The DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.
Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus
Open accessAbstract
Objective: To evaluate the performance of the FreeStyle Libre Flash continuous glucose monitoring (FSL-CGM) system against established central laboratory methods. Research design and methods: 20 subjects (8 type 1 diabetes mellitus, 12 type 2 diabetes mellitus) were analyzed. FSL-CGM sensor measurements (inserted in arm and abdomen) were compared with capillary blood glucose results analyzed with StatStrip as semigold standard. The glucose response after a standardized oral glucose load was measured by FSL-CGM and capillary samples analyzed by perchloric acid hexokinase (PCA-HK) method, StatStrip and FSL test strip (FSLC), and a commonly used CGM system (iPro2). Results: FSL-CGM arm sensor readings showed 85.5% of paired readings falling within Clarke Error Grid (ISO 15197:2013) zone A when compared with StatStrip. For FSL-CGM abdomen and FSLC, these percentages were 64% and 98%, respectively. The overall correlation of FSL-CGM in the arm and the StatStrip indicates a performance with lower results with the FSL-CGM in the arm than expected based on the StatStrip in the lower glucose ranges, and higher results than expected in the higher ranges. Following a standardized glucose load, a slower rise in glucose level was observed for FSL-CGM arm as compared with PCA-HK, StatStrip, FSLC, and iPro2 during the first 45–60 min after glucose load ingestion. Conclusions: Certain matters need attention while using the FSL-CGM in daily life including the observed lower values in the lower ranges, and the underestimation of the effect of a meal on glucose response. These effects of such deviations can partly be overcome by optimizing the available user instructions. Trial registration number: TC5348; results.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
Bailey, T., Bode, B. W., Christiansen, M. P., Klaff, L. J., & Alva, S. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0378 · Cited by 685 · cited 1×
“Independent accuracy assessments of the FSL-CGM are scarce; previous assessments were performed comparing the FSL-CGM to capillary blood evaluated with a point of care (POC) measurement.(10) Whether such a comparison can be seen as sufficient accurate remains to be seen, however.”
MethodPrice, M. J., Ades, A., Soldan, K., Welton, N. J., Macleod, J., Simms, I., DeAngelis, D., Turner, K. M., & Horner, P. J. (2016) · Health Technology Assessment · doi:10.3310/hta20220 · Cited by 494
Delays in Minimally Invasive Continuous Glucose Monitoring Devices: A Review of Current Technology
Keenan, D. B., Mastrototaro, J. J., Voskanyan, G., & Steil, G. M. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300528 · Cited by 198
Poolsup, N., Suksomboon, N., & Kyaw, A. M. (2013) · Diabetology & Metabolic Syndrome · doi:10.1186/1758-5996-5-39 · Cited by 179
The Accuracy of Point-of-Care Glucose Measurements
Rebel, A., Rice, M. A., & Fahy, B. G. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600228 · Cited by 144
Burnett, D. R., Huyett, L. M., Zisser, H. C., Doyle, F. J., & Mensh, B. D. (2014) · Diabetes · doi:10.2337/db13-1649 · Cited by 52
Cited by
Charleer, S., De Block, C., Van Huffel, L., Broos, B., Fieuws, S., Nobels, F., Mathieu, C., & Gillard, P. (2020) · Diabetes Care · doi:10.2337/dc19-1610 · Cited by 261
7. Diabetes Technology: Standards of Medical Care in Diabetes—2019
American Diabetes Association (2019) · Diabetes Care · doi:10.2337/dc19-s007 · Cited by 179
Fokkert, M., van Dijk, P., Edens, M., Barents, E., Mollema, J., Slingerland, R., Gans, R., & Bilo, H. (2019) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2019-000809 · Cited by 163
Self-citationAlva, S., Bailey, T., Brazg, R., Budiman, E. S., Castorino, K., Christiansen, M. P., Forlenza, G., Kipnes, M., Liljenquist, D. R., & Liu, H. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820958754 · Cited by 164
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
Self-citation
Metadata
- Authors
- Fokkert, M. J., van Dijk, P. R., Edens, M. A., Abbes, S., de Jong, D., Slingerland, R. J., & Bilo, H. J. G.
- Venue
- BMJ Open Diabetes Research & Care
- Publisher
- BMJ
- Published
- 2017-02
- Volume
- 5
- Issue
- 1
- Pages
- e000320
- ISSN
- 2052-4897 (ISSN-L)
- OpenAlex
- W2587847154
- Cited by
- 209
- References
- 6
- FWCI
- 19.11
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus. https://app.cochat.ai/papers/10.1136/bmjdrc-2016-000320
- The paper
- Fokkert, M. J., van Dijk, P. R., Edens, M. A., Abbes, S., de Jong, D., Slingerland, R. J., & Bilo, H. J. G. (2017). Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus. BMJ Open Diabetes Research & Care. https://doi.org/10.1136/bmjdrc-2016-000320
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.