Clinical Implications of Accuracy Measurements of Continuous Glucose Sensors
Open accessAbstract
The accuracy of a continuous glucose monitor (CGM) now supports its use by persons with diabetes and clinicians caring for them. This article reviews measures of CGM accuracy, factors contributing to accuracy, comparative accuracy assessment, clinical implications of CGM sensor accuracy, and recent clinical trials that have demonstrated the utility of CGMs.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Minimally invasive continuous glucose monitors have been clinically validated but are difficult to embed in consumer-grade wearables and mostly function as a stand-alone product.
“One example of biochemical sensors are the minimally invasive continuous glucose monitors that have been clinically validated but are difficult to embed in consumer-grade wearables and mostly function as a stand-alone product31.”
From Smart wearable devices in cardiovascular care: Where we are and how to move forward (2021) · cites this paper as [31]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667 · see citing works
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435 · see citing works
Beck, R. W., Riddlesworth, T., Ruedy, K., Ahmann, A., Bergenstal, R., Haller, S., Kollman, C., Kruger, D., McGill, J. B., Polonsky, W., et al. (2017) · JAMA · doi:10.1001/jama.2016.19975 · Cited by 1,185 · cited 1× · see citing works
“The Diamond trial 32 demonstrated a significantly greater (0.6”
InfluentialBolinder, J., Antuna, R., Geelhoed-Duijvestijn, P., Kröger, J., & Weitgasser, R. (2016) · The Lancet · doi:10.1016/s0140-6736(16)31535-5 · Cited by 893 · see citing works
Lind, M., Polonsky, W., Hirsch, I. B., Heise, T., Bolinder, J., Dahlqvist, S., Schwarz, E., Ólafsdóttir, A. F., Frid, A., Wedel, H., et al. (2017) · JAMA · doi:10.1001/jama.2016.19976 · Cited by 705 · see citing works
The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
Bailey, T., Bode, B. W., Christiansen, M. P., Klaff, L. J., & Alva, S. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0378 · Cited by 685 · see citing works
Self-citationThreshold-based insulin-pump interruption for reduction of hypoglycemia
Bergenstal, R. M., Klonoff, D. C., Garg, S. K., Bode, B. W., Meredith, M., Slover, R. H., Ahmann, A. J., Welsh, J. B., Lee, S. W., & Kaufman, F. R. (2013) · New England Journal of Medicine · doi:10.1056/nejmoa1303576 · Cited by 652 · see citing works
Haak, T., Hanaire, H., Ajjan, R., Hermanns, N., Riveline, J.-P., & Rayman, G. (2017) · Diabetes Therapy · doi:10.1007/s13300-016-0223-6 · Cited by 623 · see citing works
Garg, S. K., Weinzimer, S. A., Tamborlane, W. V., Buckingham, B. A., Bode, B. W., Bailey, T. S., Brazg, R. L., Ilany, J., Slover, R. H., Anderson, S. M., et al. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0421 · Cited by 585 · see citing works
Self-citationA history of blood glucose meters and their role in self-monitoring of diabetes mellitus
Clarke, S. F., & Foster, J. R. (2012) · British Journal of Biomedical Science · doi:10.1080/09674845.2012.12002443 · Cited by 456 · see citing works
The effect of continuous glucose monitoring in well-controlled type 1 diabetes
Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group (2009) · Diabetes Care · doi:10.2337/dc09-0108 · Cited by 399 · see citing works
Evaluating the Accuracy of Continuous Glucose-Monitoring Sensors
Kovatchev, B. P., Gonder-Frederick, L. A., Cox, D. J., & Clarke, W. L. (2004) · Diabetes Care · doi:10.2337/diacare.27.8.1922 · Cited by 312 · see citing works
Consensus Statement on Self-Monitor ing of Blood Glucose
(1987) · Diabetes Care · doi:10.2337/diacare.10.1.95 · Cited by 247 · see citing works
Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors
Kovatchev, B. P., Anderson, S. M., Heinemann, L., & Clarke, W. L. (2008) · Diabetes Care · doi:10.2337/dc07-2401 · Cited by 236 · see citing works
Assessing Sensor Accuracy for Non-Adjunct Use of Continuous Glucose Monitoring
Kovatchev, B. P., Patek, S. D., Ortiz, E. A., & Breton, M. D. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0272 · Cited by 234 · see citing works
Continuous glucose monitoring sensors: past, present and future algorithmic challenges
Facchinetti, A. (2016) · Sensors · doi:10.3390/s16122093 · Cited by 176 · see citing works
Clinical accuracy of a continuous glucose monitoring system with an advanced algorithm
Bailey, T. S., Chang, A., & Christiansen, M. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814559746 · Cited by 168 · see citing works
Self-citationA Comparative Effectiveness Analysis of Three Continuous Glucose Monitors
Damiano, E. R., McKeon, K., El-Khatib, F. H., Zheng, H., Nathan, D. M., & Russell, S. J. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814532203 · Cited by 139 · see citing works
Christiansen, M. P., Bailey, T. S., Watkins, E., Liljenquist, D. R., Price, D. A., Nakamura, K., Boock, R. J., & Peyser, T. A. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0077 · Cited by 136 · see citing works
Self-citationReproducibility of Glucose Measurements Using the Glucose Sensor
Metzger, M., Leibowitz, G., Wainstein, J., Gläser, B., & Raz, I. (2002) · Diabetes Care · doi:10.2337/diacare.25.7.1185 · Cited by 134 · see citing works
Kropff, J., Bruttomesso, D., Doll, W., Farret, A., Galasso, S., Luijf, Y. M., Mader, J. K., Place, J., Boscari, F., Pieber, T. R., et al. (2014) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12378 · Cited by 132 · see citing works
JDRF CGM Study Group (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2007.0302 · Cited by 115 · see citing works
Amperometric Glucose Sensors: Sources of Error and Potential Benefit of Redundancy
Castle, J. R., & Ward, W. K. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400127 · Cited by 95 · see citing works
Rate-of-Change Dependence of the Performance of Two CGM Systems During Induced Glucose Swings
Pleus, S., Schoemaker, M., Morgenstern, K., Schmelzeisen‐Redeker, G., Haug, C., Link, M., Zschornack, E., & Freckmann, G. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815578716 · Cited by 84 · see citing works
Analysis of Time Lags and Other Sources of Error of the DexCom SEVEN Continuous Glucose Monitor
Kamath, A., Mahalingam, A., & Brauker, J. H. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0060 · Cited by 69 · see citing works
Obermaier, K., Schmelzeisen‐Redeker, G., Schoemaker, M., Klötzer, H.-M., Kirchsteiger, H., Eikmeier, H., & del Re, L. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700404 · Cited by 66 · see citing works
King, C., Anderson, S. M., Breton, M. D., Clarke, W. L., & Kovatchev, B. P. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100302 · Cited by 65 · see citing works
Characterizing Accuracy and Precision of Glucose Sensors and Meters
Rodbard, D. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814541810 · Cited by 65 · see citing works
Performance Comparison of CGM Systems
Kirchsteiger, H., Heinemann, L., Freckmann, G., Lodwig, V., Schmelzeisen‐Redeker, G., Schoemaker, M., & del Re, L. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815586013 · Cited by 45 · see citing works
FDA Advisory Panel Votes to Recommend Non-Adjunctive Use of Dexcom G5 Mobile CGM
(2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.07252.mr · Cited by 43 · see citing works
Pettus, J., & Edelman, S. V. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0369 · Cited by 38 · see citing works
Limitations of Statistical Measures of Error in Assessing the Accuracy of Continuous Glucose Sensors
Kollman, C., Wilson, D. M., Wysocki, T., Tamborlane, W. V., & Beck, R. W. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.665 · Cited by 30 · see citing works
Acciaroli, G., Vettoretti, M., Facchinetti, A., Sparacino, G., & Cobelli, C. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0088 · Cited by 22 · see citing works
The Need for Performance Standards for Continuous Glucose Monitors
Nichols, J. H., & Klonoff, D. C. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100114 · Cited by 13 · see citing works
D’Archangelo, M. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200228 · Cited by 10 · see citing works
Flagged first, then by citations
Research works printed but not linked: 1
- Colwell JA, Dudley JD, McDonald JM, et al.: Consensus Statement on Self-Monitoring of Blood Glucose. Diabetes Care 1987;10:95-99.
Cited by
Sorted by citations
Smart wearable devices in cardiovascular care: Where we are and how to move forward
Bayoumy, K., Gaber, M., Elshafeey, A., Mhaimeed, O., Dineen, E. H., Marvel, F. A., Martin, S. S., Muse, E. D., Turakhia, M. P., Tarakji, K. G., et al. (2021) · Nature Reviews Cardiology · doi:10.1038/s41569-021-00522-7 · Cited by 829 · see citing works
Cappon, G., Vettoretti, M., Sparacino, G., & Facchinetti, A. (2019) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2019.0121 · Cited by 477 · see citing works
Pullano, S. A., Greco, M., Bianco, M. G., Foti, D., Brunetti, A., & Fiorillo, A. S. (2022) · Theranostics · doi:10.7150/thno.64035 · Cited by 210 · see citing works
Measures of Accuracy for Continuous Glucose Monitoring and Blood Glucose Monitoring Devices
Freckmann, G., Pleus, S., Grady, M., Setford, S., & Levy, B. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818812062 · Cited by 202 · see citing works
Continuous glucose monitoring: Current use in diabetes management and possible future applications
Vettoretti, M., Cappon, G., Acciaroli, G., Facchinetti, A., & Sparacino, G. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818774078 · Cited by 96 · see citing works
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86 · see citing works
Hypoglycaemia and its management in primary care setting
Ibrahim, M., Baker, J., Cahn, A., Eckel, R. H., El Sayed, N. A., Fischl, A. H., Gaede, P., Leslie, R. D., Pieralice, S., Tuccinardi, D., et al. (2020) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.3332 · Cited by 21 · see citing works
- Authors
- Bailey, T. S.
- Affiliations
- AMCR Institute (US)
- Corresponding author
- Bailey, Timothy S.
- ORCID
- Bailey, Timothy S.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2017-05
- Volume
- 19
- Issue
- 2_suppl
- Pages
- S51-S54
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- PMC
- PMC5444490
- OpenAlex
- W2619117035
- Cited by
- 60
- References
- 35
- Citations by year
- 2026: 7 · 2025: 5 · 2024: 1 · 2023: 7 · 2022: 7 · 2021: 6 · 2020: 5 · 2019: 9 · 2018: 10 · 2017: 3
- FWCI
- 5.84
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Diabetes Technology & Therapeutics (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Keywords
- continuous glucose monitoring, CGM accuracy, diabetes management, accuracy assessment, clinical trials
- MeSH terms
- Data Accuracy, Blood Glucose, Diabetes Mellitus, Humans, Blood Glucose Self-Monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review