Assessing the Accuracy of Continuous Glucose Monitoring (CGM) Calibrated With Capillary Values Using Capillary or Venous Glucose Levels as a Reference
Open accessThe DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.
Abstract
Background: Using the standard venous reference for the evaluation of continuous glucose monitoring (CGM) systems could possibly negatively affect measured CGM accuracy since CGM are generally calibrated with capillary glucose and venous and capillary glucose concentrations differ. We therefore aimed to quantify the effect of using capillary versus venous glucose reference samples on estimated accuracy in capillary calibrated CGM. Methods: We evaluated 41 individuals with type 1 diabetes mellitus (T1DM) using the Dexcom G4 CGM system over 6 days. Patients calibrated their CGM devices with capillary glucose by means of the HemoCue system. During 2 visits, capillary and venous samples were simultaneously measured by HemoCue and compared to concomitantly obtained CGM readings. The mean absolute relative difference (MARD) was calculated using capillary and venous reference samples. Results: Venous glucose values were 0.83 mmol/L (15.0 mg/dl) lower than capillary values over all glycemic ranges, P < .0001. Below 4 mmol/l (72 mg/dl), the difference was 1.25 mmol/l (22.5 mg/dl), P = .0001, at 4-10 mmol/l (72-180 mg/dl), 0.67 mmol/l (12.0 mg/dl), P < .0001 and above 10 mmol/l (180 mg/dl), 0.95 mmol/l (17.1 mg/dl), P < .0001. MARD was 11.7% using capillary values as reference compared to 13.7% using venous samples, P = .037. Below 4 mmol/l (72 mg/dl) MARD was 16.6% and 31.8%, P = .048, at 4-10 mmol/l (72-180 mg/dl) 12.1% and 12.6%, P = .32, above 10 mmol/l (180 mg/dl) 8.7% and 9.2%, P = .82. Conclusion: Using capillary glucose concentrations as reference to evaluate the accuracy of CGM calibrated with capillary samples is associated with a lower MARD than using venous glucose as the reference. Capillary glucose concentrations were significantly higher than venous in all glycemic ranges.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Glycemic Control and Excess Mortality in Type 1 Diabetes
Lind, M., Svensson, A.-M., Kosiborod, M., Gudbjörnsdottir, S., Pivodic, A., Wedel, H., Dahlqvist, S., Clements, M., & Rosengren, A. (2014) · New England Journal of Medicine · doi:10.1056/nejmoa1408214 · Cited by 964
Self-citationEstimated life expectancy in a Scottish cohort with type 1 diabetes, 2008-2010
Livingstone, S. J., Levin, D., Looker, H. C., Lindsay, R. S., Wild, S. H., Joss, N., Leese, G., Leslie, P., McCrimmon, R. J., Metcalfe, W., et al. (2015) · JAMA · doi:10.1001/jama.2014.16425 · Cited by 621
Pickup, J. C., Freeman, S. C., & Sutton, A. J. (2011) · BMJ · doi:10.1136/bmj.d3805 · Cited by 541
the SWITCH Study Group, Battelino, T., Conget, I., Olsen, B., Schütz-Fuhrmann, I., Hommel, E., Hoogma, R., Schierloh, U., Sulli, N., & Bolinder, J. (2012) · Diabetologia · doi:10.1007/s00125-012-2708-9 · Cited by 494
Continuous glucose monitoring systems for type 1 diabetes mellitus
Langendam, M., Luijf, Y. M., Hooft, L., DeVries, J. H., Mudde, A. H., & Scholten, R. J. (2012) · Cochrane Database of Systematic Reviews · doi:10.1002/14651858.cd008101.pub2 · Cited by 262
Self-citationLiu, D., Moberg, E., Kollind, M., Lins, P. E., Adamson, U., & Macdonald, I. A. (1992) · Diabetologia · doi:10.1007/bf00400932 · Cited by 162 · cited 3×
“Nonetheless, it has been noted(17,18) that external validity of these data are limited since patients generally use capillary glucose for calibration in daily life and venous glucose content can differ significantly from capillary glucose.(12,22,27) Previous studies have described differences in measured accuracy of C…”
BackgroundMethod
Cited by
Reddy, M., Jugnee, N., El Laboudi, A., Spanudakis, E., Anantharaja, S., & Oliver, N. (2018) · Diabetic Medicine · doi:10.1111/dme.13561 · Cited by 245
Ólafsdóttir, A. F., Attvall, S., Sandgren, U., Dahlqvist, S., Pivodic, A., Skrtic, S., Theodorsson, E., & Lind, M. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0392 · Cited by 173
Self-citationHeinemann, L., Schoemaker, M., Schmelzeisen-Redecker, G., Hinzmann, R., Kassab, A., Freckmann, G., Reiterer, F., & Del Re, L. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819855670 · Cited by 170
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
Self-citationLimits to the Evaluation of the Accuracy of Continuous Glucose Monitoring Systems by Clinical Trials
Schrangl, P., Reiterer, F., Heinemann, L., Freckmann, G., & Del Re, L. (2018) · Biosensors · doi:10.3390/bios8020050 · Cited by 46
“Therefore, the effect of a bias in the reference measurements was not considered throughout this paper (see [53] for the effect on MARD of using different BG meters for calibrating the CGM and for drawing the paired reference measurements for the performance assessment, and see [51] for an overview of the clinical imp…”
Method
Metadata
- Authors
- Andelin, M., Kropff, J., Matuleviciene, V., Joseph, J. I., Attvall, S., Theodorsson, E., Hirsch, I. B., Imberg, H., Dahlqvist, S., Klonoff, D., Haraldsson, B., DeVries, J. H., & Lind, M.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2016-01-24
- Volume
- 10
- Issue
- 4
- Pages
- 876-884
- ISSN
- 1932-2968 (ISSN-L)
- OpenAlex
- W2300183148
- Cited by
- 28
- References
- 6
- FWCI
- 4.04
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Assessing the Accuracy of Continuous Glucose Monitoring (CGM) Calibrated With Capillary Values Using Capillary or Venous Glucose Levels as a Reference. https://app.cochat.ai/papers/10.1177/1932296815626724
- The paper
- Andelin, M., Kropff, J., Matuleviciene, V., Joseph, J. I., Attvall, S., Theodorsson, E., Hirsch, I. B., Imberg, H., Dahlqvist, S., Klonoff, D., Haraldsson, B., DeVries, J. H., & Lind, M. (2016). Assessing the Accuracy of Continuous Glucose Monitoring (CGM) Calibrated With Capillary Values Using Capillary or Venous Glucose Levels as a Reference. Journal of Diabetes Science and Technology. https://doi.org/10.1177/1932296815626724
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.