Brachial and Radial Systolic Blood Pressure Are Not the Same: Evidence to Support the Popeye Phenomenon
Abstract
Radial intra-arterial blood pressure (BP) is sometimes used as the reference standard for validation of brachial cuff BP devices. Moreover, there is an emerging wearables market seeking to measure BP at the wrist. However, radial systolic BP may differ when compared with brachial; yet some authors have labeled these differences as a fictional Popeye phenomenon. Indeed, differences between brachial and radial systolic BP have never been accurately quantified, and this was the aim of this study. Intra-arterial BP was measured consecutively at the brachial and radial artery in 180 participants undergoing coronary angiography (aged 61±10 years; 69% men). On average, radial systolic BP was 5.5 mm Hg higher than brachial systolic BP. Only 43% of participants had radial systolic BP within ±5 mm Hg of brachial. Additionally, 46%, 19%, and 13% of participants had radial systolic BP >5, between 5 and 10, and between 10 and 15 mm Hg higher than brachial respectively. A further 14% of participants had radial systolic BP >15 mm Hg higher than brachial, representing the so-called Popeye phenomenon. Finally, 11% of participants had radial systolic BP >5 mm Hg lower than brachial. In conclusion, radial systolic BP is not representative of brachial systolic BP, with most participants having a radial systolic BP >5 mm Hg higher than brachial and many with differences >15 mm Hg. Therefore, validation testing of BP devices utilizing intra-arterial BP as the reference standard should use intra-arterial BP measured at the same site as the brachial cuff or wearable device.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Radial artery pressure is 5 to 15 mmHg higher than brachial artery pressure.
“Radial artery pressure has been shown to be 5 to 15 mmHg50 higher than the brachial artery, where blood pressure is often measured using the sphygmomanometer.”
From Synthetic photoplethysmography (PPG) of the radial artery through parallelized Monte Carlo and its correlation to body mass index (BMI) (2021) · cites this paper as [50]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Forouzanfar, M. H., Alexander, L., Anderson, H. R., Bachman, V. F., Biryukov, S., Brauer, M., Burnett, R., Casey, D., Coates, M. M., Cohen, A., et al. (2015) · The Lancet · doi:10.1016/s0140-6736(15)00128-2 · Cited by 2,701 · see citing works
Pauca, A. L., O’Rourke, M. F., & Kon, N. D. (2001) · Hypertension · doi:10.1161/hy1001.096106 · Cited by 1,179 · see citing works
McDONALD'S BLOOD FLOW IN ARTERIES: THEORETICAL, EXPERIMENTAL AND CLINICAL PRINCIPLES
Czuwala, P. J. (1991) · Echocardiography · doi:10.1111/j.1540-8175.1991.tb01207.x · Cited by 829 · see citing works
Direct Blood Pressure Measurement —Dynamic Response Requirements
Gardner, R. M. (1981) · Anesthesiology · doi:10.1097/00000542-198103000-00010 · Cited by 522 · see citing works
Role of Pulse Pressure Amplification in Arterial Hypertension
Avolio, A. P., Bortel, L. M. V., Boutouyrie, P., Cockcroft, J. R., McEniery, C. M., Protogerou, A. D., Roman, M. J., Safar, M. E., Segers, P., & Smulyan, H. (2009) · Hypertension · doi:10.1161/hypertensionaha.109.134379 · Cited by 517 · see citing works
Kroeker, E. J., & Wood, E. H. (1955) · Circulation Research · doi:10.1161/01.res.3.6.623 · Cited by 425 · see citing works
Accuracy of Cuff-Measured Blood Pressure
Picone, D. S., Schultz, M. G., Otahal, P., Aakhus, S., Al-Jumaily, A. M., Black, J. A., Bos, W. J., Chambers, J. B., Chen, C.-H., Cheng, H.-M., et al. (2017) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2017.05.064 · Cited by 277 · see citing works
Self-citationRowell, L. B., Brengelmann, G. L., Blackmon, J. R., Bruge, R. A., & Murray, J. A. (1968) · Circulation · doi:10.1161/01.cir.37.6.954 · Cited by 257 · see citing works
Sharman, J. E., Avolio, A. P., Baulmann, J., Benetos, A., Blacher, J., Blizzard, C. L., Boutouyrie, P., Chen, C.-H., Chowienczyk, P., Cockcroft, J. R., et al. (2017) · European Heart Journal · doi:10.1093/eurheartj/ehw632 · Cited by 231 · see citing works
Self-citationAmplification of the Pressure Pulse in the Upper Limb in Healthy, Middle-Aged Men and Women
Segers, P., Mahieu, D., Kips, J. G., Rietzschel, E. R., Buyzere, M. D., Bacquer, D. D., Bekaert, S., Backer, G. D., Gillebert, T. C., Verdonck, P., et al. (2009) · Hypertension · doi:10.1161/hypertensionaha.109.133009 · Cited by 187 · see citing works
Noninvasive Assessment of Local Pulse Pressure
Verbeke, F., Segers, P., Heireman, S., Vanholder, R., Verdonck, P., & Bortel, L. M. V. (2005) · Hypertension · doi:10.1161/01.hyp.0000166723.07809.7e · Cited by 169 · see citing works
FDA Regulation of Mobile Medical Apps
Shuren, J. E., Patel, B., & Gottlieb, S. L. (2018) · JAMA · doi:10.1001/jama.2018.8832 · Cited by 155 · see citing works
Bilo, G., Zorzi, C., Ochoa Munera, J. E., Torlasco, C., Giuli, V., & Parati, G. (2015) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000124 · Cited by 143 · see citing works
Lakhal, K., Ehrmann, S., Martin, M., Faiz, S., Réminiac, F., Cinotti, R., Capdevila, X., Asehnoune, K., Blanloeil, Y., Rozec, B. R., et al. (2015) · British Journal of Anaesthesia · doi:10.1093/bja/aev304 · Cited by 143 · see citing works
Formation of Peripheral Pulse Contour in Man
Remington, J. W., & Wood, E. H. (1956) · Journal of Applied Physiology · doi:10.1152/jappl.1956.9.3.433 · Cited by 130 · see citing works
Nair, D., Tan, S.-Y., Gan, H. ‐., Lim, S.-F., Tan, J., Zhu, M., Gao, H., Chua, N.-H., Peh, W. L., & Mak, K.-H. (2007) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1002306 · Cited by 105 · see citing works
Blood Pressure Measurement: Retrospective and Prospective Views
Smulyan, H., & Safar, M. E. (2011) · American Journal of Hypertension · doi:10.1038/ajh.2011.22 · Cited by 101 · see citing works
New Approaches to Evaluating and Monitoring Blood Pressure
Goldberg, E. M., & Levy, P. D. (2016) · Current Hypertension Reports · doi:10.1007/s11906-016-0650-9 · Cited by 90 · see citing works
Ribezzo, S., Spina, E., Bartolomeo, S. D., & Sanson, G. (2014) · The Scientific World JOURNAL · doi:10.1155/2014/353628 · Cited by 82 · see citing works
Impact of Atrial Fibrillation on the Accuracy of Oscillometric Blood Pressure Monitoring
Pagonas, N., Schmidt, S. C., Eysel, J., Compton, F., Hoffmann, C., Seibert, F. S., Hilpert, J. W., Tschöpe, C., Zidek, W., & Westhoff, T. H. (2013) · Hypertension · doi:10.1161/hypertensionaha.113.01426 · Cited by 82 · see citing works
Kroeker, E. J., & Wood, E. H. (1956) · Journal of Applied Physiology · doi:10.1152/jappl.1956.8.5.483 · Cited by 62 · see citing works
Brachial artery tonometry and the Popeye phenomenon
Adji, A. I., & O’Rourke, M. F. (2012) · Journal of Hypertension · doi:10.1097/hjh.0b013e328354e859 · Cited by 49 · see citing works
Picone, D. S., Schultz, M. G., Peng, X., Black, J. A., Dwyer, N., Roberts‐Thomson, P., Chen, C., Cheng, H., Pucci, G., Wang, J., et al. (2018) · Hypertension · doi:10.1161/hypertensionaha.117.10696 · Cited by 47 · see citing works
Self-citationBrachial-to-radial SBP amplification
Picone, D. S., Climie, R. E., Ahuja, K., Keske, M. A., & Sharman, J. E. (2015) · Journal of Hypertension · doi:10.1097/hjh.0000000000000637 · Cited by 39 · see citing works
Self-citationBlood pressure assessment in severe obesity: Validation of a forearm approach
Leblanc, M., Croteau, S., Ferland, A., Bussières, J. S., Cloutier, L., Hould, F., Biertho, L. D., Moustarah, F., Marceau, S., & Poirier, P. (2013) · Obesity · doi:10.1002/oby.20458 · Cited by 29 · see citing works
Davies, J. E., Shanmuganathan, M., Francis, D. P., et al. (2009) · Hypertension · doi:10.1161/hypertensionaha.109.146498 · Cited by 27 · see citing works
Peng, X., Schultz, M. G., Picone, D. S., Black, J. A., Dwyer, N., Roberts‐Thomson, P., Davies, J. E., & Sharman, J. E. (2017) · Journal of Hypertension · doi:10.1097/hjh.0000000000001400 · Cited by 26 · see citing works
Self-citationBiases in the measurement of arterial pressure
Keith, J. C., FINNIE, Watts, D. G., et al. (1984) · Critical Care Medicine · doi:10.1097/00003246-198411000-00009 · Cited by 23 · see citing works
Flawed Measurement of Brachial Tonometry for Calculating Aortic Pressure?
O’Rourke, M. F., & Takazawa, K. (2009) · Hypertension · doi:10.1161/hypertensionaha.109.139915 · Cited by 20 · see citing works
Ganio, M. S., Brothers, R. M., Lucas, R. A. I., Hastings, J. L., & Crandall, C. G. (2011) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.00247.2011 · Cited by 19 · see citing works
The accuracy of rapid oscillometric blood pressure determination.
Gorback, M. S., Quill, T. J., & Da, G. (1990) · PubMed · Cited by 17 · see citing works
A comparison of intra-arterial and auscultatory blood pressure readings.
Chyun, D. (1985) · PubMed · Cited by 15 · see citing works
Brachial-to-radial systolic blood pressure amplification in patients with type 2 diabetes mellitus
Climie, R. E., Picone, D. S., Keske, M. A., & Sharman, J. E. (2015) · Journal of Human Hypertension · doi:10.1038/jhh.2015.101 · Cited by 12 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Boonya-ananta, T., Rodriguez, A. J., Ajmal, A., Du Le, V. N., Hansen, A. K., Hutcheson, J. D., & Ramella-Roman, J. C. (2021) · Scientific Reports · doi:10.1038/s41598-021-82124-4 · Cited by 70 · see citing works
Measurements of arterial pressure and flow in vivo
Sharman, J. E., Segers, P., & Chirinos, J. A. (2022) · Textbook of Arterial Stiffness and Pulsatile Hemodynamics in Health and Disease · doi:10.1016/b978-0-323-91391-1.00002-9 · Cited by 3 · see citing works
Self-citation
- Authors
- Armstrong, M. K., Schultz, M. G., Picone, D. S., Black, J. A., Dwyer, N., Roberts-Thomson, P., & Sharman, J. E.
- Affiliations
- University of Tasmania (AU)
- Royal Hobart Hospital (AU)
- Department of Cardiology, Royal Hobart Hospital, Hobart, Australia (J.A.B., N.D., P.R.-T.).
- Corresponding author
- Sharman, James E.
- ORCID
- Venue
- Hypertension
- Published
- 2019-05
- Volume
- 73
- Issue
- 5
- Pages
- 1036-1041
- ISSN
- 0194-911X (ISSN-L)
- 1524-4563
- OpenAlex
- W2927070984
- Cited by
- 59
- References
- 34
- Citations by year
- 2026: 2 · 2025: 3 · 2024: 5 · 2023: 8 · 2022: 11 · 2021: 12 · 2020: 12 · 2019: 6
- FWCI
- 7.41
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Hypertension (publishedVersion)
- PubMed (publishedVersion)
- UTAS Research Repository (submittedVersion)
- Figshare (submittedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- brachial blood pressure, systolic blood pressure, intra-arterial blood pressure
- MeSH terms
- Blood Pressure, Blood Pressure Determination, Brachial Artery, Female, Humans, Hypertension, Male, Middle Aged, Radial Artery
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review