‘There were more wires than him’: the potential for wireless patient monitoring in neonatal intensive care
Open accessAbstract
Background: The neonatal intensive care unit (NICU) can be one of the most stressful hospital environments. Alongside providing intensive clinical care, it is important that parents have the opportunity for regular physical contact with their babies because the neonatal period is critical for parent–child bonding. At present, monitoring technology in the NICU requires multiple wired sensors to track each baby's vital signs. This study describes the experiences that parents and nurses have with the current monitoring methods, and reports on their responses to the concept of a wireless monitoring system. Methods: Semistructured interviews were conducted with six parents, each of whom had babies on the unit, and seven nurses who cared for those babies. The interviews initially focused on the participants’ experiences of the current wired system and then on their responses to the concept of a wireless system. The transcripts were analysed using a general inductive approach to identify relevant themes. Results: Participants reported on physical and psychological barriers to parental care, the ways in which the current system obstructed the efficient delivery of clinical care and the perceived benefits and risks of a wireless system. The parents and nurses identified that the wires impeded baby–parent bonding; physically and psychologically. While a wireless system was viewed as potentially enabling greater interaction, staff and parents highlighted potential concerns, including the size, weight and battery life of any new device. Conclusions: The many wires required to safely monitor babies within the NICU creates a negative environment for parents at a critical developmental period, in terms of physical and psychological interactions. Nurses also experience challenges with the existing system, which could negatively impact the clinical care delivery. Developing a wireless system could overcome these barriers, but there remain challenges in designing a device suitable for this unique environment.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Existing technologies for continuous vital sign monitoring require multiple electrodes and interfaces attached to the skin with adhesive tapes and connected by wires to electronic processing systems often tethered to the wall.
“Continuous monitoring of vital signs is essential, yet existing technologies require the use of multiple electrodes and interfaces attached to the skin with adhesive tapes and connected by wires to electronic processing systems that are often tethered to the wall.”
From Skin-interfaced biosensors for advanced wireless physiological monitoring in neonatal and pediatric intensive-care units (2020) · cites this paper as [3]
The Introduction (page unavailable) describes multiple vital-sign sensors in a neonatal intensive care unit connected by wires to a monitoring unit outside the incubator. The Results (page unavailable) describe adhesive ECG and temperature probes, but say the pulse oximeter is secured with Velcro. The paper does not establish that the monitoring units are tethered to a wall or that all the interfaces use adhesive tape. (LLM, conf=0.87)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Using thematic analysis in psychology
Braun, V., & Clarke, V. (2006) · Qualitative Research in Psychology · doi:10.1191/1478088706qp063oa · Cited by 194,414 · see citing works
Qualitative research and evaluation methods
Patton, M. Q. (1980) · Cited by 35,637 · see citing works
A General Inductive Approach for Analyzing Qualitative Evaluation Data
Thomas, D. R. (2006) · American Journal of Evaluation · doi:10.1177/1098214005283748 · Cited by 10,858 · see citing works
Kvale, S. (2007) · doi:10.4135/9781849208963 · Cited by 2,182 · see citing works
Born too soon : the Global action report on preterm birth
Howson, C. P., Kinney, M., & Lawn, J. E. (2012) · Cited by 1,928 · see citing works
Analysing and presenting qualitative data
Burnard, P., Gill, P., Stewart, K., Treasure, E. T., & Chadwick, B. L. (2008) · BDJ · doi:10.1038/sj.bdj.2008.292 · Cited by 1,156 · see citing works
Kangaroo mother care: 25 years after
Charpak, N., Gabriel Ruiz, J., Zupan, J., Cattaneo, A., Figueroa, Z., Tessier, R., Cristo, M., Anderson, G., Ludington, S., Mendoza, S., et al. (2005) · Acta Paediatrica · doi:10.1111/j.1651-2227.2005.tb01930.x · Cited by 364 · see citing works
Parenting in the Neonatal Intensive Care Unit
Cleveland, L. M. (2008) · Journal of Obstetric, Gynecologic & Neonatal Nursing · doi:10.1111/j.1552-6909.2008.00288.x · Cited by 333 · see citing works
Kangaroo Care Is Effective in Diminishing Pain Response in Preterm Neonates
Johnston, C. C., Stevens, B., Pinelli, J. M., Gibbins, S., Filion, F., Jack, A., Steele, S., Boyer, K., & Veilleux, A. (2003) · Archives of Pediatrics and Adolescent Medicine · doi:10.1001/archpedi.157.11.1084 · Cited by 314 · see citing works
Charpak, N., Ruíz, J. G., Zita Figueroa de, C., & Charpak, Y. (1997) · PEDIATRICS · doi:10.1542/peds.100.4.682 · Cited by 296 · see citing works
Valberg, P. A., van Deventer, T. E., & Repacholi, M. H. (2006) · Environmental Health Perspectives · doi:10.1289/ehp.9633 · Cited by 222 · see citing works
Als, H., & McAnulty, G. B. (2011) · Current Women s Health Reviews · doi:10.2174/157340411796355216 · Cited by 216 · see citing works
Current knowledge about skin-to-skin (kangaroo) care for preterm infants.
Anderson, G. C. (1991) · PubMed · Cited by 157 · see citing works
Russell, G., Sawyer, A., Rabe, H., Abbott, J., Gyte, G. M., Duley, L., & Ayers, S. (2014) · BMC Pediatrics · doi:10.1186/1471-2431-14-230 · Cited by 144 · see citing works
Kommers, D. R., Oei, S. G., Chen, W., Feijs, L. M. G., & Oetomo, S. B. (2015) · Acta Paediatrica · doi:10.1111/apa.13254 · Cited by 56 · see citing works
The meaning of technology in an intensive care unit—an interview study
Wikström, A., Cederborg, A., & Johanson, M. (2007) · Intensive and Critical Care Nursing · doi:10.1016/j.iccn.2007.03.003 · Cited by 51 · see citing works
Wireless transmission design for health monitoring at neonatal intensive care units
Chen, W. R., Nguyen, S. T., Coops, R., Oetomo, S. B., & Feijs, L. M. G. (2009) · International Symposium on Applied Sciences in Biomedical and Communication Technologies · doi:10.1109/isabel.2009.5373671 · Cited by 48 · see citing works
Parental interaction with infants treated with medical technology
Lantz, B., & Ottosson, C. (2012) · Scandinavian Journal of Caring Sciences · doi:10.1111/j.1471-6712.2012.01061.x · Cited by 31 · see citing works
Technology in neonatal intensive care – a study on parents’ experiences*
Jämsä, K., & Jämsä, T. J. (1998) · Technology and Health Care · doi:10.3233/thc-1998-6401 · Cited by 28 · see citing works
Flagged first, then by citations
Research works printed but not linked: 1
- [12]World Health Organization. Guidelines on optimal feeding of low birth-weight infants in low- and middle-income countries (in IRIS). Geneva: World Health Organization, 2011. 26042325
Cited by
Sorted by citations
Chung, H. U., Rwei, A. Y., Hourlier-Fargette, A., Xu, S., Lee, K., Dunne, E. C., Xie, Z., Liu, C., Carlini, A., Kim, D. H., et al. (2020) · Nature Medicine · doi:10.1038/s41591-020-0792-9 · Cited by 482 · see citing works
From lab to life: how wearable devices can improve health equity
Walter, J. R., Xu, S., & Rogers, J. A. (2024) · Nature Communications · doi:10.1038/s41467-023-44634-9 · Cited by 108 · see citing works
Jinkins, K. R., Li, S., Arafa, H., Jeong, H., Lee, Y. J., Wu, C., Campisi, E., Ni, X., Cho, D., Huang, Y. Y., et al. (2022) · Science Advances · doi:10.1126/sciadv.abo0537 · Cited by 95 · see citing works
Smart bioelectronic pacifier for real-time continuous monitoring of salivary electrolytes
Lim, H., Lee, S. M., Park, S., Choi, C., Kim, H., Kim, J., Mahmood, M., Lee, Y.-K., Kim, J., & Yeo, W. (2022) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2022.114329 · Cited by 72 · see citing works
Kwak, S. S., Yoo, S., Avila, R. O., Chung, H. U., Jeong, H., Liu, C. Y., Vogl, J. L., Kim, J., Yoon, H., Park, Y., et al. (2021) · Advanced Materials · doi:10.1002/adma.202103974 · Cited by 70 · see citing works
Skin-interfacing wearable biosensors for smart health monitoring of infants and neonates
Zhou, L., Guess, M., Kim, K. R., & Yeo, W. (2024) · Communications Materials · doi:10.1038/s43246-024-00511-6 · Cited by 59 · see citing works
Neu, M., Klawetter, S., Greenfield, J. C., Roybal, K. L., Scott, J. L., & Hwang, S. S. (2019) · Advances in Neonatal Care · doi:10.1097/anc.0000000000000671 · Cited by 42 · see citing works
Lee, W. H., Lee, Y., Na, J. Y., Kim, S. H., Lee, H. J., Lim, Y., Cho, S. H., Cho, S. H., & Park, H. (2020) · PLoS ONE · doi:10.1371/journal.pone.0243939 · Cited by 35 · see citing works
The newborn delivery room of tomorrow: emerging and future technologies
Batey, N., Henry, C., Garg, S., Wagner, M., Malhotra, A., Valstar, M., Smith, T. A., Sharkey, D., Niemuth, M., Küster, H., et al. (2022) · Pediatric Research · doi:10.1038/s41390-022-01988-y · Cited by 33 · see citing works
Camera fusion for real-time temperature monitoring of neonates using deep learning
Lyra, S., Rixen, J., Heimann, K., Karthik, S., Joseph, J., Jayaraman, K., Orlikowsky, T. W., Sivaprakasam, M., Leonhardt, S. K., & Antink, C. H. (2022) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-022-02561-9 · Cited by 30 · see citing works
Advanced portable preterm baby incubator
Shaib, M., Rashid, M., Hamawy, L., Arnout, M., Majzoub, I. E., & Zaylaa, A. J. (2017) · International Conference on Advances in Biomedical Engineering (ICABME) · doi:10.1109/icabme.2017.8167522 · Cited by 29 · see citing works
Continuous Camera-Based Premature-Infant Monitoring Algorithms for NICU
Nagy, Á., Földesy, P., Jánoki, I. G., Terbe, D., Siket, M., Szabó, M., Varga, J., & Zarándy, Á. (2021) · Applied Sciences · doi:10.3390/app11167215 · Cited by 26 · see citing works
McKee, J. L., Kaufman, M. C., Gonzalez, A. K., Fitzgerald, M. P., Massey, S. L., Fung, F. W., Kessler, S. K., Witzman, S., Abend, N. S., & Helbig, I. (2023) · The Lancet Digital Health · doi:10.1016/s2589-7500(23)00004-3 · Cited by 25 · see citing works
Bitter Flavored, Soft Composites for Wearables Designed to Reduce Risks of Choking in Infants
Cho, D., Li, R., Jeong, H., Li, S., Wu, C., Tzavelis, A., Yoo, S., Kwak, S. S., Huang, Y. Y., & Rogers, J. A. (2021) · Advanced Materials · doi:10.1002/adma.202103857 · Cited by 24 · see citing works
Bäcke, P., Hjelte, B., Hellström-Westas, L. K., Ågren, J., & Blomqvist, Y. T. (2020) · Acta Paediatrica · doi:10.1111/apa.15431 · Cited by 23 · see citing works
Heat and mass transfer of preterm neonates nursed inside incubators - A review
Hannouch, A., Lemenand, T., Khoury, K. E., & Habchi, C. (2020) · Thermal Science and Engineering Progress · doi:10.1016/j.tsep.2020.100553 · Cited by 22 · see citing works
Ruíz-Zafra, Á., Precioso, D., Salvador, B., Lubián‐López, S. P., Jiménez, J., Benavente‐Fernández, I., Pigueiras-del-Real, J., Gómez‐Ullate, D., & Gontard, L. C. (2023) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2023.3240245 · Cited by 18 · see citing works
Intelligence Sparse Sensor Network for Automatic Early Evaluation of General Movements in Infants
Bao, B., Zhang, S., Li, H., Cui, W., Guo, K., Zhang, Y., Yang, K., Liu, S., Tong, Y., Zhu, J., et al. (2024) · Advanced Science · doi:10.1002/advs.202306025 · Cited by 17 · see citing works
Bioelectronics for targeted pain management
Flavin, M. T., Foppiani, J., Paul, M. A., Alvarez, A. H., Foster, L., Gavlasova, D., Ma, H., Rogers, J. A., & Lin, S. J. (2025) · Nature Reviews Electrical Engineering · doi:10.1038/s44287-025-00177-3 · Cited by 18 · see citing works
The use of wireless sensors in the neonatal intensive care unit: a study protocol
Sénéchal, E., Radeschi, D. J., Tao, L., Lv, S., Jeanne, E., Kearney, R. E., Shalish, W., & SantʼAnna, G. (2023) · PeerJ · doi:10.7717/peerj.15578 · Cited by 15 · see citing works
Zaylaa, A. J., Rashid, M., Shaib, M., & Majzoub, I. E. (2018) · Journal of Healthcare Engineering · doi:10.1155/2018/8937985 · Cited by 18 · see citing works
A review: Recent advancements in sensor technology for non-invasive neonatal health monitoring
Mishra, S. K., Khouqeer, G. A., Aamna, B., Alodhayb, A. N., Ibrahim, S. J. A., Hooda, M. K., & Jayaswal, G. (2023) · Biosensors and Bioelectronics X · doi:10.1016/j.biosx.2023.100332 · Cited by 18 · see citing works
Wireless monitoring devices in hospitalized children: a scoping review
Sénéchal, E., Jeanne, E., Tao, L., Kearney, R. E., Shalish, W., & SantʼAnna, G. (2023) · European Journal of Pediatrics · doi:10.1007/s00431-023-04881-w · Cited by 15 · see citing works
Multi-modal body part segmentation of infants using deep learning
Voss, F., Brechmann, N., Lyra, S., Rixen, J., Leonhardt, S. K., & Antink, C. H. (2023) · BioMedical Engineering OnLine · doi:10.1186/s12938-023-01092-0 · Cited by 12 · see citing works
A Neonatal Phantom for Vital Signs Simulation
Lyra, S., Voss, F., Coenen, A., Blase, D., Badiola, I., Uguz, D. U., Leonhardt, S. K., & Antink, C. H. (2021) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2021.3108066 · Cited by 11 · see citing works
Zhou, L., Aljiffry, A., Lee, Y. J., Matthews, J., Seitter, B., Soltis, I., Huang, Y., Maher, K., & Yeo, W.-H. (2025) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2025.117372 · Cited by 9 · see citing works
Ahmed, F., Sufyan, A., Hussain, S., Rehman, A., Amin, S., & Salam, Y. (2021) · International Conference on Computing, Electronic and Electrical Engineering (ICE Cube) · doi:10.1109/icecube53880.2021.9628245 · Cited by 8 · see citing works
Scholten, A. W. J., van Leuteren, R. W., WAAL, C. G. D., de Jongh, F. H., van Kaam, A. H., & Hutten, G. J. (2022) · Physiological Measurement · doi:10.1088/1361-6579/ac69a9 · Cited by 8 · see citing works
Stovall, S. G., George, R. G., Lara, M. T., Gainous, K. O., Kitchens, R. F., & Hilton, C. L. (2024) · OTJR Occupational Therapy Journal of Research · doi:10.1177/15394492241271220 · Cited by 6 · see citing works
Zhou, L., Liu, P., Liu, J., Yuan, W., Wu, Z., Xu, D., Hu, B., Shao, Y., Lu, Y., Huang, N., et al. (2025) · Nature Communications · doi:10.1038/s41467-025-67413-0 · Cited by 10 · see citing works
Forehead monitoring of heart rate in neonatal intensive care
Stockwell, S. J., Kwok, T. C., Morgan, S. P., Sharkey, D., & Hayes‐Gill, B. (2023) · Frontiers in Physiology · doi:10.3389/fphys.2023.1127419 · Cited by 6 · see citing works
Comparison of Wired and Wireless Heart Rate Monitoring in the Neonatal Intensive Care Unit
Radeschi, D. J., Sénéchal, E., Tao, L., Lv, S., Shalish, W., SantʼAnna, G., & Kearney, R. E. (2023) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc40787.2023.10340972 · Cited by 6 · see citing works
Peterson, J., Jennings, C., & Mahaveer, A. (2024) · Frontiers in Pediatrics · doi:10.3389/fped.2024.1355777 · Cited by 6 · see citing works
Sénéchal, E., Radeschi, D. J., Lv, S., Jeanne, E., Ruiz, A. S., Tao, L., Dulmage, B., Shalish, W., Kearney, R. E., & SantʼAnna, G. (2025) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2025.1555882 · Cited by 6 · see citing works
Sénéchal, E., Maximov, A. G., Jeanne, E., Radeschi, D. J., Azevedo, V. M. G. D. O., Gross, A., Shalish, W., Kearney, R. E., & SantʼAnna, G. (2025) · Pediatric Research · doi:10.1038/s41390-025-04469-0 · Cited by 8 · see citing works
Ledin, E. R., Fasterius, L., Björling, G., Eriksson, A., & Mattsson, J. Y. (2024) · Disability and Rehabilitation · doi:10.1080/09638288.2024.2428372 · Cited by 4 · see citing works
Yui, Y., Pazandak, C. C., Barbeau, D. Y., Spencer, M., Scala, M. L., & Givrad, S. (2026) · Journal of Perinatology · doi:10.1038/s41372-025-02526-4 · Cited by 3 · see citing works
Hardware-in-the-loop setup for contactless skin servo control of neonatal incubators
Voss, F., Grünter, P., Wolski, J., Leonhardt, S. K., & Lueken, M. (2023) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2023.105628 · Cited by 3 · see citing works
Understanding the Challenges Nurses Encounter with Monitoring Technologies in a NICU
Venkatasubramanian, K., Ranalli, T.-M., Kirupaharan, P., Solanki, D., & Mankodiya, K. (2023) · International Journal of Human-Computer Interaction · doi:10.1080/10447318.2023.2278297 · Cited by 3 · see citing works
Feasibility of a Wireless Vital Signal Monitoring System in the NICU
Radeschi, D. J., Sénéchal, E., Tao, L., Lv, S., Shalish, W., SantʼAnna, G., & Kearney, R. E. (2023) · Wearable and Implantable Body Sensor Networks · doi:10.1109/bsn58485.2023.10331194 · Cited by 3 · see citing works
Sénéchal, E., Radeschi, D. J., Kearney, R. E., Shalish, W., & SantʼAnna, G. (2025) · American Journal of Perinatology · doi:10.1055/a-2604-8329 · Cited by 2 · see citing works
Overcoming Small Data Limitations in Video-Based Infant Respiration Estimation
Song, L., Bishnoi, H., Manne, S. K. R., Ostadabbas, S., Taylor, B. J., & Wan, M. (2026) · IEEE Winter Conference on Applications of Computer Vision · doi:10.1109/wacv61042.2026.00613 · Cited by 2 · see citing works
Internet of Things based Hi-Tech Incubator for New Borns
S, P. K., Keerthana, S., Dharini, R., Bhavadharani Deetchanya, S., & Shalini, D. (2023) · International Conference on Communication and Electronics Systems (ICCES) · doi:10.1109/icces57224.2023.10192880 · Cited by 1 · see citing works
Reimagining apnea monitoring in the neonatal ICU
Jeanne, E., Alvaro, R. E., & Shalish, W. (2024) · Current Opinion in Pediatrics · doi:10.1097/mop.0000000000001432 · Cited by 1 · see citing works
Accuracy of a Noncontact Breathing Frequency Monitoring Technology in the Neonatal ICU
Ren, J., Chen, X., Yao, N., Xiao, G., Yang, C., & Nie, C. (2025) · Respiratory Care · doi:10.1177/19433654251391770 · Cited by 1 · see citing works
Zhang, P., Zhuang, J., Li, T., Fan, D., Lei, Q., & Lin, D. (2025) · Scientific Reports · doi:10.1038/s41598-025-12927-2 · Cited by 1 · see citing works
Syamimi, N., & Yahud, S. (2021) · Applied Research and Smart Technology (ARSTech) · doi:10.23917/arstech.v2i1.185 · Cited by 1 · see citing works
A Wireless Neonatal Intensive Care Unit: Fiction or Closer Reality?
Sénéchal, E., Chung, H. U., Rogers, J. A., & SantʼAnna, G. (2025) · Neonatology · doi:10.1159/000547056 · Cited by 1 · see citing works
Examining Use of Dual Point Temperature Display in Servo-Control With Early Preterm Infants
Dail, R. B., Tavakoli, A. S., Moseley, S. L., Bergstedt, V. B., Kuehn, D., & Chang, W. (2021) · Advances in Neonatal Care · doi:10.1097/anc.0000000000000827
Seizure prediction in 1117 neonates leveraging EMR-embedded standardized EEG reportingPreprint
McKee, J. L., Kaufman, M. C., Gonzalez, A. K., Fitzgerald, M. P., Massey, S. L., Fung, F. W., Kessler, S. K., Witzman, S., Abend, N. S., & Helbig, I. (2022) · medRxiv · doi:10.1101/2022.06.03.22275975
James: Neonatal RDS case study
Hanks, S., & Maguire, L. (2023) · Journal of Neonatal Nursing · doi:10.1016/j.jnn.2023.10.016
Camera Fusion for Improving Body Part Segmentation of Preterm Infants
Voss, V., Koesseler, K., Leonhardt, S. K., & Lueken, M. (2024) · doi:10.47952/gro-publ-200
Skin Servo Control for Neonatal Incubators: A Novel Approach using Infrared Thermography
Voss, F., Leonhardt, S. K., & Lueken, M. (2024) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2024.11.108
Test Bench for Measuring the Effects of Contact Pressure in cECG Measurements*
Prüßmann, J., Leonhardt, S. K., & Lueken, M. (2024) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2024.11.020
Portable Incubator for Premature Infants
(2025) · Journal on Materials and its Characterization · doi:10.46632/jmc/4/2/2
Sénéchal, E., Radeschi, D. J., Jeanne, E., Ruiz, A. S., Dulmage, B., Shalish, W., Kearney, R. E., & SantʼAnna, G. (2025) · Sensors · doi:10.3390/s25185647 · Cited by 3 · see citing works
Radeschi, D. J., Sénéchal, E., Shalish, W., SantʼAnna, G., & Kearney, R. E. (2025) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc58623.2025.11254046
Spatial temperature monitoring of preterm infants using a multi-modal camera setup
Voss, F., Lyra, S., Noehl, C., Brasche, M., Heimann, K., Hensel, L., Orlikowsky, T. W., Leonhardt, S. K., & Lueken, M. (2026) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-026-03575-3
- Authors
- Bonner, O., Beardsall, K., Crilly, N., & Lasenby, J.
- Affiliations
- University of Cambridge (GB)
- Cambridge University Hospitals NHS Foundation Trust (GB)
- Corresponding author
- Bonner, Oliver
- Venue
- BMJ Innovations
- Publisher
- BMJ
- Published
- 2017-01-04
- Volume
- 3
- Issue
- 1
- Pages
- 12-18
- ISSN
- 2055-642X (ISSN-L)
- 2055-8074
- PMC
- PMC5293857
- OpenAlex
- W2569995179
- Cited by
- 61
- References
- 19
- Citations by year
- 2026: 3 · 2025: 12 · 2024: 10 · 2023: 16 · 2022: 6 · 2021: 7 · 2020: 4 · 2019: 1 · 2018: 1 · 2017: 1
- FWCI
- 1.96
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by
- Locations
- BMJ Innovations (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Cambridge University Engineering Department Publications Database (submittedVersion)
- Apollo (University of Cambridge) (submittedVersion, cc-by)
- Apollo (cc-by)
- Funders
- University of Cambridge
- Engineering and Physical Sciences Research Council (1503900, EP/L015889/1, L015889)
- Topics
- Health Sciences › Medicine › Surgery › Healthcare Technology and Patient Monitoring
- Health Sciences › Medicine › Pediatrics, Perinatology and Child Health › Infant Development and Preterm Care
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- wireless patient monitoring, neonatal intensive care, parent-infant bonding, vital signs monitoring, parental experiences, nurses' experiences, medical device design
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review