Performance of a factory-calibrated real-time continuous glucose monitoring system utilizing an automated sensor applicator
Open accessAbstract
Background This study assessed the accuracy of a factory-calibrated 10-day real-time continuous glucose monitoring (CGM) system (G6), which includes an automated sensor applicator. Methods Seventy-six participants with insulin-treated diabetes were enrolled at four U.S. sites as part of a larger study of G6 system performance. In-clinic visits for frequent comparative blood glucose measurements using a reference instrument (YSI) were conducted on days 1, 4-5, 7, and/or 10 of system use. Accuracy evaluation included the proportion of CGM values that were within ±20% of YSI reference value for glucose levels >100 mg/dL and ±20 mg/dL for YSI glucose levels ≤100 mg/dL (%20/20), the analogous %15/15 and %30/30, and the mean absolute relative difference (MARD) between temporally matched CGM and YSI values. Participants calibrated the systems once daily. Raw sensor data were reprocessed using assigned sensor codes and a factory-calibration algorithm. Results Reprocessed data from 62 participants (25 adults and 37 children and adolescents of ages 6-17 years; 3532 YSI-CGM pairs) were analyzed. The G6 system's overall %20/20 was 93.9% (adults, 92.5%; children and adolescents, 96.2%), its %15/15 was 83.3% (adults, 78.3%; children and adolescents, 91.1%), and its MARD was 9.0% (adults, 9.8%; children and adolescents, 7.7%). Overall day-1 %20/20 accuracy was 92.2%, %15/15 was 81.5%, and MARD was 9.3%. Accuracy was maintained across 10 days of use and various glucose concentration ranges i…
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The reported mean absolute relative difference (MARD) for the Dexcom G6 in accuracy studies comparing sensor glucose values with reference blood glucose values is 9.0%.
“The reported mean absolute relative difference (MARD) for the Dexcom G6 in accuracy studies comparing sensor glucose values with reference blood glucose values is 9.0% (9).”
From Continuous Glucose Monitoring Profiles in Healthy Nondiabetic Participants: A Multicenter Prospective Study (2019) · cites this paper as [9]
Cites
Continuous glucose monitoring and intensive treatment of type 1 diabetes
(2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805017 · Cited by 1,559 · see citing works
Beck, R. W., Riddlesworth, T., Ruedy, K., Ahmann, A., Bergenstal, R., Haller, S., Kollman, C., Kruger, D., McGill, J. B., Polonsky, W., et al. (2017) · JAMA · doi:10.1001/jama.2016.19975 · Cited by 1,185 · see citing works
Beck, R. W., Riddlesworth, T. D., Ruedy, K., Ahmann, A., Haller, S., Kruger, D., McGill, J. B., Polonsky, W., Price, D., Aronoff, S., et al. (2017) · Annals of Internal Medicine · doi:10.7326/m16-2855 · Cited by 635 · see citing works
Polonsky, W. H., Hessler, D., Ruedy, K. J., & Beck, R. W. (2017) · Diabetes Care · doi:10.2337/dc17-0133 · Cited by 346 · see citing works
Reddy, M., Jugnee, N., El Laboudi, A., Spanudakis, E., Anantharaja, S., & Oliver, N. (2018) · Diabetic Medicine · doi:10.1111/dme.13561 · Cited by 245 · see citing works
Continuous glucose monitoring sensors: past, present and future algorithmic challenges
Facchinetti, A. (2016) · Sensors · doi:10.3390/s16122093 · Cited by 176 · see citing works
Šoupal, J., Petruželková, L., Flekač, M., Pelcl, T., Matoulek, M., Daňková, M., Škrha, J., Svačina, Š., & Prázný, M. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0171 · Cited by 126 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Continuous glucose monitoring and metrics for clinical trials: an international consensus statement
Battelino, T., Alexander, C. M., Amiel, S. A., Arreaza-Rubin, G., Beck, R. W., Bergenstal, R. M., Buckingham, B. A., Carroll, J., Ceriello, A., Chow, E., et al. (2023) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(22)00319-9 · Cited by 831 · see citing works
Self-citationShah, V. N., DuBose, S. N., Li, Z., Beck, R. W., Peters, A. L., Weinstock, R. S., Kruger, D., Tansey, M., Sparling, D., Woerner, S., et al. (2019) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2018-02763 · Cited by 315 · see citing works
Self-citationAdvances in technology for management of type 1 diabetes
Beck, R. W., Bergenstal, R. M., Laffel, L. M., & Pickup, J. C. (2019) · The Lancet · doi:10.1016/s0140-6736(19)31142-0 · Cited by 242 · see citing works
Self-citationAccuracy and Safety of Dexcom G7 Continuous Glucose Monitoring in Adults with Diabetes
Garg, S. K., Kipnes, M., Castorino, K., Bailey, T. S., Akturk, H. K., Welsh, J. B., Christiansen, M. P., Balo, A. K., Brown, S. A., Reid, J. L., et al. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0011 · Cited by 179 · see citing works
Self-citationAlva, S., Bailey, T., Brazg, R., Budiman, E. S., Castorino, K., Christiansen, M. P., Forlenza, G., Kipnes, M., Liljenquist, D. R., & Liu, H. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820958754 · Cited by 164 · see citing works
Davis, G. M., Spanakis, E. K., Migdal, A. L., Singh, L. G., Albury, B., Urrutia, M. A., Zamudio-Coronado, K. W., Scott, W. H., Doerfler, R., Lizama, S., et al. (2021) · Diabetes Care · doi:10.2337/dc20-2856 · Cited by 160 · see citing works
Continuous glucose monitoring in the intensive care unit during the COVID-19 pandemic
Agarwal, S., Mathew, J., Davis, G. M., Shephardson, A., Levine, A., Louard, R., Urrutia, A., Perez-Guzman, C., Umpierrez, G. E., Peng, L., et al. (2021) · Diabetes Care · doi:10.2337/dc20-2219 · Cited by 138 · see citing works
A clinical guide to advanced diabetes devices and closed-loop systems using the CARES paradigm
Messer, L. H., Berget, C., & Forlenza, G. P. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0105 · Cited by 100 · see citing works
Longo, R. R., Elias, H., Khan, M., & Seley, J. J. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211008446 · Cited by 89 · see citing works
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86 · see citing works
Pemberton, J. S., Wilmot, E. G., Barnard‐Kelly, K., Leelarathna, L., Oliver, N., Randell, T., Taplin, C. E., Choudhary, P., & Adolfsson, P. (2023) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.14962 · Cited by 44 · see citing works
Time in range in pregnancy: is there a role?
Wyckoff, J. A., & Brown, F. M. (2021) · Diabetes Spectrum · doi:10.2337/ds20-0103 · Cited by 39 · see citing works
Cummings, C., Jiang, A., Sheehan, A., Ferraz‐Bannitz, R., Puleio, A., Simonson, D. C., Dreyfuss, J. M., & Patti, M. E. (2023) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.15096 · Cited by 27 · see citing works
A Head-to-Head Comparison Study of the First-Day Performance of Two Factory-Calibrated CGM Systems
Denham, D. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819895505 · Cited by 18 · see citing works
Meng, R., Gu, T., Yang, F., Liu, J., Sun, Q., & Zhu, D. (2021) · Diabetes Therapy · doi:10.1007/s13300-021-01171-2 · Cited by 11 · see citing works
- Authors
- Shah, V. N., Laffel, L. M., Wadwa, R. P., & Garg, S. K.
- Affiliations
- University of Colorado Anschutz (US)
- Joslin Diabetes Center (US)
- Corresponding author
- Laffel, Lori M.
- Garg, Satish K.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2018-06
- Volume
- 20
- Issue
- 6
- Pages
- 428-433
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- PMC
- PMC6422005
- OpenAlex
- W2809055900
- Cited by
- 197
- References
- 7
- Citations by year
- 2026: 20 · 2025: 19 · 2024: 24 · 2023: 27 · 2022: 22 · 2021: 26 · 2020: 24 · 2019: 27 · 2018: 8
- FWCI
- 16.90
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Diabetes Technology & Therapeutics (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Dexcom
- National Institute of Diabetes and Digestive and Kidney Diseases (P30DK036836)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- continuous glucose monitoring, factory calibration, Dexcom G6, sensor accuracy, mean absolute relative difference, insulin-treated diabetes, children and adolescents
- MeSH terms
- Adolescent, Adult, Blood Glucose, Calibration, Child, Computer Systems, Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2, Female, Humans, Male, Middle Aged, Reference Values, Blood Glucose Self-Monitoring, Reproducibility of Results, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review