Accurate estimation of entropy in very short physiological time series: the problem of atrial fibrillation detection in implanted ventricular devices
The DOI matches its Crossref record; 20 of the 20 references it cites were found in a registry or an index.
Abstract
Entropy estimation is useful but difficult in short time series. For example, automated detection of atrial fibrillation (AF) in very short heart beat interval time series would be useful in patients with cardiac implantable electronic devices that record only from the ventricle. Such devices require efficient algorithms, and the clinical situation demands accuracy. Toward these ends, we optimized the sample entropy measure, which reports the probability that short templates will match with others within the series. We developed general methods for the rational selection of the template length m and the tolerance matching r. The major innovation was to allow r to vary so that sufficient matches are found for confident entropy estimation, with conversion of the final probability to a density by dividing by the matching region volume, 2 r m. The optimized sample entropy estimate and the mean heart beat interval each contributed to accurate detection of AF in as few as 12 heartbeats. The final algorithm, called the coefficient of sample entropy (COSEn), was developed using the canonical MIT-BIH database and validated in a new and much larger set of consecutive Holter monitor recordings from the University of Virginia. In patients over the age of 40 yr old, COSEn has high degrees of accuracy in distinguishing AF from normal sinus rhythm in 12-beat calculations performed hourly. The most common errors are atrial or ventricular ectopy, which increase entropy despite sinus rhythm, and atrial flutter, which can have low or high entropy states depending on dynamics of atrioventricular conduction.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A Mathematical Theory of Communication
Shannon, C. E. (1948) · Bell System Technical Journal · doi:10.1002/j.1538-7305.1948.tb01338.x · Cited by 88,486
Physiological time-series analysis using approximate entropy and sample entropy
Richman, J. S., & Moorman, J. R. (2000) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.2000.278.6.h2039 · Cited by 7,880 · cited 3×
“An important aspect of the SampEn algorithm is that self-matches are not counted (19, 20).”
BackgroundSelf-citationGo, A. S., Hylek, E. M., Phillips, K. A., Chang, Y., Henault, L. E., Selby, J. V., & Singer, D. E. (2001) · JAMA · doi:10.1001/jama.285.18.2370 · Cited by 6,205
Approximate entropy as a measure of system complexity.
Pincus, S. M. (1991) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.88.6.2297 · Cited by 5,811 · cited 1×
“The original recipe has been to select r as 20% of the SD of each time series segment, based on the “preliminary” conclusions of Pincus (14) in 1991 for implementing approximate entropy (ApEn) calculations.”
MethodErgodic theory of chaos and strange attractors
Eckmann, J. P., & Ruelle, D. (1985) · Reviews of Modern Physics · doi:10.1103/revmodphys.57.617 · Cited by 4,941
Multiscale entropy analysis of biological signals
Costa, M., Goldberger, A. L., & Peng, C.-K. (2005) · Physical Review E · doi:10.1103/physreve.71.021906 · Cited by 2,482
Estimation of the Kolmogorov entropy from a chaotic signal
Grassberger, P., & Procaccia, I. (1983) · Physical Review A · doi:10.1103/physreva.28.2591 · Cited by 1,440
Physiological time-series analysis: what does regularity quantify?
Pincus, S. M., & Goldberger, A. L. (1994) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1994.266.4.h1643 · Cited by 1,315
Sample entropy analysis of neonatal heart rate variability
Lake, D. E., Richman, J. S., Griffin, M. P., & Moorman, J. R. (2002) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.00069.2002 · Cited by 1,233
Self-citationPikkujämsä, S. M., Mäkikallio, T. H., Sourander, L. B., Räihä, I. J., Puukka, P., Skyttä, J., Peng, C.-K., Goldberger, A. L., & Huikuri, H. V. (1999) · Circulation · doi:10.1161/01.cir.100.4.393 · Cited by 416
Cited by
The Virtual Brain: a simulator of primate brain network dynamics
Sanz Leon, P., Knock, S. A., Woodman, M. M., Domide, L., Mersmann, J., McIntosh, A. R., & Jirsa, V. (2013) · Frontiers in Neuroinformatics · doi:10.3389/fninf.2013.00010 · Cited by 523
A deep learning approach for real-time detection of atrial fibrillation
Andersen, R. S., Peimankar, A., & Puthusserypady, S. (2019) · Expert Systems with Applications · doi:10.1016/j.eswa.2018.08.011 · Cited by 402
Detecting atrial fibrillation by deep convolutional neural networks
Xia, Y., Wulan, N., Wang, K., & Zhang, H. (2018) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2017.12.007 · Cited by 349
A Batteryless 19 $\mu$W MICS/ISM-Band Energy Harvesting Body Sensor Node SoC for ExG Applications
Zhang, Y., Zhang, F., Shakhsheer, Y., Silver, J. D., Klinefelter, A., Nagaraju, M., Boley, J., Pandey, J., Shrivastava, A., Carlson, E. J., et al. (2013) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2012.2221217 · Cited by 326
Assessing the complexity of short-term heartbeat interval series by distribution entropy
Li, P., Liu, C., Li, K., Zheng, D., Liu, C., & Hou, Y. (2015) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-014-1216-0 · Cited by 260
New-Onset Atrial Fibrillation in the Critically Ill*
Moss, T. J., Calland, J. F., Enfield, K. B., Gomez-Manjarres, D. C., Ruminski, C., DiMarco, J. P., Lake, D. E., & Moorman, J. R. (2017) · Critical Care Medicine · doi:10.1097/ccm.0000000000002325 · Cited by 172
“≥ 30 seconds) or (2) not AF, which was primarily comprised of sinus rhythm with varying degrees of ectopy.”
BackgroundSelf-citationXiong, Z., Nash, M. P., Cheng, E., Fedorov, V. V., Stiles, M. K., & Zhao, J. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aad9ed · Cited by 161
Moss, T. J., Lake, D. E., Calland, J. F., Enfield, K. B., Delos, J. B., Fairchild, K. D., & Moorman, J. R. (2016) · Critical Care Medicine · doi:10.1097/ccm.0000000000001738 · Cited by 98
Self-citationSmart-watches: a potential challenger to the implantable loop recorder?
Carpenter, A., & Frontera, A. (2016) · Europace · doi:10.1093/europace/euv427 · Cited by 86
“Gold standard AF detection algorithms such as those used by implantable devices tend to use a combination of Shannon Entropy, co-efficient of variation(8) and the co-efficient of sample entropy.(9)”
MethodDetection of Congestive Heart Failure Based on LSTM-Based Deep Network via Short-Term RR Intervals
Wang, L., & Zhou, X. (2019) · Sensors · doi:10.3390/s19071502 · Cited by 85
“It has been used to detect cardiovascular diseases such as atrial fibrillation (AF) and achieved good results [15].”
BackgroundEerikäinen, L. M., Bonomi, A. G., Schipper, F., Dekker, L. R. C., Vullings, R., de Morree, H. M., & Aarts, R. M. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aad2c0 · Cited by 61
Ambulatory cardiac bio-signals: From mirage to clinical reality through a decade of progress
Periyaswamy, T., & Balasubramanian, M. (2019) · International Journal of Medical Informatics · doi:10.1016/j.ijmedinf.2019.07.007 · Cited by 14
Metadata
- Authors
- Lake, D. E., & Moorman, J. R.
- Venue
- American Journal of Physiology-Heart and Circulatory Physiology
- Publisher
- American Physiological Society
- Published
- 2011-01
- Volume
- 300
- Issue
- 1
- Pages
- H319-H325
- ISSN
- 0363-6135 (ISSN-L)
- 1522-1539
- OpenAlex
- W2140005438
- Cited by
- 318
- References
- 20
- FWCI
- 5.11
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.