Noninvasive ventilatory monitoring with bellows pneumographs in supine subjects
The DOI matches its Crossref record.
Abstract
We compared simultaneous measurements of ventilatory movements obtained by a bellows pneumograph (BP) and a respiratory inductive plethysmograph (RIP) vs. integrated volume from a pneumotachograph in seven healthy volunteers during different respiratory patterns. The purpose of this study was to assess if a computer-aided calibration procedure could improve the accuracy of BP, a simple semiquantitative method for noninvasive ventilatory monitoring in supine subjects. Both devices were repeatedly calibrated against a pneumotachograph in the same posture with a computer-aided least-squares method. One calibration maneuver was sufficient to achieve a minimal relative difference in volume measurements between the RIP as well as the BP and the pneumotachograph of less than 1.2 +/- 4.5 (SD) %. The accuracy remained in this range during the subsequent calibrations with time (1 h) and after body movements. However, this difference increased significantly with both devices when the subjects were studied in the lateral decubitus position. The present study indicates that despite theoretical advantages of the RIP, the BP, when properly calibrated, has similar performances in supine subjects for monitoring ventilation; it is simpler and less expensive than the RIP and is devoid of electrical drift and artifacts.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Grossman, P., & Taylor, E. W. (2007) · Biological Psychology · doi:10.1016/j.biopsycho.2005.11.014 · Cited by 1,133
Cardiovascular, electrodermal, and respiratory response patterns to fear‐ and sadness‐inducing films
Kreibig, S. D., Wilhelm, F. H., Roth, W. T., & Gross, J. J. (2007) · Psychophysiology · doi:10.1111/j.1469-8986.2007.00550.x · Cited by 495
Grossman, P., Karemaker, J., & Wieling, W. (1991) · Psychophysiology · doi:10.1111/j.1469-8986.1991.tb00412.x · Cited by 471
Emotions and respiratory patterns: review and critical analysis
Boiten, F. A., Frijda, N. H., & Wientjes, C. J. E. (1994) · International Journal of Psychophysiology · doi:10.1016/0167-8760(94)90027-2 · Cited by 462
A comparison of three quantification methods for estimation of respiratory sinus arrhythmia.
Grossman, P., van Beek, J., & Wientjes, C. (1990) · Psychophysiology · doi:10.1111/j.1469-8986.1990.tb03198.x · Cited by 381
The effects of emotional behaviour on components of the respiratory cycle.
Boiten, F. A. (1998) · Biological Psychology · doi:10.1016/s0301-0511(98)00025-8 · Cited by 205
Physiologic instability in panic disorder and generalized anxiety disorder
Wilhelm, F. H., Trabert, W., & Roth, W. T. (2001) · Biological Psychiatry · doi:10.1016/s0006-3223(00)01000-3 · Cited by 180
Breathing Patterns During Spontaneous Speech
Winkworth, A. L., Davis, P. J., Adams, R. D., & Ellis, E. (1995) · Journal of Speech, Language, and Hearing Research · doi:10.1044/jshr.3801.124 · Cited by 175
Wilhelm, F. H., Grossman, P., & Coyle, M. A. (2004) · Instrument Society of America · doi:10.5451/unibas-ep9419 · Cited by 62
Gerlach, A. L., Spellmeyer, G., Vögele, C., Huster, R., Stevens, S., Hetzel, G., & Deckert, J. (2006) · Psychosomatic Medicine · doi:10.1097/01.psy.0000203284.53066.4b · Cited by 53
“Raw signals of the two bands were converted to lung volume change using calibration regression weights established by spirometry and the least-squares method (41).”
Method
Metadata
- Authors
- Morel, D. R., Forster, A., & Suter, P. M.
- Venue
- Journal of Applied Physiology
- Publisher
- American Physiological Society
- Published
- 1983-08-01
- Volume
- 55
- Issue
- 2
- Pages
- 598-606
- ISSN
- 0161-7567 (ISSN-L)
- 8750-7587
- 1522-1601
- OpenAlex
- W2419591712
- Cited by
- 42
- References
- 0
- FWCI
- 2.08
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.