Evaluation of Linearity and Interference Effect on SMBG and POCT Devices, Showing Drastic High Values, Low Values, or Error Messages
Open accessThe DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
Background: When a patient with diabetes measures blood glucose levels using a simple blood glucose meter, an error message or abnormal value may be displayed because of interfering substances, potentially leading to unnecessary medical treatment. Methods: Here, we tested 10 self-monitoring of blood glucose (SMBG) and point-of-care testing (POCT) devices to investigate under what conditions they fail to measure blood glucose levels or display abnormal values instead. We also evaluated the influence and linearity of several well-known interfering substances and hematocrit (Ht). Each adjusted sample was measured by each device twice and the values compared with the reference values. Results: Linearity was favorable in all but one model. The impact of the interfering substances and Ht has been confirmed in many SMBG and POCT devices. In particular, some models showed significant pseudo-high or -low levels for samples that contained ascorbic acid, which is used widely in infusion solutions. Three devices showed error messages to avoid presenting a false value. Conclusions: Since pseudo-high levels may result in the excessive administration of insulin and lead to adverse events, this report serves as a critical reminder of the importance of device choice for specific indications. Three new devices have safety functions that prevent the showing of false values.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Many meters incorporate changes that eliminate or greatly ameliorate most interferences, but interferences persist (137,138).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [137]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Miller, K. M., Beck, R. W., Bergenstal, R. M., Goland, R. S., Haller, M. J., McGill, J. B., Rodriguez, H., Simmons, J. H., Hirsch, I. B., & for the T1D Exchange Clinic Network (2013) · Diabetes Care · doi:10.2337/dc12-1770 · Cited by 505
Ziegler, R., Heidtmann, B., Hilgard, D., Hofer, S., Rosenbauer, J., Holl, R., & DPV-Wiss-Initiative, F. T. (2011) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2010.00650.x · Cited by 334
Grant, R. W., & Kirkman, M. S. (2015) · Diabetes Care · doi:10.2337/dc14-2142 · Cited by 190
Harashima, S., Fukushima, T., Sasaki, M., Nishi, Y., Fujimoto, S., Ogura, M., Yamane, S., Tanaka, D., Harada, N., Hamasaki, A., et al. (2012) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.2363 · Cited by 30
Sekido, K., Sekido, T., Kaneko, A., Hosokawa, M., Sato, A., Sato, Y., Yamazaki, M., & Komatsu, M. (2017) · Endocrine Journal · doi:10.1507/endocrj.ej17-0193 · Cited by 23
Kim, S. K., Hahm, J. R., Kim, H., Kim, S., Jung, T. S., Jung, J. H., Lee, H. R., & Kim, D. R. (2013) · Yonsei Medical Journal · doi:10.3349/ymj.2013.54.5.1289 · Cited by 16
Interference by Pralidoxime (PAM) salts in clinical laboratory tests
Nagase, S., Kohguchi, K., Tohyama, K., Watanabe, M., & Iwatani, Y. (2012) · Clinica Chimica Acta · doi:10.1016/j.cca.2012.11.017 · Cited by 4
Research works printed but not linked: 14
- American Diabetes Association, 2015, Diabetes Care, 38
- Kawashima H., 2016, J Med Pharm Sci, 73, 211
- Usami M, 2013, J Clin Lab Instr Reag, 36, 207
- Nagatomo R, 2013, Practice, 30, 103
- Tsuchida K, 2017, Jap J Med Tech, 66, 110
- Fujisaki N, 2013, J Jap Diabet Soc, 56, 666
- Miura F, 2009, J Jap Diabet Soc, 52, 865
- Yamazaki, 2009, J Clin Lab Instr Reag, 32, 707
- David B., 2013, POCT12-A3 Point-of-Care Blood Glucose Testing in Acute and Chronic Care Facilities; Approved Guideline
- Mitsui T, 2013, J New Remed Clin, 62, 2194
- Yoshikawa Y, 2006, J Clin Lab Instr Reag, 29, 573
- Koide K, 2017, Diabetes, 66, A183
- Miura F, 2009, Diabetes Japan, 52, 865
- Sano S, 2005, J Clin Lab Instr Reag, 28, 349
Cited by
7. Diabetes Technology: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s007 · Cited by 358
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
7. Diabetes Technology: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s007 · Cited by 348
7. Diabetes Technology: Standards of Care in Diabetes — 2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301
7. Diabetes Technology: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281
Metadata
- Authors
- Sai, S., Urata, M., & Ogawa, I.
- Affiliations
- PL Hospital (JP)
- Corresponding author
- Sai, Shizuki
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2019-01-19
- Volume
- 13
- Issue
- 4
- Pages
- 734-743
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC6610602
- OpenAlex
- W2910826564
- Semantic Scholar
- 1a7b6ee76dc15f6f8d4eb7785fd71d9ca3aa9f7e
- Cited by
- 20
- References
- 7
- Citations by year
- 2026: 2 · 2025: 3 · 2024: 5 · 2023: 4 · 2022: 1 · 2021: 3 · 2020: 1 · 2019: 1
- FWCI
- 1.03
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Patient Safety and Medication Errors
- Keywords
- self-monitoring of blood glucose, point-of-care testing, blood glucose meters, measurement linearity, interfering substances, vitamin C, hematocrit, error messages, diabetes mellitus
- MeSH terms
- Blood Glucose, Hematocrit, Humans, Blood Glucose Self-Monitoring, Point-of-Care Systems
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Evaluation of Linearity and Interference Effect on SMBG and POCT Devices, Showing Drastic High Values, Low Values, or Error Messages. https://app.cochat.ai/papers/10.1177/1932296818821664
- The paper
- Sai, S., Urata, M., & Ogawa, I. (2019). Evaluation of Linearity and Interference Effect on SMBG and POCT Devices, Showing Drastic High Values, Low Values, or Error Messages. Journal of Diabetes Science and Technology. https://doi.org/10.1177/1932296818821664
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.