Do the New FDA Guidance Documents Help Improving Performance of Blood Glucose Monitoring Systems Compared With ISO 15197?
Open accessAbstract
Recently, the Food and Drug Administration (FDA) published guidance documents for point-of-care testing (POCT) and over-the-counter (OTC) blood glucose monitoring systems (BGMS). These are expected to improve weaknesses of the already established ISO 15197:2013 standard accuracy evaluation for OTC BGMS; however, the proposed criteria and procedures rather raise new questions. As an example, the OTC guidance stipulates tighter accuracy criteria in the low glycemic range than the POCT guidance for no obvious reason. Furthermore, a clear definition of requirements for the reference method is missing under consideration of the strict accuracy criteria. External surveillance of BGMS after introduction to the market is an issue that is addressed neither in the FDA documents, nor in ISO 15197, but in the Clinical and Laboratory Standards Institute (CLSI) guideline POCT12-A3.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Procedures might be used that exhibit variations between devices, reagent system, or calibrator batches.
“Therefore, reference measurement procedures from different manufacturers, as well as, for example, comparison measurement procedures used in independent postmarket evaluations may provide systematically different results, 21 or procedures might be used that exhibit variations between devices, reagent system, or calibrator batches.22-24”
From Impact of Two Different Reference Measurement Procedures on Apparent System Accuracy of 18 CE-Marked Current-Generation Blood Glucose Monitoring Systems (2022) · cites this paper as [23]
The “Measurement of Reference Values” section (page not supplied) reports measurement differences between devices of the same model. The “Testing Procedures” section requires testing three reagent-system lots, but does not say those lots produce different results; our copy also does not establish variation between calibrator batches. (LLM, conf=0.86)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620 · see citing works
Klonoff, D. C., Lias, C., Vigersky, R. A., Clarke, W. L., Parkes, J. L., Sacks, D. B., Kirkman, M. S., Kovatchev, B. P., Panel, T. E. G., Aurand, L. A., et al. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814539589 · Cited by 175 · see citing works
Plasma glucose measurement with the Yellow Springs Glucose 2300 STAT and the Olympus AU640
Twomey, P. J. (2004) · Journal of Clinical Pathology · doi:10.1136/jcp.2003.013417 · Cited by 75 · see citing works
Performance of Cleared Blood Glucose Monitors
Klonoff, D. C., & Prahalad, P. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815584797 · Cited by 59 · see citing works
Freckmann, G., Link, M., Schmid, C., Pleus, S., Baumstark, A., & Haug, C. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0085 · Cited by 46 · see citing works
Self-citation(2015) · doi:10.1177/1932296815580161 · Cited by 31 · see citing works
Self-citation(2016) · doi:10.1177/1932296816634356 · Cited by 23 · see citing works
Development of the Diabetes Technology Society Blood Glucose Monitor System Surveillance Protocol
Klonoff, D. C., Lias, C., Beck, S. E., Parkes, J. L., Kovatchev, B. P., Vigersky, R. A., Arreaza-Rubín, G., Burk, R. D., Kowalski, A. J., Little, R. R., et al. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815614587 · Cited by 20 · see citing works
Why the New FDA Glucose Meter POCT Guidance Is Disappointing
Krouwer, J. S. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816682763 · Cited by 7 · see citing works
Flagged first, then by citations
Research works printed but not linked: 3
- Bailey TS, Klaff LJ, Wallace JF, et al. Fundamental impor- tance of reference glucose analyzer accuracy for evaluating the performance of blood glucose monitoring systems (BGMSs). J Diabetes Sci Technol. 2016;10(4):872-875.
- Bundesärztekammer. Richtlinie der Bundesärztekammer zur Qualitätssicherung laboratoriumsmedizinischer Untersuchungen. Deutsches Ärzteblatt. 2014;111(38):1583-1618.
- Link M, Schmid C, Pleus S, et al. System accuracy evaluation of four systems for self-monitoring of blood glucose follow- ing ISO 15197 using a glucose oxidase and a hexokinase-based comparison method. J Diabetes Sci Technol. 2015;14(9): 1041-1050.
Cited by
Sorted by citations
Pullano, S. A., Greco, M., Bianco, M. G., Foti, D., Brunetti, A., & Fiorillo, A. S. (2022) · Theranostics · doi:10.7150/thno.64035 · Cited by 210 · see citing works
Investigation of the Accuracy of 18 Marketed Blood Glucose Monitors
Klonoff, D. C., Parkes, J. L., Kovatchev, B. P., Kerr, D., Bevier, W. C., Brazg, R. L., Christiansen, M., Bailey, T. S., Nichols, J. H., & Kohn, M. A. (2018) · Diabetes Care · doi:10.2337/dc17-1960 · Cited by 144 · see citing works
Self-measurement of Blood Glucose and Continuous Glucose Monitoring – Is There Only One Future?
Heinemann, L., & Stuhr, A. (2018) · touchREVIEWS in Endocrinology · doi:10.17925/ee.2018.14.2.24 · Cited by 24 · see citing works
MacLeod, K., Katz, L. B., & Cameron, H. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818790228 · Cited by 19 · see citing works
Freckmann, G., Baumstark, A., Jendrike, N., Mende, J., Schauer, S., Link, M., Pleus, S., & Haug, C. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820948873 · Cited by 12 · see citing works
Self-citationImpact of Glucose Measuring Systems and Sample Type on Diagnosis Rates of Diabetes Mellitus
Blaurock, M., Kallner, A., Menzel, S., Masuch, A., Nauck, M. A., & Petersmann, A. (2018) · Diabetes Therapy · doi:10.1007/s13300-018-0495-0 · Cited by 11 · see citing works
Aberer, F., Theiler‐Schwetz, V., Ziko, H., Hausegger, B., Wiederstein‐Grasser, I., Hochfellner, D. A., Eller, P., Tomberger, G., Ellmerer, M., Mader, J. K., et al. (2020) · Scientific Reports · doi:10.1038/s41598-020-63659-4 · Cited by 9 · see citing works
Lyon, A. W., & Lyon, M. E. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819889639 · Cited by 7 · see citing works
Zafra‐Tanaka, J. H., Beran, D., Vetter, B., Sampath, R., & Bernabé‐Ortiz, A. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296821997909 · Cited by 5 · see citing works
Thorpe, G. H. G. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818799365 · Cited by 3 · see citing works
Do We Need the Replacement of YSI 2300? A View from the Clinical Laboratory
Spanou, L., & Makris, K. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211007170 · Cited by 3 · see citing works
Analytical performance of SMBG systems
Ahn, D. T., & Klonoff, D. C. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816714-4.00002-8
Management of Point-of-care Testing
Park, C.-E. (2024) · Korean Journal of Clinical Laboratory Science · doi:10.15324/kjcls.2024.56.4.295
- Authors
- Freckmann, G., Baumstark, A., & Pleus, S.
- Affiliations
- Universität Ulm (DE)
- Corresponding author
- Freckmann, Guido
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2017-06-06
- Volume
- 11
- Issue
- 6
- Pages
- 1240-1246
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC5951049
- OpenAlex
- W2622212158
- Cited by
- 13
- References
- 9
- Citations by year
- 2024: 1 · 2021: 3 · 2020: 3 · 2019: 1 · 2018: 5
- FWCI
- 1.82
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Health Information Management › Healthcare Regulation
- Keywords
- blood glucose monitoring systems, FDA guidance, ISO 15197, point-of-care testing, accuracy assessment, diabetes mellitus
- MeSH terms
- Blood Glucose, Documentation, Humans, Predictive Value of Tests, Time Factors, United States, United States Food and Drug Administration, Blood Glucose Self-Monitoring, Reproducibility of Results, Biomarkers, Practice Guidelines as Topic, Monitoring, Ambulatory, Device Approval
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review