Interferences With CGM Systems: Practical Relevance?
Open accessThe DOI matches its Crossref record; 20 of the 20 references it cites were found in a registry or an index.
Abstract
Most systems for continuous glucose monitoring (CGM) are based on a specific electrochemical methodology for measuring glucose in interstitial subcutaneous fluid (ISF). The enzymes needed for the enzymatic reaction are located at the tip of the filaments that are inserted into the subcutaneous tissue using needles. However, depending on the particular enzyme chemistry and the voltage applied to measure the electron flow in relation to the glucose concentration in the ISF around the sensor tip, other substances can also influence the measurements taken by CGM systems, that is, shift the “glucose” data in one direction or the other. These interfering substances can be endogenous substances, as well as drugs or food components circulating in the human body. Despite the potential impact of interfering substances on CGM accuracy, there are few publications focusing on the interference of drugs and other substances on CGM system measurements.1-3 In addition, the CGM systems tested in these studies are mostly outdated. Manufacturers are required to perform interference studies during the clinical development process for each new CGM system and product generation to gain regulatory approval. However, the data generated by most of these studies are not published. Although interferences are well known and relevant when it comes to systems for self-measurement of blood glucose (SMBG), once again, publications are sparse.4 In this editorial, the publicly available information about interfering substances and how they influence glucose measurements from various CGM systems, manufactured by different companies, are summarized, with a focus on the clinical situation of people with diabetes (PwD). The subsequent summary is not comprehensive and will not cover all available CGM systems.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Medtronic Guardian 3, Medtronic Guardian 4 and Dexcom 6 are not recommended for people taking hydroxycarbamide (hydroxyurea), which can cause falsely elevated glucose readings.32”
From Blood glucose monitoring devices: current considerations (2023) · cites this paper as [32]
In the citing paper
“Hydroxyurea, used for myeloproliferative disorders and hematologic conditions, is one of the most recently identified interfering substances that cause a temporary increase in sensor glucose values discrepant from actual glucose values (129-134).”
From 7. Diabetes Technology: Standards of Care in Diabetes—2025 (2025) · cites this paper as [129]
Citations
Cites
Vitamin C pharmacokinetics: Implications for oral and intravenous use
Padayatty, S. J., Sun, H., Wang, Y., Riordan, H. D., Hewitt, S. M., Katz, A., Wesley, R. A., & Levine, M. (2004) · Annals of Internal Medicine · doi:10.7326/0003-4819-140-7-200404060-00010 · Cited by 992
Continuous glucose monitors and automated insulin dosing systems in the hospital consensus guideline
Galindo, R. J., Umpierrez, G. E., Rushakoff, R. J., Basu, A., Lohnes, S., Nichols, J. H., Spanakis, E. K., Espinoza, J., Palermo, N. E., Awadjie, D. G., et al. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820954163 · Cited by 170
Self-citationCerullo, G., Negro, M., Parimbelli, M., Pecoraro, M., Perna, S., Liguori, G., Rondanelli, M., Cena, H., & D’Antona, G. (2020) · Frontiers in Immunology · doi:10.3389/fimmu.2020.574029 · Cited by 158
Effect of Acetaminophen on CGM Glucose in an Outpatient Setting
Maahs, D. M., DeSalvo, D., Pyle, L., Ly, T., Messer, L., Clinton, P., Westfall, E., Wadwa, R. P., & Buckingham, B. (2015) · Diabetes Care · doi:10.2337/dc15-1096 · Cited by 102
Plasma ascorbic acid: measurement, stability and clinical utility revisited
Chung, W. Y., Chung, J. K.-O., Szeto, Y. T., Tomlinson, B., & Benzie, I. F. F. (2001) · Clinical Biochemistry · doi:10.1016/s0009-9120(01)00270-3 · Cited by 98
Basu, A., Veettil, S., Dyer, R., Peyser, T., & Basu, R. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0410 · Cited by 93
Continuous Glucose Monitor Interference With Commonly Prescribed Medications: A Pilot Study
Basu, A., Slama, M. Q., Nicholson, W. T., Langman, L., Peyser, T., Carter, R., & Basu, R. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817697329 · Cited by 77
Heinemann, L. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0076 · Cited by 62
Self-citationResistance to Acetaminophen Interference in a Novel Continuous Glucose Monitoring System
Calhoun, P., Johnson, T. K., Hughes, J., Price, D., & Balo, A. K. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818755797 · Cited by 58
Vitamin C deficiency in dialysis patients--are we perceiving the tip of an iceberg?
Handelman, G. J. (2006) · Nephrology Dialysis Transplantation · doi:10.1093/ndt/gfl534 · Cited by 55
Research works printed but not linked: 1
- Schiavon M, 2018, Annu Int Conf IEEE Eng Med Biol Soc, 2018, 159
Cited by
Continuous glucose monitoring and metrics for clinical trials: an international consensus statement
Battelino, T., Alexander, C. M., Amiel, S. A., Arreaza-Rubin, G., Beck, R. W., Bergenstal, R. M., Buckingham, B. A., Carroll, J., Ceriello, A., Chow, E., et al. (2023) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(22)00319-9 · Cited by 831
7. Diabetes Technology: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s007 · Cited by 348
7. Diabetes Technology: Standards of Care in Diabetes — 2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301
7. Diabetes Technology: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281
Shaw, J. L. V., Bannuru, R. R., Beach, L. A., ElSayed, N. A., Freckmann, G., Füzéry, A. K., Fung, A. W. S., Gilbert, J. D., Yun, H., Korpi-Steiner, N. L., et al. (2024) · Diabetes Care · doi:10.2337/dci24-0073 · Cited by 78
Accuracy and Potential Interferences of Continuous Glucose Monitoring Sensors in the Hospital
Bellido, V., Freckman, G., Pérez, A., & Galindo, R. J. (2023) · Endocrine Practice · doi:10.1016/j.eprac.2023.06.007 · Cited by 57
Hagerf, B. V., Protuš, M., Németová, L., Mráz, M., Kieslichová, E., Uchytilová, E., Indrova, V., Lelito, J., Girman, P., Haluzík, M., et al. (2024) · Diabetes Care · doi:10.2337/dc23-1663 · Cited by 49
Consensus Report on the Use of Continuous Glucose Monitoring in Chronic Kidney Disease and Diabetes
Rhee, C. M., Gianchandani, R. Y., Kerr, D. E., Philis‐Tsimikas, A., Kovesdy, C. P., Stanton, R. C., Drincic, A., Galindo, R. J., Kalantar‐Zadeh, K., Neumiller, J. J., et al. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241292041 · Cited by 41
Ahn, J., Yang, Y., & Park, G. (2024) · Geriatric Nursing · doi:10.1016/j.gerinurse.2024.06.041 · Cited by 38
Hutchins, K., Betts, J. A., Thompson, D., Hengist, A., & Gonzalez, J. T. (2025) · American Journal of Clinical Nutrition · doi:10.1016/j.ajcnut.2025.02.024 · Cited by 33
The Use of Continuous Glucose Monitoring to Diagnose Stage 2 Type 1 Diabetes
Mader, J. K., Wong, J. C., Freckmann, G., Garcia‐Tirado, J., Hirsch, I. B., Johnson, S. B., Kerr, D. E., Kim, S. H., Lal, R., Montaser, E., et al. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251333441 · Cited by 32
Self-citationOganesova, Z., Pembérton, J. H., & Brown, A. C. (2024) · Diabetic Medicine · doi:10.1111/dme.15369 · Cited by 31
Expanding the horizon of continuous glucose monitoring into the future of pediatric medicine
Morales-Dopico, L., & MacLeish, S. A. (2024) · Pediatric Research · doi:10.1038/s41390-024-03573-x · Cited by 26
Using continuous glucose monitoring data in daily clinical practice
Martens, T. W., Simonson, G. D., & Bergenstal, R. M. (2024) · Cleveland Clinic Journal of Medicine · doi:10.3949/ccjm.91a.23090 · Cited by 24
Drug interference with biochemical laboratory tests
Stanimirov, B., Sekeruš, V., Đanić, M., Pavlović, N., Mikov, M., Stankov, K., & Katanić, J. (2023) · Biochemia Medica · doi:10.11613/bm.2023.020601 · Cited by 21
The Performance of Continuous Glucose Monitoring During the Intraoperative Period: A Scoping Review
Lim, H. A., Kim, M.-J., Kim, N. J., Huh, J., Jeong, J.-O., Hwang, W., & Choi, H. (2024) · Journal of Clinical Medicine · doi:10.3390/jcm13206169 · Cited by 21
Managing discordance between HbA 1c and glucose management indicator
Lenters‐Westra, E., Fokkert, M. J., Kilpatrick, E. S., Schleicher, E. D., Pilla, S. J., English, E., & van Dijk, P. R. (2025) · Diabetic Medicine · doi:10.1111/dme.70023 · Cited by 21
Pfützner, A., Jensch, H., Cardinal, C., Srikanthamoorthy, G., Riehn, E., & Thomé, N. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221095573 · Cited by 20
“16 Only few reports exist in the current literature about such issues.”
BackgroundHughes, M. S., Pasquel, F. J., Davis, G. M., Lal, R., Buckingham, B. A., Boughton, C. K., Brown, S. A., & Bally, L. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0343 · Cited by 14
Blood glucose monitoring devices: current considerations
Sly, B., & Taylor, J. (2023) · Australian Prescriber · doi:10.18773/austprescr.2023.013 · Cited by 11
“32 People using continuous glucose monitoring should always have access to capillary blood glucose monitoring for correlation of hypoglycaemia; this is useful when symptoms are not aligned with continuous blood glucose monitor measurements, when accuracy may be compromised or if there is a…”
BackgroundWrench, E., Subar, D. A., Bampouras, T. M., Lauder, R. M., & Gaffney, C. J. (2024) · Experimental Physiology · doi:10.1113/ep091433 · Cited by 11
van Baal, L., Heinemann, L., Reinold, J., von Conta, J., Bahnsen, F. H., Kleesiek, J., Führer-Sakel, D., & Tan, S. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251334633 · Cited by 11
Self-citationHelleputte, S., Podlogar, T., & Gonzalez, J. T. (2025) · Performance Nutrition · doi:10.1186/s44410-025-00013-7 · Cited by 11
Update on Patient Self-Testing with Portable and Wearable Devices: Advantages and Limitations
Lippi, G., Pighi, L., & Mattiuzzi, C. (2024) · Diagnostics · doi:10.3390/diagnostics14182037 · Cited by 11
Jensch, H., Setford, S. J., Thomé, N., Srikanthamoorthy, G., Weingärtner, L., Grady, M., Holt, E., & Pfützner, A. (2025) · Sensors · doi:10.3390/s25071985 · Cited by 11
Continuous glucose monitoring as a tool in early-stage type 1 diabetes
Carr, A. L. J., Sutton, H., Agesen, R. M., Bonifacio, E., Bosi, E., Gillard, P., Swaby, R., Vercauteren, J., & Besser, R. E. J. (2026) · Diabetologia · doi:10.1007/s00125-026-06707-4 · Cited by 11
Challenges of Wearable Biosensors and Ways to Overcome Them
Tarasov, S. E., Plekhanova, Y. V., Reshetilov, A., Melenkov, S., & Saltanov, I. (2026) · Biosensors · doi:10.3390/bios16030159 · Cited by 10
Diabetes technology: A primer for clinicians
Ambalavanan, J., Isaacs, D. M., & Lansang, M. C. (2024) · Cleveland Clinic Journal of Medicine · doi:10.3949/ccjm.91a.23073 · Cited by 8
Practical Approaches to Continuous Glucose Monitoring in Primary Care: A UK-Based Consensus Opinion
Fernando, K., Alabraba, V., Welsh, J. B., Seidu, S., Strain, W. D., Bell, H. E., Davies, S., & Evans, M. (2025) · Diabetes Therapy · doi:10.1007/s13300-025-01706-x · Cited by 7
Criteria for Personalised Choice of a Continuous Glucose Monitoring System: An Expert Opinion
Molfetta, S. D., Rossi, A., Boscari, F., Irace, C., Laviola, L., & Bruttomesso, D. (2024) · Diabetes Therapy · doi:10.1007/s13300-024-01654-y · Cited by 7
Hybrid closed loop technology in emergency surgery in a person with type 1 diabetes
Afridi, H., Olsen, P., Levy, N., & Dhatariya, K. (2025) · Anaesthesia Reports · doi:10.1002/anr3.70003 · Cited by 7
Hagerf, B. V., Protuš, M., Németová, L., Kieslichová, E., Uchytilová, E., Mráz, M., Girman, P., Švirlochová, V., Franeková, J., & Jabor, A. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231194643 · Cited by 7
Šimunović, M., Kumrić, M., Rušić, D., Šimunović, M. P., & Božić, J. (2024) · Diagnostics · doi:10.3390/diagnostics14242801 · Cited by 7
Accuracy and reliability of a continuous glucose monitoring system with a focus on hypoglycaemia
van Baal, L., Raess, L., Mathew, A., Lahner, H., Unger, N., & Führer, D. (2025) · Diabetes Obesity and Metabolism · doi:10.1111/dom.70408 · Cited by 6
Accuracy of continuous glucose monitoring in critically ill patients
Santamaria, J. D., Selers, E., & Reid, D. (2025) · Critical Care and Resuscitation · doi:10.1016/j.ccrj.2025.100118 · Cited by 5
Setford, S. J. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251377026 · Cited by 5
Pfützner, A., Tencer, B., Stamm, B., Mehta, M., Sharma, P., Gilyazev, R., Jensch, H., Thomé, N., & Huth, M. (2023) · Sensors · doi:10.3390/s23094541 · Cited by 5
Conti, M., Meneghini, E., Fumagalli, G., Guidoni, F., Bertuzzi, F., & Pintaudi, B. (2023) · Acta Diabetologica · doi:10.1007/s00592-023-02166-0 · Cited by 5
Holt, E., Nguyen, H., Bispham, J. A., Liu, J., CHAPMAN, K. S. M., & Grady, M. (2023) · Clinical Diabetes · doi:10.2337/cd23-0005 · Cited by 5
Setford, S. J. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251377008 · Cited by 4
Суплотова, Л. А., Alieva, O. O., & Ibragimova, L. I. (2024) · Problems of Endocrinology · doi:10.14341/probl13535 · Cited by 4
Tienstra, M., Been, R. A., de Boer, J. W., Gans, R. O. B., van Beek, A. P., Kooistra, H., Wiersma, V. R., Mustafa, O. G., Choudhary, P., van Meerten, T., et al. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2025.0232 · Cited by 4
Seaser, J., & Sherman, B. (2024) · JEM Reports · doi:10.1016/j.jemrpt.2024.100136 · Cited by 3
Sebastian, S. A., Padda, I., Bimal, T., Varma, S., Collins, J. S. P., Ahmad, Z., & Safarova, M. S. (2025) · Disease-a-Month · doi:10.1016/j.disamonth.2025.102043 · Cited by 3
Inpatient Use of Automated Insulin Delivery Systems
Chadalawada, L., Parab, R., Feeley, J. M., Hughes, M. S., Pasquel, F. J., & Davis, G. M. (2026) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251412442 · Cited by 4
Wang, Y., Li, S., Lu, J., Feng, K., Huang, X., Hu, F., Sun, M., Zou, Y., Li, Y., Huang, W., et al. (2024) · Journal of Endocrinological Investigation · doi:10.1007/s40618-024-02393-4 · Cited by 3
Les médicaments anti-diabétiques : ce que le médecin anesthésiste réanimateur doit savoir
Joubert, M. (2023) · Anesthésie & Réanimation · doi:10.1016/j.anrea.2023.04.001 · Cited by 3
Mondtt-Blanchard, D., Giraudo, F., & Sánchez, R. (2025) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2025.02.006 · Cited by 2
Ladd, J. M., Gandhi, K. K., Friesner-Gephart, K., Hoffman, R. P., Okafor, M. J., Heinzman, C., Gariepy, C. E., Rasmussen, S. K., Nathan, J. D., & Freeman, A. J. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2025.0010 · Cited by 2
Setford, S. J. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251377027 · Cited by 2
Protuš, M., Hagerf, B. V., Jabor, A., Franeková, J., Németová, L., Uchytilová, E., Indrova, V., Beckova, J., Macek, A., Dolečková, M., et al. (2026) · Therapeutic Advances in Endocrinology and Metabolism · doi:10.1177/20420188251405372 · Cited by 2
Pavlou, P., Leelarathna, L., Avari, P., Hussain, S., Johnston, T., Hyslop, R., Pender, S., & Karalliedde, J. J. (2026) · Expert Review of Endocrinology & Metabolism · doi:10.1080/17446651.2026.2619097 · Cited by 2
Perioperatives Management von Menschen mit Diabetes mellitus: CGM im Krankenhaus
Reger-Tan, S., Lee-Barkey, Y. H., Stratmann, B., Klose, J. M., & Heinemann, L. (2026) · Die Diabetologie · doi:10.1007/s11428-025-01409-8 · Cited by 1
Self-citationvan Baal, L., Lahner, H., Pavlović, J., Moeller, L. C., Unger, N., Führer-Sakel, D., & Mathew, A. (2026) · Endocrine Connections · doi:10.1530/ec-25-0709 · Cited by 1
Daultrey, H. E., Oliver, N. S., Chakera, A. J., Libardi, A. D. L. C., Iwuji, C. C., & Gasparrini, A. (2026) · Diabetes Care · doi:10.2337/dc25-2383 · Cited by 1
Amouyal, C., Benhamou, P., Carlier, A., Marchand, L., Smati, S., & Villard, O. (2026) · Médecine des Maladies Métaboliques · doi:10.1016/j.mmm.2026.06.008 · Cited by 1
Schleicher, E. D., Müller‐Wieland, D., Nauck, M., Merz, T., Heinemann, L., & Freckmann, G. (2026) · Diabetologie und Stoffwechsel · doi:10.1055/a-2896-0406 · Cited by 1
Self-citationYoo, J. H., Moon, S. J., Lee, D. Y., Kwon, S. Y., Kim, K., Lee, J., Cho, Y., Kim, S. R., & Yun, J. (2026) · Journal of Korean Diabetes · doi:10.4093/jkd.2026.27.3.103 · Cited by 1
Vitebskaya, A. V., & Telegina, K. P. (2024) · Meditsinskiy sovet = Medical Council · doi:10.21518/ms2024-431 · Cited by 1
Ojaimi, R. E., Cheisson, G., Cosson, E., Ichaï, C., Jacqueminet, S., Catargi, B., Ouattara, A., Tauveron, I., Valensi, P. E., & Benhamou, D. (2024) · Anaesthesia Critical Care & Pain Medicine · doi:10.1016/j.accpm.2024.101468 · Cited by 1
Dynamic Interference Testing Results with the Dexcom G6 Continuous Glucose Monitoring Device
Thomé, N., Jensch, H., Srikanthamoorthy, G., Weingärtner, L., Grady, M., Setford, S. J., Holt, E., & Pfützner, A. (2025) · Diabetes Technology and Obesity Medicine · doi:10.1089/dtom.2024.0009 · Cited by 1
Ambulatory glucose profile in a patient with type 1 diabetes and deferred insulin treatment
SKUPIEN, J. K. (2025) · Polskie Archiwum Medycyny Wewnętrznej · doi:10.20452/pamw.16988
Scurvy in Northern Florida: A Report of 2 Cases and a Historical Overview
Bohuslavskyi, D., Bryant, D., Modzelewski, T., & Ammar, H. (2025) · Annals of Internal Medicine Clinical Cases · doi:10.7326/aimcc.2025.0202
Development of Continuous Glucose Monitoring Systems for Better Diabetes Management
Sharma, S., Mohite, R., Keyravan, P. J., Jha, A., Deshmukh, K., Ansari, D., Valikarimwala, H., Khatavkar, A., Purohit, B., & Srivastava, R. (2026) · Springer eBooks · doi:10.1007/978-981-95-5519-2_9
Mader, J. K., & Heinemann, L. (2026) · Journal of Diabetes Science and Technology · doi:10.1177/19322968261427022
Self-citationIkram, M., Ahmad, B., Melhaoui, O. E., Abdelkrim, D., Brahim, H., & Bkiyar, H. (2026) · Lecture notes in networks and systems · doi:10.1007/978-3-032-18894-6_23
Glycation gap size, stability over time and correlates in people with Type 1 diabetes
Begbie, T. N., Huang, M. L., Nanayakkara, N., Ronald, C. W., Cohen, N. D., & Jenkins, A. J. (2026) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2026.113378
Protuš, M., Hagerf, B. V., Jabor, A., Franeková, J., Németová, L., Uchytilová, E., Indrova, V., Beckova, J., Macek, A., Dolečková, M., et al. (2026) · Diabetes Technology & Therapeutics · doi:10.1177/15209156261480872
Hemoglobinopathies and diabetes mellitus: from pathophysiology to clinical practice
Manganas, K., Tzeravini, E., Delicou, S., & Tentolouris, A. (2026) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2026.113581
Brazg, R., Mitchell, T., Fei, J., ZHENG, J., Gao, F., Gao, A., Zhu, S., Flacke, F., Shi, L., & Strange, P. (2026) · Diabetes Therapy · doi:10.1007/s13300-026-01918-9
Guidance for Continuous Glucose Monitor Use in Canadian Hospitals
Shaw, J. L. V., Brinc, D., Fung, A. W. S., Houlden, R. L., Garon-Mailer, A., Maks, L., Moore, S., Philp, L. C., Rezvanpour, A., Sun, C., et al. (2026) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2026.09.009
Yoo, J. H., Moon, S. J., Lee, W., Kwon, S. Y., Kim, K., Lee, J., Cho, Y., Kim, S., & Yun, J. (2026) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2026.0923
Metadata
- Authors
- Heinemann, L.
- Affiliations
- Science-Consulting in Diabetes GmbH, Kaarst, Germany
- Corresponding author
- Heinemann, Lutz
- ORCID
- Heinemann, Lutz
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2021-12-15
- Volume
- 16
- Issue
- 2
- Pages
- 271-274
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC8861798
- OpenAlex
- W4200166937
- Cited by
- 65
- References
- 20
- Citations by year
- 2026: 16 · 2025: 18 · 2024: 19 · 2023: 9 · 2022: 3
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Health Sciences › Medicine › Surgery › Healthcare Technology and Patient Monitoring
- Keywords
- continuous glucose monitoring, sensor interference, diabetes management
- MeSH terms
- Blood Glucose, Diabetes Mellitus, Type 1, Humans, Insulin Infusion Systems, Blood Glucose Self-Monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.