An overlooked aspect on metabolic acidosis at birth: blood gas analyzers calculate base deficit differently
Open accessAbstract
OBJECTIVE: Metabolic acidosis (MA) at birth is commonly defined as umbilical cord arterial pH < 7.0 plus base deficit (BD) ≥ 12.0 mmol/L. Base deficit is not a measured entity but is calculated from pH and Pco(2) values, with the hemoglobin (Hb) concentration [Hb] included in the calculation algorithm as a fixed or measured value. Various blood gas analyzers use different algorithms, indicating variations in the MA diagnosis. The objective was therefore to calculate the prevalence of MA in blood and extracellular fluid with algorithms from three blood gas analyzer brands relative to the Clinical and Laboratory Standards Institute (CLSI) algorithm. DESIGN: Comparative study. SETTING: University hospital. SAMPLE: Arterial cord blood from 15 354 newborns. MAIN OUTCOME MEASURE: Prevalence of MA. METHODS: Blood was analyzed in a Radiometer ABL 735 analyzer. Base deficit was calculated post hoc with algorithms from CLSI and Corning and Roche blood gas analyzers, and with measured and fixed (9.3 mmol/L) values of [Hb]. RESULTS: The prevalence of BD ≥12.0 mmol/L in blood was with the CLSI algorithm 1.97%, Radiometer 5.18%, Corning 3.84% and Roche 3.29% (CLSI vs. other; McNemar test, p < 0.000001). Likewise, MA prevalences were 0.58, 0.66, 0.64 and 0.64%, respectively (p≤ 0.02). Base deficit ≥ 12.0 mmol/L and MA rates were lower in extracellular fluid than in blood (p≤ 0.002). Algorithms with measured or fixed Hb concentration made no differences to MA rates (p≥ 0.1). CONCLUSIONS: The neonatal metabolic acidosis rate varied significantly with blood gas analyzer brand and fetal fluid compartment for calculation of BD.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
When calculating with different equations and in different BD compartments, we found an interanalyzer difference in diagnosing BD >12.0 mmol/L of up to 426%.
“When calculating with different equations and in different BD compartments (provided below), we found an interanalyzer difference in diagnosing BD >12.0 mmol/L of up to 426%.”
From Umbilical cord pH, blood gases, and lactate at birth: normal values, interpretation, and clinical utility. (2023) · cites this paper as [24]
The Results and Discussion distinguish interanalyzer from intra-analyzer comparisons. The maximum interanalyzer difference was 163% for BD in blood and 210% for BD in extracellular fluid; 426% was the maximum intra-analyzer difference between compartments. (LLM, conf=0.94)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
VassarStats : website for statistical computation
VassarStats (2010) · Medical Entomology and Zoology · Cited by 438 · see citing works
Amer‐Wåhlin, I., Hellsten, C., Norén, H., Hagberg, H., Herbst, A., Kjellmer, I., Lilja, H., Lindoff, C., Månsson, M., Mårtensson, L., et al. (2001) · The Lancet · doi:10.1016/s0140-6736(01)05703-8 · Cited by 397 · see citing works
Self-citationThreshold of metabolic acidosis associated with newborn complications
Low, J. A., Lindsay, B. G., & Derrick, E. J. (1997) · American Journal of Obstetrics and Gynecology · doi:10.1016/s0002-9378(97)70080-2 · Cited by 316 · see citing works
Westgate, J., Harris, M., Curnow, J. S. H., & Greene, K. R. (1993) · American Journal of Obstetrics and Gynecology · doi:10.1016/0002-9378(93)90273-l · Cited by 234 · see citing works
ACOG Committee Opinion No. 348: Umbilical Cord Blood Gas and Acid-Base Analysis
(2006) · Obstetrics & Gynecology · doi:10.1097/00006250-200611000-00058 · Cited by 211 · see citing works
Ojala, K., Vääräsmäki, M., Mäkikallio, K., Valkama, M., & Tekay, A. (2006) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/j.1471-0528.2006.00886.x · Cited by 164 · see citing works
Westerhuis, M. E. M. H., Visser, G. H. A., Moons, K. G. M., van Beek, E., Benders, M. J. N. L., Bijvoet, S. M., Dessel, H. J. H. M. V., Drogtrop, A. P., van Geijn, H. P., Graziosi, G. C. M., et al. (2010) · Obstetrics and Gynecology · doi:10.1097/aog.0b013e3181dfffd6 · Cited by 124 · see citing works
Wiberg, N., Källén, K., & Olofsson, P. (2008) · BJOG: An International Journal of Obstetrics & Gynaecology · doi:10.1111/j.1471-0528.2008.01708.x · Cited by 121 · see citing works
Self-citationWiberg, N., Källén, K., Herbst, A., & Olofsson, P. (2010) · Acta Obstetricia Et Gynecologica Scandinavica · doi:10.3109/00016349.2010.513426 · Cited by 92 · cited 1× · see citing works
“In a recent study, we found that lactic acid (lactate) had a better accuracy than BDblood to reflect a five minute Apgar score < 7, both with regard to crude values and to gestational age-adjusted values (24).”
Wiberg, N., Källén, K., & Olofsson, P. (2006) · American Journal of Obstetrics and Gynecology · doi:10.1016/j.ajog.2006.05.043 · Cited by 69 · see citing works
Self-citationSTAN in clinical practice—The outcome of 2 years of regular use in the city of Gothenburg
Norén, H., Blad, S., Carlsson, A., et al. (2006) · American Journal of Obstetrics and Gynecology · doi:10.1016/j.ajog.2006.01.108 · Cited by 67 · see citing works
Kofstad, J. (2001) · Clinica Chimica Acta · doi:10.1016/s0009-8981(01)00427-2 · Cited by 64 · see citing works
Luttkus, A. K., Norén, H., Stupin, J. H., et al. (2004) · Journal of Perinatal Medicine · doi:10.1515/jpm.2004.121 · Cited by 58 · see citing works
Accuracy of base excess—An in vitro evaluation of the Van Slyke equation
Morgan, T. J., Clark, C., & Endre, Z. H. (2000) · Critical Care Medicine · doi:10.1097/00003246-200008000-00041 · Cited by 54 · see citing works
Frasch, M. G., Mansano, R. Z., Gagnon, R. J., et al. (2008) · American Journal of Obstetrics and Gynecology · doi:10.1016/j.ajog.2008.10.022 · Cited by 53 · see citing works
How to assess fetal metabolic acidosis front cord samples
Rosén, K. G., & Murphy, K. W. (1991) · jpme · doi:10.1515/jpme.1991.19.3.221 · Cited by 43 · see citing works
Wiberg, N., Källén, K., & Olofsson, P. (2006) · Early Human Development · doi:10.1016/j.earlhumdev.2005.12.005 · Cited by 39 · cited 1× · see citing works
“Base deficit values in umbilical cord blood vary not only with the calculation algorithm, but are also influenced by gestational age, length of labor, fetal fluid compartment for calculation and time of blood sampling after birth (1,10,11).”
BackgroundSelf-citationNorén, H., Luttkus, A. K., Stupin, J. H., et al. (2007) · Journal of Perinatal Medicine · doi:10.1515/jpm.2007.097 · Cited by 32 · see citing works
STAN, a clinical audit: the outcome of 2 years of regular use in the city of Varberg, Sweden
Welin, A.-K., Norén, H., Odeback, A., et al. (2007) · Acta Obstetricia Et Gynecologica Scandinavica · doi:10.1080/00016340701416903 · Cited by 22 · see citing works
Amer‐Wåhlin, I., & Rosén, K. G. (2006) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/j.1471-0528.2006.01098.x · Cited by 9 · see citing works
Westgate, J. (1993) · OpenGrey (Institut de l'Information Scientifique et Technique) · doi:10.24382/4871 · Cited by 8 · see citing works
Roemer, V. M. (2011) · Zeitschrift für Geburtshilfe und Neonatologie · doi:10.1055/s-0031-1271756 · Cited by 6 · see citing works
Flagged first, then by citations
Research works printed but not linked: 1
- Norén H, Luttkus AK, Stupin JH, Blad S, Arulkumaran S, Erkkola R, et al. Fetal scalp pH and ST analysis of the fetal ECG as an adjunct to cardiotocography to predict fetal acidosis in labor. J Perinat Med. 2007;35:408-14.
Cited by
Sorted by citations
Olofsson, P. (2023) · American Journal of Obstetrics and Gynecology · doi:10.1016/j.ajog.2022.07.001 · Cited by 104 · see citing works
Self-citationZaigham, M., Källén, K., Maršál, K., & Olofsson, P. (2020) · Acta Paediatrica · doi:10.1111/apa.14925 · Cited by 9 · see citing works
Self-citation
- Authors
- MOKARAMI, P., WIBERG, N., & OLOFSSON, P.
- Affiliations
- Lund University (SE)
- Venue
- Acta Obstetricia Et Gynecologica Scandinavica
- Publisher
- Wiley
- Published
- 2012-03-12
- Volume
- 91
- Issue
- 5
- Pages
- 574-579
- ISSN
- 0001-6349 (ISSN-L)
- 1600-0412
- OpenAlex
- W2158982304
- Cited by
- 28
- References
- 22
- Citations by year
- 2026: 1 · 2025: 1 · 2024: 4 · 2023: 2 · 2021: 3 · 2020: 2 · 2019: 5 · 2016: 3 · 2015: 3 · 2014: 2 · 2013: 2
- FWCI
- 1.77
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Acta Obstetricia Et Gynecologica Scandinavica (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Lunds Universitet
- Region Skåne
- Topics
- Health Sciences › Medicine › Nephrology › Renal function and acid-base balance
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Clinical Biochemistry › Metabolism and Genetic Disorders
- Keywords
- neonatal metabolic acidosis, base deficit, umbilical cord blood, blood gas analyzer, calculation algorithm, hemoglobin levels, extracellular fluid, cordocentesis, neonates
- MeSH terms
- Acidosis, Algorithms, Blood Gas Analysis, Fetal Blood, Humans, Infant, Newborn, Prevalence, Extracellular Fluid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review