Glucose Sensing in the Peritoneal Space Offers Faster Kinetics Than Sensing in the Subcutaneous Space
Open accessThe DOI matches its Crossref record; 47 of the 47 references it cites were found in a registry or an index.
Abstract
The paramount goal in the treatment of type 1 diabetes is the maintenance of normoglycemia. Continuous glucose monitoring (CGM) technologies enable frequent sensing of glucose to inform exogenous insulin delivery timing and dosages. The most commonly available CGMs are limited by the physiology of the subcutaneous space in which they reside. The very same advantages of this minimally invasive approach are disadvantages with respect to speed. Because subcutaneous blood flow is sensitive to local fluctuations (e.g., temperature, mechanical pressure), subcutaneous sensing can be slow and variable. We propose the use of a more central, physiologically stable body space for CGM: the intraperitoneal space. We compared the temporal response characteristics of simultaneously placed subcutaneous and intraperitoneal sensors during intravenous glucose tolerance tests in eight swine. Using compartmental modeling based on simultaneous intravenous sensing, blood draws, and intraarterial sensing, we found that intraperitoneal kinetics were more than twice as fast as subcutaneous kinetics (mean time constant of 5.6 min for intraperitoneal vs. 12.4 min for subcutaneous). Combined with the known faster kinetics of intraperitoneal insulin delivery over subcutaneous delivery, our findings suggest that artificial pancreas technologies may be optimized by sensing glucose and delivering insulin in the intraperitoneal space.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures





Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“As regards IP sensing, we simulated the model proposed by Burnett et al 142 which has an average time delay of 0.68 minutes and a time constant of 5.6 minutes.”
From Developing the UVA/Padova Type 1 Diabetes Simulator: Modeling, Validation, Refinements, and Utility (2023) · cites this paper as [142]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
Hypoglycemia in the Diabetes Control and Complications Trial
The Diabetes Control and Complications Trial Research Group (1997) · Diabetes · doi:10.2337/diab.46.2.271 · Cited by 746
The barrier of hypoglycemia in diabetes
Cryer, P. E. (2008) · Diabetes · doi:10.2337/db08-1084 · Cited by 725
Methods for reducing biosensor membrane biofouling
Wisniewski, N., & Reichert, M. (2000) · Colloids and Surfaces B: Biointerfaces · doi:10.1016/s0927-7765(99)00148-4 · Cited by 543 · cited 1×
“Thirdly, subcutaneous kinetics have been variably reported to be moderately slow (17–25) and likely worsen with implantation time as the encapsulation develops (26,27).”
BackgroundManagement of occlusion and thrombosis associated with long-term indwelling central venous catheters
Baskin, J. L., Pui, C.-H., Reiss, U., Wilimas, J. A., Metzger, M. L., Ribeiro, R. C., & Howard, S. C. (2009) · The Lancet · doi:10.1016/s0140-6736(09)60220-8 · Cited by 445
Boyne, M. S., Silver, D. M., Kaplan, J., & Saudek, C. D. (2003) · Diabetes · doi:10.2337/diabetes.52.11.2790 · Cited by 404
A Tale of Two Compartments: Interstitial Versus Blood Glucose Monitoring
Cengiz, E., & Tamborlane, W. V. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0002 · Cited by 390 · cited 1×
“Further, all measures of subcutaneous kinetics are expected to worsen over implantation time, and subcutaneous sensor performance has been shown to be susceptible to decreases in peripheral perfusion (29,30).”
BackgroundAn Artificial Endocrine Pancreas
Albisser, A. M., Leibel, B. S., Ewart, T. G., Davidovac, Z., Botz, C. K., & Zingg, W. (1974) · Diabetes · doi:10.2337/diab.23.5.389 · Cited by 379
Decreased Epinephrine Responses to Hypoglycemia during Sleep
Jones, T. W., Porter, P., Sherwin, R. S., Davis, E. A., O'Leary, P., Frazer, F., Byrne, G., Stick, S., & Tamborlane, W. V. (1998) · New England Journal of Medicine · doi:10.1056/nejm199806043382303 · Cited by 301
Tanenberg, R. J., Newton, C. A., & Drake, A. J. (2010) · Endocrine Practice · doi:10.4158/ep09260.cr · Cited by 225 · cited 1×
“This is especially vexing during sleep, since sleeping on SQ sensors can cause large inaccuracies (6), and sleeping patients are at high risk for hypoglycemic complications (6-16).”
Background
Cited by
Fokkert, M. J., van Dijk, P. R., Edens, M. A., Abbes, S., de Jong, D., Slingerland, R. J., & Bilo, H. J. G. (2017) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2016-000320 · Cited by 209
A review of optical methods for continuous glucose monitoring
Jernelv, I. L., Mileńko, K., Fuglerud, S. S., Hjelme, D. R., Ellingsen, R., & Aksnes, A. (2018) · Applied Spectroscopy Reviews · doi:10.1080/05704928.2018.1486324 · Cited by 145
Latres, E., Finan, D. A., Greenstein, J. L., Kowalski, A. J., & Kieffer, T. J. (2019) · Cell Metabolism · doi:10.1016/j.cmet.2019.02.007 · Cited by 133
Scholten, K., & Meng, E. F. (2018) · International Journal of Pharmaceutics · doi:10.1016/j.ijpharm.2018.02.022 · Cited by 108
Realizing a Closed-Loop (Artificial Pancreas) System for the Treatment of Type 1 Diabetes
Lal, R., Ekhlaspour, L., Hood, K. K., & Buckingham, B. A. (2019) · Endocrine Reviews · doi:10.1210/er.2018-00174 · Cited by 116
Metadata
- Authors
- Burnett, D. R., Huyett, L. M., Zisser, H. C., Doyle, F. J., & Mensh, B. D.
- Affiliations
- Theranova (United States) (US)
- University of California, Santa Barbara (US)
- Howard Hughes Medical Institute (US)
- Corresponding author
- Mensh, Brett D.
- Venue
- Diabetes
- Publisher
- American Diabetes Association
- Published
- 2014-06-14
- Volume
- 63
- Issue
- 7
- Pages
- 2498-2505
- ISSN
- 0012-1797 (ISSN-L)
- 1939-327X
- PMC
- PMC4066336
- OpenAlex
- W2162507602
- Cited by
- 52
- References
- 46
- Citations by year
- 2026: 1 · 2025: 3 · 2024: 2 · 2023: 4 · 2022: 2 · 2021: 3 · 2020: 3 · 2019: 8 · 2018: 6 · 2017: 8 · 2016: 5 · 2015: 6 · 2014: 1
- FWCI
- 5.37
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Diabetes (publishedVersion, cc-by-nc-nd)
- PubMed Central (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- Funders
- National Institute of Diabetes and Digestive and Kidney Diseases (R43DK093140, R43-DK-93140)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Pancreatic function and diabetes
- Keywords
- continuous glucose monitoring, type 1 diabetes, glucose kinetics, intravenous glucose tolerance test, compartmental modeling, artificial pancreas, pigs
- MeSH terms
- Animals, Ascitic Fluid, Diabetes Mellitus, Female, Glucose, Glucose Tolerance Test, Insulin, Insulin Infusion Systems, Kinetics, Peritoneal Cavity, Swine, Blood Glucose Self-Monitoring, Biosensing Techniques, Pancreas, Artificial, Subcutaneous Tissue
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Glucose Sensing in the Peritoneal Space Offers Faster Kinetics Than Sensing in the Subcutaneous Space. https://app.cochat.ai/papers/10.2337/db13-1649
- The paper
- Burnett, D. R., Huyett, L. M., Zisser, H. C., Doyle, F. J., & Mensh, B. D. (2014). Glucose Sensing in the Peritoneal Space Offers Faster Kinetics Than Sensing in the Subcutaneous Space. Diabetes. https://doi.org/10.2337/db13-1649
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.