Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and Thrombus
Open accessAbstract
Background— An open-irrigated radiofrequency (RF) ablation catheter was developed to measure contact force (CF). Three optical fibers measure microdeformation of the catheter tip. The purpose of this study was to (1) validate the accuracy of CF sensor (CFS) (bench test); and (2) determine the relationship between CF and tissue temperatures, lesion size, steam pop, and thrombus during RF ablation using a canine thigh muscle preparation. Methods and Results— CFS measurements (total 1409) from 2 catheters in 3 angles (perpendicular, parallel, and 45�) were compared with a certified balance (range, 0 to 50 g). CFS measurements correlated highly ( R 2 ≥0.988; mean error, ≤1.0 g). In 10 anesthetized dogs, a skin cradle over the thigh muscle was superfused with heparinized blood at 37�C. A 7F catheter with 3.5-mm saline-irrigated electrode and CFS (Endosense) was held perpendicular to the muscle at CF of 2, 10, 20, 30, and 40 g. RF was delivered (n=100) for 60 seconds at 30 or 50 W (irrigation 17 or 30 mL/min). Tissue temperature (3 and 7 mm depths), lesion size, thrombus, and steam pop increased significantly with increasing CF at each RF power. Lesion size was greater with applications of lower power (30 W) and greater CF (30 to 40 g) than at high power (50 W) with lower CF (2 to 10 g). Conclusions— This novel ablation catheter, which accurately measures CF, confirmed CF is a major determinant of RF lesion size. Steam pop and thrombus incidence also increases with CF. CFS in an open-irrigated ablation catheter that may optimize the selection of RF power and application time to maximize lesion formation and reduce the risk of steam pop and thrombus.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Nakagawa, H., Yamanashi, W. S., Pitha, J. V., Arruda, M., Wang, X., Ohtomo, K., Beckman, K. J., McClelland, J. H., Lazzara, R., & Jackman, W. M. (1995) · Circulation · doi:10.1161/01.cir.91.8.2264 · Cited by 625
Self-citationNakagawa, H., Shah, N., Matsudaira, K., Overholt, E., Chandrasekaran, K., Beckman, K. J., Spector, P., Calame, J. D., Rao, A., Hasdemir, C., et al. (2001) · Circulation · doi:10.1161/01.cir.103.5.699 · Cited by 380
Self-citationHAINES, D. E. (1991) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.1991.tb01353.x · Cited by 264
Yokoyama, K., Nakagawa, H., Wittkampf, F. H. M., Pitha, J. V., Lazzara, R., & Jackman, W. M. (2006) · Circulation · doi:10.1161/circulationaha.105.540062 · Cited by 243
Self-citationNakagawa, H., Wittkampf, F. H. M., Yamanashi, W. S., Pitha, J. V., Imai, S., Campbell, B., Arruda, M., Lazzara, R., & Jackman, W. M. (1998) · Circulation · doi:10.1161/01.cir.98.5.458 · Cited by 208
Self-citationZHOU, L., KEANE, D., REED, G., & RUSKIN, J. (1999) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.1999.tb00719.x · Cited by 185
Efficacy and safety of septal and left-atrial linear ablation for atrial fibrillation
Jaı̈s, P., Shah, D. C., Haı̈ssaguerre, M., Takahashi, A., Lavergne, T., Hocini, M., Garrigue, S., Barold, S. S., Le Métayer, P., & Clémenty, J. (1999) · The American Journal of Cardiology · doi:10.1016/s0002-9149(99)00714-6 · Cited by 173
Self-citationCalkins, H. G., Prystowsky, E. N., Carlson, M. D., Klein, L. S., Saul, J. P., & Gillette, P. C. (1994) · Circulation · doi:10.1161/01.cir.90.3.1279 · Cited by 140
Matsudaira, K., Nakagawa, H., Wittkampf, F. H. M., Yamanashi, W. S., Imai, S., Pitha, J. V., Lazzara, R., & Jackman, W. M. (2003) · Pacing and Clinical Electrophysiology · doi:10.1046/j.1460-9592.2003.t01-1-00173.x · Cited by 93
Self-citationYamane, T., Jaı̈s, P., Shah, D. C., Hocini, M., Peng, J. T., Deisenhofer, I., Clémenty, J., & Haı̈ssaguerre, M. (2000) · Circulation · doi:10.1161/01.cir.102.21.2565 · Cited by 85
Self-citation
Cited by
Calkins, H., Hindricks, G., Cappato, R., Kim, Y.-H., Saad, E. B., Aguinaga, L., Akar, J. G., Badhwar, V., Brugada, J., Camm, J., et al. (2017) · Heart Rhythm · doi:10.1016/j.hrthm.2017.05.012 · Cited by 2,755
Self-citationCalkins, H., Kuck, K. H., Cappato, R., Brugada, J., Camm, A. J., Chen, S.-A., Crijns, H. J. G., Damiano, R. J., Davies, D. W., DiMarco, J., et al. (2012) · Europace · doi:10.1093/europace/eus027 · Cited by 2,722
Self-citationCalkins, H., Kuck, K. H., Cappato, R., Brugada, J., Camm, A. J., Chen, S.-A., Crijns, H. J. G., Damiano, R. J., Davies, D. W., DiMarco, J., et al. (2012) · Heart Rhythm · doi:10.1016/j.hrthm.2011.12.016 · Cited by 1,456
Self-citationTzeis, S., Gerstenfeld, E. P., Kalman, J., Saad, E. B., Shamloo, A. S., Andrade, J. G., Barbhaiya, C. R., Baykaner, T., Boveda, S., Calkins, H., et al. (2024) · Heart Rhythm · doi:10.1016/j.hrthm.2024.03.017 · Cited by 298
Medical Smart Textiles Based on Fiber Optic Technology: An Overview
Massaroni, C., Saccomandi, P., & Schena, E. (2015) · Journal of Functional Biomaterials · doi:10.3390/jfb6020204 · Cited by 172
Metadata
- Authors
- Yokoyama, K., Nakagawa, H., Shah, D. C., Lambert, H., Leo, G., Aeby, N., Ikeda, A., Pitha, J. V., Sharma, T., Lazzara, R., & Jackman, W. M.
- Venue
- Circulation: Arrhythmia and Electrophysiology
- Published
- 2008-12
- Volume
- 1
- Issue
- 5
- Pages
- 354-362
- ISSN
- 1941-3084 (ISSN-L)
- 1941-3149
- OpenAlex
- W2130349647
- Cited by
- 624
- References
- 13
- FWCI
- 11.88
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and Thrombus. https://app.cochat.ai/papers/10.1161/circep.108.803650
- The paper
- Yokoyama, K., Nakagawa, H., Shah, D. C., Lambert, H., Leo, G., Aeby, N., Ikeda, A., Pitha, J. V., Sharma, T., Lazzara, R., & Jackman, W. M. (2008). Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and Thrombus. Circulation: Arrhythmia and Electrophysiology. https://doi.org/10.1161/circep.108.803650
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.