Feasibility and performance of continuous glucose monitoring to guide computerized insulin infusion therapy in cardiovascular intensive care unit
Open accessAbstract
Background: We evaluated the feasibility of real-time continuous glucose monitoring (CGM) for titrating continuous intravenous insulin infusion (CII) to manage hyperglycemia in postoperative individuals in the cardiovascular intensive care unit and assessed their accuracy, nursing acceptance, and postoperative individual satisfaction. Methods: Dexcom G6 CGM devices were applied to 59 postsurgical patients with hyperglycemia receiving CII. A hybrid approach combining CGM with periodic point-of-care blood glucose (POC-BG) tests with two phases (initial-ongoing) of validation was used to determine CGM accuracy. Mean and median absolute relative differences and Clarke Error Grid were plotted to evaluate the CGM accuracy. Surveys of nurses and patients on the use of CGMs experience were conducted and results were analyzed. Results: In this cohort (mean age 64, 32% female, 32% with diabetes) with 864 paired POC-BG and CGM values analyzed, mean and median absolute relative difference between POC-BG and CGM values were 13.2% and 9.8%, respectively. 99.7% of paired CGM and POC-BG were in Zones A and B of the Clarke Error Grid. Responses from nurses reported CGMs being very or quite convenient (n = 28; 93%) and it was favored over POC-BG testing (n = 28; 93%). Majority of patients (n = 42; 93%) reported their care process using CGM as being good or very good. Conclusion: This pilot study demonstrates the feasibility, accuracy, and nursing convenience of adopting CGM via a hybrid approach for insulin titration in postoperative settings. These findings provide robust rationale for larger confirmatory studies to evaluate the benefit of CGM in postoperative care to improve workflow, enhance health outcomes, and cost-effectiveness.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Hospital CGM was successfully used in conjunction with periodic POC monitoring in hybrid protocols used for intravenous insulin titrations, decreasing the burden of hourly POC testing.
“However, hospital CGM was successfully used in conjunction with periodic POC monitoring in hybrid protocols used for intravenous insulin titrations, decreasing the burden of hourly POC testing (54)(55)(56).”
From 16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2026 (2026) · cites this paper as [56]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Galindo, R. J., Aleppo, G., Klonoff, D. C., Spanakis, E. K., Agarwal, S., Vellanki, P., Olson, D. E., Umpierrez, G. E., Davis, G. M., & Pasquel, F. J. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820932903 · Cited by 155 · see citing works
Continuous glucose monitoring in the intensive care unit during the COVID-19 pandemic
Agarwal, S., Mathew, J., Davis, G. M., Shephardson, A., Levine, A., Louard, R., Urrutia, A., Perez-Guzman, C., Umpierrez, G. E., Peng, L., et al. (2021) · Diabetes Care · doi:10.2337/dc20-2219 · Cited by 138 · see citing works
Perioperative management of hyperglycemia and diabetes in cardiac surgery patients
Galindo, R. J., Fayfman, M., & Umpierrez, G. E. (2018) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2017.10.005 · Cited by 95 · see citing works
Davis, G. M., Faulds, E., Walker, T., Vigliotti, D., Rabinovich, M., Hester, J., Peng, L., McLean, B., Hannon, P., Poindexter, N., et al. (2021) · Diabetes Care · doi:10.2337/dc20-2085 · Cited by 94 · see citing works
Continuous glucose monitoring in the operating room and cardiac intensive care unit
Perez-Guzman, M. C., Duggan, E., Gibanica, S., Cardona, S., Corujo-Rodriguez, A., Faloye, A., Halkos, M., Umpierrez, G. E., Peng, L., Davis, G. M., et al. (2021) · Diabetes Care · doi:10.2337/dc20-2386 · Cited by 90 · see citing works
Sadhu, A. R., Serrano, I. A., Xu, J., Nisar, T., Lucier, J., Pandya, A. R., & Patham, B. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820964264 · Cited by 84 · see citing works
Faulds, E. R., Boutsicaris, A., Sumner, L., Jones, L., McNett, M., Smetana, K. S., May, C. C., Buschur, E., Exline, M. C., Ringel, M. D., et al. (2021) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgab409 · Cited by 71 · see citing works
Flagged first, then by citations
Research works printed but not linked: 24
- Schmeltz LR, DeSantis AJ, Thiyagarajan V, et al. Reduction of surgical mortality and morbidity in diabetic patients undergo- ing cardiac surgery with a combined intravenous and subcu- taneous insulin glucose management strategy. Diabetes Care. 2007;30(4):823-828. doi:10.2337/dc06-2184 · DOI
- Moghissi ES, Korytkowski MT, DiNardo M, et al. American Association of Clinical Endocrinologists and American Diabetes Association consensus statement on inpatient glycemic control. Diabetes Care. 2009;32(6):1119-1131. doi:10.2337/dc09-9029 · DOI · Paper page
- Kotagal M, Symons RG, Hirsch IB, et al. Perioperative hyper- glycemia and risk of adverse events among patients with and without diabetes. Ann Surg. 2015;261(1):97-103.
- Furnary AP, Gao G, Grunkemeier GL, et al. Continuous insu- lin infusion reduces mortality in patients with diabetes under- going coronary artery bypass grafting. J Thorac Cardiovasc Surg. 2003;125(5):1007-1021. doi:10.1067/mtc.2003.181 · DOI · Paper page
- Schnipper JL, Magee M, Larsen K, Inzucchi SE, Maynard G; Society of Hospital Medicine Glycemic Control Task Force. Society of Hospital Medicine Glycemic Control Task Force summary: practical recommendations for assessing the impact of glycemic control efforts. J Hosp Med. 2008;3(5)(suppl):66- 75. doi:10.1002/jhm.356 · DOI
- ElSayed NA, Aleppo G, Aroda VR, et al. 16. Diabetes care in the hospital: standards of care in diabetes-2023. Diabetes Care. 2022;46(suppl 1):S267-S278. doi:10.2337/dc23-S016 · DOI
- Rice MJ, Smith JL, Coursin DB. Glucose measurement in the ICU: regulatory intersects reality. Crit Care Med. 2017;45(4):741-743.
- McCall AL, Lieb DC, Gianchandani R, et al. Management of individuals with diabetes at high risk for hypoglycemia: an endocrine society clinical practice guideline. J Clin Endocrinol Metab. 2023;108(3):529-562. doi:10.1210/clinem/dgac596 · DOI
- Preiser J-C, Brunkhorst F. Tight glucose control and hypogly- cemia. Crit Care Med. 2008;36(4):1391; author reply 1391- 1392.
- Stamou SC, Nussbaum M, Carew JD, et al. Hypoglycemia with intensive insulin therapy after cardiac surgery: predisposing factors and association with mortality. J Thorac Cardiovasc Surg. 2011;142(1):166-173.
- D'Ancona G, Bertuzzi F, Sacchi L, et al. Iatrogenic hypoglyce- mia secondary to tight glucose control is an independent deter- minant for mortality and cardiac morbidity. Eur J Cardiothorac Surg. 2011;40(2):360-366. doi:10.1016/j.ejcts.2010.11.065 · DOI
- Kovatchev BP, Patek SD, Ortiz EA, Breton MD. Assessing sensor accuracy for nonadjunct use of continuous glucose monitoring. Diabetes Technol Ther. 2015;17(3):177-186.
- American Diabetes Association Professional Practice Committee. 7. Diabetes technology: standards of care in dia- betes-2024. Diabetes Care. 2023;47(suppl 1):S126-S144. doi:10.2337/dc24-S007 · DOI · Paper page
- Boeder S, Kobayashi E, Ramesh G, Serences B, Kulasa K, Majithia AR. Accuracy and glycemic efficacy of continuous glucose monitors in critically ill COVID-19 patients: a retro- spective study. J Diabetes Sci Technol. 2023;17(3):642-648.
- Reutrakul S, Genco M, Salinas H, et al. Feasibility of inpatient continuous glucose monitoring during the COVID-19 pandemic: early experience. Diabetes Care. 2020;43(10):e137-e138.
- Faulds ER, Jones L, McNett M, et al. Facilitators and barriers to nursing implementation of continuous glucose monitoring (CGM) in critically ill patients with COVID-19. Endocr Pract. 2021;27(4):354-361.
- Shelden D, Ateya M, Jensen A, Arnold P, Bellomo T, Gianchandani R. Improving hospital glucometrics, work- flow, and outcomes with a computerized intravenous insulin dose calculator built into the electronic health record. J Diabetes Sci Technol. 2021;15(2):271-278. doi:10.1177/1932296820974767 · DOI
- Clarke WL, Cox D, Gonder-Frederick LA, Carter W, Pohl SL. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care. 1987;10(5):622-628.
- Ancona P, Eastwood GM, Lucchetta L, Ekinci EI, Bellomo R, Mårtensson J. The performance of flash glucose monitor- ing in critically ill patients with diabetes. Crit Care Resusc. 2017;19(2):167-174.
- Rabiee A, Andreasik V, Abu-Hamdah R, et al. Numerical and clinical accuracy of a continuous glucose monitoring system during intravenous insulin therapy in the surgical and burn intensive care units. J Diabetes Sci Technol. 2009;3(4):951- 959.
- Gómez AM, Umpierrez GE, Muñoz OM, et al. Continuous glucose monitoring versus capillary point-of-care testing for inpatient glycemic control in type 2 diabetes patients hos- pitalized in the general ward and treated with a basal bolus insulin regimen. J Diabetes Sci Technol. 2016;10(2):325- 329.
- Schaupp L, Donsa K, Neubauer KM, et al. Taking a closer look-continuous glucose monitoring in noncritically ill hospi- talized patients with type 2 diabetes mellitus under basal-bolus insulin therapy. Diabetes Technol Ther. 2015;17(9):611-618. doi:10.1089/dia.2014.0343 · DOI
- Dumitrascu AG, Perry MF, Boone RJ, et al. Continuous glucose monitoring for patients with COVID-19 pneumo- nia: initial experience at a tertiary care center. Endocr Pract. 2023;29(3):155-161. doi:10.1016/j.eprac.2022.12.012 · DOI
- Finn E, Schlichting L, Grau L, Douglas IS, Pereira RI. Real-world accuracy of CGM in inpatient critical and noncritical care settings at a safety-net hospital. Diabetes Care. 2023;46(10):1825-1830. doi:10.2337/dc23-0089 · DOI
Cited by
Sorted by citations
16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2026
American Diabetes Association Professional Practice Committee for Diabetes*, Bajaj, M., McCoy, R. G., Balapattabi, K., Bannuru, R. R., Bellini, N. J., Bennett, A. K., Beverly, E. A., Briggs Early, K., ChallaSivaKanaka, S., et al. (2026) · Diabetes Care · doi:10.2337/dc26-s016 · Cited by 80 · see citing works
- Authors
- Ang, L., Lin, Y. K., Schroeder, L. F., Huang, Y., DeGeorge, C. A., Arnold, P., Akanbi, F., Knotts, S., DuBois, E., Desbrough, N., Qu, Y., Freeman, R., Esfandiari, N. H., Pop-Busui, R., & Gianchandani, R.
- Affiliations
- University of Michigan (US)
- Michigan State University (US)
- Division of Metabolism, Endocrinology & Diabetes, Department of Internal Medicine, University of Michigan, Ann Arbor, MI, USA
- Cedars-Sinai Medical Center (US)
- Corresponding author
- Ang, Lynn
- ORCID
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2024-04-02
- Volume
- 18
- Issue
- 3
- Pages
- 562-569
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC11089859
- OpenAlex
- W4393716705
- Cited by
- 11
- References
- 7
- Citations by year
- 2026: 5 · 2025: 4 · 2024: 2
- Type
- conference abstract
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- JDRF (Center of Excellence at University of Michigan)
- National Institute of Diabetes and Digestive and Kidney Diseases (K23DK129724, R01DK107956, R01DK11672, U01DK119083)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Surgery › Healthcare Technology and Patient Monitoring
- Keywords
- continuous glucose monitoring, continuous intravenous insulin infusion, hyperglycemia management, postoperative care, cardiovascular intensive care unit, CGM accuracy, Clarke error grid, insulin dose adjustment, patient satisfaction, Dexcom G6
- MeSH terms
- Continuous Glucose Monitoring, Aged, Blood Glucose, Diabetes Mellitus, Feasibility Studies, Female, Humans, Hyperglycemia, Hypoglycemic Agents, Infusions, Intravenous, Insulin, Insulin Infusion Systems, Intensive Care Units, Male, Middle Aged, Blood Glucose Self-Monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review