Determinants of Lesion Size During Radiofrequency Catheter Ablation: The Role of Electrode‐Tissue Contact Pressure and Duration of Energy Delivery
The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
Determinants of Radiofrequency Lesion Size. Introduction: Radiofrequency (RF) catheter ablation is a promising modality in the management of cardiac arrhythmias, but the optimum protocol for energy delivery has not yet been determined. The purpose of this study was to examine the effects of varying electrode‐tissue contact pressure and varying duration of RF energy delivery on the size of the resultant lesion in an in vitro model of isolated perfused and superfused canine myocardium. Methods and Results: RF power (500 kHz) was delivered through the tip of a specially‐designed thermistor‐tipped 1.6‐mm diameter electrode mounted on an adjustable fulcrum so that contact force could be varied. RF power was adjusted to maintain a constant electrode‐tissue interface temperature of 80°C during lesion production. Seventy‐nine lesions were created with a 90‐second RF energy delivery, and a contact force that was varied between 0 and 400 newtons (N) (0‐41 g). Lesions produced with a small contact force (10 N) were significantly larger than those with a contact force of zero (width 5.5 mm vs 3.8 mm, P <0.0002), but not significantly different from those produced with the maximum contact force of 400 N (width 6.5 mm, P = NS). However, the greater contact force significantly decreased the power required to maintain a constant electrode‐tissue interface temperature. Ninety‐six lesions were then created using a constant contact force, but duration of energy delivery was varied from 10 to 600 seconds. Lesion size grew monoexponentially with time. The t 1/2 of lesion growth was 7.6 and 9.6 seconds for depth and width, respectively. Conclusion: Thus, close electrode‐tissue contact is essential for adequate lesion formation during RF ablation of myocardium, but increasing contact force does not significantly increase lesion size if power is adjusted to maintain a constant electrode‐tissue interface temperature. In order to approach steady‐state and maximize lesion size, duration of RF energy delivery should be at least 40 seconds.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
High frequency alternating current ablation of an accessory pathway in humans
Borggrefe, M., Budde, T., Podczeck, A., & Breithardt, G. (1987) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(87)80200-0 · Cited by 285
Catheter ablation of the atrioventricular junction with radiofrequency energy.
Langberg, J. J., Chin, M. C., Rosenqvist, M., Cockrell, J., Dullet, N., Van Hare, G., Griffin, J. C., & Scheinman, M. M. (1989) · Circulation · doi:10.1161/01.cir.80.6.1527 · Cited by 245
Epstein, L. M., Scheinman, M. M., Langberg, J. J., Chilson, D. A., Goldberg, H., & Griffin, J. C. (1989) · Circulation · doi:10.1161/01.cir.80.4.757 · Cited by 126
Ring, M. E., Huang, S. K. S., Gorman, G. D., & Graham, A. R. (1989) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1989.tb06155.x · Cited by 75
Cited by
Calkins, H., Hindricks, G., Cappato, R., Kim, Y.-H., Saad, E. B., Aguinaga, L., Akar, J. G., Badhwar, V., Brugada, J., Camm, J., et al. (2017) · Heart Rhythm · doi:10.1016/j.hrthm.2017.05.012 · Cited by 2,755
Self-citationCalkins, H., Hindricks, G., Cappato, R., Kim, Y.-H., Saad, E. B., Aguinaga, L., Akar, J. G., Badhwar, V., Brugada, J., Camm, J., et al. (2018) · EP Europace · doi:10.1093/europace/eux274 · Cited by 1,164
Self-citationYokoyama, K., Nakagawa, H., Shah, D. C., Lambert, H., Leo, G., Aeby, N., Ikeda, A., Pitha, J. V., Sharma, T., Lazzara, R., et al. (2008) · Circulation: Arrhythmia and Electrophysiology · doi:10.1161/circep.108.803650 · Cited by 624
Reddy, V. Y., Shah, D., Kautzner, J., Schmidt, B., Saoudi, N., Herrera, C., Jaïs, P., Hindricks, G., Peichl, P., Yulzari, A., et al. (2012) · Heart Rhythm · doi:10.1016/j.hrthm.2012.07.016 · Cited by 484
Taghji, P., El Haddad, M., Phlips, T., Wolf, M., Knecht, S., Vandekerckhove, Y., Tavernier, R., Nakagawa, H., & Duytschaever, M. (2018) · JACC: Clinical Electrophysiology · doi:10.1016/j.jacep.2017.06.023 · Cited by 350
Metadata
- Authors
- HAINES, D. E.
- Venue
- Journal of Cardiovascular Electrophysiology
- Publisher
- Wiley
- Published
- 1991-12
- Volume
- 2
- Issue
- 6
- Pages
- 509-515
- ISSN
- 1045-3873 (ISSN-L)
- 1540-8167
- OpenAlex
- W1995216474
- Cited by
- 264
- References
- 4
- FWCI
- 6.12
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.