Modeling of Calibration Effectiveness and Blood-to-Interstitial Glucose Dynamics as Potential Confounders of the Accuracy of Continuous Glucose Sensors during Hyperinsulinemic Clamp

Open access
King, Christopher, Anderson, Stacey M., Breton, Marc, Clarke, William L., Kovatchev, Boris P. · 2007 · Journal of Diabetes Science and Technology
Cited by 66

The DOI matches its Crossref record.

Abstract

Background: Models of the dynamics of interstitial fluid-based continuous glucose sensors imply a variable sensor deviation from reference blood glucose (BG), depending on both sensor calibration procedure and BG dynamics. These effects could have a significant effect on the cross-interpretation of nonidentical accuracy studies. Methods: Hyperinsulinemic euglycemic and hypoglycemic clamps were performed on 39 subjects with type 1 diabetes wearing the Medtronic Continuous Glucose Monitoring System®. Sensor calibration and interstitial glucose (IG) dynamics were modeled and analyzed as potential confounders of sensor deviation from reference BG. Results: The mean absolute deviation (MAD) of sensor data was 20.9 mg/dl during euglycemia and 24.5 mg/dl during descent into and recovery from hypoglycemia. Computer-generated recalibration reduced MAD to 10.6 and 14.6 mg/dl, respectively. Modeling of IG dynamics reduced the MAD further to 10.0 and 10.4 mg/dl (using idiosyncratic parameters) or to 10.6 and 11.5 mg/dl (using model parameters common for all subjects), respectively. Conclusions: The sensor MAD from reference is strongly influenced by the choice of calibration points. Thus, cross-experiment comparisons of sensor accuracy are likely to be heavily dependent on the employed calibration procedures. Demanding calibration points substantially differing in value was found to improve calibration effectiveness. Simulation using existing IG models and population parameters reduced the bias resulting from BG—IG dynamics.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1177/193229680700100302

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 66

Metadata

Authors
King, C., Anderson, S. M., Breton, M., Clarke, W. L., & Kovatchev, B. P.
Venue
Journal of Diabetes Science and Technology
Published
2007-05
Volume
1
Issue
3
Pages
317-322
ISSN
1932-2968 (ISSN-L)
OpenAlex
W2163274210
Cited by
66
References
0
Type
conference abstract
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Modeling of Calibration Effectiveness and Blood-to-Interstitial Glucose Dynamics as Potential Confounders of the Accuracy of Continuous Glucose Sensors during Hyperinsulinemic Clamp. https://app.cochat.ai/papers/10.1177/193229680700100302
The paper
King, C., Anderson, S. M., Breton, M., Clarke, W. L., & Kovatchev, B. P. (2007). Modeling of Calibration Effectiveness and Blood-to-Interstitial Glucose Dynamics as Potential Confounders of the Accuracy of Continuous Glucose Sensors during Hyperinsulinemic Clamp. Journal of Diabetes Science and Technology. https://doi.org/10.1177/193229680700100302

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.