Use and Accuracy of Inpatient CGM During the COVID-19 Pandemic: An Observational Study of General Medicine and ICU Patients
Open accessThe DOI matches its Crossref record; 24 of the 24 references it cites were found in a registry or an index.
Abstract
Background: Continuous glucose monitoring (CGM) is widely used in the outpatient setting for people with diabetes and has been limited to investigational use only for the inpatient population. In April 2020, the US FDA exercised enforcement discretion for the temporary use of inpatient CGM during the pandemic, thus hospitals were presented the opportunity to implement this technology. Methods: We sought to investigate the accuracy of CGM in hospitalized patients on general care floors and the intensive care unit (ICU) in attempts to decrease healthcare professional exposure to COVID-19 and ultimately improve glycemic management of patients affected by COVID-19. Point of care (POC) and laboratory (Lab) glucose values were matched with simultaneous CGM glucose values and measures of accuracy were performed to evaluate the safety and usability of CGM in this population. Our data are presented drawing a distinction between POC and Lab as reference glucose sources. Results: In 808 paired samples obtained from 28 patients (10 ICU, 18 general floor), overall mean absolute relative difference (MARD) for all patients using either POC or Lab as reference was 13.2%. When using POC as the reference glucose MARD was 13.9% and using Lab glucose as reference 10.9%. Using both POC and Lab reference glucose pairs the overall MARD for critical care patients was 12.1% and for general floor patients 14%. Conclusion: We determined, with proper protocols and safeguards in place, use of CGM in the hospitalized patient is a reasonable alternative to standard of care to achieve the goal of reducing healthcare professional exposure. Further study is necessary to validate safety, accuracy, and efficacy of this technology. Investigation and analysis are necessary for the development of protocols to utilize CGM trend arrows, alerts, and alarms.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Data on the safety and efficacy of real-time CGM use in the hospital, particularly with implementation of remote monitoring (e.g., a glucose telemetry system), has grown (47,50-52).”
From 16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2026 (2026) · cites this paper as [51]
In the citing paper
“Studies have been published assessing the effectiveness of this approach, which may ultimately lead to the approved use of CGM for monitoring hospitalized individuals (277,286-295).”
From 7. Diabetes Technology: Standards of Care in Diabetes—2025 (2025) · cites this paper as [289]
Citations
Cites
International Organization for Standardization
(2011) · SpringerReference · doi:10.1007/springerreference_16847 · Cited by 12,534
Understanding bland altman analysis
Giavarina, D. (2015) · Biochemia Medica · doi:10.11613/bm.2015.015 · Cited by 4,706
International consensus on use of continuous glucose monitoring
Danne, T., Nimri, R., Battelino, T., Bergenstal, R. M., Close, K. L., DeVries, J. H., Garg, S., Heinemann, L., Hirsch, I., Amiel, S. A., et al. (2017) · Diabetes Care · doi:10.2337/dc17-1600 · Cited by 2,069
Management of diabetes and hyperglycemia in hospitals.
Clement, S., Braithwaite, S. S., Magee, M. F., Ahmann, A., Smith, E. P., Schafer, R. G., Hirsch, I. B., & on behalf of the Diabetes in Hospitals Writing Committee (2004) · Diabetes Care · doi:10.2337/diacare.27.2.553 · Cited by 1,266
Hypoglycemia and clinical outcomes in patients with diabetes hospitalized in the general ward
Turchin, A., Matheny, M. E., Shubina, M., Scanlon, J. V., Greenwood, B., & Pendergrass, M. L. (2009) · Diabetes Care · doi:10.2337/dc08-2127 · Cited by 453 · cited 4×
“1140 Journal of Diabetes Science and Technology 16(5)”
BackgroundInfluentialTime Lag of Glucose From Intravascular to Interstitial Compartment in Humans
Basu, A., Dube, S., Slama, M., Errazuriz, I., Amezcua, J. C., Kudva, Y. C., Peyser, T., Carter, R. E., Cobelli, C., & Basu, R. (2013) · Diabetes · doi:10.2337/db13-1132 · Cited by 205
Prone Positioning in Awake, Nonintubated Patients With COVID-19 Hypoxemic Respiratory Failure
Thompson, A. E., Ranard, B. L., Wei, Y., & Jelic, S. (2020) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2020.3030 · Cited by 203
Measures of Accuracy for Continuous Glucose Monitoring and Blood Glucose Monitoring Devices
Freckmann, G., Pleus, S., Grady, M., Setford, S., & Levy, B. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818812062 · Cited by 202
Shah, V. N., Laffel, L. M., Wadwa, R. P., & Garg, S. K. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0143 · Cited by 197
Galindo, R. J., Migdal, A. L., Davis, G. M., Urrutia, M. A., Albury, B., Zambrano, C., Vellanki, P., Pasquel, F. J., Fayfman, M., Peng, L., et al. (2020) · Diabetes Care · doi:10.2337/dc19-2073 · Cited by 181
Research works printed but not linked: 2
- Guidance Document, 2005, Fed Regist, 1988, 1
- 10.1177/1932296814539590 · DOI · Paper page
Cited by
7. Diabetes Technology: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s007 · Cited by 348
Self-citation7. Diabetes Technology: Standards of Care in Diabetes — 2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301
Self-citation7. Diabetes Technology: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281
Self-citation16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s016 · Cited by 221
Self-citation16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s016 · Cited by 195
Self-citation16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Galindo, R. J., Hilliard, M. E., Johnson, E. L., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s016 · Cited by 167
Self-citationMansour, M., Darweesh, M. S., & Soltan, A. (2024) · Alexandria Engineering Journal · doi:10.1016/j.aej.2024.01.021 · Cited by 193
16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2026
American Diabetes Association Professional Practice Committee for Diabetes*, Bajaj, M., McCoy, R. G., Balapattabi, K., Bannuru, R. R., Bellini, N. J., Bennett, A. K., Beverly, E. A., Briggs Early, K., ChallaSivaKanaka, S., et al. (2026) · Diabetes Care · doi:10.2337/dc26-s016 · Cited by 80
A Consensus Statement for Continuous Glucose Monitoring Metrics for Inpatient Clinical Trials
Spanakis, E. K., Cook, C. B., Kulasa, K. M., Aloi, J., Bally, L., Davis, G. M., Dungan, K. M., Galindo, R. J., Mendez, C. E., Pasquel, F. J., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231191104 · Cited by 93
Shaw, J. L. V., Bannuru, R. R., Beach, L. A., ElSayed, N. A., Freckmann, G., Füzéry, A. K., Fung, A. W. S., Gilbert, J. D., Yun, H., Korpi-Steiner, N. L., et al. (2024) · Diabetes Care · doi:10.2337/dci24-0073 · Cited by 78
Self-citationAvari, P., Lumb, A. N., Flanagan, D., Rayman, G. A., Misra, S., Dhatariya, K., & Choudhary, P. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221137338 · Cited by 76
Accuracy and Potential Interferences of Continuous Glucose Monitoring Sensors in the Hospital
Bellido, V., Freckman, G., Pérez, A., & Galindo, R. J. (2023) · Endocrine Practice · doi:10.1016/j.eprac.2023.06.007 · Cited by 57
Espinoza, J., Xu, N. Y., Nguyen, K. T., & Klonoff, D. C. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211058148 · Cited by 46
Finn, E., Schlichting, L., Grau, L., Douglas, I. S., & Pereira, R. I. (2023) · Diabetes Care · doi:10.2337/dc23-0089 · Cited by 45
“Diabetes Care 2023;46(10):1825–1830 | https://doi.”
BackgroundResultInfluentialWearable devices for patient monitoring in the intensive care unit
Angelucci, A., Greco, M., Cecconi, M., & Aliverti, A. (2025) · Intensive Care Medicine Experimental · doi:10.1186/s40635-025-00738-8 · Cited by 60
Hagerf, B. V., Protuš, M., Németová, L., Mráz, M., Kieslichová, E., Uchytilová, E., Indrova, V., Lelito, J., Girman, P., Haluzík, M., et al. (2024) · Diabetes Care · doi:10.2337/dc23-1663 · Cited by 49
“(18) in more stable physiological conditions in the outpatient setting and in studies performed perioperatively (16,17) and slightly more favorable than some of the reports of the use of CGM during the COVID-19 pandemic (9,10,13,14).”
BackgroundResultBellido, V., Aguilera, E., Cardona‐Hernandez, R., Díaz-Soto, G., de Villar, N. G. P., César, M. J. P., & Ampudia‐Blasco, F. J. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221088601 · Cited by 47
Accuracy of continuous glucose monitoring systems in intensive care unit patients: a scoping review
Nielsen, C. G., Grigonyte‐Daraskeviciene, M., Olsen, M. T., Møller, M. H., Nørgaard, K., Perner, A., Mårtensson, J., Pedersen‐Bjergaard, U., Kristensen, P. L., & Bestle, M. H. (2024) · Intensive Care Medicine · doi:10.1007/s00134-024-07663-6 · Cited by 47
“Following the temporary FDA approval of the use of CGM systems during the COVID-19 pandemic, several clinical studies using a hybrid protocol approach were conducted [19, 44, 48, 52, 125, 126].”
InfluentialZelada, H., Perez‐Guzman, M. C., Cherñavvsky, D. R., & Galindo, R. J. (2023) · Endocrine Connections · doi:10.1530/ec-23-0180 · Cited by 37
Lee, M. Y., Seav, S. M., Ongwela, L., Lee, J. J., Aubyrn, R., Cao, F. Y., Kalinsky, A., Aparicio Ramos, O., Gu, Y., Kingston, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-0626 · Cited by 32
Continuous Glucose Monitoring in the Intensive Care Unit
ARROYO, L. I. G., Faulds, E. R., Perez‐Guzman, M. C., Davis, G. M., Dungan, K. M., & Pasquel, F. J. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231169522 · Cited by 36
BOEDER, S. C., Kobayashi, E., Ramesh, G., Serences, B., Kulasa, K. M., & Majithia, A. R. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221113865 · Cited by 31
Implementation of Continuous Glucose Monitoring in Critical Care: A Scoping Review
Faulds, E. R., Dungan, K. M., & McNett, M. (2023) · Current Diabetes Reports · doi:10.1007/s11892-023-01503-5 · Cited by 30
COVID-19 and Diabetes Outcomes: Rationale for and Updates from the CORONADO Study
Smati, S., Tramunt, B., Wargny, M., Gourdy, P., Hadjadj, S., & Cariou, B. (2022) · Current Diabetes Reports · doi:10.1007/s11892-022-01452-5 · Cited by 32
Nursing Perspectives on the Use of Continuous Glucose Monitoring in the Intensive Care Unit
Faulds, E. R., Dungan, K. M., McNett, M., JONES, L. G., Poindexter, N., Exline, M. C., Pattison, J., & Pasquel, F. J. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231170616 · Cited by 28
Friman, O., Soltani, N., Lind, M., Zetterqvist, P., Balintescu, A., Perner, A., Oldner, A., Rooyackers, O., & Mårtensson, J. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0035 · Cited by 25
Continuous Glucose Monitoring in Older Adults: What We Know and What We Have Yet to Learn
Price, C. E., Callahan, K. E., Aloi, J., & Usoh, C. O. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241234651 · Cited by 25
“…monitoring with CGM proved to be exceptionally useful in the inpatient setting for patients with COVID-19 and hyperglycemia, as this technology aided in targeting adequate glycemic control under strict isolation precautions while reducing healthcare professional exposure for checking FSBG. 13,14”
MethodPerioperative use and accuracy of continuous glucose monitoring: A systematic review
Carlier, L., de Ponthaud, C., Jacqueminet, S., Phan, F., Gaujoux, S., & Amouyal, C. (2025) · Diabetes Obesity and Metabolism · doi:10.1111/dom.16583 · Cited by 31
“18,62 Usually, MARD is considered accurate when 10% for ambulatory use, and new sensors meet this performance.”
Multi-Horizon Glucose Prediction Across Populations With Deep Domain Generalization
Zhu, T., Afentakis, I., Li, K., Armiger, R., Hill, N. E., Oliver, N. S., & Georgiou, P. (2024) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2024.3428921 · Cited by 26
“…T1D care; it has increasingly been adopted across a diverse range of clinical scenarios,includingitsuseinhospitalsettings[8],[9],especially duringtheCOVID-19pandemic[10],[11],andamongindivid-ualswithoutdiabetes[12],suchasaidingindietarystrategies aimedatamelioratingmetaboliccomplications[13].”
Bann, S. A., Hercus, J. C., Atkins, P., Alkhairy, A., Loyal, J. P., Sekhon, M., & Thompson, D. J. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0074 · Cited by 21
“35 We demonstrated that by using a simple calibration protocol with sensor validation at 2, 12, and 24 h post-insertion, the accuracy of the G6 was signi fi cantly improved from an MARD of 13.19% (5.11, 19.03) for uncalibrated participants to an MARD of 9.42% (3.03, 13.33) for calibrated…”
InfluentialWang, Y., Li, S., Lu, J., Feng, K., Huang, X., Hu, F., Sun, M., Zou, Y., Li, Y., Huang, W., et al. (2024) · Diabetologia · doi:10.1007/s00125-024-06136-1 · Cited by 23
“Furthermore, the use of CGM in hospitals has been accelerated due to the COVID-19 pandemic [9–11].”
Dumitrascu, A. G., Perry, M. F., Boone, R. J., Guzman, M. P., Chirilă, R. M., McNally, A. W., Colibaseanu, D., Meek, S., Ball, C. T., White, L. J., et al. (2022) · Endocrine Practice · doi:10.1016/j.eprac.2022.12.012 · Cited by 15
Continuous glucose monitoring (CGM) in a non-Icu hospital setting: The patient's journey
Irace, C., Coluzzi, S., Cianni, G. D., Forte, E., Landi, F., Rizzo, M. R., Sesti, G., Succurro, E., & Consoli, A. (2023) · Nutrition Metabolism and Cardiovascular Diseases · doi:10.1016/j.numecd.2023.06.021 · Cited by 19
Longo, R. R., & Joshi, R. (2022) · Diabetes Spectrum · doi:10.2337/dsi22-0009 · Cited by 20
Self-citationAccuracy of a Continuous Glucose Monitor During Pediatric Type 1 Diabetes Inpatient Admissions
Cobry, E. C., Pyle, L., WATERMAN, L. A., Forlenza, G. P., Towers, L., KARAMI, A. J., Jost, E., Berget, C., & Wadwa, R. P. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0375 · Cited by 15
Cardio‐renal‐metabolic disease in primary care setting
Ibrahim, M. A., Ba-Essa, E. M., Baker, J. C., Cahn, A., Ceriello, A., Cosentino, F., Davies, M. J., Eckel, R. H., Gaal, L. V., Gæde, P., et al. (2023) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.3755 · Cited by 14
Accuracy of a Real-Time Continuous Glucose Monitor in Pediatric Diabetic Ketoacidosis Admissions
WATERMAN, L. A., Pyle, L., Forlenza, G. P., Towers, L., KARAMI, A. J., Jost, E., Berget, C., Wadwa, R. P., & Cobry, E. C. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0542 · Cited by 16
Kalogeropoulou, M., Thomson, L., & Beardsall, K. (2022) · Archives of Disease in Childhood Fetal & Neonatal · doi:10.1136/archdischild-2022-324593 · Cited by 11
Shang, J., Yuan, Z., Zhang, Z., Zhou, Q., Zou, Y., & Wang, W. (2025) · Journal of Medical Internet Research · doi:10.2196/67012 · Cited by 12
“Several reports indicate that diabetes and hyperglycemia are linked to worse morbidity and mortality in individuals with COVID-19 [9,10].”
Kim, Y. C., Wolf, R. M., Busin, K., & Vanderhoek, S. M. (2023) · Pediatric Anesthesia · doi:10.1111/pan.14732 · Cited by 9
Dexcom G7 Accuracy and Reproducibility in the Intensive Care Unit
Baran, J. D., Spanbauer, C., Bais, R., Chiang, H.-H., Chao, J. H., Khakpour, D., Panfil, P., Pasquel, F. J., Ullal, J., Gligorijevic, N., et al. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2025.0184 · Cited by 10
“…in perspective, the mean APACHE II score for the surgical critically ill patients studied in the 2001 study from Van den Berghe et al. was 9, 24 whereas the mean score from the Normoglycemia in Intensive Care Evaluation – Survival Using Glucose Algorithm Regulation (NICE-SUGAR) study…”
Faulds, E. R., Hester, J. C., Badakhshi, Y., Miller, J. D., Basil, R. C., Chandra, S., Chang, A. S., Garcia, M., JONES, L. G., Pintor-Jimenez, K. A., et al. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251343108 · Cited by 11
“3,9,11 We pooled data from three academic institutions that used CGM during the COVID-19 pandemic with the aim of examining accuracy of CGM in the critical care environment and evaluating CGM performance during periods of hemodynamic instability.”
Ullal, J., Spanbauer, C., Baran, J. D., Bais, R., Chiang, H.-H., Chao, J. H., Khakpour, D., Panfil, P., Pasquel, F. J., Gligorijevic, N., et al. (2025) · Diabetes Technology & Therapeutics · doi:10.1177/15209156251368933 · Cited by 8
Prince, N., Ramsay, T., Shorr, R., RABASA-LHORET, R. P. R., & Sun, C. (2026) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251412482 · Cited by 7
“For both Longo et al 30 and Wang et al 19 studies, the proportion of the hypoglycemic range CGM-CBG or CGM-LabBG pairs that occurred in the PLWD admitted to non-critical care wards was not reported.”
InfluentialHa, E. Y., Chung, S. M., Park, I. R., Lee, Y. Y., Choi, E. Y., & Moon, J. S. (2022) · Frontiers in Endocrinology · doi:10.3389/fendo.2022.869451 · Cited by 9
Real-World Performance of Personal Continuous Glucose Monitors During Hospitalization
Dumitrascu, A. G., Perry, M. C., Chindris, A.-M., Chirilă, R. M., Boone, R. J., Blumenfeld, S. G., Hodge, D. O., White, L. J., Gnanapandithan, K., Cruz, D. A. D. L., et al. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0639 · Cited by 8
“12 In contrast we analyzed CGM-POC pairs from all times of the day, including periods of rapid 22 Finn, 19 and Longo 23 found this value to be 13.2%, 11.4%, and 10.9%, respectively.”
InfluentialManagement of Hyperglycemia in Hospitalized Patients
Gianchandani, R. Y., Wei, M., & Demidowich, A. P. (2024) · Annals of Internal Medicine · doi:10.7326/annals-24-02754 · Cited by 12
Wang, R., Kyi, M., Krishnamoorthi, B., Tjahyadi, J., Connell, A., Chiang, C. Y., Renouf, D., Barmanray, R., & Fourlanos, S. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0604 · Cited by 11
Rayan, M. N., Faulds, E. R., Lee, B., McNett, M., Exline, M. C., Lin, C. J., JONES, L. G., Rasul, A., & Dungan, K. M. (2025) · CHEST Critical Care · doi:10.1016/j.chstcc.2025.100193 · Cited by 5
Agarwal, S., Demidowich, A. P., Soliman, D., Umpierrez, G. E., & Galindo, R. J. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251364276 · Cited by 7
Supporting the Use of a Person’s Own Diabetes Technology in the Inpatient Setting
Pattison, J., Dungan, K. M., & Faulds, E. R. (2022) · Diabetes Spectrum · doi:10.2337/dsi22-0011 · Cited by 5
Monitoring Devices for glucose in critical patients during the COVID-19 pandemic
Xue, X., Wang, Y., Wang, J., Zhao, J., Tang, Y., Song, X., & Xu, C. (2022) · International Healthcare Review (online) · doi:10.56226/ihr.v1i2.29 · Cited by 5
Endocrinology in the Time of COVID-19: A Rapid Evolution of Knowledge and Care
Rizvi, A. A., Stoian, A. P., Lessan, N. G., & Rizzo, M. (2021) · Medicina · doi:10.3390/medicina57080805 · Cited by 5
“However, an unintended and rather salubrious consequence of this action was to utilize CGM for full therapeutic use, including initiation in the inpatient setting and transitioning to discharge and outpatient use [21].”
BackgroundMensah, I. K., Azimi, V., & Farnsworth, C. W. (2025) · Clinical Chemistry · doi:10.1093/clinchem/hvaf080 · Cited by 4
Müssig, K., Gallwitz, B., Haak, T., Kellerer, M., & Siegel, E. G. (2025) · Diabetologie und Stoffwechsel · doi:10.1055/a-2592-9700 · Cited by 7
Shimizu, N., Hayashi, A., Ito, S., Suzuki, A., Fujishima, R., Matoba, K., Wada, T., Takano, K., Katagiri, M., & Shichiri, M. (2021) · Endocrine Journal · doi:10.1507/endocrj.ej21-0540 · Cited by 4
Volčanšek, Š., Janež, A., & Srpčič, M. (2026) · Diabetology · doi:10.3390/diabetology7010006 · Cited by 3
“Integration with Electronic Health Record (EHR) and protocols was reported as achievable, with remote monitoring also minimizing personnel exposure in infectious disease settings [7,28,131–133].”
Müssig, K., Gallwitz, B., Haak, T., Kellerer, M., & Siegel, E. G. (2024) · Diabetologie und Stoffwechsel · doi:10.1055/a-2251-2772 · Cited by 3
Müssig, K., Gallwitz, B., Haak, T., Kellerer, M., & Siegel, E. G. (2025) · Diabetes aktuell · doi:10.1055/a-2500-1766 · Cited by 5
Comparing glucose monitoring methods: efficiency insights in a simulated hospital setting
Davasgaium, A., Robbins, T., Leca, B. M., Epure, A., Sankar, S., & Randeva, H. S. (2025) · Frontiers in Clinical Diabetes and Healthcare · doi:10.3389/fcdhc.2025.1517161 · Cited by 3
Müssig, K., Gallwitz, B., Haak, T., Kellerer, M., & Siegel, E. G. (2024) · Diabetologie und Stoffwechsel · doi:10.1055/a-2312-1160 · Cited by 3
Hospital Diabetes Meeting 2022
Huang, J. K., Yeung, A. M., Nguyen, K. T., Xu, N. Y., Preiser, J., Rushakoff, R. J., Seley, J. J., Umpierrez, G. E., Wallia, A., Drincic, A., et al. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221110878 · Cited by 4
Self-citationLiu, Y., Zhang, J., Chun, X., Gao, Y., Yao, R., Liang, Y., Zhu, L., He, Y., & Huang, W. (2024) · Endocrine Practice · doi:10.1016/j.eprac.2024.06.004 · Cited by 3
Pañero‐Moreno, M., Muxella-Roson, M., Dubè-Llobet, G., Cabrerizo-Jimenez, M., Guix‐Comellas, E. M., Ordozgoiti, A. V., & Marsà-Perez, N. (2025) · Enfermería Intensiva (English ed ) · doi:10.1016/j.enfi.2025.500534 · Cited by 5
Protuš, M., Hagerf, B. V., Jabor, A., Franeková, J., Németová, L., Uchytilová, E., Indrova, V., Beckova, J., Macek, A., Dolečková, M., et al. (2026) · Therapeutic Advances in Endocrinology and Metabolism · doi:10.1177/20420188251405372 · Cited by 2
“…of perioperative CGM use, 25–27 as well as reports evaluating exceptional applications of CGM in intensive care during the COVID-19 pandemic, 19–24 demonstrate that the accuracy of contemporary CGM systems has improved and that they might be suitable for broader use beyond their originally…”
Yeung, A. M., Dirisanala, S., Abraham, A., Huang, J. K., Brennan, G. M., Urrutia, M. A., Baran, J. D., Nguyen, K. T., Xu, N. Y., Shang, T., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231184448 · Cited by 2
Type 1 Diabetes Management in the Hospital Setting
Granados, A., & Carrillo‐Iregui, A. (2024) · Pediatrics in Review · doi:10.1542/pir.2022-005645 · Cited by 3
Müssig, K., Gallwitz, B., Haak, T., Kellerer, M., & Siegel, E. G. (2025) · Die Diabetologie · doi:10.1007/s11428-024-01276-9 · Cited by 1
Logothetis, G., Avramidis, K., Konstantaki, E., Matziou, V. N., & Doupis, J. (2025) · touchREVIEWS in Endocrinology · doi:10.17925/ee.2025.21.1.7 · Cited by 1
“27 Only one study reported the systematic bias, which illustrates the agreement between rtCGM-POC pairs, while Sadhu et al. used the Bland–Altman plot for their analysis.”
InfluentialRamesh, G., Kobayashi, E., Sharma, N., Majithia, A. R., Kulasa, K. M., & BOEDER, S. C. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251338865 · Cited by 2
Müssig, K., Gallwitz, B., Haak, T., Kellerer, M., & Siegel, E. G. (2025) · Experimental and Clinical Endocrinology & Diabetes · doi:10.1055/a-2490-5208 · Cited by 1
Perioperatives Management von Menschen mit Diabetes mellitus: CGM im Krankenhaus
Reger-Tan, S., Lee-Barkey, Y. H., Stratmann, B., Klose, J. M., & Heinemann, L. (2026) · Die Diabetologie · doi:10.1007/s11428-025-01409-8 · Cited by 1
Cho, S., Vigers, T., Pyle, L. L., Franklin, A. R. K., Sopfe, J., Jeney, F., & Forlenza, G. P. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0246 · Cited by 1
Pañero‐Moreno, M., Guix‐Comellas, E. M., & Ordozgoiti, A. V. (2024) · Nursing in Critical Care · doi:10.1111/nicc.13198 · Cited by 4
Breakthrough technology for in-hospital glucose monitoring
Kerr, D. E., & Klonoff, D. C. (2022) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(22)00104-8
Application of the New “Points in Range” Metrics in the Assessment of In-Hospital Glycemic Control
Rosinha, P. M. O., Teixeira, S., Amaral, C., & Cardoso, M. H. (2023) · Cureus · doi:10.7759/cureus.34278 · Cited by 1
“After April 2020, when the Food and Drug Administration (FDA) made the use of CGM available in hospitalized COVID-19 patients, several studies sought to assess its accuracy and safety in inpatient settings with favorable results, although there is still not enough evidence to prove the cost-benefit of its more indiscr…”
BackgroundThe Use of Diabetes Technology in Persons with Diabetes and Coronavirus 2019
Szmuilowicz, E. D., & Aleppo, G. (2023) · Contemporary Endocrinology · doi:10.1007/978-3-031-28536-3_11
NUOLATINIŲ GLIUKOZĖS JUTIKLIŲ PRITAIKYMAS SERGANTIEMS NE TIK CUKRINIU DIABETU
Bakutis, J. (2023) · Health Sciences · doi:10.35988/sm-hs.2023.181
Faulds, E. R., Lee, B., Rasul, A., JONES, L. G., McNett, M., Exline, M. C., Shoben, A. B., Lin, C. J., Madan, R., & Dungan, K. M. (2024) · Research Square · doi:10.21203/rs.3.rs-4243392/v1
Continuous Glucose Monitoring in Acutely and Critically Ill Patients
Pal, A. D. (2024) · Journal of PeriAnesthesia Nursing · doi:10.1016/j.jopan.2024.05.019
Tan, S. H. E., Qi, P. G. J., Yeoh, E. C. K., & Liu, A. Y. L. (2026) · Journal of Diabetes and its Complications · doi:10.1016/j.jdiacomp.2026.109267 · Cited by 1
González-Aguirre, J. E., Medina-Ramirez, H. A., Soto-Monciváis, B., Sanchez-Ruiz, S. R., Garcia-Arroyo, C., Moreno-Hoyos-Abril, J. F., & Náñez-Terreros, H. (2026) · Archives of Medical Research · doi:10.1016/j.arcmed.2026.103382
Müssig, K., Gallwitz, B., Haak, T., Kellerer, M., & Siegel, E. G. (2026) · Die Diabetologie · doi:10.1007/s11428-026-01424-3
Accuracy of a Continuous Glucose Monitor in the Intensive Care Unit
Cucchi, E. W., Ali, A. H., Guilarte-Walker, Y., BARTON, B. A., Malkani, S., & Mordes, J. (2026) · Diabetes Technology & Therapeutics · doi:10.1177/15209156261423933 · Cited by 5
“17 A retrospective single-center analysis of glucose concentrations in critically ill patients during the COVID-19 pandemic using the Dexcom G6 (cid:1) reported a MARD of 16.3% when compared with POC-FSBG and 12.1% when compared with laboratory plasma glucose concentrations.”
InfluentialHemodynamic Monitoring in the Cardiovascular Intensive Care Unit
Hicks, M. H., Acharya, S. K., Rajkumar, K. P., Gaillard, J. P., Crimi, E., & Khanna, A. K. (2026) · Cardiology Clinics · doi:10.1016/j.ccl.2026.05.002
Jiang, Z., Zhang, Y., Dong, Z., Wang, C., & Wang, J. (2026) · Frontiers in Endocrinology · doi:10.3389/fendo.2026.1921773
“Several mechanisms could plausibly explain an association between CGM-guided management and clinical outcomes, but none was tested as a mediator in this study (20, 21).”
Ruiz, J., Buros‐Stein, A., Wheelock, A., Stewart, B., Bernique, A. S., Jost, E., Garvey, K. C., Astley, C. M., & Cobry, E. C. (2026) · Diabetes Technology & Therapeutics · doi:10.1177/15209156261486139
Diabetes im Krankenhaus – Praxisempfehlung der DDG
Müssig, K., Gallwitz, B., Haak, T., Kellerer, M., & Siegel, E. G. (2026) · Diabetologie und Stoffwechsel · doi:10.1055/a-2884-3481
“…Patienten mit einem Diabetes unter Insulin-therapie eine bestehende Therapie mit Insulinpen, Insulinpumpe oder AID (automated insulin delivery)-System fortgeführt werden, sofern eine sichere Durchführung bei ausgeglichener Stoff-wechsellage gewährleistet und dokumentiert ist [25, 26, 27, 28, 29].”
Metadata
- Authors
- Longo, R. R., Elias, H., Khan, M., & Seley, J. J.
- Affiliations
- Lahey Hospital and Medical Center (US)
- Weill Cornell Medicine, New York, NY, USA
- Corresponding author
- Longo, Rebecca Rick
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2021-05-10
- Volume
- 16
- Issue
- 5
- Pages
- 1136-1143
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC9445343
- OpenAlex
- W3160205675
- Cited by
- 89
- References
- 24
- Citations by year
- 2026: 12 · 2025: 21 · 2024: 22 · 2023: 16 · 2022: 15 · 2021: 3
- FWCI
- 10.45
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Keywords
- continuous glucose monitoring, COVID-19 pandemic, glycemic control, CGM accuracy, intensive care unit, medical inpatients, healthcare worker exposure, mean absolute relative difference
- MeSH terms
- COVID-19, Blood Glucose, Critical Care, Humans, Inpatients, Intensive Care Units, Blood Glucose Self-Monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.