Recommendations for Blood Pressure Measurement in Humans and Experimental Animals: Part 1: Blood Pressure Measurement in Humans: A Statement for Professionals From the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research
Open accessAbstract
Accurate measurement of blood pressure is essential to classify individuals, to ascertain blood pressure–related risk, and to guide management. The auscultatory technique with a trained observer and mercury sphygmomanometer continues to be the method of choice for measurement in the office, using the first and fifth phases of the Korotkoff sounds, including in pregnant women. The use of mercury is declining, and alternatives are needed. Aneroid devices are suitable, but they require frequent calibration. Hybrid devices that use electronic transducers instead of mercury have promise. The oscillometric method can be used for office measurement, but only devices independently validated according to standard protocols should be used, and individual calibration is recommended. They have the advantage of being able to take multiple measurements. Proper training of observers, positioning of the patient, and selection of cuff size are all essential. It is increasingly recognized that office measurements correlate poorly with blood pressure measured in other settings, and that they can be supplemented by self-measured readings taken with validated devices at home. There is increasing evidence that home readings predict cardiovascular events and are particularly useful for monitoring the effects of treatment. Twenty-four-hour ambulatory monitoring gives a better prediction of risk than office measurements and is useful for diagnosing white-coat hypertension. There is increasing evidence that a failure of blood pressure to fall during the night may be associated with increased risk. In obese patients and children, the use of an appropriate cuff size is of paramount importance.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Many studies utilized 24-hour (or 48-hour) ambulatory BP measurements, considered to be the most accurate method to diagnose hypertension and the best predictor of future cardiovascular risk.154”
From Treatment of Adult Obstructive Sleep Apnea With Positive Airway Pressure: An American Academy of Sleep Medicine Systematic Review, Meta-Analysis, and GRADE Assessment (2019) · cites this paper as [154]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“Blood pressure recorded at each visit was the mean of two or more BP measurements obtained using the recommended measurement guidelines from the American Heart Association [22].”
In the citing paper
“As with best-practice measurement of BP under resting conditions, operators must follow a standardized protocol to measure exercise BP because deviation from recommended protocols will increase the likelihood of BP measurement error [49].”
From The Identification and Management of High Blood Pressure Using Exercise Blood Pressure: Current Evidence and Practical Guidance (2022) · cites this paper as [49]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Items 201–226 of 226
Accuracy of prehospital sphygmomanometers
Jones, J. S., Ramsey, W., & Hetrick, T. (1987) · Journal of Emergency Medicine · doi:10.1016/0736-4679(87)90006-0 · Cited by 26
Ambulatory monitoring and self-monitoring of blood pressure during pregnancy
Waugh, J. J. S., Halligan, A. W. F., & Shennan, A. (2000) · Blood Pressure Monitoring · doi:10.1097/00126097-200002000-00002 · Cited by 26
Abe, H., Yokouchi, M., Saitoh, F., Deguchi, F., Kimura, G., Kojima, S., Yoshimi, H., Ito, K., Kuramochi, M., & Ikeda, M. (1987) · PubMed · Cited by 25
Whitcomb, B. L. (1995) · Archives of Family Medicine · doi:10.1001/archfami.4.5.419 · Cited by 24
Measurement of blood pressure and heart rate by telemetry in conscious unrestrained marmosets
Schnell, C., & Wood, J. M. (1995) · Laboratory Animals · doi:10.1258/002367795781088180 · Cited by 24
Utility of blood pressure monitoring outside of the clinic setting.
Appel, L. J., Robinson, K., Güallar, E., Erlinger, T. P., Masood, S. O., Jehn, M., Fleisher, L. A., Powe, N., & Bass, E. B. (2002) · PubMed Central · Cited by 23
Self-citationAn evaluation of the Vita-Stat automatic blood pressure measuring device.
Polk, B. F., ROSNER, B. A., Feudo, R., & VanDenburgh, M. (1980) · Hypertension · doi:10.1161/01.hyp.2.2.221 · Cited by 22
A Community-based Evaluation of the Vita-Stat Automatic Blood Pressure Recorder
Salaita, K., Whelton, P. K., & Seidler, A. J. (1990) · American Journal of Hypertension · doi:10.1093/ajh/3.5.366 · Cited by 22
Assessment of a blood pressure measurement training programme for lay observers
Graciani, A., Banegas, J. R., López‐García, E., Díez‐Gañán, L., & Rodríguez‐Artalejo, F. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200210000-00001 · Cited by 22
Auscultatory Measurement of Blood Pressure Performed by the Doctor on Duty
Nielsen, P. E., Oxenbøll, B., Astvad, K., & Gyntelberg, F. (1975) · Acta Medica Scandinavica · doi:10.1111/j.0954-6820.1975.tb19502.x · Cited by 21
Adams, C., Burke, V., & Beilin, L. J. (2002) · Journal of Clinical Epidemiology · doi:10.1016/s0895-4356(01)00492-9 · Cited by 21
Clinical Evaluation of a Finger Oscillometric Blood Pressure Device
Sesler, J. M., Munroe, W. P., & McKenney, J. M. (1991) · DICP · doi:10.1177/106002809102501204 · Cited by 21
Tracking of 24-hour and casual blood pressure
O’Sullivan, J. J., Derrick, G., & Foxall, R. (2000) · Journal of Hypertension · doi:10.1097/00004872-200018090-00004 · Cited by 20
Whelton, P. K., Thompson, S. G., Barnes, G., & Miall, W. E. (1983) · American Journal of Epidemiology · doi:10.1093/oxfordjournals.aje.a113514 · Cited by 18
Mapping Baroreceptor Function to Genome: A Mathematical Modeling Approach
Kendziorski, C. M., Cowley, A. W., Greene, A. S., Salgado, H. C., Jacob, H. J., & Tonellato, P. J. (2002) · Genetics · doi:10.1093/genetics/160.4.1687 · Cited by 17
Adverse Effect of Helicopter Flight on the Ability to Palpate Carotid Pulses
Hunt, R. C., Carroll, R. G., Whitley, T. W., Bryan-Berge, D., & Dufresne, D. A. (1994) · Annals of Emergency Medicine · doi:10.1016/s0196-0644(94)70129-6 · Cited by 17
Is early diagnosis of hypertension a function of cuff width?
Arcuri, E. A. M., Santos, J. L. F., & Silva, M. R. E. (1989) · Journal of Hypertension · doi:10.1097/00004872-198900076-00026 · Cited by 15
Wessig, K., Hollinger, S., Schmalzhaf, I., & Lenz, T. (2000) · Blood Pressure Monitoring · doi:10.1097/00126097-200008000-00008 · Cited by 14
Noninvasive versus invasive blood pressure measurement in normotensive and hypotensive baboons.
Hartford, C. G., Marcos, E. F., & Rogers, G. (1996) · PubMed · Cited by 10
[Reliability of the measurement of arterial pressure in the forearm in the obese subject].
Warembourg, A., Poncelet, P., & Carre, A. G. (1987) · PubMed · Cited by 8
Suggestions to bring electronic protocol system into compliance.
Garnett, N. L., & Gipson, C. A. (2001) · PubMed · Cited by 7
Adaptation to restraint in the rat.
Popovic, V. (1988) · PubMed · Cited by 7
Bicycle accidents in childhood.
Worrell, J. (1987) · BMJ · doi:10.1136/bmj.294.6587.1614-b · Cited by 6
HALOTHANE IMMUNOSUPPRESSION IN TUMOR-BEARING MICE
Lewis, R. E., Cruse, J. M., & Hazlewood, J. (1979) · Anesthesiology · doi:10.1097/00000542-197909001-00252 · Cited by 6
National guidelines are needed to provide sanitary facilities in primary schools
Finlay, F., & Jones, R. (2001) · BMJ · doi:10.1136/bmj.323.7309.398 · Cited by 6
Long-Term Blood-Pressure Monitoring in Unrestrained Animals
Griffin, K. A., Abu‐Amarah, I., & BIDANI, A. K. (2003) · Humana Press eBooks · doi:10.1385/1-59259-392-5:329 · Cited by 6
Flagged first, then by citations
Research works printed but not linked: 35
- Verdecchia P, Schillaci G, Borgioni C, Ciucci A, Gattobigio R, Porcellati C. Nocturnal pressure is the true pressure. Blood Press Monit. 1996; 1 (Suppl 2): S81–S85.
- 10.1161/circ.103.9.1245 · DOI
- 10.1161/hyp.36.5.801 · DOI
- 10.1161/hyp.25.6.1276 · DOI
- 10.1161/circ.100.4.354 · DOI
- 10.1161/circ.103.9.1245 · DOI
- Pickering TG. White coat hypertension. In: Izzo JL Black HR eds. Hypertension Primer third edition. Dallas: American Heart Association; 2003.
- 10.1161/circ.102.10.1139 · DOI
- Low PA. Clinical Autonomic Disorders: Evaluation and Management second ed. Philadelphia: Lippincott-Raven; 1997.
- 10.1161/str.27.8.1316 · DOI
- 10.1161/circ.8.2.232 · DOI
- 10.1161/circ.88.5.8222141 · DOI
- Cates EM, Schlussel YR, James GD, Pickering TG. A validation study of the Spacelabs 90207 ambulatory blood pressure monitor. J Ambul Monitor. 1990; 3: 149–154.
- Penaz J. Photo-electric measurement of blood pressure volume and flow in the finger. Digest Tenth International Conference Medical Biological Engineering. Dresden; 1973:104.
- Ware RW, Laenger CJ. Indirect blood pressure measurement by Doppler ultrasonic kinetoarteriography. Proc 20th Annual Conference Engineering Medical Biology. 1967; 9: 27–30.
- Wonka F, Thummler M, Schoppe A. Clinical test of a blood pressure measurement device with a wrist cuff. Blood Press Monit. 1996; 1: 361–366.
- 10.1161/hyp.36.4.484 · DOI
- Grim CE, Grim CM, Li J. Entering medical students who say they have been trained to take blood pressure do not follow American Heart Association guidelines. Am J Hypertens. 1999; 12.
- 10.1001/archinte.1984.00350190188032 · DOI
- Beevers DG, Beevers M, Zezulka AV, Hornby R, Herod J. Standards for blood pressure measuring devices. BMJ (Clin Res Ed). 1987; 294: 1614.
- O’Brien E, Atkins N, O’Malley K. Selecting the correct bladder according to distribution of arm circumference in the population (abstr). Hypertension. 1993; 11: 1149–1150.
- 10.1161/circ.68.4.6616774 · DOI
- Ljungvall P, Thorvinger B, Thulin T. The influence of a heart level pillow on the result of blood pressure measurement. J Hum Hypertens. 1989; 3: 471–474.
- Rose G. Standardisation of observers in blood pressure measurement. Lancet. 1965; 10: 673–674.
- 10.1161/circ.28.1.20 · DOI
- 10.1001/jama.1967.03130150084015 · DOI
- 10.1161/hyp.35.5.1032 · DOI
- Myers MG. Self-measurement of blood pressure at home: the potential for reporting bias. Blood Press Monit. 1998; 3 (suppl 1): S19–S22.
- Imai Y, Ohkubo T, Tsuji I, Nagai K, Satoh H, Hisamichi S, Abe K. Prognostic value of ambulatory and home blood pressure measurements in comparison to screening blood pressure measurements: a pilot study in Ohasama. Blood Press Monit. 1996; 1 (Suppl 2): S51–S58.
- 10.1161/hyp.32.2.255 · DOI
- Verdecchia P Bentivoglia M Providenza M Savino K Corea L. Reliability of home self-recorded arterial pressure in essential hypertension in relation to the stage of the disease. In: Germano G ed. Blood pressure recording in the clinical management of hypertension. Rome: Ediziono Pozzi; 1985: 40–42.
- 10.1161/circ.98.18.1892 · DOI
- 10.1161/hyp.31.2.712 · DOI
- Prineas RJ Elkwiry ZM. Epidemiology and measurement of high blood pressure in children and adolescents. In: Loggie JMH ed. Pediatric and Adolescent Hypertension. Boston MA: Blackwell Scientific Publications; 1992: 91–103.
- Brown MA Davis GK. Hypertension in pregnancy. In: Mancia G Chalmers J Julius S Saruta T Weber MA Ferrari AU Wilkinson IB ed. Manual of Hypertension. London: Harcourt Publishers Limited; 2002: 579–597.
Cited by
Sorted by citations
Measurement of Blood Pressure in Humans: A Scientific Statement From the American Heart Association
Muntner, P., Shimbo, D., Carey, R. M., Charleston, J. B., Gaillard, T., Misra, S., Myers, M. G., Ogedegbe, G., Schwartz, J. E., Townsend, R. R., et al. (2019) · Hypertension · doi:10.1161/hyp.0000000000000087 · Cited by 1,512
European Society of hypertension position paper on ambulatory blood pressure monitoring
O’Brien, E., Parati, G., Stergiou, G., Asmar, R., Beilin, L., Bilo, G., Clement, D., de la Sierra, A., de Leeuw, P., Dolan, E., et al. (2013) · Journal of Hypertension · doi:10.1097/hjh.0b013e328363e964 · Cited by 1,419
Self-citationAstrup, A., Rössner, S., Van Gaal, L., Rissanen, A., Niskanen, L., Al Hakim, M., Madsen, J., Rasmussen, M. F., & Lean, M. E. (2009) · The Lancet · doi:10.1016/s0140-6736(09)61375-1 · Cited by 1,163
Addressing Current Criticism Regarding the Value of Self-Report Dietary Data
Subar, A. F., Freedman, L. S., Tooze, J. A., Kirkpatrick, S. I., Boushey, C., Neuhouser, M. L., Thompson, F. E., Potischman, N., Guenther, P. M., Tarasuk, V., et al. (2015) · The Journal of Nutrition · doi:10.3945/jn.115.219634 · Cited by 1,047
Assessment and management of blood-pressure variability
Parati, G., Ochoa, J. E., Lombardi, C., & Bilo, G. (2013) · Nature Reviews Cardiology · doi:10.1038/nrcardio.2013.1 · Cited by 853
Metadata
- Authors
- Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J.
- ORCID
- Venue
- Hypertension
- Published
- 2005-01
- Volume
- 45
- Issue
- 1
- Pages
- 142-161
- ISSN
- 0194-911X (ISSN-L)
- 1524-4563
- OpenAlex
- W2160801689
- Semantic Scholar
- cc608ac67b0966145e814f721be04e1b0f95eeb3
- Cited by
- 2,051
- References
- 226
- Citations by year
- 2026: 34 · 2025: 43 · 2024: 48 · 2023: 61 · 2022: 79 · 2021: 77 · 2020: 101 · 2019: 85 · 2018: 113 · 2017: 144 · 2016: 166 · 2015: 161 · 2014: 104 · 2013: 114 · 2012: 133
- FWCI
- 39.42
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Hypertension (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- blood pressure measurement, auscultatory method, Korotkoff sounds, mercury sphygmomanometer, aneroid sphygmomanometer, oscillometric method, home blood pressure monitoring, ambulatory blood pressure monitoring, white coat hypertension, cuff size, observer training, pregnant women, laboratory animals, hypertension risk prediction
- MeSH terms
- Adult, Aged, Animals, Auscultation, Automation, Blood Pressure Determination, Calibration, Cardiovascular Diseases, Child, Circadian Rhythm, Clinical Competence, Diagnostic Errors, Equipment Design, Female, Humans, Hypertension, Male, Middle Aged, Obesity, Organ Specificity, Oscillometry, Posture, Predictive Value of Tests, Pregnancy, Self Care, Stress, Psychological, Blood Pressure Monitors, Blood Pressure Monitoring, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review