Performance of a factory-calibrated real-time continuous glucose monitoring system utilizing an automated sensor applicator

Open access
Shah, Viral N., Laffel, Lori M., Wadwa, R. Paul, Garg, Satish K.·2018·Diabetes Technology & Therapeutics

Abstract

Background This study assessed the accuracy of a factory-calibrated 10-day real-time continuous glucose monitoring (CGM) system (G6), which includes an automated sensor applicator. Methods Seventy-six participants with insulin-treated diabetes were enrolled at four U.S. sites as part of a larger study of G6 system performance. In-clinic visits for frequent comparative blood glucose measurements using a reference instrument (YSI) were conducted on days 1, 4-5, 7, and/or 10 of system use. Accuracy evaluation included the proportion of CGM values that were within ±20% of YSI reference value for glucose levels >100 mg/dL and ±20 mg/dL for YSI glucose levels ≤100 mg/dL (%20/20), the analogous %15/15 and %30/30, and the mean absolute relative difference (MARD) between temporally matched CGM and YSI values. Participants calibrated the systems once daily. Raw sensor data were reprocessed using assigned sensor codes and a factory-calibration algorithm. Results Reprocessed data from 62 participants (25 adults and 37 children and adolescents of ages 6-17 years; 3532 YSI-CGM pairs) were analyzed. The G6 system's overall %20/20 was 93.9% (adults, 92.5%; children and adolescents, 96.2%), its %15/15 was 83.3% (adults, 78.3%; children and adolescents, 91.1%), and its MARD was 9.0% (adults, 9.8%; children and adolescents, 7.7%). Overall day-1 %20/20 accuracy was 92.2%, %15/15 was 81.5%, and MARD was 9.3%. Accuracy was maintained across 10 days of use and various glucose concentration ranges i…

Integrity signals

Registry

Crossref · 10.1089/dia.2018.0143

Cites

7 of 7 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

We couldn’t retrieve this paper’s full text, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Shah, V. N., Laffel, L. M., Wadwa, R. P., & Garg, S. K.
Affiliations
  • University of Colorado Anschutz (US)
  • Joslin Diabetes Center (US)
Corresponding author
  • Laffel, Lori M.
  • Garg, Satish K.
Venue
Diabetes Technology & Therapeutics
Published
2018-06
Volume
20
Issue
6
Pages
428-433
ISSN
  • 1520-9156 (ISSN-L)
  • 1557-8593
OpenAlex
W2809055900
Cited by
197
References
7
Citations by year
2026: 20 · 2025: 19 · 2024: 24 · 2023: 27 · 2022: 22 · 2021: 26 · 2020: 24 · 2019: 27 · 2018: 8
FWCI
16.90
Type
journal article
Language
en
Open access
Locations
  • Diabetes Technology & Therapeutics (publishedVersion)
  • PubMed Central (publishedVersion)
  • PubMed (publishedVersion)
Funders
  • Dexcom
  • National Institute of Diabetes and Digestive and Kidney Diseases (P30DK036836)
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
Keywords
continuous glucose monitoring, factory calibration, Dexcom G6, sensor accuracy, mean absolute relative difference, insulin-treated diabetes, children and adolescents
MeSH terms
Adolescent, Adult, Blood Glucose, Calibration, Child, Computer Systems, Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2, Female, Humans, Male, Middle Aged, Reference Values, Blood Glucose Self-Monitoring, Reproducibility of Results, Young Adult

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review