Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology
Open accessAbstract
Citation information : McAlinden C, Khadka J & Pesudovs K. Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology. Ophthalmic Physiol Opt 2011, 31 , 330–338. doi: 10.1111/j.1475‐1313.2011.00851.x The ever‐expanding choice of ocular metrology and imaging equipment has driven research into the validity of their measurements. Consequently, studies of the agreement between two instruments or clinical tests have proliferated in the ophthalmic literature. It is important that researchers apply the appropriate statistical tests in agreement studies. Correlation coefficients are hazardous and should be avoided. The ‘limits of agreement’ method originally proposed by Altman and Bland in 1983 is the statistical procedure of choice. Its step‐by‐step use and practical considerations in relation to optometry and ophthalmology are detailed in addition to sample size considerations and statistical approaches to precision (repeatability or reproducibility) estimates.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“This was consistent with recommendations from McAlinden et al that sample sizes for agreement studies should include at least 100 subjects(McAlinden et al., 2011).”
From Acceptability and Concordance of Self- Versus Clinician- Sampling for HPV Testing among Rural South Indian Women (2021) · cites this paper as McAlinden et al., 2011
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A Coefficient of Agreement for Nominal Scales
Cohen, J. (1960) · Educational and Psychological Measurement · doi:10.1177/001316446002000104 · Cited by 42,398
Intraclass correlations: Uses in assessing rater reliability.
Shrout, P. E., & Fleiss, J. L. (1979) · Psychological Bulletin · doi:10.1037/0033-2909.86.2.420 · Cited by 24,532
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619
A concordance correlation coefficient to evaluate reproducibility.
Lin, L. I.-K. (1989) · Biometrics · doi:10.2307/2532051 · Cited by 8,785
Scheffé, H. (1960) · Soil Science · doi:10.1097/00010694-196006000-00016 · Cited by 5,625
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Comprehensive Textbook of Psychiatry
Schwab, J. J. (1968) · Psychosomatics · doi:10.1016/s0033-3182(68)71813-2 · Cited by 1,291
Measurement reliability and agreement in psychiatry
Shrout, P. E. (1998) · Statistical Methods in Medical Research · doi:10.1177/096228029800700306 · Cited by 1,211
Clinical biostatistics: LIV. The biostatistics of concordance
Kramer, M. S., & Feinstein, A. R. (1981) · Clinical Pharmacology & Therapeutics · doi:10.1038/clpt.1981.18 · Cited by 1,087
Improving the Reliability of the Pelli-Robson Contrast Sensitivity Test
Bullimore, M. A., Elliott, D. B., & Bailey, I. L. (1991) · Noninvasive Assessment of the Visual System · doi:10.1364/navs.1991.md12 · Cited by 257
Cited by
Sorted by citations
Items 201–300 of 433
Li, X., Chang, P., Li, Z., Qian, S., Zhu, Z., Wang, Q., & Yun-E, Z. (2020) · Expert Review of Medical Devices · doi:10.1080/17434440.2020.1848541 · Cited by 11
Castillo-Peinado, L. S., Calderón-Santiago, M., & Priego‐Capote, F. (2020) · Talanta · doi:10.1016/j.talanta.2020.121692 · Cited by 11
Tañá‐Rivero, P., Tello-Elordi, C., Orts-Vila, P., Tañá-Sanz, P., & Tañá-Sanz, S. (2023) · Clinical ophthalmology · doi:10.2147/opth.s409132 · Cited by 10
Ruiz‐belda, C., Rodrigo, F., & Pinero, D. P. (2017) · Clinical and Experimental Optometry · doi:10.1111/cxo.12623 · Cited by 10
Interexaminer reproducibility for subjective refractions for an ametropic participant
Mathebula, S. D., & RUBIN, A. M. (2022) · BMJ Open Ophthalmology · doi:10.1136/bmjophth-2021-000954 · Cited by 10
Comprehensive Primary Eye Care: A Comparison Between an In-Person Eye Exam and a Tele-Eye Care Exam
Blais, N., Tousignant, B., & Hanssens, J.-M. (2024) · Clinical Optometry · doi:10.2147/opto.s436659 · Cited by 10
Ferrer‐Blasco, T., Esteve-Taboada, J. J., Monsálvez-Romín, D., Aloy, M. Á., Adsuara, J. E., Cerdá-Durán, P., & Montés‐Micó, R. (2017) · International Ophthalmology · doi:10.1007/s10792-017-0813-y · Cited by 10
Validation of posterior corneal curvature measurements with color light-emitting diode topography
Pinero, D. P., Camps, V. J., Sáiz, D. D. F., García, C., & Caballero, M. T. (2019) · European Journal of Ophthalmology · doi:10.1177/1120672119870738 · Cited by 10
Orbscan 3 Versus Pentacam HR: Evaluating the Possible Interchangeable Use of Various Parameters
Gharieb, H. M., Othman, I. S., & Elkitkat, R. S. (2020) · Cornea · doi:10.1097/ico.0000000000002257 · Cited by 10
Curtis, R., Hazari, H., Eden, K., Hopman, W. M., Irrcher, I., & Bona, M. D. (2020) · Journal of Telemedicine and Telecare · doi:10.1177/1357633x20960628 · Cited by 10
Validation of keratometric measurements obtained with a new integrated aberrometry-topography system
Shneor, E., Millodot, M., Zyroff, M., & Gordon‐Shaag, A. (2012) · Journal of Optometry · doi:10.1016/j.optom.2012.03.003 · Cited by 10
Eye laterality: a comprehensive analysis in refractive surgery candidates
Linke, S. J., Druchkiv, V., Steinberg, J., Richard, G. W., & Katz, T. R. (2013) · Acta Ophthalmologica · doi:10.1111/aos.12040 · Cited by 10
Relations entre les paramètres de la réfraction : sphère, cylindre et axe
Touzeau, O., Gaujoux, T., Bullet, J., Allouch, C., Borderie, V. M., & Laroche, L. (2012) · Journal Français d Ophtalmologie · doi:10.1016/j.jfo.2011.12.006 · Cited by 10
Wang, Q., Huang, Y., Gao, R., Chen, H., Song, B., Tu, R., Zhao, Y., McAlinden, C., & Huang, J. (2020) · Journal of Cataract & Refractive Surgery · doi:10.1097/j.jcrs.0000000000000072 · Cited by 10
Self-citationGonzález-Arellanes, R., Urquídez-Romero, R., Rodríguez‐Tadeo, A., Esparza‐Romero, J., Estrada, R. O. M., Ramírez‐López, E., Robles-Sardín, A. E., Pacheco-Moreno, B. I., & Alemán-Mateo, H. (2020) · Clinical Nutrition · doi:10.1016/j.clnu.2020.12.020 · Cited by 10
Bubb, L., Mathews, D., Oehring, D., & Harper, R. A. (2021) · Eye · doi:10.1038/s41433-021-01394-4 · Cited by 10
Mohidin, N., Low, Y. C., Mohd-Ali, B., Shahimin, M. M., & Arif, N. (2022) · F1000Research · doi:10.12688/f1000research.110560.1 · Cited by 9
Chow, A. H. Y., Mungalsingh, M. A., Thai, D., Selimos, Z., Yan, S. K., Xu, H., & JONES, D. A. (2024) · Ophthalmic and Physiological Optics · doi:10.1111/opo.13332 · Cited by 9
Assessment of corneal epithelial thickness mapping by spectral-domain optical coherence tomography
Tañá‐Rivero, P., Orts-Vila, P., Tañá-Sanz, P., Ramos-Alzamora, M., & Montés‐Micó, R. (2024) · Frontiers in Medicine · doi:10.3389/fmed.2024.1459636 · Cited by 9
Facchin, A., & Boccardo, L. (2024) · Optometry and Vision Science · doi:10.1097/opx.0000000000002184 · Cited by 9
Chen, S., Gao, R., McAlinden, C., Ye, J., Wang, Y., Chen, M., Huang, J., Sun, Y., & Yu, A. (2020) · Journal of Ophthalmology · doi:10.1155/2020/9739878 · Cited by 9
Self-citationMinimal detectable change of the disc‐fovea angle for ocular torsion assessment
Piedrahita‐Alonso, E., Valverde-Megías, A., Martin‐García, B., Hernández‐García, E., & de Liaño, R. G. (2021) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12897 · Cited by 9
Calvo‐Maroto, A. M., Llorente‐González, S., Bezunartea, J., Hurtado-Ceña, F. J., Berrozpe‐Villabona, C., Bilbao‐Malavé, V., Pinero, D. P., Barrio‐Barrio, J., & Recalde, S. (2022) · Children · doi:10.3390/children9010088 · Cited by 9
Ametropia detection using a novel, compact wavefront autorefractor
Hernández, C. S., Gil, A., Zaytouny, A., Casares, I., Poderoso, J., de Lara, A., Wehse, A., Dave, S. R., Lim, D., Lage, E., et al. (2023) · Ophthalmic and Physiological Optics · doi:10.1111/opo.13263 · Cited by 9
Ogbuehi, K. C., Mucke, S., & Osuagwu, U. L. (2012) · Ophthalmic and Physiological Optics · doi:10.1111/j.1475-1313.2012.00945.x · Cited by 9
Sánchez, I., Carmona, J., Gonzalez-Puertas, S., Valiente, A., & Martín, R. (2019) · Eye & Contact Lens Science & Clinical Practice · doi:10.1097/icl.0000000000000655 · Cited by 9
Comparison of ocular parameters of two biometric measurement devices in highly myopic eyes
Guo, X. C. (2019) · International Journal of Ophthalmology · doi:10.18240/ijo.2019.10.05 · Cited by 9
Corneal thickness evaluation in healthy eyes: Comparison between two different Scheimpflug devices
Rosa, N., Bernardo, M. D., Pepe, A., Vitiello, L., Marotta, G., Imparato, R., & Capasso, L. (2020) · PLoS ONE · doi:10.1371/journal.pone.0243370 · Cited by 9
García‐Marqués, J. V., Macedo-de-Araújo, R. J., Cerviño, A., García‐Lázaro, S., & González‐Méijome, J. M. (2022) · Contact Lens and Anterior Eye · doi:10.1016/j.clae.2021.101571 · Cited by 9
McAlinden, C., Bao, F., Guo, Y., Yu, X., Lu, W., Chen, H., Wang, Q., & Huang, J. (2016) · Journal of Cataract & Refractive Surgery · doi:10.1016/j.jcrs.2016.01.043 · Cited by 9
Self-citationDomínguez‐Vicent, A., Trussardi, C., Wogatai, U., Montés‐Micó, R., & Brautaset, R. (2016) · Journal of Refractive Surgery · doi:10.3928/1081597x-20160721-03 · Cited by 9
Clinical relationship of meibometry with ocular symptoms and tear film stability
García‐Resúa, C., Pena‐Verdeal, H., Giraldez, M. J., & Yebra‐Pimentel, E. (2017) · Contact Lens and Anterior Eye · doi:10.1016/j.clae.2017.07.003 · Cited by 9
Vision in high-level football officials
Baptista, A. M. G., Serra, P. M., McAlinden, C., & Barrett, B. T. (2017) · PLoS ONE · doi:10.1371/journal.pone.0188463 · Cited by 9
Self-citationThe development and evaluation of the new Ocular Surface Disease Index-6
Pult, H., & Wolffsohn, J. S. (2018) · Contact Lens and Anterior Eye · doi:10.1016/j.clae.2018.04.032 · Cited by 9
Ferrer‐Blasco, T., Esteve-Taboada, J. J., Domínguez‐Vicent, A., Aloy, M. Á., Adsuara, J. E., Mimica, P., & Montés‐Micó, R. (2019) · Arquivos Brasileiros de Oftalmologia · doi:10.5935/0004-2749.20190020 · Cited by 9
The Role of Visual Performance in Fine Motor Skills
Granados‐Delgado, P., Casares‐López, M., Martino, F., Anera, R. G., & Castro, J. J. (2024) · Life · doi:10.3390/life14111354 · Cited by 8
Ruiz-Mesa, R., Ruiz-Santos, M., Blanch-Ruiz, J., & Jiménez-Nieto, A. (2022) · Clinical ophthalmology · doi:10.2147/opth.s351845 · Cited by 8
Definitions for Keratoconus Progression and Their Impact on Clinical Practice
Koppen, C., Jiménez‐García, M., Kreps, E. O., Dhubhghaill, S. N., Rozema, J. J., & on behalf of The REDCAKE Study Group (2023) · Eye & Contact Lens Science & Clinical Practice · doi:10.1097/icl.0000000000001038 · Cited by 8
Effect of digital ocular massage on intraocular pressure and Schlemm’s canal dimensions
Wu, T. H. Y., Lau, H. H., Lai, C. H. Y., Wong, R. W. L., Wong, A. K. W., & Lam, A. K. C. (2024) · Scientific Reports · doi:10.1038/s41598-024-56748-1 · Cited by 8
Huo, Y., Xie, R., Chen, X., Li, S., Zou, H., Liu, Y., & Wang, Y. (2024) · Scientific Reports · doi:10.1038/s41598-024-74497-z · Cited by 8
Coloured filters show gender differences and poor repeatability
Elliott, D. B., & Wood, J. M. (2017) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12417 · Cited by 8
Evaluation of the “Freiburg Acuity VEP” on Commercial Equipment
Bach, M., & Farmer, J. D. (2019) · Documenta Ophthalmologica · doi:10.1007/s10633-019-09726-2 · Cited by 8
Beato, J. N. (2020) · International Journal of Ophthalmology · doi:10.18240/ijo.2020.06.10 · Cited by 8
Measurement properties of reading tests in subjects with maculopathy
Brussee, T., Alagahgi, B., van Nispen, R. M. A., & van Rens, G. H. M. B. (2020) · Acta Ophthalmologica · doi:10.1111/aos.14574 · Cited by 8
Domínguez‐Vicent, A., Kensén, J., Ramsay, M. W., Brautaset, R., & Venkataraman, A. P. (2021) · Frontiers in Medicine · doi:10.3389/fmed.2021.764236 · Cited by 8
Evaluation of a visual acuity eHealth tool in patients with cataract
Wanten, J. C., Bauer, N. J. C., Claessens, J. L. J., van Amelsfort, T., Berendschot, T. T. J. M., Wisse, R. P. L., & Nuijts, R. M. M. A. (2022) · Journal of Cataract & Refractive Surgery · doi:10.1097/j.jcrs.0000000000001108 · Cited by 9
Neveu, P., Priot, A.-E., Philippe, M., Fuchs, P., & Roumes, C. (2015) · Clinical and Experimental Optometry · doi:10.1111/cxo.12311 · Cited by 8
Bergin, C., Guber, I., Hashemi, K. K., & Majo, F. (2015) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.14-16091 · Cited by 8
Anterior Segment Biometry and Their Correlation with Corneal Biomechanics in Caucasian Children
Bueno-Gimeno, I., Martínez-Albert, N., Gené‐Sampedro, A., & Gregori, E. E. (2018) · Current Eye Research · doi:10.1080/02713683.2018.1539181 · Cited by 8
Akagün, N., & Altıparmak, U. E. (2025) · Ophthalmic and Physiological Optics · doi:10.1111/opo.13527 · Cited by 7
Feng, Y., Nitter, T., Bertelsen, G., & Stojanović, A. (2024) · Clinical and Experimental Ophthalmology · doi:10.1111/ceo.14423 · Cited by 7
Jiménez, R., Redondo, B., Davies, L. N., & Vera, J. (2019) · Optometry and Vision Science · doi:10.1097/opx.0000000000001406 · Cited by 7
Less is more: optimal recording time for measuring the steady-state accommodative response
Redondo, B., Vera, J., Molina, R., & Jiménez, R. (2021) · Clinical and Experimental Optometry · doi:10.1080/08164622.2021.2003692 · Cited by 7
Development and validation of a new C hinese reading chart for children
Cheung, J. P. Y., Liu, D. S. K., Lam, C. C.-C., & Cheong, A. M. Y. (2015) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12228 · Cited by 7
Monsálvez-Romín, D., Domínguez‐Vicent, A., Esteve-Taboada, J. J., Montés‐Micó, R., & Ferrer‐Blasco, T. (2019) · Arquivos Brasileiros de Oftalmologia · doi:10.5935/0004-2749.20190041 · Cited by 7
The Cambridge Colour Test: Reliability of discrimination trivectors in colour space
Hasrod, N., & RUBIN, A. M. (2019) · African vision and eye health · doi:10.4102/aveh.v78i1.451 · Cited by 7
Venkataraman, A. P., Andersson, J., Fivelsdal, L., Nilsson, M., & Domínguez‐Vicent, A. (2021) · PLoS ONE · doi:10.1371/journal.pone.0247670 · Cited by 7
Utility Values in the Visually Impaired
Gothwal, V. K., & Bagga, D. K. (2013) · Optometry and Vision Science · doi:10.1097/opx.0b013e318291063a · Cited by 7
Martínez-Albert, N., Esteve-Taboada, J. J., & Montés‐Micó, R. (2017) · Expert Review of Medical Devices · doi:10.1080/17434440.2017.1281739 · Cited by 7
MacGregor, S., Bykhovskaya, Y., Lindsay, R. G., Lechner, J., Haritunians, T., Henders, A., Siscovick, D. S., Anand, S. P., Aldave, A. J., Dj, C., et al. (2011) · Cited by 7
Agreement analysis: further statistical insights
Cesana, B. M., & Antonelli, P. (2012) · Ophthalmic and Physiological Optics · doi:10.1111/j.1475-1313.2012.00916.x · Cited by 7
Pilling, R. F., Allen, L. C., Anketell, P. M., Bullaj, R., Harwood, J., & Little, S. (2023) · British and Irish Orthoptic Journal · doi:10.22599/bioj.290 · Cited by 6
Salvador‐Roger, R., Esteve-Taboada, J. J., Venkataraman, A. P., & Domínguez‐Vicent, A. (2025) · Ophthalmic and Physiological Optics · doi:10.1111/opo.13482 · Cited by 6
Doyle, M., O’Dwyer, V. M., Moore, M., & Harrington, S. C. (2025) · Clinical and Experimental Optometry · doi:10.1080/08164622.2025.2542321 · Cited by 6
Validation of a Web-Based Platform for Online Training in Point-of-Care Diaphragm Ultrasound
Dugar, S. P., Mehkri, O., Li, M., Hastings, A., Siuba, M., Vashisht, R., Thiruchelvam, N., Wong, J., Vorona, S., Duggal, A., et al. (2022) · ATS Scholar · doi:10.34197/ats-scholar.2021-0057br · Cited by 6
Espinilla, O. G., Gallegos-Cocho, I., Sánchez, I., Cañadas, P., & Martín, R. (2022) · Clinical and Experimental Optometry · doi:10.1080/08164622.2021.2006042 · Cited by 6
Clinical validation of a novel smartphone application for measuring best corrected visual acuity
Ogino, M., Salmerón-Campillo, R. M., Hunter, S., Hussey, V. M., Suh, D. W., Gore, R. A., López‐Gil, N., Jaskulski, M., & Piña-Miguelsanz, D. (2023) · Journal of Optometry · doi:10.1016/j.optom.2023.01.001 · Cited by 7
Ortiz-Toquero, S., Fuente, C., Auladell, C., & Arnalich‐Montiel, F. (2023) · Life · doi:10.3390/life13071474 · Cited by 6
Aldaba, M., Ochando, P., Vila‐Vidal, N., Viñuela-Navarro, V., València, L. G., & Pérez-Corral, J. (2023) · Clinical and Experimental Optometry · doi:10.1080/08164622.2023.2277287 · Cited by 6
Jiménez‐García, M., Dhubhghaill, S. N., Hershko, S., Koppen, C., & Rozema, J. J. (2021) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12821 · Cited by 6
Optical Coherence Tomography Analysis of Retinal Layers in Celiac Disease
Vitiello, L., Bernardo, M. D., Erra, L., Rocca, F. D., Rosa, N., & Ciacci, C. (2022) · Journal of Clinical Medicine · doi:10.3390/jcm11164727 · Cited by 6
Said, A. G. A., Pinero, D. P., & Shneor, E. (2022) · Ophthalmic and Physiological Optics · doi:10.1111/opo.13066 · Cited by 6
Bai, M., Chen, J., Xu, Y., Feng, J. H., Ma, R., Jia, H., Xu, H., Liang, G., & Wei, H. (2023) · Scientific Reports · doi:10.1038/s41598-023-48975-9 · Cited by 7
An indirect method to compare the reference centres for corneal measurements
Xu, J., Bao, J., Lü, F., & He, J. C. (2011) · Ophthalmic and Physiological Optics · doi:10.1111/j.1475-1313.2011.00880.x · Cited by 6
Papadopoulou, D. N., Mangioris, G., Petropoulos, I. K., Mendrinos, E., Mavropoulos, A., & Pournaras, C. J. (2012) · Acta Ophthalmologica · doi:10.1111/j.1755-3768.2011.02374.x · Cited by 6
Huang, J., Liao, N., Savini, G., Li, Y., Bao, F., Ye, Y., Yu, A., & Wang, Q. (2014) · Cornea · doi:10.1097/ico.0000000000000329 · Cited by 6
Development of the Japanese version of the Visual Discomfort Scale
Imaizumi, S., Koyama, S., & Tanno, Y. (2018) · PLoS ONE · doi:10.1371/journal.pone.0191094 · Cited by 6
Comparison between two devices in the detection of corneal thickness changes after cataract surgery
Bernardo, M. D., Vitiello, L., Abbinante, G., Pascale, I. D., Capasso, L., Marotta, G., & Rosa, N. (2021) · Scientific Reports · doi:10.1038/s41598-021-86158-6 · Cited by 6
Martínez‐Alberquilla, I., Svensson, S., Ruiz‐Alcocer, J., Madrid‐Costa, D., Domínguez‐Vicent, A., & Venkataraman, A. P. (2024) · Scientific Reports · doi:10.1038/s41598-024-73206-0 · Cited by 5
Samolov, B., van de Moosdijk, S., Venkataraman, A. P., & Domínguez‐Vicent, A. (2025) · Journal of Optometry · doi:10.1016/j.optom.2025.100535 · Cited by 5
Tzamalis, A. K., Tsiampali, C., Prousali, E., Mataftsi, A., Ziakas, N. G., & Kynigopoulos, M. (2021) · Optometry and Vision Science · doi:10.1097/opx.0000000000001781 · Cited by 5
Clinical Evaluation Of a 0.05 D-step Binocular Wavefront Optometer in Young Adults in China
Cheng, M., Chen, X., Lei, Y., Li, B., Jiang, Y., Xu, Y., Zhou, X., & Wang, X. (2023) · Clinical and Experimental Optometry · doi:10.1080/08164622.2023.2175603 · Cited by 5
Xiong, Y., Lin, Y., Zhao, Z., Wang, H., & Zhang, G. (2023) · BMC Ophthalmology · doi:10.1186/s12886-023-02962-w · Cited by 5
Nieto, L. V., la Rosa, A. L., González-García, M. J., & López‐Miguel, A. (2023) · Ophthalmic and Physiological Optics · doi:10.1111/opo.13243 · Cited by 5
Enders, C., Baeurle, F., Lang, G. E., Dreyhaupt, J., Trick, S., Kilani, A., Lang, G. K., & Werner, J. U. (2019) · Klinische Monatsblätter für Augenheilkunde · doi:10.1055/a-0983-2271 · Cited by 5
Agreement in central corneal thickness measurement between Corvis ST and ocular response analyzer
Salouti, R., Razeghinejad, R., Masihpour, N., Ghoreyshi, M., & Nowroozzadeh, M. H. (2020) · International Ophthalmology · doi:10.1007/s10792-020-01436-y · Cited by 5
Placide, J., da Silva, H. V. N., McCabe, S. E., Ronquillo, Y. C., & Moshirfar, M. (2021) · Expert Review of Medical Devices · doi:10.1080/17434440.2021.2012153 · Cited by 5
Ocular anterior segment and corneal parameters evaluation in celiac disease
Bernardo, M. D., Vitiello, L., Gagliardi, M., Capasso, L., Rosa, N., & Ciacci, C. (2022) · Scientific Reports · doi:10.1038/s41598-022-06058-1 · Cited by 5
Validation of the DYOP visual acuity test
Gantz, L., Paritzky, D., Wunch, I., Kageyama, A. H., Wolf, N., Cherny, C., & Rosenfield, M. (2022) · Journal of Optometry · doi:10.1016/j.optom.2022.09.006 · Cited by 5
Zahavi, O., Domínguez‐Vicent, A., Brautaset, R., & Venkataraman, A. P. (2021) · Applied Sciences · doi:10.3390/app11031250 · Cited by 5
Reliability of retinal vessel calibre measurements using a retinal oximeter
Heitmar, R., & Kalitzeos, A. A. (2015) · BMC Ophthalmology · doi:10.1186/s12886-015-0174-0 · Cited by 5
Elliott, D. B. (2014) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12153 · Cited by 5
Martínez-Albert, N., Nieto-Fernández, J. C., & García‐Marqués, J. V. (2023) · Journal of Refractive Surgery · doi:10.3928/1081597x-20230202-02 · Cited by 4
Lal, B., Alonso‐Caneiro, D., Read, S. A., & Carkeet, A. (2024) · Ophthalmic and Physiological Optics · doi:10.1111/opo.13357 · Cited by 4
Recchioni, A., Venkataraman, A. P., Rauz, S., & Domínguez‐Vicent, A. (2024) · Frontiers in Medicine · doi:10.3389/fmed.2024.1385294 · Cited by 4
Romano, D., Coco, G., Borgia, A., Calza, S., Kaye, S. B., Gadhvi, K. A., Semeraro, F., & Romano, V. (2025) · Journal of Clinical Medicine · doi:10.3390/jcm14020439 · Cited by 4
Lei, C. S., Li, X., Xu, H., Ning, R., Wang, Y., Savini, G., Schiano‐Lomoriello, D., Zhou, X., Li, K., & Huang, J. (2025) · Journal of Refractive Surgery · doi:10.3928/1081597x-20250317-03 · Cited by 4
Doostparast, A., Semnani, F., Ghandhari, M., Ghandhari, M., Khosronejad, A. H., Ahmadi, A., & Eslampoor, A. (2026) · Scientific Reports · doi:10.1038/s41598-026-42204-9 · Cited by 4
Baskaran, A., Britto, T., Sowndher, R. T., & Thomas, P. A. (2021) · Saudi Journal of Ophthalmology · doi:10.4103/1319-4534.325776 · Cited by 4
Differences in stereoacuity between crossed and uncrossed disparities reduce with practice
Clayton, R., & Siderov, J. (2022) · Ophthalmic and Physiological Optics · doi:10.1111/opo.13040 · Cited by 4
Metadata
- Authors
- McAlinden, C., Khadka, J., & Pesudovs, K.
- Affiliations
- Flinders University (AU)
- Corresponding author
- McAlinden, Colm
- Venue
- Ophthalmic and Physiological Optics
- Published
- 2011-05-26
- Volume
- 31
- Issue
- 4
- Pages
- 330-338
- ISSN
- 0275-5408 (ISSN-L)
- 1475-1313
- OpenAlex
- W2126376404
- Cited by
- 436
- References
- 49
- Citations by year
- 2026: 22 · 2025: 24 · 2024: 33 · 2023: 30 · 2022: 37 · 2021: 42 · 2020: 47 · 2019: 33 · 2018: 22 · 2017: 28 · 2016: 27 · 2015: 27 · 2014: 20 · 2013: 15 · 2012: 22
- FWCI
- 5.01
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Ophthalmic and Physiological Optics (publishedVersion)
- PubMed (publishedVersion)
- Adelaide Research & Scholarship (AR&S) (University of Adelaide) (submittedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Health Sciences › Medicine › Ophthalmology › Ophthalmology and Visual Impairment Studies
- Keywords
- repeatability, reproducibility, limits of agreement, correlation coefficient, ophthalmic instruments, sample size, measurement precision, optometry
- MeSH terms
- Humans, Ophthalmology, Optometry, Vision, Ocular, Reproducibility of Results, Qualitative Research
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review