Effect of Repeated Doses of Acetaminophen on a Continuous Glucose Monitoring System with Permselective Membrane
Open accessThe DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
Continuous glucose monitoring systems use an electrochemical methodology for measuring glucose in interstitial fluid (ISF). Dexcom G6 uses a measurement voltage where other common compounds may get oxidized. The diffusion membrane can block some of the interfering compounds; however, not all compounds can be blocked due to their molecular mass. Dexcom G4 and G5 systems showed significant interference from acetaminophen.1-3 Dexcom introduced a permselective membrane in G6 that blocks the diffusion of acetaminophen to the electrode surface thereby minimizing its effect.4 In the clinical study performed by Dexcom, the study participants were asked not to take any acetaminophen from the day before sensor insertion until the in-clinic session, where study subjects were asked to take 1 g of acetaminophen and were followed for 8 hours. The study’s objective was met in that the observed mean interference effect was 3.1 ± 4.8 mg/dL, which was statistically significantly lower than the performance goal of 10 mg/dL. The authors concluded that this interference effect would have a negligible effect on bolus insulin administered. The authors argued that because the temporal profile of interference with G4 sensors follows that of ISF and plasma acetaminophen concentrations,2 studies of higher or repeated doses were unnecessary. Since the interference from acetaminophen is reduced by using a permselective membrane, its effectiveness should be demonstrated by challenging the membrane with re…
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Hydroxyurea, used for myeloproliferative disorders and hematologic conditions, is one of the most recently identified interfering substances that cause a temporary increase in sensor glucose values discrepant from actual glucose values (129-134).”
From 7. Diabetes Technology: Standards of Care in Diabetes—2025 (2025) · cites this paper as [134]
Citations
Cites
Effect of Acetaminophen on CGM Glucose in an Outpatient Setting
Maahs, D. M., DeSalvo, D., Pyle, L., Ly, T., Messer, L., Clinton, P., Westfall, E., Wadwa, R. P., & Buckingham, B. (2015) · Diabetes Care · doi:10.2337/dc15-1096 · Cited by 102
Basu, A., Veettil, S., Dyer, R., Peyser, T., & Basu, R. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0410 · Cited by 93 · cited 2×
“518 Journal of Diabetes Science and Technology 15(2)”
Continuous Glucose Monitor Interference With Commonly Prescribed Medications: A Pilot Study
Basu, A., Slama, M. Q., Nicholson, W. T., Langman, L., Peyser, T., Carter, R., & Basu, R. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817697329 · Cited by 77
Resistance to Acetaminophen Interference in a Novel Continuous Glucose Monitoring System
Calhoun, P., Johnson, T. K., Hughes, J., Price, D., & Balo, A. K. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818755797 · Cited by 58
Cited by
Continuous glucose monitoring and metrics for clinical trials: an international consensus statement
Battelino, T., Alexander, C. M., Amiel, S. A., Arreaza-Rubin, G., Beck, R. W., Bergenstal, R. M., Buckingham, B. A., Carroll, J., Ceriello, A., Chow, E., et al. (2023) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(22)00319-9 · Cited by 831
7. Diabetes Technology: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s007 · Cited by 348
7. Diabetes Technology: Standards of Care in Diabetes — 2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301
7. Diabetes Technology: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281
Continuous glucose monitoring: The achievement of 100 years of innovation in diabetes technology
Galindo, R. J., & Aleppo, G. (2020) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2020.108502 · Cited by 133
Tauschmann, M., Cardona‐Hernandez, R., DeSalvo, D. J., Hood, K. K., Лаптев, Д. Н., Olinder, A. L., Wheeler, B. J., & Smart, C. E. (2025) · Hormone Research in Paediatrics · doi:10.1159/000543156 · Cited by 83
“New technologies such as machine learning combined with CGM could help T1D risk detection without hospital visits or laboratory tests [117].”
Avari, P., Lumb, A. N., Flanagan, D., Rayman, G. A., Misra, S., Dhatariya, K., & Choudhary, P. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221137338 · Cited by 76
Interferences with CGM systems: practical relevance?
Heinemann, L. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211065065 · Cited by 65
Landscape of Continuous Glucose Monitoring (CGM) and Integrated CGM: Accuracy Considerations
Bailey, T. S., & Alva, S. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0236 · Cited by 56
Accuracy and Potential Interferences of Continuous Glucose Monitoring Sensors in the Hospital
Bellido, V., Freckman, G., Pérez, A., & Galindo, R. J. (2023) · Endocrine Practice · doi:10.1016/j.eprac.2023.06.007 · Cited by 57
Consensus Report on the Use of Continuous Glucose Monitoring in Chronic Kidney Disease and Diabetes
Rhee, C. M., Gianchandani, R. Y., Kerr, D. E., Philis‐Tsimikas, A., Kovesdy, C. P., Stanton, R. C., Drincic, A., Galindo, R. J., Kalantar‐Zadeh, K., Neumiller, J. J., et al. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241292041 · Cited by 41
“127 Similarly, per Abbott Medical Affairs, FreeStyle Libre 2 Plus sensors have been modified to minimize interference by ascorbic acid.”
Cummings, C., Jiang, A., Sheehan, A., Ferraz‐Bannitz, R., Puleio, A., Simonson, D. C., Dreyfuss, J. M., & Patti, M. E. (2023) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.15096 · Cited by 27
“CGM readings can be affected by nonglycaemic factors, such as changes in blood flow and compression at sensor sites,(59) and interfering medications such as acetaminophen (for the Dexcom G4 device).(60) We did not require verification of sensor glucose values via capillary glucose monitoring, so it is possible that so…”
BackgroundIntegrating Continuous Glucose Monitoring Into Clinical Practices and Patients' Lives
Hirsch, I. B., & Miller, E. M. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0233 · Cited by 25
Ying, L. P., Yin, O. X., Quan, O. W., Jain, N., Mayuren, J., Pandey, M., Gorain, B., & Candasamy, M. (2024) · International Journal of Diabetes in Developing Countries · doi:10.1007/s13410-024-01349-x · Cited by 23
“Not only that, the administration of acetaminophen (> 1000 mg) will affect the Dexcom® CGM reading [70, 71].”
Pfützner, A., Jensch, H., Cardinal, C., Srikanthamoorthy, G., Riehn, E., & Thomé, N. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221095573 · Cited by 20
“11 At this initial protocol development stage, our approach was to explore the overall suitability of the designed in vitro setup to provide valid interference results rather than simulating daily clinical circumstances.”
BackgroundJensch, H., Setford, S. J., Thomé, N., Srikanthamoorthy, G., Weingärtner, L., Grady, M., Holt, E., & Pfützner, A. (2025) · Sensors · doi:10.3390/s25071985 · Cited by 11
Challenges of Wearable Biosensors and Ways to Overcome Them
Tarasov, S. E., Plekhanova, Y. V., Reshetilov, A., Melenkov, S., & Saltanov, I. (2026) · Biosensors · doi:10.3390/bios16030159 · Cited by 10
Continuous Glucose Monitoring Using the Dexcom G6 in Cardiac Surgery During the Postoperative Period
Insler, S. R., Wakefield, B. J., Debs, A., Brake, K., Nwosu, I., Isaacs, D. M., Bena, J. F., & Lansang, M. C. (2024) · Endocrine Practice · doi:10.1016/j.eprac.2024.04.015 · Cited by 10
Zelnick, L. R., Trikudanathan, S., Hall, Y. N., Ayers, E., Anderson, L., Ashford, N. K., Jones, E., Hoofnagle, A. N., de Boer, I. H., & Hirsch, I. B. (2025) · Diabetes Technology & Therapeutics · doi:10.1177/15209156251368934 · Cited by 7
“41,42 These issues are reported to have been resolved in the G6 system, 43,44 and the Dexcom G6 manufacture notice advises that acetaminophen interference is limited when drug exposure is 4 g/day.”
Phillips, S., Setford, S. J., Grady, M., Liu, Z., & Cameron, H. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211042352 · Cited by 7
Setford, S. J. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251377026 · Cited by 5
Holt, E., Nguyen, H., Bispham, J. A., Liu, J., CHAPMAN, K. S. M., & Grady, M. (2023) · Clinical Diabetes · doi:10.2337/cd23-0005 · Cited by 5
“An increasing number of interferents are being identi fi ed, to date including acetaminophen, salicylic acid, galactose, xy-lose, ascorbic acid, hydroxyurea, mannitol, tetracycline, ethanol, red wine, lisinopril, albuterol, and atenolol (18 – 26).”
BackgroundSetford, S. J. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251377008 · Cited by 4
Lamine, F., Pruijm, M. T., Bahon, V., & Zanchi, A. (2022) · Kidney and Dialysis · doi:10.3390/kidneydial2030042 · Cited by 3
Yoo, J. H., Moon, S. J., Lee, D. Y., Kwon, S. Y., Kim, K., Lee, J., Cho, Y., Kim, S. R., & Yun, J. (2026) · Journal of Korean Diabetes · doi:10.4093/jkd.2026.27.3.103 · Cited by 1
Vitebskaya, A. V., & Telegina, K. P. (2024) · Meditsinskiy sovet = Medical Council · doi:10.21518/ms2024-431 · Cited by 1
Dynamic Interference Testing Results with the Dexcom G6 Continuous Glucose Monitoring Device
Thomé, N., Jensch, H., Srikanthamoorthy, G., Weingärtner, L., Grady, M., Setford, S. J., Holt, E., & Pfützner, A. (2025) · Diabetes Technology and Obesity Medicine · doi:10.1089/dtom.2024.0009 · Cited by 1
“20 22 – 25 There is a need for cost-effective, large-scale, in vitro substance screening with CGM sensors for potential clinically relevant interfering substances and to effectively rule out noninterfering substances.”
Interfering substances with glucose monitoring systems and their clinical implications
Anelli, V. (2023) · The Journal of AMD · doi:10.36171/jamd23.26.3.3
Frontino, G., Rigamonti, A., & Riccardo, B. (2025) · Diabetes Obesity and Metabolism · doi:10.1111/dom.16459
Batir, D. M., Boev, M., Gurău, G., Matei, M. N., Nechita, A., & Maftei, N.-M. (2026) · IntechOpen eBooks · doi:10.5772/intechopen.1014475
Yoo, J. H., Moon, S. J., Lee, W., Kwon, S. Y., Kim, K., Lee, J., Cho, Y., Kim, S., & Yun, J. (2026) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2026.0923
Metadata
- Authors
- Denham, D.
- Affiliations
- Clinical Trials of Texas (US)
- Corresponding author
- Denham, Douglas
- ORCID
- Denham, Douglas
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2020-08-29
- Volume
- 15
- Issue
- 2
- Pages
- 517-518
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC8256057
- OpenAlex
- W3082759272
- Cited by
- 28
- References
- 4
- Citations by year
- 2026: 2 · 2025: 7 · 2024: 5 · 2023: 5 · 2022: 5 · 2021: 3 · 2020: 1
- FWCI
- 2.64
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- abbott diabetes care
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Physical Sciences › Chemistry › Analytical Chemistry › Analytical Chemistry and Sensors
- Keywords
- paracetamol, continuous glucose monitoring, permselective membranes, repeated dosing, drug interference, glucose sensor
- MeSH terms
- Acetaminophen, Blood Glucose, Diabetes Mellitus, Type 1, Humans, Blood Glucose Self-Monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.