Büehren, T., & Bissmann, W. (2013) · Journal of Cataract & Refractive Surgery · doi:10.1016/j.jcrs.2013.03.034 · Cited by 22
Huelle, J. O., Katz, T. R., Druchkiv, V., Pahlitzsch, M., Steinberg, J., Richard, G. W., & Linke, S. J. (2014) · British Journal of Ophthalmology · doi:10.1136/bjophthalmol-2013-304786 · Cited by 22
Interchangeability among five devices that measure anterior eye distances
Domínguez‐Vicent, A., Pérez‐Vives, C., Ferrer‐Blasco, T., Albarrán‐Diego, C., & Montés‐Micó, R. (2015) · Clinical and Experimental Optometry · doi:10.1111/cxo.12247 · Cited by 21
Huerva, V., Ascaso, F. J., Soldevila, J., & Lavilla, L. (2014) · Journal of Cataract & Refractive Surgery · doi:10.1016/j.jcrs.2013.10.045 · Cited by 21
Hu, A., Qiao, L.-Y., Zhang, Y., Cai, X., Li, L., & Wan, X. H. (2015) · Europe PMC (PubMed Central) · doi:10.3980/j.issn.2222-3959.2015.05.34 · Cited by 21
Corneal Power Measurement Obtained by Fourier-Domain Optical Coherence Tomography
Wang, Q., Hua, Y., Savini, G., Chen, H., Bao, F., Lin, S., Lu, W., & Huang, J. (2015) · Cornea · doi:10.1097/ico.0000000000000564 · Cited by 21
Image-guided lens extraction surgery: a systematic review
Panagiotopoulou, E.-K., Ntonti, P., Gkika, M. G., Konstantinidis, A., Perente, İ., Dardabounis, D., Ioannakis, K., & Labiris, G. (2019) · International Journal of Ophthalmology · doi:10.18240/ijo.2019.01.21 · Cited by 21
Domínguez-Vicent, A., Brautaset, R., & Venkataraman, A. P. (2019) · PLoS ONE · doi:10.1371/journal.pone.0221466 · Cited by 21
Scleral shape and its correlation with corneal parameters in keratoconus
Dhaese, S. E., Kreps, E. O., & Consejo, A. (2020) · Contact Lens and Anterior Eye · doi:10.1016/j.clae.2020.08.010 · Cited by 21
Effect of age in the ciliary muscle during accommodation: Sectorial analysis
Domínguez‐Vicent, A., Monsálvez-Romín, D., Esteve-Taboada, J. J., Montés‐Micó, R., & Ferrer‐Blasco, T. (2018) · Journal of Optometry · doi:10.1016/j.optom.2018.01.001 · Cited by 20
Yağc, R., Kulak, A. E., Güler, E., Tenlik, A., Gürağaç, F. B., & Hepşen, İ. F. (2015) · Cornea · doi:10.1097/ico.0000000000000480 · Cited by 20
Masmali, A. M., Alanazi, S. A., Almagren, B., & El-Hiti, G. A.-R. (2019) · Clinical Optometry · doi:10.2147/opto.s206904 · Cited by 20
Mazzoni, A., Nordin, K., Berntsen, S., Demmelmaier, I., & Igelström, H. (2017) · BMC Sports Science Medicine and Rehabilitation · doi:10.1186/s13102-017-0072-2 · Cited by 23
Evaluation of the Performance of Algorithm-Based Methods for Subjective Refraction
Venkataraman, A. P., Sirak, D., Brautaset, R., & Domínguez‐Vicent, A. (2020) · Journal of Clinical Medicine · doi:10.3390/jcm9103144 · Cited by 19
Habbe, K. J., Фрингс, А., Saad, A., & Geerling, G. (2023) · PLoS ONE · doi:10.1371/journal.pone.0279977 · Cited by 19
Ohlendorf, A., Leube, A., & Wahl, S. (2016) · Healthcare · doi:10.3390/healthcare4030041 · Cited by 19
Agreement of white-to-white measurements with the IOLMaster 700, Atlas 9000, and Sirius systems
Ferrer‐Blasco, T., Esteve-Taboada, J. J., Martínez-Albert, N., Alfonso, J. F., & Montés‐Micó, R. (2018) · Expert Review of Medical Devices · doi:10.1080/17434440.2018.1481745 · Cited by 19
Kirsch, K., Detilleux, J., Serteyn, D., & Sandersen, C. (2019) · PLoS ONE · doi:10.1371/journal.pone.0211104 · Cited by 19
Özcan, Y., Özçalışkan, Ş., Balci, S. Y., & Artunay, Ö. (2020) · Photodiagnosis and Photodynamic Therapy · doi:10.1016/j.pdpdt.2020.102008 · Cited by 19
Garzón, N., García‐Montero, M., López‐Artero, E., Poyales, F., & Albarrán‐Diego, C. (2019) · Journal of Cataract & Refractive Surgery · doi:10.1016/j.jcrs.2019.04.017 · Cited by 19
Comparability and repeatability of different methods of corneal astigmatism assessment
Ferreira, T. B., & Ribeiro, F. J. (2017) · Clinical ophthalmology · doi:10.2147/opth.s146730 · Cited by 18
Wonneberger, W., Sterner, B., MacLean, U., Claesson, M., & Zetterberg, M. (2018) · Cornea · doi:10.1097/ico.0000000000001513 · Cited by 18
Impact of Lens Material on Objective Refraction in Eyes with Trifocal Diffractive Intraocular Lenses
Garzón, N., Poyales, F., García‐Montero, M., Vega, F., Millán, M. S., & Albarrán‐Diego, C. (2021) · Current Eye Research · doi:10.1080/02713683.2021.1946563 · Cited by 18
Martínez-García, M. D. M., Ruiz-Cárdenas, J. D., & Rabinovich, R. (2017) · COPD Journal of Chronic Obstructive Pulmonary Disease · doi:10.1080/15412555.2017.1358257 · Cited by 18
Validity of automated refraction after segmented refractive multifocal intraocular lens implantation
Albarrán‐Diego, C., Muñoz, G., Rohrweck, S., García‐Lázaro, S., & Albero, J. R. (2017) · International Journal of Ophthalmology · doi:10.18240/ijo.2017.11.15 · Cited by 18
Comparison of Anterior Segment Measurements Obtained by Three Different Devices in Healthy Eyes
FUENTE, C. L. D. L., Sánchez-Cano, A. I., Segura, F. J., & Pinilla, I. (2014) · BioMed Research International · doi:10.1155/2014/498080 · Cited by 18
Thirunavukarasu, A. J., Hassan, R., Limonard, A., & Savant, S. (2023) · PLoS ONE · doi:10.1371/journal.pone.0281847 · Cited by 18
Koech, A. C., Mwaniki, I. G., Mutunga, J. M., Mukhanya, M. W., Mwadime, E., Ochieng, M. C., Mwashigadi, G., Mistry, H. D., Craik, R., von Dadelszen, P., et al. (2024) · Frontiers in Global Women s Health · doi:10.3389/fgwh.2024.1427261 · Cited by 17
Bao, F., Savini, G., Shu, B., Zhu, S., Gao, R., Dang, G., Yu, A., Wang, Q., & Huang, J. (2017) · Journal of Refractive Surgery · doi:10.3928/1081597x-20170606-01 · Cited by 17
Vera, J., Redondo, B., Molina, R., Koulieris, G. A., & Jiménez, R. (2019) · Current Eye Research · doi:10.1080/02713683.2019.1688837 · Cited by 17
Tang, W. C., Tang, Y. Y., & Lam, C. S. Y. (2013) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12098 · Cited by 17
Visual acuity and contrast sensitivity screening with a new iPad application
Rodríguez‐Vallejo, M., Llorens‐Quintana, C., Furlan, W. D., & Monsoriu, J. A. (2016) · Displays · doi:10.1016/j.displa.2016.06.001 · Cited by 17
Gómez-Gómez, A., Talens‐Estarelles, C., Alcocer-Yuste, P., & Nieto, J. C. (2021) · Journal of Glaucoma · doi:10.1097/ijg.0000000000001878 · Cited by 16
Domínguez‐Vicent, A., Venkataraman, A. P., Dalin, A., Brautaset, R., & Montés‐Micó, R. (2023) · Eye and Vision · doi:10.1186/s40662-023-00343-4 · Cited by 16
Association between Contrast Sensitivity and Reading with Macular Pathology
Brussee, T., Berg, T., van Nispen, R. M. A., de Boer, I., & van Rens, G. H. M. B. (2018) · Optometry and Vision Science · doi:10.1097/opx.0000000000001178 · Cited by 16
Comparison of two devices to simulate vision with intraocular lenses
Wahl, S., Song, C., & Ohlendorf, A. (2019) · Clinical ophthalmology · doi:10.2147/opth.s188890 · Cited by 16
Cup‐to‐disc and arteriole‐to‐venule ratios in children aged 6–7 and 12–13 years
McClelland, J. F., O’Donoghue, L., McIntyre, M., & Saunders, K. J. (2011) · Ophthalmic and Physiological Optics · doi:10.1111/j.1475-1313.2011.00878.x · Cited by 16
Arias‐Álvarez, M., Tomás‐Grasa, C., Sopeña‐Pinilla, M., Fernández‐Espinosa, G., Alonso, S. B., Acha, J., Mena, D. R., & Pinilla, I. (2024) · Scientific Reports · doi:10.1038/s41598-024-54099-5 · Cited by 15
Zhang, X.-F. (2020) · International Journal of Ophthalmology · doi:10.18240/ijo.2020.04.15 · Cited by 15
Domínguez‐Vicent, A., Al‐Soboh, L., Brautaset, R., & Venkataraman, A. P. (2020) · Journal of Clinical Medicine · doi:10.3390/jcm9103061 · Cited by 15
Comparison of a new optical biometry with an optical low-coherence reflectometry for ocular biometry
Güler, E., Kulak, A. E., Totan, Y., Yuvarlak, A., & Hepşen, İ. F. (2016) · Contact Lens and Anterior Eye · doi:10.1016/j.clae.2016.06.001 · Cited by 15
Intereye Symmetry in Bietti Crystalline Dystrophy
Liu, Z., Ayton, L. N., O’Hare, F., Arslan, J., Hu, M. L., Noar, A. P., Wang, J.-H., Hickey, D. G., McGuinness, M. B., Vincent, A. L., et al. (2021) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2021.07.009 · Cited by 14
Muzyka−Woźniak, M., Oleszko, A., & Grzybowski, A. E. (2022) · Journal of Clinical Medicine · doi:10.3390/jcm11236921 · Cited by 14
Wang, H., Zhu, L., Pang, C., & Fan, Q. (2024) · BMC Ophthalmology · doi:10.1186/s12886-024-03448-z · Cited by 14
Salouti, R., Kamalipour, A., Masihpour, N., Zamani, M., Ghoreyshi, M., Salouti, K., & Nowroozzadeh, M. H. (2020) · Journal of Cataract & Refractive Surgery · doi:10.1097/j.jcrs.0000000000000252 · Cited by 14
Assessment of anterior segment measurements using a high-resolution imaging device
Montés‐Micó, R., Tañá‐Rivero, P., Aguilar-Córcoles, S., & Ruíz-Mesa, R. (2020) · Expert Review of Medical Devices · doi:10.1080/17434440.2020.1816463 · Cited by 14
Ocular surface physiology and aqueous tear secretion in cats of diverse cephalic conformations
Oksa-Minalto, J., Maggs, D. J., Akimova, J., Ilgaža, A., & Sebbag, L. (2022) · Veterinary Ophthalmology · doi:10.1111/vop.13051 · Cited by 14
Kliner, D., Wang, L., Winger, D. G., Follansbee, W. P., & Soman, P. (2015) · Journal of Nuclear Cardiology · doi:10.1007/s12350-015-0071-1 · Cited by 14
Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes
FUENTE, C. L. D. L., Sánchez-Cano, A. I., Segura, F. J., Fuentes‐Broto, L., & Pinilla, I. (2014) · BioMed Research International · doi:10.1155/2014/808646 · Cited by 14
The measurement properties of mentoring relationship quality scales for mentoring programs
Ferro, A., Wells, S., Speechley, K. N., Lipman, E., & DeWit, D. (2014) · Prevention Science · doi:10.1007/s11121-013-0392-8 · Cited by 13
Tañá-Sanz, P., Rodríguez-Carrillo, M. D., Ruiz-Santos, M., Montés‐Micó, R., Ruíz-Mesa, R., & Tañá‐Rivero, P. (2021) · Expert Review of Medical Devices · doi:10.1080/17434440.2021.2008908 · Cited by 13
Statistical Perspectives on Reproducibility: Definitions and Challenges
Simkus, A., Coolen‐Maturi, T., Coolen, F. P. A., & Bendtsen, C. (2025) · Journal of Statistical Theory and Practice · doi:10.1007/s42519-025-00459-x · Cited by 14
Comparison of specular microscopy and ultrasound pachymetry before and after cataract surgery
López‐Miguel, A., Sanchidrián, M., Fernández, I., Holgueras, A., & Maldonado, M. J. (2016) · Graefe s Archive for Clinical and Experimental Ophthalmology · doi:10.1007/s00417-016-3537-1 · Cited by 13
Repeatability and agreement of intraocular pressure measurement among three tonometers
pavón, I. S., Cañadas, P., & Martín, R. (2020) · Clinical and Experimental Optometry · doi:10.1111/cxo.13043 · Cited by 13
Shu, B., Bao, F., Savini, G., Lu, W., Tu, R., Chen, H., Song, B., Wang, Q., & Huang, J. (2020) · Eye and Vision · doi:10.1186/s40662-020-00177-4 · Cited by 13
Pyramidal Aberrometry in Wavefront-Guided Myopic LASIK
Фрингс, А., Hassan, H., & Allan, B. D. (2020) · Journal of Refractive Surgery · doi:10.3928/1081597x-20200519-03 · Cited by 13
Comparison, within-session repeatability and normative data of three phoria tests
Facchin, A., & Maffioletti, S. (2020) · Journal of Optometry · doi:10.1016/j.optom.2020.05.007 · Cited by 13
Espinilla, O. G., Gallegos-Cocho, I., Sánchez, I., Cañadas, P., & Martín, R. (2021) · Clinical and Experimental Optometry · doi:10.1080/08164622.2021.1914511 · Cited by 13
Yoga for Healthy Aging: Science or Hype?
Madhivanan, P., Krupp, K., Waechter, R. L., & Shidhaye, R. R. (2021) · Advances in Geriatric Medicine and Research · doi:10.20900/agmr20210016 · Cited by 13
An evaluation of the IOLMaster 700 and its agreement with the IOLMaster v3 in children
Leighton, R. E., Breslin, K. M. M., Saunders, K. J., & McCullough, S. J. (2021) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12918 · Cited by 13
Rossi, G. C. M., Bettio, F., González‐Pérez, M., Briola, A., Pasinetti, G. L. M., & Scudeller, L. (2022) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph19052517 · Cited by 13
Huelle, J. O., Katz, T. R., Draeger, J., Pahlitzsch, M., Druchkiv, V., Steinberg, J., Richard, G. W., & Linke, S. J. (2013) · Graefe s Archive for Clinical and Experimental Ophthalmology · doi:10.1007/s00417-012-2246-7 · Cited by 13
Esteve-Taboada, J. J., Ferrer‐Blasco, T., Aloy, M. Á., Adsuara, J. E., Cerdá-Durán, P., Mimica, P., & Montés‐Micó, R. (2017) · Graefe s Archive for Clinical and Experimental Ophthalmology · doi:10.1007/s00417-017-3801-z · Cited by 13
Martínez-Albert, N., Esteve-Taboada, J. J., & Montés‐Micó, R. (2017) · Graefe s Archive for Clinical and Experimental Ophthalmology · doi:10.1007/s00417-017-3832-5 · Cited by 13
Pinero, D. P., Soto-Negro, R., Ruiz-Fortes, P., Pérez-Cambrodí, R. J., & Fukumitsu, H. (2019) · Cornea · doi:10.1097/ico.0000000000002034 · Cited by 13
Associations Between Spatial and Temporal Contrast Sensitivity and Reading
Brussee, T., Berg, T., van Nispen, R. M. A., & van Rens, G. H. M. B. (2016) · Optometry and Vision Science · doi:10.1097/opx.0000000000001030 · Cited by 13
Valero, B., Fénolland, J.-R., Rosenberg, R., Sendon, D., Mesnard, C., Sigaux, M., Giraud, J., & Renard, J. P. (2017) · Journal Français d Ophtalmologie · doi:10.1016/j.jfo.2017.06.008 · Cited by 13
Venkataraman, A. P., Domínguez‐Vicent, A., Selin, P., Brautaset, R., & Montés‐Micó, R. (2023) · Journal of Cataract & Refractive Surgery · doi:10.1097/j.jcrs.0000000000001380 · Cited by 12
Breher, K., Ohlendorf, A., & Wahl, S. (2019) · Scientific Reports · doi:10.1038/s41598-019-56915-9 · Cited by 12
Change in optical axial length after cataract surgery: segmental method vs composite method
Goto, S., Maeda, N., Noda, T., Ohnuma, K., Iehisa, I., Koh, S., & Nishida, K. (2020) · Journal of Cataract & Refractive Surgery · doi:10.1097/j.jcrs.0000000000000149 · Cited by 12
Performance of a new device for the clinical determination of light discomfort
Montés‐Micó, R., Cerviño, A., Martínez-Albert, N., García‐Marqués, J. V., & Marié, S. (2020) · Expert Review of Medical Devices · doi:10.1080/17434440.2020.1837623 · Cited by 12
Ruiß, M., Findl, O., Draschl, P., Harrer, A., & Hirnschall, N. D. (2021) · Journal of Cataract & Refractive Surgery · doi:10.1097/j.jcrs.0000000000000583 · Cited by 12
Wolfenden, G., James, F., Hung, L. H. T., Bruce, M., & Thompson, M. (2022) · Journal of Small Animal Practice · doi:10.1111/jsap.13491 · Cited by 12
Hernández, C. S., Gil, A., Casares, I., Poderoso, J., Wehse, A., Dave, S. R., Lim, D., Sánchez-Montañés, M. A., & Lage, E. (2022) · Journal of Optometry · doi:10.1016/j.optom.2022.03.001 · Cited by 12
Qiao, L.-Y., Zhang, X., Jan, C., Li, X., Li, M., & Wang, H. (2019) · Science China Life Sciences · doi:10.1007/s11427-018-9484-6 · Cited by 12
Changes in symmetry of anterior chamber following routine cataract surgery in non-glaucomatous eyes
Lee, H., Zukaite, I., Juniat, V. A. R., Dimitry, M. E. S., Lewis, A., & Nanavaty, M. A. (2019) · Eye and Vision · doi:10.1186/s40662-019-0144-3 · Cited by 12
Choroidal Differences between Spectral and Swept-source Domain Technologies
Pinilla, I., Sánchez-Cano, A. I., Insa, G., Bartolomé, I., Perdices, L., & Orduna, E. (2020) · Current Eye Research · doi:10.1080/02713683.2020.1795883 · Cited by 12
Tañá‐Rivero, P., Rodríguez-Carrillo, M. D., Ruiz-Santos, M., García-Tomás, B., & Montés‐Micó, R. (2021) · Journal of Refractive Surgery · doi:10.3928/1081597x-20201013-01 · Cited by 11
Horizontal and anterior chamber diameter for phakic intraocular lens sizing
Oleszko, A., Marek, J., & Muzyka−Woźniak, M. (2020) · Clinical and Experimental Optometry · doi:10.1111/cxo.13101 · Cited by 11
Accuracy of image analysis for linear zoometric measurements in dromedary camels
Gherissi, D. E., Lamraoui, R., Chacha, F., & Gaouar, S. B. S. (2022) · Tropical Animal Health and Production · doi:10.1007/s11250-022-03242-3 · Cited by 11
Sánchez, I., Ortiz-Toquero, S., & Martín, R. (2018) · Eye & Contact Lens Science & Clinical Practice · doi:10.1097/icl.0000000000000486 · Cited by 11
Kortuem, C., Marx, T., Altpeter, E. K., Trauzettel‐Klosinski, S., & Kuester-Gruber, S. (2020) · Ophthalmic Research · doi:10.1159/000509687 · Cited by 11
Hazari, H., Curtis, R., Eden, K., Hopman, W. M., Irrcher, I., & Bona, M. D. (2020) · Journal of Telemedicine and Telecare · doi:10.1177/1357633x20960640 · Cited by 11
Metadata
- Authors
- McAlinden, C., Khadka, J., & Pesudovs, K.
- Affiliations
- Flinders University (AU)
- Corresponding author
- McAlinden, Colm
- Venue
- Ophthalmic and Physiological Optics
- Published
- 2011-05-26
- Volume
- 31
- Issue
- 4
- Pages
- 330-338
- ISSN
- 0275-5408 (ISSN-L)
- 1475-1313
- OpenAlex
- W2126376404
- Cited by
- 436
- References
- 49
- Citations by year
- 2026: 22 · 2025: 24 · 2024: 33 · 2023: 30 · 2022: 37 · 2021: 42 · 2020: 47 · 2019: 33 · 2018: 22 · 2017: 28 · 2016: 27 · 2015: 27 · 2014: 20 · 2013: 15 · 2012: 22
- FWCI
- 5.01
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Ophthalmic and Physiological Optics (publishedVersion)
- PubMed (publishedVersion)
- Adelaide Research & Scholarship (AR&S) (University of Adelaide) (submittedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Health Sciences › Medicine › Ophthalmology › Ophthalmology and Visual Impairment Studies
- Keywords
- repeatability, reproducibility, limits of agreement, correlation coefficient, ophthalmic instruments, sample size, measurement precision, optometry
- MeSH terms
- Humans, Ophthalmology, Optometry, Vision, Ocular, Reproducibility of Results, Qualitative Research
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review