Prototype of a Nanostructured Sensing Contact Lens for Noninvasive Intraocular Pressure Monitoring
The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
PURPOSE: To present the application of a new sensor based on a flexible, highly piezoresistive, nanocomposite, all-organic bilayer (BL) adapted to a contact lens (CL) for non-invasive monitoring intraocular pressure (IOP). METHODS: A prototype of a sensing CL, adapted to a pig eyeball, was tested on different enucleated pig eyes. A rigid, gas-permeable CL was designed as a doughnut shape with a 3-mm hole, where the BL film-based sensor was incorporated. The sensor was a polycarbonate film coated with a polycrystalline layer of the highly piezoresistive molecular conductor β-(ET)₂I₃, which can detect deformations caused by pressure changes of 1 mm Hg. The pig eyeballs were subjected to controlled-pressure variations (low-pressure transducer) to register the electrical resistance response of the CL sensor to pressure changes. Similarly, a CL sensor was designed according to the anatomic characteristics of the eye of a volunteer on the research team. RESULTS: A good correlation (r² = 0.99) was demonstrated between the sensing CL electrical response, and IOP (mm Hg) changes in pig eyes, with a sensitivity of 0.4 Ω/mm Hg. A human eye test also showed the high potential of this new sensor (IOP variations caused by eye massage, blinking, and eye movements were registered). CONCLUSIONS: A new nanostructured sensing CL for continuous monitoring of IOP was validated in an in vitro model (porcine eyeball) and in a human eye. This prototype has adequate sensitivity to continuously monitor IOP. This device will be useful for glaucoma diagnosis and treatment.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Wireless contact lens sensor for intraocular pressure monitoring: assessment on enucleated pig eyes
Leonardi, M., Pitchon, E. M., Bertsch, A., Renaud, P., & Mermoud, A. (2009) · Acta Ophthalmologica · doi:10.1111/j.1755-3768.2008.01404.x · Cited by 300
The parameters of the porcine eyeball
Sanchez, I., Martin, R., Ussa, F., & Fernandez-Bueno, I. (2011) · Graefe's Archive for Clinical and Experimental Ophthalmology · doi:10.1007/s00417-011-1617-9 · Cited by 298
Self-citationFirst steps toward noninvasive intraocular pressure monitoring with a sensing contact lens
Leonardi, M., Leuenberger, P., Bertrand, D., Bertsch, A., & Renaud, P. (2004) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.04-0015 · Cited by 244
Cited by
Yetisen, A. K., Martinez‐Hurtado, J. L., Ünal, B., Khademhosseini, A., & Butt, H. (2018) · Advanced Materials · doi:10.1002/adma.201706910 · Cited by 569
Wearable Biofuel Cells: A Review
Bandodkar, A. J., & Wang, J. (2016) · Electroanalysis · doi:10.1002/elan.201600019 · Cited by 193
Capacitive contact lens sensor for continuous non-invasive intraocular pressure monitoring
Chen, G.-Z., Chan, I.-S., & Lam, D. C. C. (2013) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2013.08.029 · Cited by 133
Metadata
- Authors
- Sánchez, I., Laukhin, V., Moya, A., Martin, R., Ussa, F., Laukhina, E., Guimera, A., Villa, R., Rovira, C., Aguiló, J., Veciana, J., & Pastor, J. C.
- Venue
- Investigative Opthalmology & Visual Science
- Published
- 2011-10-21
- Volume
- 52
- Issue
- 11
- Pages
- 8310
- ISSN
- 0146-0404 (ISSN-L)
- 1552-5783
- OpenAlex
- W1966647970
- Cited by
- 48
- References
- 3
- FWCI
- 2.01
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Prototype of a Nanostructured Sensing Contact Lens for Noninvasive Intraocular Pressure Monitoring. https://app.cochat.ai/papers/10.1167/iovs.10-7064
- The paper
- Sánchez, I., Laukhin, V., Moya, A., Martin, R., Ussa, F., Laukhina, E., Guimera, A., Villa, R., Rovira, C., Aguiló, J., Veciana, J., & Pastor, J. C. (2011). Prototype of a Nanostructured Sensing Contact Lens for Noninvasive Intraocular Pressure Monitoring. Investigative Opthalmology & Visual Science. https://doi.org/10.1167/iovs.10-7064
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.