Hemoglobin Point-of-Care Testing: The HemoCue System
Open accessThe DOI matches its Crossref record; 18 of the 18 references it cites were found in a registry or an index.
Abstract
Besides the use of traditional laboratory resources, the diagnosis of anemia can also be accomplished by assessing hemoglobin (Hb) concentration with point-of-care testing (POCT) devices such as the HemoCue test systems. In several situations, these devices might suitably replace traditional laboratory testing, including several areas of health care where a very rapid Hb measurement might be required to make immediate therapeutic decisions. The use of these devices, however, should fulfill some basic criteria, including economic, clinical, and regulatory issues; appropriate training of the users and knowledge of test requirements, performance, limitations, and potential interferences; the use of venous and arterial sampling, when possible; and a rigorous quality assessment, which should be under the responsibility of laboratory professionals. Because of its optimal performance along with the fact that the HemoCue is probably one of the most commonly used devices worldwide, the aim of this article is to review the literature data about the performance of this test system as compared with laboratory reference testing estimations and according to the biological matrix.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients*
Frasca, D., Dahyot-Fizelier, C., Catherine, K., Levrat, Q., Debaene, B., & Mimoz, O. (2011) · Critical Care Medicine · doi:10.1097/ccm.0b013e3182227e2d · Cited by 178
Hemoglobin estimation by the HemoCue® portable hemoglobin photometer in a resource poor setting
Nkrumah, B., Nguah, S. B., Sarpong, N., Dekker, D., Idriss, A., May, J., & Adu-Sarkodie, Y. (2011) · BMC Clinical Pathology · doi:10.1186/1472-6890-11-5 · Cited by 173 · cited 3×
“Compared Accuracy and Reproducibility Type of Sample Rechner et al.7 Coulter Max-M — + Arterial Paddle8 Coulter STKS — + Capillary, venous, arterial Gomez-Simon et al.11 Technicon H3 analyzer — + Venous Morris et al.9 Abbott Cell-Dyn 3500 B-Hemoglobin – + Capillary, venous Van de Louw et al.10 Cell counter Pentra 120…”
BackgroundMethodLamhaut, L., Apriotesei, R., Combes, X., Lejay, M., Carli, P., & Vivien, B. (2011) · Anesthesiology · doi:10.1097/aln.0b013e3182270c22 · Cited by 160 · cited 1×
“5 g/L.(17) Another group compared the accuracy of the HemoCue (model not reported) with a Beckman Coulter co-oximeter in arterial blood samples of patients who received general anesthesia for spine surgery.”
ResultA Comparison of Three Methods of Hemoglobin Monitoring in Patients Undergoing Spine Surgery
Miller, R. D., Ward, T. A., Shiboski, S. C., & Cohen, N. H. (2011) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31820eecd1 · Cited by 153 · cited 1×
“7%) obtained in these patients, the HemoCue values exhibited a bias lower than 10 g/L as compared with the co-oximeter.(18) This led the authors to conclude that arterial HemoCue values might be virtually interchangeable with those obtained with standard co-oximetry.”
ResultAn azide-methemoglobin method for hemoglobin determination in blood.
Vanzetti, G. (1966) · PubMed · Cited by 141
Lardi, A. M., Hirst, C., Mortimer, A. J., & McCollum, C. N. (1998) · Anaesthesia · doi:10.1046/j.1365-2044.1998.00328.x · Cited by 82 · cited 1×
“45 g/dL, which led the authors to conclude that the HemoCue provided Hb results comparable with the laboratory reference method and was hence useful for nearpatient testing.(6) Another study was carried out to determine whether HemoCue (the specific model was not specified) may be useful in a neonatal unit to measure…”
ResultAn iron treatment trial in an Aboriginal community: Improving non‐adherence
Kruske, S., Ruben, A. R., & Brewster, D. (1999) · Journal of Paediatrics and Child Health · doi:10.1046/j.1440-1754.1999.t01-1-00351.x · Cited by 60
Gómez-Simón, A., Navarro‐Núñez, L., Pérez‐Ceballos, E., Lozano, M. L., Candela, M., Cascales, A., Martı́nez, C., Corral, J., Vicente, V., & Rivera, J. (2007) · Transfusion and Apheresis Science · doi:10.1016/j.transci.2007.01.010 · Cited by 54 · cited 3×
“Compared Accuracy and Reproducibility Type of Sample Rechner et al.7 Coulter Max-M — + Arterial Paddle8 Coulter STKS — + Capillary, venous, arterial Gomez-Simon et al.11 Technicon H3 analyzer — + Venous Morris et al.9 Abbott Cell-Dyn 3500 B-Hemoglobin – + Capillary, venous Van de Louw et al.10 Cell counter Pentra 120…”
BackgroundMethodEvaluation of the Haemoglobin Colour Scale and comparison with the HemoCue haemoglobin assay.
Paddle, J. (2002) · Europe PMC (PubMed Central) · Cited by 48
Determination of capillary hemoglobin levels using the HemoCue system in intensive care patients
Séguin, P., Kleiber, A., Chanavaz, C., Morçet, J., & Mallédant, Y. (2010) · Journal of Critical Care · doi:10.1016/j.jcrc.2010.08.012 · Cited by 45 · cited 3×
“Compared Accuracy and Reproducibility Type of Sample Rechner et al.7 Coulter Max-M — + Arterial Paddle8 Coulter STKS — + Capillary, venous, arterial Gomez-Simon et al.11 Technicon H3 analyzer — + Venous Morris et al.9 Abbott Cell-Dyn 3500 B-Hemoglobin – + Capillary, venous Van de Louw et al.10 Cell counter Pentra 120…”
BackgroundResultReliability of HemoCue in patients with gastrointestinal bleeding
de Louw, A. V., Lasserre, N., Drouhin, F., Thierry, S., Lécuyer, L., Caen, D., & Tenaillon, A. (2006) · Intensive Care Medicine · doi:10.1007/s00134-006-0461-6 · Cited by 44 · cited 4×
“Compared Accuracy and Reproducibility Type of Sample Rechner et al.7 Coulter Max-M — + Arterial Paddle8 Coulter STKS — + Capillary, venous, arterial Gomez-Simon et al.11 Technicon H3 analyzer — + Venous Morris et al.9 Abbott Cell-Dyn 3500 B-Hemoglobin – + Capillary, venous Van de Louw et al.10 Cell counter Pentra 120…”
BackgroundMethodInfluentialBanfi, G., Lombardi, G., Colombini, A., & Lippi, G. (2010) · Clinica Chimica Acta · doi:10.1016/j.cca.2010.03.039 · Cited by 40 · cited 1×
“Sports medicine is another ideal context for POCT, where rapid test results might guide the application of specific training regimens(28) and, even more important, might support antidoping testing when carried out with rigorous preanalytical and analytical requirements.(29,30) There is, for example, a wide experience…”
Self-citationReliability of capillary hemoglobin screening under routine conditions
Ziemann, M., Lizardo, B., Geusendam, G., & Schlenke, P. (2011) · Transfusion · doi:10.1111/j.1537-2995.2011.03183.x · Cited by 37 · cited 3×
“Compared Accuracy and Reproducibility Type of Sample Rechner et al.7 Coulter Max-M — + Arterial Paddle8 Coulter STKS — + Capillary, venous, arterial Gomez-Simon et al.11 Technicon H3 analyzer — + Venous Morris et al.9 Abbott Cell-Dyn 3500 B-Hemoglobin – + Capillary, venous Van de Louw et al.10 Cell counter Pentra 120…”
BackgroundMethodLaboratory testing in pharmacies
Lippi, G., Plebani, M., Favaloro, E. J., & Trenti, T. (2010) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2010.184 · Cited by 35
Self-citationGupta, A., Wrench, I. J., Feast, M., & Alderson, J. D. (2008) · Anaesthesia · doi:10.1111/j.1365-2044.2007.05400.x · Cited by 26 · cited 3×
“Gupta et al.(12) Sysmex XE 2100 Hb 301 + Venous Frasca et al.”
BackgroundResultLaboratory medicine and natural disasters: are we ready for the challenge?
(2010) · doi:10.1515/cclm.2010.148 · Cited by 23 · cited 1×
“The use of POCT devices also represents the best option in the unfortunate circumstance of natural disasters, where there is a compelling need to convey laboratory technologies that can be easily transported, installed, and appropriately used outside the traditional laboratory environment.(27) Sports medicine is anoth…”
Richards, N. A., Boyce, H., & Yentis, S. M. (2009) · International Journal of Obstetric Anesthesia · doi:10.1016/j.ijoa.2009.05.010 · Cited by 16 · cited 2×
“HemoCue was hence considered a quick and reliable method for Hb assessment in both the acute and stable phases of gastrointestinal bleeding.14 On the other hand, Richards et al.15 compared Hb values measured in venous and capillary samples (toe and thumb) in patients undergoing caesarean section under neuraxial anesth…”
BackgroundSpectrophotometry of Hemoglobin and Hemoglobin Derivatives
van Kampen, E. J., & Zijlstra, W. G. (1983) · Advances in clinical chemistry · doi:10.1016/s0065-2423(08)60401-1
Cited by
Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings
Whitehead, R. D., Mei, Z., Mapango, C., & Jefferds, M. E. D. (2019) · Annals of the New York Academy of Sciences · doi:10.1111/nyas.14124 · Cited by 188
Chaudhary, E., George, F., Saji, A., Dey, S., Ghosh, S., Thomas, T., Kurpad, A. V., Sharma, S., Singh, N., Agarwal, S., et al. (2023) · Nature Communications · doi:10.1038/s41467-023-42709-1 · Cited by 92
Mehta, U., Dey, S., Chowdhury, S., Ghosh, S., Hart, J. E., & Kurpad, A. (2021) · Environmental Epidemiology · doi:10.1097/ee9.0000000000000125 · Cited by 62
Metadata
- Authors
- Sanchis-Gomar, F., Cortell-Ballester, J., Pareja-Galeano, H., Banfi, G., & Lippi, G.
- Venue
- SLAS Technology
- Publisher
- Elsevier BV
- Published
- 2013-06
- Volume
- 18
- Issue
- 3
- Pages
- 198-205
- ISSN
- 2472-6303 (ISSN-L)
- OpenAlex
- W2137301465
- Cited by
- 158
- References
- 18
- FWCI
- 4.08
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.