Hematocrit Interference of Blood Glucose Meters for Patient Self-Measurement

Open access
Ramljak, Sanja, Lock, John Paul, Schipper, Christina, Musholt, Petra B., Forst, Thomas, Lyon, Martha, Pfützner, Andreas · 2013 · Journal of Diabetes Science and Technology
Cited by 72

The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.

Abstract

Background: Abnormal hematocrit levels may interfere with glucose readings of patient self-assessment blood glucose (BG) meters. The aim of this laboratory investigation was to assess the potential influence of hematocrit variations on a variety of BG meters applying different measurement technologies. Methods: Venous heparinized blood was manipulated to contain three different BG concentrations (50–90, 120–180, and 280–350 mg/dl) and five different hematocrit levels (25%, 35%, 45%, 55%, and 65%). After careful oxygenation to normal blood oxygen pressure (65–100 mmHg), each sample was measured (eight times) with the following devices: Accu-Chek® Aviva, Nano, and Active, Breeze®2 and Contour®, FreeStyle Freedom Lite®, GlucoDr. auto™, Glucofix® mio Plus, GlucoLab™, GlucoMen® LX Plus, Nova Max® Link, Nova Max® Plus, OneTouch® Ultra®2 and Verio®, On Call® Plus and Platinum, Optium Xceed®, Precision Xceed®, and TaiDoc Fora TD-4227. A YSI 2300 STAT Plus™ glucose analyzer served as reference method. Stability to hematocrit influence was assumed, with <10% mean glucose result bias between the highest and lowest hematocrit levels. Results: Six of the investigated meters showed a stable performance in this investigation: Accu-Chek Active (7%), Glucofix mio Plus (5%), GlucoMen LX Plus (4%), NovaMax Plus (4%), Nova Max Link (7%), and OneTouch Verio (3%). All other meters failed this hematocrit interference test, with FreeStyle Freedom Lite (11%), and On Call Platinum 12%) being the better devices and On Call Plus (68%), GlucoLab (51%), TaiDoc Fora TD-4227 (39%), and Breeze 2 (38%) showing the worst performance. Conclusions: Hematocrit may affect BG meter performance in daily routine. In case of interference, low hematocrit values (<35%) result in too high readings. Our results encourage use of meters that are not affected by hematocrit interference.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1177/193229681300700123

Cites

3 of 3 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 3 references

Cited by

See all 72

Metadata

Authors
Ramljak, S., Lock, J. P., Schipper, C., Musholt, P. B., Forst, T., Lyon, M., & Pfützner, A.
Venue
Journal of Diabetes Science and Technology
Published
2013-01
Volume
7
Issue
1
Pages
179-189
ISSN
1932-2968 (ISSN-L)
OpenAlex
W2169834664
Cited by
72
References
3
FWCI
3.67
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Hematocrit Interference of Blood Glucose Meters for Patient Self-Measurement. https://app.cochat.ai/papers/10.1177/193229681300700123
The paper
Ramljak, S., Lock, J. P., Schipper, C., Musholt, P. B., Forst, T., Lyon, M., & Pfützner, A. (2013). Hematocrit Interference of Blood Glucose Meters for Patient Self-Measurement. Journal of Diabetes Science and Technology. https://doi.org/10.1177/193229681300700123

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.