Recommendations for Blood Pressure Measurement in Humans and Experimental Animals: Part 1: Blood Pressure Measurement in Humans: A Statement for Professionals From the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research
Open accessAbstract
Accurate measurement of blood pressure is essential to classify individuals, to ascertain blood pressure–related risk, and to guide management. The auscultatory technique with a trained observer and mercury sphygmomanometer continues to be the method of choice for measurement in the office, using the first and fifth phases of the Korotkoff sounds, including in pregnant women. The use of mercury is declining, and alternatives are needed. Aneroid devices are suitable, but they require frequent calibration. Hybrid devices that use electronic transducers instead of mercury have promise. The oscillometric method can be used for office measurement, but only devices independently validated according to standard protocols should be used, and individual calibration is recommended. They have the advantage of being able to take multiple measurements. Proper training of observers, positioning of the patient, and selection of cuff size are all essential. It is increasingly recognized that office measurements correlate poorly with blood pressure measured in other settings, and that they can be supplemented by self-measured readings taken with validated devices at home. There is increasing evidence that home readings predict cardiovascular events and are particularly useful for monitoring the effects of treatment. Twenty-four-hour ambulatory monitoring gives a better prediction of risk than office measurements and is useful for diagnosing white-coat hypertension. There is increasing evidence that a failure of blood pressure to fall during the night may be associated with increased risk. In obese patients and children, the use of an appropriate cuff size is of paramount importance.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Many studies utilized 24-hour (or 48-hour) ambulatory BP measurements, considered to be the most accurate method to diagnose hypertension and the best predictor of future cardiovascular risk.154”
From Treatment of Adult Obstructive Sleep Apnea With Positive Airway Pressure: An American Academy of Sleep Medicine Systematic Review, Meta-Analysis, and GRADE Assessment (2019) · cites this paper as [154]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“Blood pressure recorded at each visit was the mean of two or more BP measurements obtained using the recommended measurement guidelines from the American Heart Association [22].”
In the citing paper
“As with best-practice measurement of BP under resting conditions, operators must follow a standardized protocol to measure exercise BP because deviation from recommended protocols will increase the likelihood of BP measurement error [49].”
From The Identification and Management of High Blood Pressure Using Exercise Blood Pressure: Current Evidence and Practical Guidance (2022) · cites this paper as [49]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Chobanian, A. V., Bakris, G. L., Black, H. R., Cushman, W. C., Green, L. A., Izzo, J. L., Jones, D. W., Materson, B. J., Oparil, S., Wright, J. T., et al. (2003) · Hypertension · doi:10.1161/01.hyp.0000107251.49515.c2 · Cited by 13,271
Self-citationPrevalence and trends in obesity among US adults, 1999–2000
Flegal, K. M. (2002) · JAMA · doi:10.1001/jama.288.14.1723 · Cited by 6,190
US Environmental Protection Agency
Liu, X. (1998) · Choice Reviews Online · doi:10.5860/choice.35-4455 · Cited by 5,887
Is Pulse Pressure Useful in Predicting Risk for Coronary Heart Disease?
Franklin, S. S., Khan, S. A., Wong, N. D., Larson, M. G., & Levy, D. A. (1999) · Circulation · doi:10.1161/01.cir.100.4.354 · Cited by 1,756
Ambulatory blood pressure. An independent predictor of prognosis in essential hypertension.
Verdecchia, P., Porcellati, C., Schillaci, G., Borgioni, C., Ciucci, A., Battistelli, M., Guerrieri, M., Gatteschi, C., Zampi, I., Santucci, A., et al. (1994) · Hypertension · doi:10.1161/01.hyp.24.6.793 · Cited by 1,671
O'Brien, E., Asmar, R., Beilin, L., Imai, Y., Mallion, J.-M., Mancia, G., Mengden, T., Myers, M., Padfield, P., Palatini, P., et al. (2003) · Journal of Hypertension · doi:10.1097/00004872-200305000-00001 · Cited by 1,655
Self-citationHuman blood pressure determination by sphygmomanometry.
Perloff, D., Grim, C. M., Flack, J. M., Frohlich, E. D., Hill, M. D., McDonald, M., & Morgenstern, B. Z. (1993) · Circulation · doi:10.1161/01.cir.88.5.2460 · Cited by 1,592
Self-citationStaessen, J. A. (1999) · JAMA · doi:10.1001/jama.282.6.539 · Cited by 1,592
Franklin, S. S., Larson, M. G., Khan, S. A., Wong, N. D., Leip, E. P., Kannel, W. B., & Levy, D. (2001) · Circulation · doi:10.1161/01.cir.103.9.1245 · Cited by 1,390
Prognostic Value of Ambulatory Blood-Pressure Recordings in Patients with Treated Hypertension
Clement, D. L., De Buyzere, M. L., De Bacquer, D. A., de Leeuw, P. W., Duprez, D. A., Fagard, R. H., Gheeraert, P. J., Missault, L. H., Braun, J. J., Six, R. O., et al. (2003) · New England Journal of Medicine · doi:10.1056/nejmoa022273 · Cited by 1,085
National High Blood Pressure Education Program Working Group on Hypertension Control in Children and Adolescents (1996) · PEDIATRICS · doi:10.1542/peds.98.4.649 · Cited by 999
Blood pressure measuring devices: recommendations of the European Society of Hypertension
O'Brien, E. T., Waeber, B., Parati, G., Staessen, J. A., & Myers, M. G. (2001) · BMJ · doi:10.1136/bmj.322.7285.531 · Cited by 906
Parati, G., Casadei, R., Groppelli, A., Di Rienzo, M., & Mancia, G. (1989) · Hypertension · doi:10.1161/01.hyp.13.6.647 · Cited by 899
Parati, G., Saul, J. P., Di Rienzo, M., & Mancia, G. (1995) · Hypertension · doi:10.1161/01.hyp.25.6.1276 · Cited by 849
Bobrie, G. (2004) · JAMA · doi:10.1001/jama.291.11.1342 · Cited by 842
Wühl, E., Witte, K. K., Soergel, M., Mehls, O., & Schaefer, F. S. (2002) · Journal of Hypertension · doi:10.1097/00004872-200210000-00019 · Cited by 816
Neaton, J. D. (1992) · Archives of Internal Medicine · doi:10.1001/archinte.1992.00400130082009 · Cited by 784
O'Brien, E. T., Pickering, T. G., Asmar, R. G., Myers, M. G., Parati, G., Staessen, J. A., Mengden, T., Imai, Y., Waeber, B., Palatini, P., et al. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200202000-00002 · Cited by 735
Self-citationEpidemiologic assessment of the role of blood pressure in stroke. The Framingham study
Kannel, W. B. (1970) · JAMA · doi:10.1001/jama.214.2.301 · Cited by 706
A Longitudinal Study of Coronary Heart Disease
Paul, O., Lepper, M. H., Phelan, W. H., DUPERTUIS, G. W., Macmillan, A., McKean, H. E., & Park, H.-B. (1963) · Circulation · doi:10.1161/01.cir.28.1.20 · Cited by 610
Sesso, H. D., Stampfer, M. J., ROSNER, B. A., Hennekens, C. H., Gaziano, J. M., Manson, J. E., & Glynn, R. J. (2000) · Hypertension · doi:10.1161/01.hyp.36.5.801 · Cited by 599
COWLEY, A. W., LIARD, J. F., & GUYTON, A. C. (1973) · Circulation Research · doi:10.1161/01.res.32.5.564 · Cited by 593
A Noninvasive Computerized Tail-Cuff System for Measuring Blood Pressure in Mice
Krege, J. H., Hodgin, J. B., Hagaman, J. R., & Smithies, O. (1995) · Hypertension · doi:10.1161/01.hyp.25.5.1111 · Cited by 587
Relation Between Nocturnal Decline in Blood Pressure and Mortality The Ohasama Study
Ohkubo, T. (1997) · American Journal of Hypertension · doi:10.1016/s0895-7061(97)00274-4 · Cited by 491
Validation in awake rats of a tail-cuff method for measuring systolic pressure.
Buñag, R. D. (1973) · Journal of Applied Physiology · doi:10.1152/jappl.1973.34.2.279 · Cited by 490
Abnormal Vascular Function and Hypertension in Mice Deficient in Estrogen Receptor β
Zhu, Y., Bian, Z., Lu, P., Karas, R. H., Bao, L., Cox, D. H., Hodgin, J. B., Shaul, P. W., Thorén, P., Smithies, O., et al. (2002) · Science · doi:10.1126/science.1065250 · Cited by 482
Recommendations for the use of home (self) and ambulatory blood pressure monitoring
Pickering, T. G. (1996) · American Journal of Hypertension · doi:10.1016/0895-7061(95)00341-x · Cited by 473
Self-citationPickering, T. G., Davidson, K. W., Gerin, W., & Schwartz, J. E. (2002) · Hypertension · doi:10.1161/01.hyp.0000038733.08436.98 · Cited by 437
Self-citationPrognostic Value of Ambulatory Blood Pressure Monitoring in Refractory Hypertension
Redón, J., Campos, C., Narciso, M. L., Rodicio, J. L., Pascual, J. M., & Ruilope, L. M. (1998) · Hypertension · doi:10.1161/01.hyp.31.2.712 · Cited by 418
24-h ambulatory blood pressure in 352 normal Danish subjects, related to age and gender*
Wiinberg, N., Høegholm, A., Christensen, H. R., Bang, L. E., Mikkelsen, K. L., Nielsen, P. E., Svendsen, T. L., Kampmann, J. P., Madsen, N. H., & Bentzon, M. W. (1995) · American Journal of Hypertension · doi:10.1016/0895-7061(95)00216-2 · Cited by 412
Recommendations on blood pressure measurement.
Petrie, J. C., O'Brien, E. T., Littler, W. A., & de Swiet, M. (1986) · BMJ · doi:10.1136/bmj.293.6547.611 · Cited by 396
Pfeffer, J. M., Pfeffer, & Frohlich, E. D. (1971) · PubMed · Cited by 390
Regulator of G-protein signaling-2 mediates vascular smooth muscle relaxation and blood pressure
Tang, M., Wang, G., Lu, P., Karas, R. H., Aronovitz, M. J., Heximer, S. P., Kaltenbronn, K. M., Blumer, K., Siderovski, D. P., Zhu, Y., et al. (2003) · Nature Medicine · doi:10.1038/nm958 · Cited by 387
Obesity-induced hypertension. Renal function and systemic hemodynamics.
Hall, J. E., Brands, M. W., Dixon, W. N., & Smith, M. J. (1993) · Hypertension · doi:10.1161/01.hyp.22.3.292 · Cited by 328
Self-citationMiura, K., Daviglus, M. L., Dyer, A. R., Liu, K., Garside, D. B., Stamler, J., & Greenland, P. (2001) · Archives of Internal Medicine · doi:10.1001/archinte.161.12.1501 · Cited by 309
Borow, K. M., & Newburger, J. W. (1982) · American Heart Journal · doi:10.1016/0002-8703(82)90403-3 · Cited by 308
ERROR IN BLOOD-PRESSURE MEASUREMENT DUE TO INCORRECT CUFF SIZE IN OBESE PATIENTS
Maxwell, M. H., Schroth, P., Waks, A. U., Karam, M., & Dornfeld, L. (1982) · The Lancet · doi:10.1016/s0140-6736(82)91163-1 · Cited by 307
Practice guidelines for primary care physicians: 2003 ESH/ESC hypertension guidelines.
Cífková, R., Erdine, S. E., Fagard, R. H., Farsang, C., Heagerty, A. M., Kiowski, W., Kjeldsen, S., Lüscher, T. F., Mallion, J. M., Mancia, G., et al. (2003) · PubMed · doi:10.1097/01.hjh.0000084773.37215.1b · Cited by 299
Cardiovascular Outcome in White-Coat Versus Sustained Mild Hypertension
Khattar, R. S., Senior, R., & Lahiri, A. (1998) · Circulation · doi:10.1161/01.cir.98.18.1892 · Cited by 298
Reliability of reporting self-measured blood pressure values by hypertensive patients.
Mengden, T., Medina, R. M. H., Beltran, B., Álvarez, E., Kraft, K., & Vetter, H. (1998) · American Journal of Hypertension · doi:10.1016/s0895-7061(98)00241-6 · Cited by 296
What is the value of home blood pressure measurement in patients with mild hypertension?
Kleinert, H. D., Harshfield, G. A., Pickering, T. G., Devereux, R. B., Sullivan, P. A., Marion, R., Mallory, W. K., & LARAGH, J. H. (1984) · Hypertension · doi:10.1161/01.hyp.6.4.574 · Cited by 295
Self-citationBarone, F. C., Knudsen, D. J., Nelson, A., FEUERSTEIN, G. Z., & Willette, R. N. (1993) · Journal of Cerebral Blood Flow & Metabolism · doi:10.1038/jcbfm.1993.87 · Cited by 291
Butz, G. M., & Davisson, R. L. (2001) · Physiological Genomics · doi:10.1152/physiolgenomics.2001.5.2.89 · Cited by 282
Fagard, R. H., Staessen, J. A., Thijs, L., Gąsowski, J., Bulpitt, C. J., Clement, D. L., de Leeuw, P. W., Dobovišek, J., Jääskivi, M., Leonetti, G., et al. (2000) · Circulation · doi:10.1161/01.cir.102.10.1139 · Cited by 281
Feasibility of ambulatory, continuous 24-hour finger arterial pressure recording.
Imholz, B. P. M., Langewouters, G. J., van Montfrans, G. A., Parati, G., van Goudoever, J., Wesseling, K. H., Wieling, W., & Mancia, G. (1993) · Hypertension · doi:10.1161/01.hyp.21.1.65 · Cited by 259
Surgery's relevance to an understanding of basic biology. Tissue repair and cellular regeneration
Dunphy, J. E. (1967) · JAMA · doi:10.1001/jama.202.2.116 · Cited by 258
Brockway, B. P., Mills, P. A., & Azar, S. H. (1991) · Clinical and Experimental Hypertension · doi:10.3109/10641969109042094 · Cited by 249
Home Monitoring Service Improves Mean Arterial Pressure in Patients with Essential Hypertension
Rogers, M. A. M., Small, D. I., Buchan, D. A., Butch, C. A., Stewart, C. M., Krenzer, B. E., & Husovsky, H. L. (2001) · Annals of Internal Medicine · doi:10.7326/0003-4819-134-11-200106050-00008 · Cited by 246
Fujii, M., Hara, H., Meng, W., Vonsattel, J. P., Huang, Z., & Moskowitz, M. A. (1997) · Stroke · doi:10.1161/01.str.28.9.1805 · Cited by 242
Arterial tonometry: Review and analysis
Drzewiecki, G. M., Melbin, J., & Noordergraaf, A. (1983) · Journal of Biomechanics · doi:10.1016/0021-9290(83)90037-4 · Cited by 236
Ohkubo, T., Imai, Y., Tsuji, I., Nagai, K., Ito, S., Satoh, H., & Hisamichi, S. (1998) · HYPERTENSION · doi:10.1161/01.hyp.32.2.255 · Cited by 235
Tail-cuff blood pressure measurement without external preheating in awake rats.
Buñag, R. D., & Butterfield, J. (1982) · Hypertension · doi:10.1161/01.hyp.4.6.898 · Cited by 230
Evaluation and applications of radiotelemetry in small laboratory animals
Kramer, K., & Kinter, L. B. (2003) · Physiological Genomics · doi:10.1152/physiolgenomics.00164.2002 · Cited by 224
Mauck, G. W., Smith, C. R., Geddes, L. A., & BOURLAND, J. D. (1980) · Journal of Biomechanical Engineering · doi:10.1115/1.3138195 · Cited by 222
Kario, K., Shimada, K., Schwartz, J. E., Matsuo, T., Hoshide, S., & Pickering, T. G. (2001) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(01)01325-0 · Cited by 222
Self-citationPractice Guidelines For Primary Care Physicians
&NA; (2003) · Journal of Hypertension · doi:10.1097/00004872-200310000-00001 · Cited by 215
Discovery of a Spontaneous Genetic Mouse Model of Preeclampsia
Davisson, R. L., Hoffmann, D. S., Butz, G. M., Aldape, G., Schlager, G., Merrill, D., Sethi, S. M., Weiss, R. M., & Bates, J. N. (2002) · Hypertension · doi:10.1161/hy02t2.102904 · Cited by 203
Acute hypertension activates mitogen-activated protein kinases in arterial wall.
Xu, Q., Liu, Y., Gorospe, M., Udelsman, R., & Holbrook, N. J. (1996) · Journal of Clinical Investigation · doi:10.1172/jci118442 · Cited by 201
Tsuji, I. (1997) · American Journal of Hypertension · doi:10.1016/s0895-7061(96)00407-4 · Cited by 195
Circadian blood pressure variation in transgenic hypertensive rats.
Lemmer, B., Mattes, A., Böhm, M., & Ganten, D. (1993) · Hypertension · doi:10.1161/01.hyp.22.1.97 · Cited by 194
Task Force II: Blood pressure measurement and cardiovacular outcome
Staessen, J. A., Asmar, R. G., Buyzere, M. D., Imai, Y., Parati, G., Shimada, K., Stergiou, G. S., Redón, J., & Verdecchia, P. (2001) · Blood Pressure Monitoring · doi:10.1097/00126097-200112000-00016 · Cited by 187
Norman, R. A., Coleman, T. G., & Dent, A. C. (1981) · Hypertension · doi:10.1161/01.hyp.3.1.119 · Cited by 186
Evidence for endothelin involvement in the response to high salt
Pollock, D. M., & Pollock, J. S. (2001) · American Journal of Physiology-Renal Physiology · doi:10.1152/ajprenal.2001.281.1.f144 · Cited by 183
Modulation of cardiovascular control mechanisms and their interaction
Persson, P. B. (1996) · Physiological Reviews · doi:10.1152/physrev.1996.76.1.193 · Cited by 182
Clinical Utility of Aortic Pulses and Pressures Calculated From Applanated Radial-Artery Pulses
Smulyan, H., Siddiqui, D. S., Carlson, R. J., London, G. M., & Safar, M. E. (2003) · Hypertension · doi:10.1161/01.hyp.0000084051.34269.a9 · Cited by 178
Prospective Evaluation of Clinical Characteristics of Orthostatic Hypotension
Low, P. A., Opfer‐Gehrking, T. L., McPhee, B. R., Fealey, R. D., Benarroch, E. E., Willner, C. L., Suárez, G. A., Proper, C. J., Felten, J. A., Huck, C. A., et al. (1995) · Mayo Clinic Proceedings · doi:10.4065/70.7.617 · Cited by 178
Inter-arm differences in blood pressure: when are they clinically significant?
Lane, D. A., Beevers, M., Barnes, N., Bourne, J., John, A., Malins, S., & Beevers, D. G. (2002) · Journal of Hypertension · doi:10.1097/00004872-200206000-00019 · Cited by 176
Oscillometric blood pressure measurement: progress and problems
van Montfrans, G. A. (2001) · Blood Pressure Monitoring · doi:10.1097/00126097-200112000-00004 · Cited by 175
Griffin, K. A., Picken, M. M., & BIDANI, A. K. (1995) · Journal of Clinical Investigation · doi:10.1172/jci118125 · Cited by 174
Mills, P. A., Huetteman, D. A., Brockway, B. P., Zwiers, L. M., Gelsema, A. J. M., Schwartz, R. S., & Kramer, K. (2000) · Journal of Applied Physiology · doi:10.1152/jappl.2000.88.5.1537 · Cited by 173
van Popele, N. M., Bos, W. J. W., de Beer, N. A. M., van der Kuip, D. A. M., Hofman, A., Grobbee, D. E., & Witteman, J. C. M. (2000) · Hypertension · doi:10.1161/01.hyp.36.4.484 · Cited by 170
A Genomic-Systems Biology Map for Cardiovascular Function
Stoll, M., Cowley, A. W., Tonellato, P. J., Greene, A. S., Kaldunski, M. L., Roman, R. J., Dumas, P., Schork, N. J., Wang, Z., & Jacob, H. J. (2001) · Science · doi:10.1126/science.1062117 · Cited by 168
Davisson, R. L., Yang, G., Beltz, T. G., Cassell, M. D., Johnson, A. K., & Sigmund, C. D. (1998) · Circulation Research · doi:10.1161/01.res.83.10.1047 · Cited by 156
Hypertensive cardiovascular damage in patients with primary autonomic failure
Vagaonescu, T. D., Saadia, D., Tuhrim, S., Phillips, R. A., & Kaufmann, H. (2000) · The Lancet · doi:10.1016/s0140-6736(99)05320-9 · Cited by 154
Role of renal hemodynamics and arterial pressure in aldosterone "escape".
Hall, J. E., Granger, J. P., Smith, M. J., & Premen, A. J. (1984) · Hypertension · doi:10.1161/01.hyp.6.2_pt_2.i183 · Cited by 153
Self-citationIndirect Measurement of Mean Blood Pressure in the Anesthetized Patient
Yelderman, M., & Ream, A. K. (1979) · Anesthesiology · doi:10.1097/00000542-197903000-00022 · Cited by 151
The reliability of vital sign measurements
Edmonds, Z. V., Mower, W. R., Lovato, L. M., & Lomeli, R. (2002) · Annals of Emergency Medicine · doi:10.1067/mem.2002.122017 · Cited by 150
The influence of restraint on blood pressure in the rat
Irvine, R. J., White, J. M., & Chan, R. (1997) · Journal of Pharmacological and Toxicological Methods · doi:10.1016/s1056-8719(97)00081-6 · Cited by 148
Autonomic cardiovascular control in conscious mice
Just, A., Faulhaber, J., & Ehmke, H. (2000) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.2000.279.6.r2214 · Cited by 148
Meneton, P., Ichikawa, I., Inagami, T., & Schnermann, J. B. (2000) · American Journal of Physiology-Renal Physiology · doi:10.1152/ajprenal.2000.278.3.f339 · Cited by 145
Cardiovascular responses to caloric restriction and thermoneutrality in C57BL/6J mice
Williams, T. D., Chambers, J. B., Henderson, R. P., Rashotte, M. E., & Overton, J. M. (2002) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.00612.2001 · Cited by 132
Measurement of Blood-pressure: Comparison of Intra-arterial and Cuff Values
Holland, W. W., & Humerfelt, S. (1964) · BMJ · doi:10.1136/bmj.2.5419.1241 · Cited by 131
Are Aneroid Sphygmomanometers Accurate in Hospital and Clinic Settings?
Canzanello, V. J., Jensen, P. L., & Schwartz, G. L. (2001) · Archives of Internal Medicine · doi:10.1001/archinte.161.5.729 · Cited by 131
INVASIVE V. NON-INVASIVE MEASUREMENT OF ARTERIAL PRESSURE
van Egmond, J., Hasenbos, M. A. W. M., & Crul, J. F. (1985) · British Journal of Anaesthesia · doi:10.1093/bja/57.4.434 · Cited by 129
Enhanced Blood Pressure Variability in eNOS Knockout Mice
Stauss, H. M., Gödecke, A., Mrowka, R., Schrader, J., & Persson, P. B. (1999) · Hypertension · doi:10.1161/01.hyp.33.6.1359 · Cited by 126
Effect of Vertical Displacement of the Arm on Indirect Blood-Pressure Measurement
Mitchell, P. L., Parlin, R. W., & Blackburn, H. (1964) · New England Journal of Medicine · doi:10.1056/nejm196407092710204 · Cited by 125
Long-term renal consequences of hypertension for normal and diseased kidneys
BIDANI, A. K., & Griffin, K. A. (2002) · Current Opinion in Nephrology & Hypertension · doi:10.1097/00041552-200201000-00011 · Cited by 123
Comparison of performance of various sphygmomanometers with intra-arterial blood-pressure readings.
Hunyor, S. N., Flynn, J. M., & Cochineas, C. (1978) · BMJ · doi:10.1136/bmj.2.6131.159 · Cited by 122
Cardiovascular Influences of α 1b -Adrenergic Receptor Defect in Mice
Vecchione, C., Fratta, L., Rizzoni, D., Notte, A., Poulet, R., Porteri, E., Frati, G., Guelfi, D., Trimarco, V., Mulvany, M. J., et al. (2002) · Circulation · doi:10.1161/01.cir.0000012750.08480.55 · Cited by 122
Terminal digit preference, random error, and bias in routine clinical measurement of blood pressure
Wen, S. W., Kramer, M. S., Hoey, J., Hanley, J. A., & Usher, R. H. (1993) · Journal of Clinical Epidemiology · doi:10.1016/0895-4356(93)90118-k · Cited by 119
Direct and indirect methods used to study arterial blood pressure
Vliet, B. N. V., Chafe, L. L., Antic, V., Schnyder‐Candrian, S., & Montani, J. (2000) · Journal of Pharmacological and Toxicological Methods · doi:10.1016/s1056-8719(00)00126-x · Cited by 119
Rubini, R., Porta, A., Baselli, G., Cerutti, S., & Paro, M. (1993) · Journal of the Autonomic Nervous System · doi:10.1016/0165-1838(93)90050-5 · Cited by 118
Carlson, S. H., & Wyss, J. M. (2000) · Hypertension · doi:10.1161/01.hyp.35.2.e1 · Cited by 117
Effects of anesthesia on cardiovascular control mechanisms.
Vatner, S. F. (1978) · Environmental Health Perspectives · doi:10.1289/ehp.7826193 · Cited by 117
Renoprotection by ACE Inhibition or Aldosterone Blockade Is Blood Pressure–Dependent
Griffin, K. A., Abu‐Amarah, I., Picken, M. M., & BIDANI, A. K. (2003) · Hypertension · doi:10.1161/01.hyp.0000049881.25304.73 · Cited by 117
Comparing compliance patterns between randomized treatments
Vrijens, B., & Goetghebeur, E. (1997) · Controlled Clinical Trials · doi:10.1016/s0197-2456(96)00235-8 · Cited by 114
Inhibition of NO Synthesis Enhances Chronic Cardiovascular and Renal Actions of Leptin
Kuo, J. J., Jones, O. B., & Hall, J. E. (2001) · Hypertension · doi:10.1161/01.hyp.37.2.670 · Cited by 114
Self-citationErrors in clinical measurement of blood pressure in obesity.
King, G. (1967) · PubMed · Cited by 114
BIDANI, A. K., Griffin, K. A., Picken, M. M., & Lansky, D. M. (1993) · American Journal of Physiology-Renal Physiology · doi:10.1152/ajprenal.1993.265.3.f391 · Cited by 113
How accurate are sphygmomanometers?
Mion, D., & Pierin, A. (1998) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1000589 · Cited by 110
Ambulatory blood pressure monitoring in normotensive children
Lurbe, E., Redón, J., Liao, Y., Tacons, J., Cooper, R., & Álvarez, V. (1994) · Journal of Hypertension · doi:10.1097/00004872-199412000-00016 · Cited by 109
Characteristics of 24 h Telemetered Blood Pressure in eNOS‐Knockout and C57Bl/6J Control Mice
Vliet, B. N. V., Chafe, L. L., & Montani, J. (2003) · The Journal of Physiology · doi:10.1113/jphysiol.2003.041897 · Cited by 109
Miscuffing: inappropriate blood pressure cuff application.
Manning, D. M., Kuchirka, C., & Kamiński, J. (1983) · Circulation · doi:10.1161/01.cir.68.4.763 · Cited by 108
Radiotelemetric BP monitoring, antihypertensives and glomeruloprotection in remnant kidney model
Griffin, K. A., Picken, M. M., & BIDANI, A. K. (1994) · Kidney International · doi:10.1038/ki.1994.361 · Cited by 106
Assessment of techniques for measurement of blood pressure in infants and children.
Elseed, A. M., Shinebourne, E. A., & Joseph, M. C. (1973) · Archives of Disease in Childhood · doi:10.1136/adc.48.12.932 · Cited by 99
Electrocardiogram in Evaluation of Resistance to Antihypertensive Therapy
Ibrahim, M. M. (1977) · Archives of Internal Medicine · doi:10.1001/archinte.1977.03630210011006 · Cited by 98
Genomic map of cardiovascular phenotypes of hypertension in female Dahl S rats
Moreno, C., Dumas, P., Kaldunski, M. L., Tonellato, P. J., Greene, A. S., Roman, R. J., Cheng, Q., Wang, Z., Jacob, H. J., & Cowley, A. W. (2003) · Physiological Genomics · doi:10.1152/physiolgenomics.00105.2003 · Cited by 98
Ueber Blutdruckmessung beim Menschen
von Recklinghausen, H. (1901) · Naunyn-Schmiedeberg s Archives of Pharmacology · doi:10.1007/bf01977772 · Cited by 97
The use of radiotelemetry in small laboratory animals: recent advances.
Kramer, K., Kinter, L. B., Brockway, B. P., Voss, H., Remie, R., & van Zutphen, B. F. M. (2001) · PubMed · Cited by 96
The mercury sphygmomanometer should be abandoned before it is proscribed
Markandu, N., Whitcher, F., Arnold, A., & Carney, C. (2000) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1000932 · Cited by 96
An assessment of the Dinamap 845
Hutton, P., Dye, J., & Roberts, C. P. (1984) · Anaesthesia · doi:10.1111/j.1365-2044.1984.tb07240.x · Cited by 95
Diurnal blood pressure variation and dietary salt in spontaneously hypertensive rats.
Calhoun, D. A., Zhu, S., Wyss, J. M., & Oparil, S. (1994) · Hypertension · doi:10.1161/01.hyp.24.1.1 · Cited by 94
Churchill, P. C., Churchill, M. C., BIDANI, A. K., Griffin, K. A., Picken, M. M., Pravenec, M., Křen, V., Lezin, E. S., Wang, J. M., Wang, N., et al. (1997) · Journal of Clinical Investigation · doi:10.1172/jci119657 · Cited by 93
Self-citationHata, Y., Kimura, Y., Muratani, H., Fukiyama, K., Kawano, Y., Ashida, T., Yokouchi, M., Imai, Y., Ozawa, T., Fujii, J., et al. (2000) · Hypertension Research · doi:10.1291/hypres.23.553 · Cited by 92
Long-term measurement of arterial blood pressure in conscious mice
Mattson, D. L. (1998) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.1998.274.2.r564 · Cited by 92
Töyry, J. P., NISKANEN, L. K., Länsimies, E., Partanen, K. P. L., & Uusitupa, M. I. J. (1996) · Stroke · doi:10.1161/01.str.27.8.1316 · Cited by 90
BIDANI, A. K., Griffin, K. A., Bakris, G. L., & Picken, M. M. (2000) · Kidney International · doi:10.1046/j.1523-1755.2000.00009.x · Cited by 89
(1996) · Journal of the Neurological Sciences · doi:10.1016/s0022-510x(96)00206-7 · Cited by 89
Alanore, A. L. B., Châtellier, G., Bobrie, G., Fofol, I., & Plouin, P. (2000) · Journal of Hypertension · doi:10.1097/00004872-200018040-00006 · Cited by 88
Reproducibility of increased blood pressure during an emergency department or urgent care visit
Backer, H., Decker, L., & Ackerson, L. (2003) · Annals of Emergency Medicine · doi:10.1067/mem.2003.151 · Cited by 88
Optimizing cuff width for noninvasive measurement of blood pressure
Marks, L. A., & Groch, A. J. (2000) · Blood Pressure Monitoring · doi:10.1097/00126097-200006000-00002 · Cited by 85
From theory into practice: arterial haemodynamics in clinical hypertension
O’Rourke, M. F. (2002) · Journal of Hypertension · doi:10.1097/00004872-200210000-00002 · Cited by 85
Mechanisms of escape from sodium retention during angiotensin II hypertension
Hall, J. E., Granger, J. P., Hester, R. L., Coleman, T. G., Smith, M. J., & Cross, R. B. (1984) · American Journal of Physiology-Renal Physiology · doi:10.1152/ajprenal.1984.246.5.f627 · Cited by 84
Self-citationLong-term blood pressure telemetry in AT2 receptor-disrupted mice
Groß, V., Milia, A. F., Plehm, R., Inagami, T., & Luft, F. C. (2000) · Journal of Hypertension · doi:10.1097/00004872-200018070-00018 · Cited by 83
Jordan, J., & Biaggioni, I. (2002) · Journal of Clinical Hypertension · doi:10.1111/j.1524-6175.2001.00516.x · Cited by 82
Hata, Y., Muratani, H., Kimura, Y., Fukiyama, K., Kawano, Y., Ashida, T. S., Yokouchi, M., Imai, Y., Ozawa, T., Fujii, J., et al. (2002) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1001301 · Cited by 81
Spontaneous blood pressure fluctuations and renal blood flow dynamics
Cupples, W. A., Novák, P., Novak, V., & Salevsky, F. C. (1996) · American Journal of Physiology-Renal Physiology · doi:10.1152/ajprenal.1996.270.1.f82 · Cited by 80
“Step” vs. “dynamic” autoregulation: implications for susceptibility to hypertensive injury
BIDANI, A. K., Hacıoğlu, R., Abu‐Amarah, I., Williamson, G. A., Loutzenhiser, R., & Griffin, K. A. (2003) · American Journal of Physiology-Renal Physiology · doi:10.1152/ajprenal.00012.2003 · Cited by 77
Influence of body and arm position on blood pressure readings
Netea, R. T., Lenders, J. W. M., Smits, P., & Thien, T. A. (2003) · Journal of Hypertension · doi:10.1097/00004872-200302000-00006 · Cited by 77
Association of the Auscultatory Gap with Vascular Disease in Hypertensive Patients
Cavallini, M. C., Roman, M. J., Blank, S. G., Pini, R., Pickering, T. G., & Devereux, R. B. (1996) · Annals of Internal Medicine · doi:10.7326/0003-4819-124-10-199605150-00003 · Cited by 76
Self-citationLimitations of current validation protocols for home blood pressure monitors for individual patients
Gerin, W., Schwartz, A. R., Schwartz, J. E., Pickering, T. G., Davidson, K. W., Bress, J., O'Brien, E. T., & Atkins, N. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200212000-00004 · Cited by 76
Self-citationEffect of back support and stethoscope head on seated blood pressure determinations.
Cushman, W. C., Cooper, K., Horne, R. A., & Meydrech, E. F. (1990) · American Journal of Hypertension · doi:10.1093/ajh/3.3.240 · Cited by 75
Turkkan, J. S. (1990) · Journal of Medical Primatology · doi:10.1111/j.1600-0684.1990.tb00452.x · Cited by 75
Overton, J. M., Williams, T. D., Chambers, J. B., & Rashotte, M. E. (2001) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.2001.280.4.r1007 · Cited by 72
The prevalence and clinical significance of nocturnal hypertension in pregnancy
Brown, M. A., Davis, G. K., & McHugh, L. A. (2001) · Journal of Hypertension · doi:10.1097/00004872-200108000-00012 · Cited by 71
Epidemiological Perspective of Body Position and Arm Level in Blood Pressure Measurement
Terént, A., & Breig-åsberg, E. (1994) · Blood Pressure · doi:10.3109/08037059409102246 · Cited by 70
Optimal size of cuff bladder for indirect measurement of arterial pressure in adults
Russell, A. E., Wing, L. M. H., Smith, S. A., Aylward, P. E., McRitchie, R. J., HASSAM, R. M., West, M., & Chalmers, J. P. (1989) · Journal of Hypertension · doi:10.1097/00004872-198908000-00002 · Cited by 68
Comparison of Direct and Indirect Arterial Pressure Measurements in Hospitalised Patients**
Breit, S., & OʼRourke, M. F. (1974) · Australian and New Zealand Journal of Medicine · doi:10.1111/j.1445-5994.1974.tb03222.x · Cited by 68
Telemetry for Cardiovascular Monitoring in a Pharmacological Study
Anderson, N., Devlin, A. M., Graham, D., Morton, J. J., Hamilton, C., Reid, J. L., Schork, N. J., & Dominiczak, A. F. (1999) · Hypertension · doi:10.1161/01.hyp.33.1.248 · Cited by 68
Spurious Hypertension in the Obese Patient
Linfors, E. W. (1984) · Archives of Internal Medicine · doi:10.1001/archinte.144.7.1482 · Cited by 67
Patterns of blood pressure variability in normotensive and hypertensive rats
Holstein‐Rathlou, N., He, J., Wagner, A. J., & Marsh, D. J. (1995) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.1995.269.5.r1230 · Cited by 66
Stewart, M. J., Gough, K., & Padfield, P. L. (1995) · Journal of Hypertension · doi:10.1097/00004872-199503000-00004 · Cited by 63
Facts and Fallacies About Measuring Blood Pressure in Rats
Buñnag, R. D. (1983) · Clinical and Experimental Hypertension · doi:10.3109/10641968309051801 · Cited by 63
The effect of crossing legs on blood pressure: a randomized single-blind cross-over study.
Peters, G. L., Binder, S. K., & Campbell, N. R. C. (1999) · Blood Pressure Monitoring · doi:10.1097/00126097-199900420-00007 · Cited by 62
Lezin, E. S., Liu, W., Wang, J., Wang, N., Křen, V., Krenová, D., Musilová, A., Zdobinska, M., Zı́dek, V., Lau, D. C., et al. (1997) · Hypertension · doi:10.1161/01.hyp.30.4.854 · Cited by 62
Evaluation of an electronic blood pressure device for epidemiological studies.
Cooper, R., Puras, Tracy, J. E., Kaufman, J. S., Asuzu, M. C., Ordúñez, P., Mufunda, J., & Sparks (1997) · PubMed · Cited by 61
Accuracy of aneroid sphygmomanometers in clinical usage: University of Michigan experience
Yarows, S. A., & Qian, K. (2001) · Blood Pressure Monitoring · doi:10.1097/00126097-200104000-00006 · Cited by 61
Neus, H., dderz, W. G., Otten, H., ddel, H. R., & Eiff, A. W. V. (1985) · Journal of Hypertension · doi:10.1097/00004872-198502000-00006 · Cited by 59
Exercising Restraint in Measuring Blood Pressure in Conscious Mice
Groß, V., & Luft, F. C. (2003) · Hypertension · doi:10.1161/01.hyp.0000060866.69947.d1 · Cited by 59
Wingfield, D., Cooke, J., Thijs, L., Staessen, J. A., FLETCHER, A. E., Fagard, R. H., & Bulpitt, C. J. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200206000-00005 · Cited by 58
Effects of home blood pressure measurement on long-term BP control.
Stahl, S. M., Kelley, C. R., Neill, P. J., Grim, C. E., & Mamlin, J. J. (1984) · American Journal of Public Health · doi:10.2105/ajph.74.7.704 · Cited by 58
Measuring blood pressure in the elderly: does atrial fibrillation increase observer variability?
Sykes, D., Dewar, R., Mohanaruban, K., Donovan, K. S., Nicklason, F., Thomas, D. M., & Fisher, D. L. (1990) · BMJ · doi:10.1136/bmj.300.6718.162 · Cited by 58
Epidemiologic Assessment of the Role of Blood Pressure in Stroke
Kannel, W. B. (1996) · JAMA · doi:10.1001/jama.1996.03540150071040 · Cited by 58
Griffin, K. A., Picken, M. M., & BIDANI, A. K. (2003) · Kidney International · doi:10.1111/j.1523-1755.2004.00356.x · Cited by 57
Isolated Diastolic Hypertension
Pickering, T. G. (2003) · Journal of Clinical Hypertension · doi:10.1111/j.1524-6175.2003.02840.x · Cited by 55
Self-citationGraves, J. W., Tibor, M., Murtagh, B., Klein, L. K., & Sheps, S. G. (2004) · Blood Pressure Monitoring · doi:10.1097/00126097-200402000-00004 · Cited by 55
Self-citationHemodynamic monitoring for 24 h in unanesthetized rats
Smith, T. L., Coleman, T. G., Stanek, K. A., & Murphy, W. R. C. (1987) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1987.253.6.h1335 · Cited by 54
Selective recording in blood pressure readings may increase subsequent mortality
Wingfield, D. (2002) · QJM · doi:10.1093/qjmed/95.9.571 · Cited by 52
van Ittersum, F. J., Wijering, R. M. J., Lambert, J., Donker, A. J. M., & Stehouwer, C. D. A. (1998) · Journal of Hypertension · doi:10.1097/00004872-199816080-00007 · Cited by 52
Measurement of blood pressure—a routine test in need of standardization
Thulin, T., Andersson, G., & Scherstén, B. (1975) · Postgraduate Medical Journal · doi:10.1136/pgmj.51.596.390 · Cited by 51
Comparative Effects of Selective T- and L-Type Calcium Channel Blockers in the Remnant Kidney Model
Griffin, K. A., Picken, M. M., Bakris, G. L., & BIDANI, A. K. (2001) · Hypertension · doi:10.1161/01.hyp.37.5.1268 · Cited by 50
Accuracy of auscultatory blood pressure measurements in hypertensive and obese subjects.
Nielsen, P. E., Larsen, B., Holstein, P., & Poulsen, H. (1983) · Hypertension · doi:10.1161/01.hyp.5.1.122 · Cited by 50
Lezin, E. S., Griffin, K. A., Picken, M. M., et al. (1999) · Hypertension · doi:10.1161/01.hyp.34.2.187 · Cited by 50
Arm position is important for blood pressure measurement
Netea, R., Lenders, J., Smits, P., & Thien, T. B. (1999) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1000720 · Cited by 49
Noninvasive blood pressure measurement in mice using pulsed Doppler ultrasound
Reddy, A. K., Taffet, G. E., Madala, S., Michael, L. H., Entman, M. L., & Hartley, C. J. (2003) · Ultrasound in Medicine & Biology · doi:10.1016/s0301-5629(02)00746-9 · Cited by 49
Genetically determined chloride-sensitive hypertension and stroke
Tanaka, M., Schmidlin, O., Yi, S.-L., Bollen, A. W., & Morris, R. C. (1997) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.94.26.14748 · Cited by 47
Evaluation of a community-based automated blood pressure measuring device.
Lewis, J. E., Boyle, E., Magharious, L., & Myers, M. G. (2002) · PubMed Central · Cited by 46
Prehospital blood pressures: Inaccuracies caused by ambulance noise?
Prasad, N. H., Brown, L. H., Ausband, S. C., Cooper-Spruill, O., Carroll, R. G., & Whitley, T. W. (1994) · The American Journal of Emergency Medicine · doi:10.1016/0735-6757(94)90025-6 · Cited by 46
Belmin, J., Visintin, J.-M., Salvatore, R., Sebban, C., & Moulias, R. (1995) · The American Journal of Medicine · doi:10.1016/s0002-9343(99)80079-5 · Cited by 45
Development of skin interfacing cannula.
Hall, C. W., Adams, L. M., & Ghidoni, J. J. (1975) · PubMed · Cited by 45
Comparison of forearm and upper arm blood pressures
Singer, A. J., Kahn, S. R., Thode, H. C., & Hollander, J. E. (1999) · Prehospital Emergency Care · doi:10.1080/10903129908958919 · Cited by 44
Graves, J. W. (2001) · Blood Pressure Monitoring · doi:10.1097/00126097-200102000-00003 · Cited by 43
Self-citationSinaiko, A. R., Gómez-Marı́n, O., & Prineas, R. J. (1990) · American Journal of Epidemiology · doi:10.1093/oxfordjournals.aje.a115706 · Cited by 42
Accuracy of telemedical home blood pressure measurement in the diagnosis of hypertension
Møller, D. S., Dideriksen, A. K., Sørensen, S., Madsen, L. D., & Pedersen, E. B. (2003) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1001584 · Cited by 42
Kurtz, T. W. (2003) · Hypertension · doi:10.1161/01.hyp.0000049882.23078.eb · Cited by 42
Self-citationReinhart, G. A., Preusser, L. C., Burke, S. E., et al. (2002) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00346.2002 · Cited by 41
Graves, J., Bailey, K., & Sheps, S. (2003) · Blood Pressure Monitoring · doi:10.1097/00126097-200312000-00001 · Cited by 38
Self-citationRenal Blood Flow Dynamics and Arterial Pressure Lability in the Conscious Rat
Pires, S. L. S., Barrès, C., Sassard, J., & Julien, C. (2001) · Hypertension · doi:10.1161/01.hyp.38.1.147 · Cited by 38
Ambulatory Blood Pressure Monitoring in Atrial Fibrillation
Lip, G. Y. H., Zarifis, J., Beevers, M., & Beevers, D. G. (1996) · The American Journal of Cardiology · doi:10.1016/s0002-9149(96)00293-7 · Cited by 37
Prevention of catheter-related infections by a new, catheter-restricted antibiotic filling technique
Palm, Ü., Boemke, W., Bayerl, D., Schnoy, N., Nc, J., & Reinhardt, H. W. (1991) · Laboratory Animals · doi:10.1258/002367791781082469 · Cited by 35
Autonomic control of blood pressure and heart rate in obese hypertensive dogs
Truett, A. A., Borne, A. T., Poincot, M. A., & West, D. B. (1996) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.1996.270.3.r541 · Cited by 35
Wright, J. C., & Looney, S. W. (1997) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1000431 · Cited by 34
Low-Dose Angiotensin II Enhances Pressor Responses Without Causing Sustained Hypertension
Pelaez, L. I., Manriquez, M. C., Nath, K. A., Romero, J. C., & Juncos, L. A. (2003) · Hypertension · doi:10.1161/01.hyp.0000085782.99773.b6 · Cited by 34
Systematic difference between blood pressure readings caused by cuff type.
Bovet, P., Hungerbühler, P., Quilindo, J., Grettve, M. L., Waeber, B., & Burnand, B. (1994) · Hypertension · doi:10.1161/01.hyp.24.6.786 · Cited by 34
Blood pressure monitoring during pregnancy
Feldman, D. M. (2001) · Blood Pressure Monitoring · doi:10.1097/00126097-200102000-00001 · Cited by 34
State of the market for devices for blood pressure measurement
O'Brien, E. T. (2001) · Blood Pressure Monitoring · doi:10.1097/00126097-200112000-00003 · Cited by 33
Measurement of systolic blood pressure using pulse oximetry during helicopter flight
Talke, P. O. (1991) · Critical Care Medicine · doi:10.1097/00003246-199107000-00018 · Cited by 33
Advantages of continuous measurement of cardiac output 24 h a day
Montani, J. P., Mizelle, H. L., Vliet, B. N. V., & Adair, T. H. (1995) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1995.269.2.h696 · Cited by 32
Influence of rate of cuff inflation and deflation on observed blood pressure by sphygmomanometry
King, G. E. (1963) · American Heart Journal · doi:10.1016/0002-8703(63)90003-6 · Cited by 30
Genetics of experimental hypertension
Dominiczak, A. F., Clark, J. S., Jeffs, B., et al. (1998) · Journal of Hypertension · doi:10.1097/00004872-199816121-00003 · Cited by 30
Jn, A., & Dh, S. (2000) · PubMed · Cited by 30
Shimamura, T., Nakajima, M., Iwasaki, T., Hayasaki‐Kajiwara, Y., Yonetani, Y., & Iwaki, K. (1999) · Journal of Hypertension · doi:10.1097/00004872-199917020-00005 · Cited by 29
Twenty-four-hour ambulatory blood pressure monitoring in atrial fibrillation
Olsen, R. B., Amlie, A., & Omvik, P. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200206000-00002 · Cited by 29
Jamieson, M. (1997) · American Journal of Hypertension · doi:10.1016/s0895-7061(96)00321-4 · Cited by 28
Skip Patterns in DINAMAP-Measured Blood Pressure in 3 Epidemiological Studies
Rose, K. M., Arnett, D. K., Ellison, R. C., & HEISS, G. M. (2000) · Hypertension · doi:10.1161/01.hyp.35.5.1032 · Cited by 28
Identification and evaluation of noninvasive blood pressure measuring techniques
HASSLER, C. R., Lutz, G. A., Linebaugh, R., & Cummings, K. D. (1979) · Toxicology and Applied Pharmacology · doi:10.1016/0041-008x(79)90312-0 · Cited by 27
Altunkan, Ş., Yildiz, S., & Azer, S. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200210000-00006 · Cited by 27
Research works printed but not linked: 35
- Verdecchia P, Schillaci G, Borgioni C, Ciucci A, Gattobigio R, Porcellati C. Nocturnal pressure is the true pressure. Blood Press Monit. 1996; 1 (Suppl 2): S81–S85.
- 10.1161/circ.103.9.1245 · DOI
- 10.1161/hyp.36.5.801 · DOI
- 10.1161/hyp.25.6.1276 · DOI
- 10.1161/circ.100.4.354 · DOI
- 10.1161/circ.103.9.1245 · DOI
- Pickering TG. White coat hypertension. In: Izzo JL Black HR eds. Hypertension Primer third edition. Dallas: American Heart Association; 2003.
- 10.1161/circ.102.10.1139 · DOI
- Low PA. Clinical Autonomic Disorders: Evaluation and Management second ed. Philadelphia: Lippincott-Raven; 1997.
- 10.1161/str.27.8.1316 · DOI
- 10.1161/circ.8.2.232 · DOI
- 10.1161/circ.88.5.8222141 · DOI
- Cates EM, Schlussel YR, James GD, Pickering TG. A validation study of the Spacelabs 90207 ambulatory blood pressure monitor. J Ambul Monitor. 1990; 3: 149–154.
- Penaz J. Photo-electric measurement of blood pressure volume and flow in the finger. Digest Tenth International Conference Medical Biological Engineering. Dresden; 1973:104.
- Ware RW, Laenger CJ. Indirect blood pressure measurement by Doppler ultrasonic kinetoarteriography. Proc 20th Annual Conference Engineering Medical Biology. 1967; 9: 27–30.
- Wonka F, Thummler M, Schoppe A. Clinical test of a blood pressure measurement device with a wrist cuff. Blood Press Monit. 1996; 1: 361–366.
- 10.1161/hyp.36.4.484 · DOI
- Grim CE, Grim CM, Li J. Entering medical students who say they have been trained to take blood pressure do not follow American Heart Association guidelines. Am J Hypertens. 1999; 12.
- 10.1001/archinte.1984.00350190188032 · DOI
- Beevers DG, Beevers M, Zezulka AV, Hornby R, Herod J. Standards for blood pressure measuring devices. BMJ (Clin Res Ed). 1987; 294: 1614.
- O’Brien E, Atkins N, O’Malley K. Selecting the correct bladder according to distribution of arm circumference in the population (abstr). Hypertension. 1993; 11: 1149–1150.
- 10.1161/circ.68.4.6616774 · DOI
- Ljungvall P, Thorvinger B, Thulin T. The influence of a heart level pillow on the result of blood pressure measurement. J Hum Hypertens. 1989; 3: 471–474.
- Rose G. Standardisation of observers in blood pressure measurement. Lancet. 1965; 10: 673–674.
- 10.1161/circ.28.1.20 · DOI
- 10.1001/jama.1967.03130150084015 · DOI
- 10.1161/hyp.35.5.1032 · DOI
- Myers MG. Self-measurement of blood pressure at home: the potential for reporting bias. Blood Press Monit. 1998; 3 (suppl 1): S19–S22.
- Imai Y, Ohkubo T, Tsuji I, Nagai K, Satoh H, Hisamichi S, Abe K. Prognostic value of ambulatory and home blood pressure measurements in comparison to screening blood pressure measurements: a pilot study in Ohasama. Blood Press Monit. 1996; 1 (Suppl 2): S51–S58.
- 10.1161/hyp.32.2.255 · DOI
- Verdecchia P Bentivoglia M Providenza M Savino K Corea L. Reliability of home self-recorded arterial pressure in essential hypertension in relation to the stage of the disease. In: Germano G ed. Blood pressure recording in the clinical management of hypertension. Rome: Ediziono Pozzi; 1985: 40–42.
- 10.1161/circ.98.18.1892 · DOI
- 10.1161/hyp.31.2.712 · DOI
- Prineas RJ Elkwiry ZM. Epidemiology and measurement of high blood pressure in children and adolescents. In: Loggie JMH ed. Pediatric and Adolescent Hypertension. Boston MA: Blackwell Scientific Publications; 1992: 91–103.
- Brown MA Davis GK. Hypertension in pregnancy. In: Mancia G Chalmers J Julius S Saruta T Weber MA Ferrari AU Wilkinson IB ed. Manual of Hypertension. London: Harcourt Publishers Limited; 2002: 579–597.
Cited by
Sorted by citations
Measurement of Blood Pressure in Humans: A Scientific Statement From the American Heart Association
Muntner, P., Shimbo, D., Carey, R. M., Charleston, J. B., Gaillard, T., Misra, S., Myers, M. G., Ogedegbe, G., Schwartz, J. E., Townsend, R. R., et al. (2019) · Hypertension · doi:10.1161/hyp.0000000000000087 · Cited by 1,512
European Society of hypertension position paper on ambulatory blood pressure monitoring
O’Brien, E., Parati, G., Stergiou, G., Asmar, R., Beilin, L., Bilo, G., Clement, D., de la Sierra, A., de Leeuw, P., Dolan, E., et al. (2013) · Journal of Hypertension · doi:10.1097/hjh.0b013e328363e964 · Cited by 1,419
Self-citationAstrup, A., Rössner, S., Van Gaal, L., Rissanen, A., Niskanen, L., Al Hakim, M., Madsen, J., Rasmussen, M. F., & Lean, M. E. (2009) · The Lancet · doi:10.1016/s0140-6736(09)61375-1 · Cited by 1,163
Addressing Current Criticism Regarding the Value of Self-Report Dietary Data
Subar, A. F., Freedman, L. S., Tooze, J. A., Kirkpatrick, S. I., Boushey, C., Neuhouser, M. L., Thompson, F. E., Potischman, N., Guenther, P. M., Tarasuk, V., et al. (2015) · The Journal of Nutrition · doi:10.3945/jn.115.219634 · Cited by 1,047
Assessment and management of blood-pressure variability
Parati, G., Ochoa, J. E., Lombardi, C., & Bilo, G. (2013) · Nature Reviews Cardiology · doi:10.1038/nrcardio.2013.1 · Cited by 853
Metadata
- Authors
- Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J.
- ORCID
- Venue
- Hypertension
- Published
- 2005-01
- Volume
- 45
- Issue
- 1
- Pages
- 142-161
- ISSN
- 0194-911X (ISSN-L)
- 1524-4563
- OpenAlex
- W2160801689
- Semantic Scholar
- cc608ac67b0966145e814f721be04e1b0f95eeb3
- Cited by
- 2,051
- References
- 226
- Citations by year
- 2026: 34 · 2025: 43 · 2024: 48 · 2023: 61 · 2022: 79 · 2021: 77 · 2020: 101 · 2019: 85 · 2018: 113 · 2017: 144 · 2016: 166 · 2015: 161 · 2014: 104 · 2013: 114 · 2012: 133
- FWCI
- 39.42
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Hypertension (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- blood pressure measurement, auscultatory method, Korotkoff sounds, mercury sphygmomanometer, aneroid sphygmomanometer, oscillometric method, home blood pressure monitoring, ambulatory blood pressure monitoring, white coat hypertension, cuff size, observer training, pregnant women, laboratory animals, hypertension risk prediction
- MeSH terms
- Adult, Aged, Animals, Auscultation, Automation, Blood Pressure Determination, Calibration, Cardiovascular Diseases, Child, Circadian Rhythm, Clinical Competence, Diagnostic Errors, Equipment Design, Female, Humans, Hypertension, Male, Middle Aged, Obesity, Organ Specificity, Oscillometry, Posture, Predictive Value of Tests, Pregnancy, Self Care, Stress, Psychological, Blood Pressure Monitors, Blood Pressure Monitoring, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review