Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study
Open accessThe DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
Background and objectives: Intermittent vital signs measurements are the current standard on hospital wards, typically recorded once every 8 hours. Early signs of deterioration may therefore be missed. Recent innovations have resulted in ‘wearable’ sensors, which may capture patient deterioration at an earlier stage. The objective of this study was to determine whether a wireless ‘patch’ sensor is able to reliably measure respiratory and heart rate continuously in high-risk surgical patients. The secondary objective was to explore the potential of the wireless sensor to serve as a safety monitor. Design: In an observational methods comparisons study, patients were measured with both the wireless sensor and bedside routine standard for at least 24 hours. Setting: University teaching hospital, single centre. Participants: Twenty-five postoperative surgical patients admitted to a step-down unit. Outcome measures: Primary outcome measures were limits of agreement and bias of heart rate and respiratory rate. Secondary outcome measures were sensor reliability, defined as time until first occurrence of data loss. Results: 1568 hours of vital signs data were analysed. Bias and 95% limits of agreement for heart rate were −1.1 (−8.8 to 6.5) beats per minute. For respiration rate, bias was −2.3 breaths per minute with wide limits of agreement (−15.8 to 11.2 breaths per minute). Median filtering over a 15 min period improved limits of agreement of both respiration and heart rate. 63% of the measurements were performed without data loss greater than 2 min. Overall data loss was limited (6% of time). Conclusions: The wireless sensor is capable of accurately measuring heart rate, but accuracy for respiratory rate was outside acceptable limits. Remote monitoring has the potential to contribute to early recognition of physiological decline in high-risk patients. Future studies should focus on the ability to detect patient deterioration on low care environments and at home after discharge.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Agreement Between Methods of Measurement with Multiple Observations Per Individual
Bland, J. M., & Altman, D. G. (2007) · Journal of Biopharmaceutical Statistics · doi:10.1080/10543400701329422 · Cited by 2,065
Smith, G. B., Prytherch, D. R., Meredith, P., Schmidt, P. E., & Featherstone, P. I. (2013) · Resuscitation · doi:10.1016/j.resuscitation.2012.12.016 · Cited by 1,077
Yasuma, F., & Hayano, J.-I. (2004) · Chest · doi:10.1378/chest.125.2.683 · Cited by 907
Cited by
Van Gelder, I. C., Rienstra, M., Bunting, K. V., Casado-Arroyo, R., Caso, V., Crijns, H. J. G. M., De Potter, T. J. R., Dwight, J., Guasti, L., Hanke, T., et al. (2024) · European Heart Journal · doi:10.1093/eurheartj/ehae176 · Cited by 3,024
How to use digital devices to detect and manage arrhythmias: an EHRA practical guide
Svennberg, E., Tjong, F., Goette, A., Akoum, N., Di Biase, L., Bordachar, P., Boriani, G., Burri, H., Conte, G., Deharo, J. C., et al. (2022) · Europace · doi:10.1093/europace/euac038 · Cited by 261
Clinical requirements of future patient monitoring in the intensive care unit: qualitative study
Poncette, A.-S., Spies, C., Mosch, L., Schieler, M., Weber-Carstens, S., Krampe, H., & Balzer, F. (2019) · JMIR Medical Informatics · doi:10.2196/13064 · Cited by 136
“Nowadays, technologies to remotely monitor patients are available, such as wireless monitoring sensors (eg, ECG, pulse oximetry [8,9], and hemoglobin [10]), noninvasive measurement of hemodynamic parameters (eg, blood pressure and cardiac output [11]), as well as mobile communication devices (eg, mobile phones and tab…”
BackgroundNoninvasive Continuous Monitoring of Vital Signs With Wearables: Fit for Medical Use?
Jacobsen, M., Dembek, T. A., Kobbe, G., Gaidzik, P. W., & Heinemann, L. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820904947 · Cited by 49
Garbern, S. C., Mbanjumucyo, G., Umuhoza, C., Sharma, V. K., Mackey, J., Tang, O., Martin, K. D., Twagirumukiza, F. R., Rosman, S. L., McCall, N., et al. (2019) · DIGITAL HEALTH · doi:10.1177/2055207619879349 · Cited by 39
“The VP was selected due to its battery life, prior use in hospitalized patients and the authors’ familiarity with the device.(17,18) Data from the VP was continuously transmitted through Bluetooth connection to a Samsung Galaxy J3 smartphone running physIQ’s mobile application and subsequently transmitted from the sma…”
Method
Metadata
- Authors
- Breteler, M. J. M., Huizinga, E., van Loon, K., Leenen, L. P. H., Dohmen, D. A. J., Kalkman, C. J., & Blokhuis, T. J.
- Venue
- BMJ Open
- Publisher
- BMJ
- Published
- 2018-02
- Volume
- 8
- Issue
- 2
- Pages
- e020162
- ISSN
- 2044-6055 (ISSN-L)
- OpenAlex
- W2790668025
- Cited by
- 124
- References
- 3
- FWCI
- 5.38
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study. https://app.cochat.ai/papers/10.1136/bmjopen-2017-020162
- The paper
- Breteler, M. J. M., Huizinga, E., van Loon, K., Leenen, L. P. H., Dohmen, D. A. J., Kalkman, C. J., & Blokhuis, T. J. (2018). Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study. BMJ Open. https://doi.org/10.1136/bmjopen-2017-020162
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.