Comparative Performance Assessment of Point-of-Care Testing Devices for Measuring Glucose and Ketones at the Patient Bedside
Open accessThe DOI matches its Crossref record; 33 of the 34 references it cites were found in a registry or an index.
Abstract
Background: Point-of-care (POC) testing devices for monitoring glucose and ketones can play a key role in the management of dysglycemia in hospitalized diabetes patients. The accuracy of glucose devices can be influenced by biochemical changes that commonly occur in critically ill hospital patients and by the medication prescribed. Little is known about the influence of these factors on ketone POC measurements. The aim of this study was to assess the analytical performance of POC hospital whole-blood glucose and ketone meters and the extent of glucose interference factors on the design and accuracy of ketone results. Methods: StatStrip glucose/ketone, Optium FreeStyle glucose/ketone, and Accu-Chek Performa glucose were also assessed and results compared to a central laboratory reference method. The analytical evaluation was performed according to Clinical and Laboratory Standards Institute (CLSI) protocols for precision, linearity, method comparison, and interference. Results: The interferences assessed included acetoacetate, acetaminophen, ascorbic acid, galactose, maltose, uric acid, and sodium. The accuracies of both Optium ketone and glucose measurements were significantly influenced by varying levels of hematocrit and ascorbic acid. StatStrip ketone and glucose measurements were unaffected by the interferences tested with exception of ascorbic acid, which reduced the higher level ketone value. The accuracy of Accu-Chek glucose measurements was affected by hematocrit, by ascorbic acid, and significantly by galactose. The method correlation assessment indicated differences between the meters in compliance to ISO 15197 and CLSI 12-A3 performance criteria. Conclusions: Combined POC glucose/ketone methods are now available. The use of these devices in a hospital setting requires careful consideration with regard to the selection of instruments not sensitive to hematocrit variation and presence of interfering substances.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“High hematocrit will increase or decrease the measured glucose, or will have minimal effect, depending on the design of the measuring system (132,133).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [133]
In the citing paper
“Several sugars-notably maltose, galactose, and xylose-falsely increase results of some glucose meters.”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [133]
Citations
Cites
Joint British Diabetes Societies guideline for the management of diabetic ketoacidosis
Savage, M. W., Dhatariya, K. K., Kilvert, A., Rayman, G., Rees, J. A. E., Courtney, C. H., Hilton, L., Dyer, P. H., Hamersley, M. S., & Societies, F. T. J. B. D. (2011) · Diabetic Medicine · doi:10.1111/j.1464-5491.2011.03246.x · Cited by 338 · cited 1×
“Recent guidelines have advocated the use of whole-blood measurement of BHB for the management of DKA and for monitoring target directed therapy.(6,7) 563351 DSTXXX10.”
MethodDiabetic ketoacidosis in children and adolescents with diabetes
Wolfsdorf, J., Craig, M. E., Daneman, D., Dunger, D., Edge, J., Lee, W., Rosenbloom, A., Sperling, M., & Hanas, R. (2009) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2009.00569.x · Cited by 313
Glucose Measurement: Confounding Issues in Setting Targets for Inpatient Management
Dungan, K., Chapman, J., Braithwaite, S. S., & Buse, J. (2007) · Diabetes Care · doi:10.2337/dc06-1679 · Cited by 292
Glucose control in the intensive care unit
Fahy, B. G., Sheehy, A. M., & Coursin, D. B. (2009) · Critical Care Medicine · doi:10.1097/ccm.0b013e3181a19ceb · Cited by 170
The Accuracy of Point-of-Care Glucose Measurements
Rebel, A., Rice, M. A., & Fahy, B. G. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600228 · Cited by 144 · cited 1×
“A wide range of endogenous and exogenous substances present in the blood of critically ill patients can donate electrons to nonspecifically initiate the electrochemical reaction and cause interference with glucose results, giving elevated values.(13) An increasingly large number of interfering factors have been associ…”
BackgroundResuscitation After Severe Burn Injury Using High-Dose Ascorbic Acid: A Retrospective Review
Kahn, S. A., Beers, R. J., & Lentz, C. W. (2011) · Journal of Burn Care & Research · doi:10.1097/bcr.0b013e318204b336 · Cited by 137
Interference of Maltose, Icodextrin, Galactose, or Xylose with Some Blood Glucose Monitoring Systems
Schleis, T. G. (2007) · Pharmacotherapy: The Journal of Human Pharmacology and Drug Therapy · doi:10.1592/phco.27.9.1313 · Cited by 102 · cited 2×
“The enzymes used in some handheld POC glucose devices have been reported not to be specific for glucose and shown to react in the presence of maltose and galactose.(14) Similarly it was recently reported that Optium BHB measurements were adversely influenced by acetoacetate, suggesting poor specificity of the BHB dehy…”
MethodResultArora, S., Henderson, S. O., Long, T., & Menchine, M. (2011) · Diabetes Care · doi:10.2337/dc10-1844 · Cited by 85
Hyperglycaemia in critically ill patients: marker or mediator of mortality?
Corstjens, A. M., van der Horst, I. C. C., Zijlstra, J. G., Groeneveld, A. B. J., Zijlstra, F., Tulleken, J. E., & Ligtenberg, J. J. M. (2006) · Critical Care · doi:10.1186/cc4957 · Cited by 68
Heinemann, L. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0076 · Cited by 62 · cited 1×
“However, it has been well reported that the accuracy of many hospital glucose meters can be affected by the biochemical and physiological changes that commonly occur in acute, chronic, and critically ill hospital patients and also by the medication prescribed.(11,12) A wide range of endogenous and exogenous substances…”
Background
Research works printed but not linked: 7
- Charles RA, 2007, Singapore Med J, 48, 986
- Clinical and Laboratory Standards Institute, 2004, Evaluation of Precision Performance of Quantitative Measuring Methods; Approved Guideline—Second Edition
- Clinical and Laboratory Standards Institute, 2013, Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline—Third Edition
- Clinical and Laboratory Standards Institute, 2005, Interference Testing in Clinical Chemistry. Approved Guideline—Second Edition
- Clinical and Laboratory Standards Institute, 2003, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
- International Organization for Standardization, 2013, In Vitro Diagnostic Test Systems—Requirements for Blood Glucose Monitoring Systems for Selftesting in Managing Diabetes Mellitus
- Clinical and Laboratory Standards Institute, 2013, Point of Care Blood Glucose Testing in Acute and Chronic Care Facilities. Approved Guideline—Third Edition
Cited by
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
Sacks, D. B., Arnold, M. A., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Clinical Chemistry · doi:10.1093/clinchem/hvad080 · Cited by 208
Dhatariya, K. (2016) · The Review of Diabetic Studies · doi:10.1900/rds.2016.13.217 · Cited by 119
Exogenous Ketone Supplements in Athletic Contexts: Past, Present, and Future
Evans, M., McClure, T. S., Koutnik, A. P., & Egan, B. (2022) · Sports Medicine · doi:10.1007/s40279-022-01756-2 · Cited by 93
“For example, recovery of βHB assayed by enzymatic technique indicated that the FreeStyle Optium POC device (Abbott Laboratories, UK) overestimates [R-βHB], whereas the StatStrip POC device (Nova Biomedical, USA) underestimated R-βHB [114].”
MethodHager, D. N., Hooper, M. H., Bernard, G. R., Busse, L. W., Ely, E. W., Fowler, A. A., Gaieski, D. F., Hall, A. J., Hinson, J. S., Jackson, J. C., et al. (2019) · Trials · doi:10.1186/s13063-019-3254-2 · Cited by 78
The performance of fl ash glucose monitoring in critically ill patients with diabetes
Ancona, P., Eastwood, G. M., Lucchetta, L., Ekinci, E. I., Bellomo, R., & Mårtensson, J. (2017) · Critical Care and Resuscitation · doi:10.1016/s1441-2772(23)00789-5 · Cited by 48
Shaw, D. M., Mérien, F., Braakhuis, A. J., Maunder, E., & Dulson, D. K. (2019) · Sports Medicine · doi:10.1007/s40279-019-01246-y · Cited by 60
Highly specific and rapid glycan based amperometric detection of influenza viruses
Cui, X., Das, A., Dhawane, A. N., Sweeney, J. A., Zhang, X., Chivukula, V., & Iyer, S. S. (2017) · Chemical Science · doi:10.1039/c6sc03720h · Cited by 47
Larsen, A. E., Bibby, B. M., & Hansen, M. (2018) · Nutrients · doi:10.3390/nu10121918 · Cited by 33
“We obtained blood by finger sticks and immediately analyzed for capillary blood (cb)-glucose (Hemocue Glucose 201 RT [17,18]), total cb-cholesterol (Accutend plus [19,20]) and cb-ketones (Freestyle Precision Neo [21]).”
MethodDetection of Enzymes, Viruses, and Bacteria Using Glucose Meters
Das, A., Cui, X., Chivukula, V., & Iyer, S. S. (2018) · Analytical Chemistry · doi:10.1021/acs.analchem.8b02960 · Cited by 28
Luethi, N., Cioccari, L., Crisman, M., Bellomo, R., Eastwood, G. M., & Mårtensson, J. (2016) · Critical Care · doi:10.1186/s13054-016-1462-7 · Cited by 30
“Yet, previous validation studies suggest that the use of point-of-care instruments may overestimate true blood β-OHB levels when hematocrit is below normal [14], which was the case in most of our patients.”
BackgroundMethodNorgren, J., Sindi, S., Matton, A., Kåreholt, I., Akenine, U., Nordin, K., Rosenborg, S., Ngandu, T., & Kivipelto, M. (2019) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.00454.2019 · Cited by 29
“We used the jitter (5) function in STATA to enhance visibility of dots with identical coordinates in Fig.”
MethodHamblin, P. S., Wong, R., Ekinci, E. I., Sztal‐Mazer, S., Balachandran, S., Frydman, A. S., Hanrahan, T. P., Hu, R., Ket, S., Moss, A., et al. (2021) · Diabetes Care · doi:10.2337/dc21-0256 · Cited by 16
Critical Care Glucose Point-of-Care Testing
Narla, S. N., Jones, M., Hermayer, K. L., & Zhu, Y. (2016) · Advances in clinical chemistry · doi:10.1016/bs.acc.2016.05.002 · Cited by 17
“ascorbic acid also falsely increased glucose concentrations measured by both GDH–NAD and GDH–PQQ assays [69].”
BackgroundMethodInfluentialHowell, A. P., Parrett, J. L., & Malcom, D. R. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819889638 · Cited by 15
“Use of LA blood glucose is recommended to POC bedside monitoring in patients receiving high-dose intravenous vitamin C.(18,19,29) Our study suggests that the varying mechanisms of blood glucose testing did have an impact in the accuracy of POC testing in the patients receiving intravenous vitamin C for the treatment o…”
BackgroundMethodMulticentric evaluation of eight glucose and four ketone blood meters
Guimont, M.-C., Desjobert, H., Fonfrède, M., Vitoux, D., Benoist, J., Launay, J., Peoc’h, K., & Lefèvre, G. (2015) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2015.07.032 · Cited by 17
“To date, they have evidenced that ketone meter’s measurements may also be modified by the hematocrit [21], but we did not explore it within the frame of the present study.”
BackgroundElectrochemical Glucose Biosensors for Diabetes Care
Ocvirk, G., Buck, H., & DuVall, S. H. (2016) · Bioanalytical reviews · doi:10.1007/11663_2016_3 · Cited by 16
“[499] 2015 Accu-Chek Performa FreeStyle OptiumTM StatStrip V Hematocrit variation, interference, and precision”
BackgroundVitamin C for Sepsis and Septic Shock
Ammar, M. A., Ammar, A. A., Condeni, M. S., & Bell, C. M. (2021) · American Journal of Therapeutics · doi:10.1097/mjt.0000000000001423 · Cited by 16
Edwards, M. G. P., Furuholmen-Jenssen, T., Søegaard, E. G. I., Thapa, S. B., & Andersen, J. R. (2024) · Frontiers in Nutrition · doi:10.3389/fnut.2024.1406366 · Cited by 11
Lachance, O., Goyer, F., Adhikari, N. K. J., Masse, M.-H., Bilodeau, J., Lamontagne, F., & Leclair, M.-A. (2021) · Journal of Medical Case Reports · doi:10.1186/s13256-021-02869-4 · Cited by 11
“While some evidence suggests that certain glucometers (for example, StatStrip [Nova Biomedical]) are more reliable than others in the context of high-dose vitamin C therapy, clinical experience with such devices remains extremely limited [12].”
BackgroundChenoweth, J. A., Dang, L. T., Gao, G., & Tran, N. K. (2020) · Clinical Toxicology · doi:10.1080/15563650.2020.1732404 · Cited by 10
Diet-induced ketosis in adult patients with subacute acquired brain injury: a feasibility study
Edwards, M. G. P., Andersen, J. R., Curtis, D. J., Riberholt, C. G., & Poulsen, I. (2024) · Frontiers in Medicine · doi:10.3389/fmed.2023.1305888 · Cited by 10
Depczynski, B., Lee, A. T. K., Varndell, W., & Chiew, A. L. (2019) · Journal of Diabetes Research · doi:10.1155/2019/7387128 · Cited by 11
“Defining the expected degree of ketosis with acute illness has becomemore important given the ability to undertake point-of-care BHB testing [5] and that sodium glucose cotransporter type 2 inhibitors (SGLT2i) may precipitate euglycaemic DKA with intercurrent illness [6].”
BackgroundMethodGlucometry When Using Vitamin C in Sepsis
Hager, D. N., Martin, G. S., Sevransky, J., & Hooper, M. H. (2018) · CHEST Journal · doi:10.1016/j.chest.2018.03.018 · Cited by 10
Establishing Pragmatic Analytical Performance Specifications for Blood Beta-Hydroxybutyrate Testing
Kilpatrick, E. S., Butler, A. E., Atkin, S. L., & Sacks, D. B. (2023) · Clinical Chemistry · doi:10.1093/clinchem/hvad020 · Cited by 10
“This observation—that the largest SDs were at the highest measured BOHB concentrations—was also found to be the case in other studies (10–12) and so there may be a particular issue in tracking BOHB in patients who have higher initial values, especially around the 6.0 mmol/L concentration indicative…”
InfluentialFlannery, A. H., Bastin, M. L. T., Magee, C. A., & Bensadoun, E. (2017) · CHEST Journal · doi:10.1016/j.chest.2017.05.023 · Cited by 9
Helmersson‐Karlqvist, J., Hammarström, K. H., Palmberg, K., & Backman‐Johansson, C. (2019) · Journal of Clinical Laboratory Analysis · doi:10.1002/jcla.22851 · Cited by 8
Rosengrave, P., Wohlrab, C., Spencer, E. S., Williman, J. A., Shaw, G., & Carr, A. C. (2022) · Critical Care and Resuscitation · doi:10.51893/2022.2.oa7 · Cited by 7
Kilpatrick, E. S., Butler, A. E., Saeed, S., Alamuddin, N., Atkin, S. L., & Sacks, D. B. (2023) · Diabetology & Metabolic Syndrome · doi:10.1186/s13098-023-01176-w · Cited by 7
“In a comparison study, the Stat Strip BOHB meter showed good correlation to a central laboratory reference method and good day-to-day reproducibility [14].”
Methodde Kreutzenberg, S. V., & Avogaro, A. (2017) · Journal of Endocrinological Investigation · doi:10.1007/s40618-017-0615-0 · Cited by 10
“The performance of whole blood tests including StatStrip Ketone has previously been validated against laboratory methods in published studies [12, 18, 19].”
MethodMeade, S. M., Girolamo, N. D., Cray, C., Aquilino, N., DiGeronimo, P. M., & Brandão, J. (2022) · American Journal of Veterinary Research · doi:10.2460/ajvr.21.11.0192 · Cited by 6
“35 Precision of POC performance was only evaluated using plasma samples.”
Bjerg, L. N., Thomsen, H. H., Madsen, J. B., & Sandfeld‐Paulsen, B. (2022) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365513.2022.2054027 · Cited by 6
“ascorbic acid and haematocrit, affecting the accuracy of the BHB estimates [22].”
BackgroundBjerg, L. N., Thomsen, H. H., & Sandfeld‐Paulsen, B. (2025) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365513.2025.2460198 · Cited by 4
“BHB can be assessed by enzymatic methods either in a routine laboratory or by point-of-care (POC) meters [13–18].”
Hamblin, P. S., Wong, R., Ekinci, E. I., Sztal‐Mazer, S., Balachandran, S., Frydman, A. S., Hanrahan, T. P., Hu, R., Ket, S., Moss, A., et al. (2021) · Clinical Endocrinology · doi:10.1111/cen.14621 · Cited by 4
Role of vitamin C in diabetic ketoacidosis: Is it ready for prime time?
Casillas, S., Pomerantz, A., Surani, S., & Varón, J. (2018) · World Journal of Diabetes · doi:10.4239/wjd.v9.i12.206 · Cited by 4
Soehnlein, C., Hirsch, S. E., Myre, B. L., Stacy, N. I., Clark, S. A., Cray, C., Aoki, D. M., Kaleel, K. L., Klingshirn, S., Toonder, M., et al. (2022) · Journal of Wildlife Diseases · doi:10.7589/jwd-d-21-00197 · Cited by 4
“Our results are similar to those in human patients, in that the POC device displayed a strong correlation with the laboratory reference method; however, interfering factors such as preanalytical error (e.g., hemolysis) or underlying disease may affect the results (Ceriotti et al. 2015).”
ResultSalcedo-Cifuentes, M., Gómez, F., Curcio, C., Corchuelo, J., & Méndez, F. (2021) · Revista Médicas UIS · doi:10.18273/revmed.v34n2-2021001 · Cited by 1
Surani, S., & Sharma, M. (2021) · Current Medical Research and Opinion · doi:10.1080/03007995.2021.1929133 · Cited by 1
Precision Medicine in the Intensive Care Unit: Identifying Opportunities and Overcoming Barriers
Palmieri, T. L., & Tran, N. K. (2019) · Annual update in intensive care and emergency medicine · doi:10.1007/978-3-030-06067-1_1
Current Treatment of DKA: A Review and Approach to Establishing a Best Practice Protocol
Qually, M. (2015) · Mspace (University of Manitoba)
Fitria, S. T., Yaswir, R., & Efrida, E. (2022) · Jurnal Kesehatan Andalas · doi:10.25077/jka.v10i3.1827 · Cited by 1
Holland, A. M., Moore, A. R., Ansley, J., Fritz, N., & Bederman, I. R. (2023) · Muscles · doi:10.3390/muscles2040025
“They are deemed appropriate for this purpose because, although not as accurate as reference methods, they provide nearly instantaneous results at a lower cost, which allows clinicians to make critical healthcare decisions based on more frequent ketone readings [22].”
Steinbach, M. K., Müller, H. N., Reichard, Y., Eulberg, K., Frömel, T., Lein, A., Peter, C., Prager, K., & Grimmler, M. (2025) · Clinica Chimica Acta · doi:10.1016/j.cca.2025.120703
Grzych, G., Pinsart, A., Defauwes, I., Campas, M., Grifnée, E., Tullio, P. D., & Cavalier, E. (2025) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2025-1384
Manoy, S., Crook, P., Gaughan, K., Vara, R., Flower, A., Oladimeji, T., Jarvis, C., Pope, A., & Pandithan, D. (2026) · Journal of Pediatric Endocrinology and Metabolism · doi:10.1515/jpem-2026-0274
“…glucose dehydrogenase and its co-factor pyrro-loquinolinequione (GDH-PQQ) as a reagent are known to falsely elevate POC glucometer readings with an increased current at the detecting electrode [8] resulting from the presence of additional disaccharide or monosaccharide sugars such as galactose [9].”
Metadata
- Authors
- Ceriotti, F., Kaczmarek, E., Guerra, E., Mastrantonio, F., Lucarelli, F., Valgimigli, F., & Mosca, A.
- Affiliations
- IRCCS Ospedale San Raffaele (IT)
- Menarini Group (Italy) (IT)
- University of Milan (IT)
- Corresponding author
- Ceriotti, Ferruccio
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2014-12-17
- Volume
- 9
- Issue
- 2
- Pages
- 268-277
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC4604596
- OpenAlex
- W2156229084
- Semantic Scholar
- bd6fed17a5bdaf987d5645a903f1747f69a1ec33
- Cited by
- 45
- References
- 34
- Citations by year
- 2026: 1 · 2025: 3 · 2024: 2 · 2023: 5 · 2022: 6 · 2021: 6 · 2020: 1 · 2019: 7 · 2018: 4 · 2017: 4 · 2016: 4 · 2015: 2
- FWCI
- 2.22
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Keywords
- point-of-care testing, glucose interference, ascorbic acid interference, hematocrit, acetoacetate, hospitalized diabetes patients, method comparison, ISO 15197, critically ill patients
- MeSH terms
- Blood Glucose, Female, Humans, Ketones, Male, Monitoring, Physiologic, Point-of-Care Systems
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.