Cuffless Blood Pressure Measurement
Open accessThe DOI matches its Crossref record; 119 of the 119 references it cites were found in a registry or an index.
Abstract
Cuffless blood pressure (BP) measurement has become a popular field due to clinical need and technological opportunity. However, no method has been broadly accepted hitherto. The objective of this review is to accelerate progress in the development and application of cuffless BP measurement methods. We begin by describing the principles of conventional BP measurement, outstanding hypertension/hypotension problems that could be addressed with cuffless methods, and recent technological advances, including smartphone proliferation and wearable sensing, that are driving the field. We then present all major cuffless methods under investigation, including their current evidence. Our presentation includes calibrated methods (i.e., pulse transit time, pulse wave analysis, and facial video processing) and uncalibrated methods (i.e., cuffless oscillometry, ultrasound, and volume control). The calibrated methods can offer convenience advantages, whereas the uncalibrated methods do not require periodic cuff device usage or demographic inputs. We conclude by summarizing the field and highlighting potentially useful future research directions.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“These devices use several different parameters and techniques to assess BP levels, including pulse transit time, pulse wave analysis, photoplethysmography and applanation tonometry, and have different intended uses (such as wearable devices for occasional use or for short-term continuous monitoring in an intensive care unit)55,56.”
From Blood pressure and its variability: classic and novel measurement techniques (2022) · cites this paper as [56]
In the citing paper
“Therefore, these devices can provide detailed information on true BP profile and behaviour over time, as well as all types of BPV (including short-term to long-term BPV6,55–57), and have great potential for use in screening, early diagnosis, close monitoring and optimal management of hypertension.”
From Blood pressure and its variability: classic and novel measurement techniques (2022) · cites this paper as [56]
In the citing paper
“Cuffless BP technologies embedded in wearable devices and smartphones can obtain multiple or even continuous measurements during routine activities for days, weeks or months, without the discomfort of cuff-induced limb compression.”
From Blood pressure and its variability: classic and novel measurement techniques (2022) · cites this paper as [56]
In the citing paper
“However, cuffless devices have specific accuracy issues that are not present in the automated cuff BP devices, such as the need for individual user calibration and their questionable accuracy in tracking BP changes after calibration55,56.”
From Blood pressure and its variability: classic and novel measurement techniques (2022) · cites this paper as [56]
Citations
Cites
Rapid object detection using a boosted cascade of simple features
Viola, P., & Jones, M. (2005) · Proceedings of the 2001 IEEE Computer Society Conference on Computer Vision and Pattern Recognition. CVPR 2001 · doi:10.1109/cvpr.2001.990517 · Cited by 18,300
MIMIC-III, a freely accessible critical care database
Johnson, A. E. W., Pollard, T. J., Shen, L., Lehman, L.-W. H., Feng, M., Ghassemi, M., Moody, B., Szolovits, P., Anthony Celi, L., & Mark, R. G. (2016) · Scientific Data · doi:10.1038/sdata.2016.35 · Cited by 8,141
Wang, H., Naghavi, M., Allen, C., Barber, R. M., Bhutta, Z. A., Carter, A., Casey, D. C., Charlson, F. J., Chen, A. Z., Coates, M. M., et al. (2016) · The Lancet · doi:10.1016/s0140-6736(16)31012-1 · Cited by 6,596
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967
McDonald's blood flow in arteries; theoretical, experimental and clinical principles
(1992) · General Pharmacology: The Vascular System · doi:10.1016/0306-3623(92)90079-y · Cited by 2,388
Schwartz, G., Tee, B. C.-K., Mei, J., Appleton, A. L., Kim, D. H., Wang, H., & Bao, Z. (2013) · Nature Communications · doi:10.1038/ncomms2832 · Cited by 2,042
Chow, C. K. (2013) · JAMA · doi:10.1001/jama.2013.184182 · Cited by 1,991
Prediction of cardiovascular risk factors from retinal fundus photographs via deep learning
Poplin, R., Varadarajan, A. V., Blumer, K., Liu, Y., McConnell, M. V., Corrado, G. S., Peng, L., & Webster, D. R. (2018) · Nature Biomedical Engineering · doi:10.1038/s41551-018-0195-0 · Cited by 1,865
Remote plethysmographic imaging using ambient light
Verkruysse, W., Svaasand, L. O., & Nelson, J. S. (2008) · Optics Express · doi:10.1364/oe.16.021434 · Cited by 1,719
Research works printed but not linked: 10
- O'Brien E, 1994, J. Hum. Hypertens., 8, 73
- Marino PL., 2007, The ICU Book
- Penáz J., 1992, Physiol. Res., 41, 5
- Taira D, 2017, Hawai'i J. Med. Public Health, 76, 314
- Bera TK., 2014, J. Med. Eng., 381251
- Bramwell JC, 1922, Proc. R. Soc. B, 93, 298
- Bouguet J-Y., 2001, Pyramidal implementation of the affine Lucas Kanade feature tracker description of the algorithm
- Liu J, 2017, IEEE J. Transl. Eng. Heal. Med., 5, 1900110
- IEEE. 2014. IEEE Standard for Wearable Cuffless Blood Pressure Measuring Devices . New York: IEEE. 38 pp.
- IEEE. 2019. IEEE Standard for Wearable, Cuffless Blood Pressure Measuring Devices - Amendment 1 . New York: IEEE. 35 pp.
Cited by
Items 101–180 of 180
Blood Pressure Assessment Across the Lifespan: Improving Clinical Research and Clinical Practice
Abdalla, M., Juraschek, S. P., Biaggioni, I., Bowling, C. B., Balijepalli, R. C., Brady, T. M., Bress, A. P., Brook, R. D., Cohen, J. B., Commodore-Mensah, Y. Y., et al. (2026) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2026.02.5124 · Cited by 2
Self-citationMai, A., Luan, Y., Loy, S. L., Qin, C., Ji, H., Obeng, A., Oladuti, M., Ejima, K., Mok, K. T., Kaur, S., et al. (2026) · Frontiers in Digital Health · doi:10.3389/fdgth.2026.1805704 · Cited by 4
“Ambulatory blood pressure monitoring algorithms are often trained on limited demographic datasets, raising concerns about generalizability across populations with varying physiological or anatomical characteristics (120 – 126).”
Telemedicine for hypertension management: where we stand, where we are headed
Omboni, S. (2022) · Connected Health · doi:10.20517/ch.2022.09 · Cited by 5
Okpechi, I. G., Ringrose, J., Padwal, R. S., & Bello, A. K. (2023) · Connected Health · doi:10.20517/ch.2022.18 · Cited by 2
Current and Developing Technologies for BP Monitoring
Gunasekaran, D., & Turner, J. M. (2023) · Current Cardiology Reports · doi:10.1007/s11886-023-01956-w · Cited by 2
“This involves shining light onto the skin and measuring the reflected or transmitted light which varies around its mean due to light absorption by hemoglobin content in pulsatile arterial blood through the cardiac cycle [6].”
BackgroundAslani, R., Dias, D., & Cunha, J. P. S. (2024) · IEEE Mediterranean Electrotechnical Conference (MELECON) · doi:10.1109/melecon56669.2024.10608712 · Cited by 2
SINGLE-CYCLE PULSE WAVEFORM RECOGNITION METHOD BASED ON SVM
Li, Y., Geng, X., YAO, F., Wang, X., Wang, Y., Zhang, Y., & Wang, H. (2024) · Journal of Mechanics in Medicine and Biology · doi:10.1142/s0219519424400876 · Cited by 2
Cheng, B. H., Huang, H., Song, Z., Wu, S., Liu, Q., & Zheng, Y. (2024) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc53108.2024.10782076 · Cited by 2
Digital advancements in hypertension management
Ramdani, S., Benkaddour, N. E. H., & Haddiya, I. (2024) · International Conference on Embedded Systems and Artificial Intelligence (ESAI) · doi:10.1109/esai62891.2024.10913561 · Cited by 4
“Consistent and frequent app usage has been linked to greater weight loss [39].”
Xu, Y., He, Z., & Wang, H. (2025) · Sensors · doi:10.3390/s25113254 · Cited by 2
UTransBPNet: A General Deep Learning Model for Cuffless Blood Pressure Estimation under ActivitiesPreprint
Zheng, Y., Liu, Q., Hong, J., Wu, S., & Zhang, Y. (2023) · doi:10.36227/techrxiv.21915195.v1 · Cited by 5
Huang, Y., He, Y., Song, Z., Gao, K., & Zheng, Y. (2024) · Connected Health And Telemedicine · doi:10.20517/chatmed.2023.23 · Cited by 1
Salika, F., Zaki, C., Harb, H., & Saux, É. (2024) · Cogent Engineering · doi:10.1080/23311916.2024.2374520 · Cited by 1
“Also, cuffless monitors (such as Tonometry (Mukkamala et al., 2022)) are under development, measuring BP waveforms Oxygen saturation (Suresh et al., 2009): In general, there are many types of oxygen sensors used for different applications, such as electro-chemical, Zirconia, Optical, Clark,…”
Hayek, M. A., Catacora, A., Lawley, M., Kum, H., & Ohsfeldt, R. L. (2024) · Health Services Insights · doi:10.1177/11786329241283797 · Cited by 1
“34 This highlights the demand for innovative cuffless BP monitoring devices that can integrate regular monitoring into an individual’s daily life, ensuring accurate diagnosis and effective long-term hypertension treatment and care.”
BackgroundReal-time classification of blood pressure changes using photoplethysmography and deep learning
Hong, J., Jin, W., Nandi, M., & Alastruey, J. (2025) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2025.108380 · Cited by 2
Continuous Cuffless Blood Pressure Estimation via Effective and Efficient Broad Learning Model
Zhang, C., Hu, P., Shen, Z., & Ding, X. (2025) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2025.3604464 · Cited by 1
Li, Y., Ling, L., Xiao, Z., Liu, J., Yang, C., Pan, J., Liu, C., Li, H., Zhou, Y., Sui, C., et al. (2025) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2025.108695 · Cited by 2
Seo, W.-D., Choi, I., Kim, G. C., Song, H. S., & Kim, Y. K. (2026) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2026.109852 · Cited by 1
Liu, J., Hu, S., Wang, Y., Wang, D., & Yang, C. (2026) · Expert Systems with Applications · doi:10.1016/j.eswa.2026.132477 · Cited by 1
Calibration: The Unseen Constraint on Pervasive Inference
Kawsar, F. (2026) · IEEE Pervasive Computing · doi:10.1109/mprv.2026.3690300 · Cited by 1
Zhang, Z., Qu, P., & Tang, H. X. (2026) · Sensors · doi:10.3390/s26154676 · Cited by 1
“Moreover, manually designed features cannot fully capture the high-dimensional, nonlinear, and multi-scale temporal patterns embedded in raw physiological signals [4,5,14,15].”
InfluentialArjomand, A., Boudesh, A., Bayatmakou, F., Krylov, G., Kent, K. B., & Mohammadi, A. (2026) · Applied Soft Computing · doi:10.1016/j.asoc.2026.116316 · Cited by 2
Selecting patients for interventional procedures to treat hypertension
Lobo, M. D., Rull, G., Saxena, M., & Kapil, V. (2023) · Blood Pressure · doi:10.1080/08037051.2023.2248276 · Cited by 3
“Presently only oscillometric monitors are validated for home BP measurement although an emerging class of cuffless, wearable BP monitors will at some point in the foreseeable future also provide valuable out of office BP metrics [21,22].”
BackgroundPearls and pitfalls of cuffless blood pressure monitoring devices
Iacobescu, R.-A., Florescu, B.-D., Dimache, A.-M., Corneanu, L. E., & Coman, A. E. (2023) · Romanian Medical Journal · doi:10.37897/rmj.2023.3.2 · Cited by 1
Kumar, R., Kumar, V., Rich, C. A., Lemmerhirt, D. F., Balendra, Fowlkes, J. B., & Sahani, A. K. (2025) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-024-03268-9 · Cited by 2
On-Device, Continuous, Cuffless, and Accelerometer-Based Blood Pressure Monitoring
Saha, S. S., Puthenveedu, M. A., & Chowdhary, M. (2025) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc58623.2025.11251861 · Cited by 1
“Table I summarizes several cuffless BP monitoring techniques [2], [4], [7]–[9], [13]–[16].”
InfluentialThe Standards of Over-the-Counter Cuffless Sphygmomanometers for Monitoring Blood PressurePreprint
Huang, M., TAMURA, T., & Umezu, S. (2024) · Preprints.org · doi:10.20944/preprints202406.0627.v1
Garrett, A., Kim, B., Gurel, N. Z., Sie, E. J., Wilson, B. K., Marsili, F., Forman, J. P., Hamburg, N. M., Boas, D. A., & Roblyer, D. M. (2024) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2024.08.16.608163 · Cited by 2
Savostin, A., Ларгин, А., Savostina, G., Риттер, Д. В., & Koshekov, A. (2023) · Вестник КазАТК · doi:10.52167/1609-1817-2023-125-2-354-362
Zaid, M., Mihail, P., Despins, L. A., Keller, J., Skubic, M., & Guidoboni, G. (2023) · Journal of Cardiology and Cardiovascular Medicine · doi:10.29328/journal.jccm.1001157
UTransBPNet: A General Deep Learning Model for Cuffless Blood Pressure Estimation under ActivitiesPreprint
Zheng, Y., Liu, Q., Hong, J., Wu, S., & Zhang, Y. (2023) · doi:10.36227/techrxiv.21915195.v2
Hipertansiyon Hastalarının Giyilebilir Kan Basıncı Cihazlarına Yönelik Görüşleri
Şahan, S., & Güler, S. (2024) · Ordu Üniversitesi Hemşirelik Çalışmaları Dergisi · doi:10.38108/ouhcd.1236826
Cai, J., Li, Z., Wen, Q., Sun, X., Wu, J., Lu, Y., He, T., Zhang, W., & Yuan, H. (2024) · Research Square · doi:10.21203/rs.3.rs-3784165/v1
Non-Invasive Systemic Blood Pressure Measurement Wait Period Error MitigationPreprint
McMahon, S., & Morrow, P. (2024) · medRxiv · doi:10.1101/2024.10.03.24314772
Novel metrics for tracking blood pressure changes incontinuous cuffless blood pressure estimations
He, S., & Bolić, M. (2024) · Scientific Reports · doi:10.1038/s41598-024-77171-6 · Cited by 1
Continuous and Cuffless Blood Pressure Estimation Using Photoplethysmography for Wearable Devices
Tiwari, S., Saini, S. K., Joshi, A. M., & Maddila, R. K. (2024) · IEEE India Conference · doi:10.1109/indicon63790.2024.10958245
“The development of PPG-based techniques, which minimize the requirement of cuffs, attracts much attention due to their continuous, non-invasive, affordability, user-friendliness, and widespread adoption in wearable healthcare technology [4], [5].”
Sung, F. J. W., Koay, Y. Y., Lim, S. S., Leong, M. C., & Hau, Y. W. (2025) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-031-97578-3_11
Application Research of Photoplethysmography technology in Wearable Device Blood Pressure Detection
Fan, P., Qu, Z., Feng, J., Li, R., Zhao, Y., & Liang, L. (2025) · International Conference on Biomedical and Intelligent Systems (IC-BIS) · doi:10.1145/3745034.3745120
Lequeux, B., & Garban, T. (2025) · La Presse Médicale Formation · doi:10.1016/j.lpmfor.2025.100689
Kokatanur, U. A., & Dandagi, P. M. (2025) · BioNanoScience · doi:10.1007/s12668-025-02193-x
Estimating Blood Pressure from PPG Signals Using a Dual Calibration Siamese Network Approach
Cárdenas, D. A. C., Dias, F. M., de Toledo, M. A. F., Oliveira, F. A. C., Ribeiro, E., Moreno, R. A., Krieger, J. E., & Gutierrez, M. A. (2025) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-031-94934-0_80
Okubai, T., Kahsay, D. T., Netsereab, T. B., & Asnake, M. A. (2025) · SAGE Open Nursing · doi:10.1177/23779608251403037
Neckband-type Self-powered Pulsewave Sensor for Continuous Blood Pressure Monitoring
Kim, T., Jaworski, D. J., Sethi, R., Huisman, E. J., Fung, K., & Park, E. J. (2025) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc58623.2025.11253314
“[6] BP is then estimated by establishing a Edward J. Park is with the Biomechatronic Systems Laboratory, School of Mechatronic Systems Engineering, and WearTech Labs at Simon Fraser University, Surrey, BC, V3T 0N1, Canada (phone: 778 778-8662; e-mail: ed_park@sfu.ca).”
Joseph, J., Kiran, V. R., Nabeel, P. M., & Chandran, D. S. (2025) · Journal of Visualized Experiments · doi:10.3791/69002
Nguyen, T., Kolla, S. S., Wang, H., Weisbecker, H., Messersmith, B., Leal, D. B. M., Kuntamukkala, R., Liu, Y., Hanson, E. D., Paterson, C., et al. (2025) · IEEE Journal on Flexible Electronics · doi:10.1109/jflex.2025.3614515
Schjerven, F. E., Steffensen, T. L., Lindseth, F., Flade, H. M., & Kirkeby‐Garstad, I. (2026) · IEEE data descriptions. · doi:10.1109/ieeedata.2025.3650442
Xu, R., Guo, J., Li, W., Xie, Z., Zhai, Y., Jiang, S., Li, Y., Xia, K., & Chen, Y. (2026) · Frontiers in Physiology · doi:10.3389/fphys.2026.1746894 · Cited by 2
Jiao, X., Gao, Q., Qi, L., Xu, L., & Qian, W. (2026) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2026.3664375 · Cited by 1
Cheng, Y.-J., Kim, E., Hahn, J., Chung, J., & Kwon, Y. (2026) · Physiological Measurement · doi:10.1088/1361-6579/ae575a
Yu, Y., & Lowe, A. (2026) · Hypertension Research · doi:10.1038/s41440-026-02651-3 · Cited by 1
Non-Invasive Blood Pressure Monitoring and Estimation using Dual-Wavelength PPG and Machine Learning
Sanneboyina, A., Naresh, S., Kuriti, J., Kollem, S., Peddakrishna, S., Gaddam, C. K., & Krishnaiah, M. (2025) · IEEE International Conference on Recent Trends in Computing and Smart Mobility (RCSM) · doi:10.1109/rcsm67767.2025.11507881
“Advancements in cuffless BP monitoring have led to various frameworks, broadly classified by Ramakrishna et al. into calibrated and uncalibrated methods [4].”
Sato, S., & Nakayama, S. (2026) · Biomedical Engineering Advances · doi:10.1016/j.bea.2026.100220
Zhang, Z., & Hu, D. (2026) · International Conference on Intelligent Computing and Signal Processing (ICSP) · doi:10.1109/icsp69961.2026.11540717
“A representative approach is pulse transit time (PTT)-based estimation [3], which measures the pulse arrival time difference between proximal and distal sensors and then estimates blood pressure through an empirical regression relationship.”
Ateş, E. S., & Ceyhun, G. (2026) · BMC Cardiovascular Disorders · doi:10.1186/s12872-026-06073-4
Machine Learning-Enabled Non-Contact Blood Pressure Estimation Using a 24 GHz FMCW Radar Sensor
Chen, C., Chang, H.-C., Kuo, T.-W., Chen, K.-W., Chen, C.-H., & Wu, C. M. (2026) · IEEE MTT-S International Microwave Biomedical Conference (IMBioC) · doi:10.1109/imbioc69142.2026.11541156
“In parallel to radar-based approaches, a large body of work has explored cuffless BP estimation using contact sensors such as electrocardiography (ECG) and photoplethysmography (PPG) [5].”
Conformal photoplethysmography sensor with ultrathin architecture for accurate health monitoring
Pan, X., Xu, Z., Wu, W., He, Z., Lu, H., Ji, S., Geng, M., Han, X., & Pan, C. (2026) · Journal of Materials Chemistry C · doi:10.1039/d6tc00782a
Noninvasive Arterial Pressure Monitoring during Anesthesia
Meidert, A. S., & Saugel, B. (2026) · Anesthesiology · doi:10.1097/aln.0000000000006070
Scherpf, M., Ernst, H., Malberg, H., & Schmidt, M. (2026) · Physiological Measurement · doi:10.1088/1361-6579/ae7fea
Choo, T. W. J., Yap, A. A. M., Kawaja, A., Chao, V. T. T., Ng, C. J., Olivo, M., & Bi, R. (2026) · medRxiv · doi:10.64898/2026.06.19.26356074
Lin, Y., Yu, Y., Zhang, H., & Ding, X. (2026) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2026.3704348
Putra, A. A. R., Ahmad, N. A., Yusof, Y. M., Azizan, A., Alnahari, A. Y., & Ansariadi (2026) · Lecture notes in computer science · doi:10.1007/978-981-92-2891-1_12
Shahverdi, E., Shahverdi, A., Kamali, S., & Herda, L. R. (2026) · Herz · doi:10.1007/s00059-026-05389-3
EarBP: Earbuds-Based Noninvasive Blood Pressure Elevation Monitoring Using Pulse Transit Time
Morshed, M. B., Rahman, M. M., Islam, M. S., Hahn, J., Zhu, L., & Desai, S. A. (2026) · IEEE Internet of Things Journal · doi:10.1109/jiot.2026.3716618
Shi, W., Mi, L., Chen, J., Liu, B., Wang, M., Dai, Z., Liu, B., Gao, R., Lin, Y., Guo, W., et al. (2026) · Advanced Science · doi:10.1002/advs.76775
“By contrast, SBP remained more sensitive to environmental and physiological perturbations and may be more appropriate for trend monitoring, consistent with broader observations in cuffless BP research [ 16 ].”
Contact-Free Blood Pressure Estimation Based on Pulse Wave Velocity Using mmWave Signals
Liu, J., Wang, J., Gao, Q., Zhu, L., Chen, X., Pan, M., & Fang, Y. M. (2026) · IEEE Internet of Things Journal · doi:10.1109/jiot.2026.3717857
Bovio, D., Turrini, A., Prandoni, A., Porro, A., Cerveri, P., & Salito, C. (2026) · Biosensors · doi:10.3390/bios16080412
“Second, peripheral optical signals like PPG are highly susceptible to motion artifacts and localized contact pressure changes, which frequently result in signal saturation, loss of physiological pulse morphology, or severe estimation drift over time [2].”
Aloui, Z., & Landry, C. (2026) · Artery Research · doi:10.1007/s44200-026-00120-3
“These PWA BP estimation models typically rely on photoplethysmography (PPG), which reflects distal blood volume changes, or tonometry, which captures peripheral arterial forces, to large datasets to learn the relationship between the extracted PWA features and SBP and diastolic BP (DBP) values [8].”
InfluentialShang, Y., & Zhang, N. (2026) · SSRN Electronic Journal · doi:10.2139/ssrn.7227883
Chandana, Sukesh Rao, M., Bhat, S. J., & Kumar, M. J. D. (2026) · IEEE Access · doi:10.1109/access.2026.3722377
Jo, H., Park, D., Lee, S., & Kwon, N. K. (2026) · Electronics · doi:10.3390/electronics15173864
Zheng, W., Xu, X., Song, Y., Cao, Y., Sun, Z., Dong, K., & Liu, Q. (2026) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2026.118419
Han, Q., Li, Y., Feng, C., Zhang, Y., Dai, Q., Wei, N., & Li, K. (2026) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2026.3731717
Lee, H., Shin, J., Lee, E. M., & Ihm, S. (2026) · Clinical Hypertension · doi:10.5646/ch.2026.32.e40
“Alternative methods such as pulse transit time (PTT) use the inverse relationship between arterial stiffness and pulse wave velocity [18,19].”
Cha, J., Kim, M., Kim, S. R., Park, J. H., Joo, H. J., & Yu, C. W. (2026) · Clinical Hypertension · doi:10.5646/ch.2026.32.e42
“Cuffless BP technologies, including photoplethysmography-based devices, have emerged as potential tools for out-of-office monitoring, although validation and implementation concerns remain [13,14].”
Torad, M. A., Ahmed, H. E., Torad, A. A. M., Taha, M. M., Horváth, A. A., & Emam, M. A. (2026) · Sensors · doi:10.3390/s26196047
Фармакологічний авангард та цифрова кардіологія в менеджменті артеріальної гіпертензії
(2026) · Український медичний часопис · doi:10.32471/umj.1680-3051.280993
Demographic subgroup-aware analysis of blood pressure predictive models
Kolawole, B., Fan, S., Lao, Y., & Koomson, V. (2026) · PLOS Digital Health · doi:10.1371/journal.pdig.0001687
Deutges, M., Schneider, H., & Redtel, H. (2026) · IEEE Access · doi:10.1109/access.2026.3743693
Shi, Z., Zhang, G., & Li, M. (2026) · IEEE Information Technology and Mechatronics Engineering Conference (ITOEC) · doi:10.1109/itoec69754.2026.11732659
“Although the prospects are broad, existing general-purpose PPG blood pressure prediction models still face significant accuracy challenges [6].”
Patil, R., Manjunath, C. N., & Nishanth, K. R. (2026) · Journal of Indian College of Cardiology · doi:10.4103/jicc.jicc_46_23
Metadata
- Authors
- Mukkamala, R., Stergiou, G. S., & Avolio, A. P.
- Affiliations
- University of Pittsburgh (US)
- National and Kapodistrian University of Athens (GR)
- Macquarie University (AU)
- Venue
- Annual Review of Biomedical Engineering
- Publisher
- Annual Reviews
- Published
- 2022-06-06
- Volume
- 24
- Issue
- 1
- Pages
- 203-230
- ISSN
- 1523-9829 (ISSN-L)
- 1545-4274
- OpenAlex
- W4220676114
- Cited by
- 180
- References
- 119
- Citations by year
- 2026: 47 · 2025: 49 · 2024: 39 · 2023: 34 · 2022: 11
- FWCI
- 2.96
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- Annual Review of Biomedical Engineering (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health
- National Heart, Lung, and Blood Institute (R01HL146470)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Keywords
- cuffless blood pressure measurement, non-invasive blood pressure monitoring, pulse transit time, pulse wave analysis, volume control, hypertension, hypotension, wearable sensing, smartphone-based monitoring
- MeSH terms
- Blood Pressure, Blood Pressure Determination, Humans, Hypertension, Oscillometry, Pulse Wave Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.