The frequency of intermittently scanned glucose and diurnal variation of glycemic metrics
Open accessAbstract
Background: The relation between the frequency of intermittently scanned continuous glucose monitoring (isCGM) and diurnal variation of time in range (TIR) and time below range (TBR) is unknown. Method: A total of 163 persons with type 1 diabetes who used isCGM had glucose data for 60 days downloaded. Mean TIR and median TBR were calculated for 15-minute periods and presented for daytime and nighttime. The values for tertiles of scanning frequency were compared. Results: The 1st tertile ( n = 53) of the population scanned <10 times; the 2nd tertile ( n = 56) 10-13 times, and the 3rd tertile ( n = 54) >13 per 24 hours. TIR (%, mean ± (SD)) increased significantly from the 1st to the 3rd scan tertile both during the day (43.8 ± 14.8, 52.0 ± 12.3, 62.1 ± 12.8) and the night (44.5 ± 17.3, 52.3 ± 18.5, 64.0 ± 13.9; P < .0001). In contrast, TBR (median, (IQR)) was not significantly associated with scan tertiles during daytime (3.5% (1.1-7.8), 4.4% (1.8-6.1), 3.5% (2.1-6.1); P = .85) or nighttime (3.8% (1.4-13.7), 5.0% (1.6-9.6), 5.7% (3.6-10.9); P = .24). In a multiple regression model, a 50% increase in 24-hour scanning frequency was associated with a 7.8 percentage point increase in TIR (95% CI, 5.6-10.0). Conclusions: Increased scanning frequency was associated with a higher TIR both during daytime and nighttime with no change in TBR.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The frequency of scanning with isCGM devices is correlated with improved outcomes.
“The frequency of scanning with isCGM devices is also correlated with improved outcomes (40-43).”
From 7. Diabetes Technology: Standards of Care in Diabetes—2025 (2025) · cites this paper as [40]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Battelino, T., Danne, T., Bergenstal, R. M., Amiel, S. A., Beck, R., Biester, T., Bosi, E., Buckingham, B. A., Cefalu, W. T., Close, K. L., et al. (2019) · Diabetes Care · doi:10.2337/dci19-0028 · Cited by 4,032 · cited 3× · see citing works
“1464 Journal of Diabetes Science and Technology 16(6)”
BackgroundInfluentialBolinder, J., Antuna, R., Geelhoed-Duijvestijn, P., Kröger, J., & Weitgasser, R. (2016) · The Lancet · doi:10.1016/s0140-6736(16)31535-5 · Cited by 893 · see citing works
6. Glycemic Targets: <i>Standards of Medical Care in Diabetes—2021</i>
American Diabetes Association (2021) · Diabetes Care · doi:10.2337/dc21-s006 · Cited by 847 · see citing works
Miller, K. M., Beck, R. W., Bergenstal, R. M., Goland, R. S., Haller, M. J., McGill, J. B., Rodriguez, H., Simmons, J. H., Hirsch, I. B., & for the T1D Exchange Clinic Network (2013) · Diabetes Care · doi:10.2337/dc12-1770 · Cited by 505 · see citing works
Charleer, S., De Block, C., Van Huffel, L., Broos, B., Fieuws, S., Nobels, F., Mathieu, C., & Gillard, P. (2020) · Diabetes Care · doi:10.2337/dc19-1610 · Cited by 261 · see citing works
Dunn, T. C., Xu, Y., Hayter, G., & Ajjan, R. A. (2018) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2017.12.015 · Cited by 239 · see citing works
Tyndall, V., Stimson, R. H., Zammitt, N. N., Ritchie, S. A., McKnight, J. A., Dover, A. R., & Gibb, F. W. (2019) · Diabetologia · doi:10.1007/s00125-019-4894-1 · Cited by 165 · see citing works
Hermanns, N., Ehrmann, D., Schipfer, M., Kröger, J., Haak, T., & Kulzer, B. (2019) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2019.03.003 · Cited by 113 · see citing works
Paris, I., Henry, C., Pirard, F., Gérard, A., & Colin, I. M. (2018) · Endocrinology, Diabetes & Metabolism · doi:10.1002/edm2.23 · Cited by 84 · see citing works
Boucher, S. E., Gray, A. R., Wiltshire, E. J., de Bock, M. I., Galland, B. C., Tomlinson, P. A., Rayns, J. A., MacKenzie, K. E., Chan, H., Rose, S., et al. (2020) · Diabetes Care · doi:10.2337/dc20-0613 · Cited by 51 · see citing works
Jangam, S. R., Dunn, T. C., Xu, Y., Hayter, G., & Ajjan, R. (2019) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2018-000611 · Cited by 29 · see citing works
Stimson, R. H., Dover, A. R., Ritchie, S. A., Wright, R. J., McKnight, J. A., Zammitt, N. N., & Gibb, F. W. (2020) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2020-001292 · Cited by 13 · see citing works
Effects of unrestricted access to flash glucose monitoring in type 1 diabetes
Hansen, K. W. (2020) · Endocrinology Diabetes & Metabolism · doi:10.1002/edm2.125 · Cited by 11 · see citing works
Self-citationGlycemic Metrics Derived From Intermittently Scanned Continuous Glucose Monitoring
Hansen, K. W., & Bibby, B. M. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820975822 · Cited by 8 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
7. Diabetes Technology: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s007 · Cited by 348 · see citing works
7. Diabetes Technology: <i>Standards of Care in Diabetes</i> — <i>2023</i>
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301 · see citing works
7. Diabetes Technology: <i>Standards of Care in Diabetes—2024</i>
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281 · see citing works
7. Diabetes Technology: <i>Standards of Medical Care in Diabetes—2022</i>
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s007 · Cited by 358 · see citing works
- Authors
- Hansen, K. W., & Bibby, B. M.
- Affiliations
- Regionshospitalet Silkeborg (DK)
- Aarhus University (DK)
- Corresponding author
- Hansen, Klavs W.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2021-05-27
- Volume
- 16
- Issue
- 6
- Pages
- 1461-1465
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC9631535
- OpenAlex
- W3165377388
- Cited by
- 20
- References
- 14
- Citations by year
- 2026: 2 · 2025: 6 · 2024: 5 · 2023: 4 · 2022: 2 · 2021: 1
- FWCI
- 2.24
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Rosa and Asta Jensen Foundation
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Keywords
- intermittently scanned continuous glucose monitoring, scanning frequency, time in range, time below range, diurnal variation, type 1 diabetes, glycemic metrics
- MeSH terms
- Blood Glucose, Diabetes Mellitus, Type 1, Glucose, Glycated Hemoglobin, Humans, Hypoglycemic Agents, Blood Glucose Self-Monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review