A Reusable Electrochemical Biosensor for Monitoring of Small Molecules (Cortisol) Using Molecularly Imprinted Polymers
The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
This paper establishes the feasibility of a reusable biosensor that can be operated and stored at room temperature, for detection of small molecules in low resource settings. The sensor was fabricated using molecularly imprinted polymers (MIP) and cortisol was chosen as a model analyte. Cortisol imprinted polymer films were prepared by electropolymerizing pyrrole on an electrode surface in the presence of cortisol. Electrochemical over-oxidation of polypyrrole (PPy) was performed for the controlled release of cortisol templates and to create cortisol specific imprinting sites. Stepwise fabrication of imprinted sensors was characterized through cyclic voltammetry (CV) and scanning electron microscopy (SEM). The sensor exhibited a detection limit of 1 pM L −1 for cortisol. A unique feature of the sensor was that cross-reactivity with prednisolone (which has 100% interference in ELISA), was minimized to 18.3% compared to ELISA. The sensitivity of the sensor remained over 90% after 7 cycles of elution/rebinding, while the sensitivity decreased by 10% after 4 weeks of storage at room temperature, suggesting the sensor can be used multiple times and used with low overhead costs in low resource settings such as agricultural fields. The sensor was used for detection of cortisol in saliva samples of farm workers; benchmarking with ELISA showed excellent correlation. These findings indicate that such a sensor can be used for in-field measurements of small molecules (e.g. cortisol).
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Charney, D. S. (2004) · American Journal of Psychiatry · doi:10.1176/appi.ajp.161.2.195 · Cited by 1,605
van Eck, M., Berkhof, H., Nicolson, N., & Sulon, J. (1996) · Psychosomatic Medicine · doi:10.1097/00006842-199609000-00007 · Cited by 617
Zangheri, M., Cevenini, L., Anfossi, L., Baggiani, C., Simoni, P., Di Nardo, F., & Roda, A. (2015) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2014.08.048 · Cited by 373
Krasowski, M. D., Drees, D., Morris, C. S., Maakestad, J., Blau, J. L., & Ekins, S. (2014) · BMC Clinical Pathology · doi:10.1186/1472-6890-14-33 · Cited by 212
Cited by
Touch‐Based Stressless Cortisol Sensing
Tang, W., Yin, L., Sempionatto, J. R., Moon, J., Teymourian, H., & Wang, J. (2021) · Advanced Materials · doi:10.1002/adma.202008465 · Cited by 302
Klinghammer, S., Voitsekhivska, T., Licciardello, N., Kim, K., Baek, C.-K., Cho, H., Wolter, K.-J., Kirschbaum, C., Baraban, L., & Cuniberti, G. (2020) · ACS Sensors · doi:10.1021/acssensors.0c02267 · Cited by 62
Portable and wearable real-time stress monitoring: A critical review
Parlak, O. (2021) · Sensors and Actuators Reports · doi:10.1016/j.snr.2021.100036 · Cited by 60
Metadata
- Authors
- Manickam, P., Pasha, S. K., Snipes, S. A., & Bhansali, S.
- Venue
- Journal of The Electrochemical Society
- Publisher
- The Electrochemical Society
- Published
- 2016-12-23
- Volume
- 164
- Issue
- 2
- Pages
- B54-B59
- ISSN
- 0013-4651 (ISSN-L)
- 1945-7111
- OpenAlex
- W2563824673
- Cited by
- 78
- References
- 4
- FWCI
- 1.68
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.