Comparison of Time‐Domain Short‐Term Heart Interval Variability Analysis Using a Wrist‐Worn Heart Rate Monitor and the Conventional Electrocardiogram

PORTO, LUIZ GUILHERME GROSSI, JUNQUEIRA Jr, LUIZ F. · 2009 · Pacing and Clinical Electrophysiology
Cited by 135

The DOI matches its Crossref record; 8 of the 8 references it cites were found in a registry or an index.

Abstract

Background: Wrist‐worn heart rate monitors have not been extensively validated for heart rate variability analysis. The purpose of this study was to compare time‐domain variability of heart interval series (R‐Ri) recorded by the Polar S810 monitor (Polar Electro Oy, Kempele, Finland) and the conventional electrocardiogram (ECG). Methods: Agreement was verified between variability indices of 5‐minute R‐Ri simultaneously recorded by both devices and processed by unique software, from 33 subjects aged 18 to 42 years, normal or with different clinical conditions, in rest supine and active standing. ECG minus Polar differences were quantified by the Bland‐Altman analysis, and tested by the one‐sample t‐test or Wilcoxon test. Results: In the supine position, the Polar overestimates (P < 0.0001) the absolute and percentage mean or median of the number (−2.00; −0.49%) and mean of R‐Ri (–1.85 ms; –0.20%) and pNN50 (−2.20%; −8.68%), and underestimates the standard deviation (SDNN) (0.32 ms; 0.59%) (P = 0.08; P = 0.02) and root mean square successive difference (RMSSD) (0.90 ms; 1.56%) (P = 0.0008; P < 0.0001). The coefficient of variation (CV) showed null difference. On standing, differences were overestimated for the number (−2.61 intervals; −0.64%) and mean of R‐Ri (−0.70 ms; −0.09%), and underestimated for rMSSD (1.70 ms; 10.84%) (P < 0.0001 to < 0.02). The SDNN, CV, and pNN50 indices did not show differences (P = 0.12 to 0.73). Conclusions: The Polar S810 monitor was feasible and reliable for recording short‐term R‐R interval series, showing excellent agreement with the ECG in providing the time‐domain indexes of heart interval variability with differences functionally not relevant. The CV showed the higher agreement in both postures, and the SDNN and pNN50 in the standing posture.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1111/j.1540-8159.2009.02175.x

Cites

8 of 8 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 8 references

Cited by

See all 135

Metadata

Authors
PORTO, L. G. G., & JUNQUEIRA Jr, L. F.
Venue
Pacing and Clinical Electrophysiology
Publisher
Wiley
Published
2008-12-23
Volume
32
Issue
1
Pages
43-51
ISSN
  • 0147-8389 (ISSN-L)
  • 1540-8159
OpenAlex
W2073792739
Cited by
135
References
8
FWCI
5.50
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Comparison of Time‐Domain Short‐Term Heart Interval Variability Analysis Using a Wrist‐Worn Heart Rate Monitor and the Conventional Electrocardiogram. https://app.cochat.ai/papers/10.1111/j.1540-8159.2009.02175.x
The paper
PORTO, L. G. G., & JUNQUEIRA Jr, L. F. (2009). Comparison of Time‐Domain Short‐Term Heart Interval Variability Analysis Using a Wrist‐Worn Heart Rate Monitor and the Conventional Electrocardiogram. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/j.1540-8159.2009.02175.x

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.