Accuracy of blood-glucose measurements using glucose meters and arterial blood gas analyzers in critically ill adult patients: systematic review
Open accessThe DOI matches its Crossref record; 8 of the 8 references it cites were found in a registry or an index.
Abstract
Introduction: Glucose control to prevent both hyperglycemia and hypoglycemia is important in an intensive care unit. Arterial blood gas analyzers and glucose meters are commonly used to measure blood-glucose concentration in an intensive care unit; however, their accuracies are still unclear. Methods: We performed a systematic literature search (January 1, 2001, to August 31, 2012) to find clinical studies comparing blood-glucose values measured with glucose meters and/or arterial blood gas analyzers with those simultaneously measured with a central laboratory machine in critically ill adult patients. Results: We reviewed 879 articles and found 21 studies in which the accuracy of blood-glucose monitoring by arterial blood gas analyzers and/or glucometers by using central laboratory methods as references was assessed in critically ill adult patients. Of those 21 studies, 11 studies in which International Organization for Standardization criteria, error-grid method, or percentage of values within 20% of the error of a reference were used were selected for evaluation. The accuracy of blood-glucose measurements by arterial blood gas analyzers and glucose meters by using arterial blood was significantly higher than that of measurements with glucose meters by using capillary blood (odds ratios for error: 0.04, P < 0.001; and 0.36, P < 0.001). The accuracy of blood-glucose measurements with arterial blood gas analyzers tended to be higher than that of measurements with glucose meters by using arterial blood (P = 0.20). In the hypoglycemic range (defined as < 81 mg/dl), the incidence of errors using these devices was higher than that in the nonhypoglycemic range (odds ratios for error: arterial blood gas analyzers, 1.86, P = 0.15; glucose meters with capillary blood, 1.84, P = 0.03; glucose meters with arterial blood, 2.33, P = 0.02). Unstable hemodynamics (edema and use of a vasopressor) and use of insulin were associated with increased error of blood glucose monitoring with glucose meters. Conclusions: Our literature review showed that the accuracy of blood-glucose measurements with arterial blood gas analyzers was significantly higher than that of measurements with glucose meters by using capillary blood and tended to be higher than that of measurements with glucose meters by using arterial blood. These results should be interpreted with caution because of the large variation of accuracy among devices. Because blood-glucose monitoring was less accurate within or near the hypoglycemic range, especially in patients with unstable hemodynamics or receiving insulin infusion, we should be aware that current blood glucose-monitoring technology has not reached a high enough degree of accuracy and reliability to lead to appropriate glucose control in critically ill patients.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Intensive versus Conventional Glucose Control in Critically Ill Patients
(2009) · New England Journal of Medicine · doi:10.1056/nejmoa0810625 · Cited by 5,119
Standards of medical care in diabetes 2010
American Diabetes Association (2010) · Diabetes Care · doi:10.2337/dc10-s011 · Cited by 3,470
Griesdale, D. E. G., de Souza, R. J., van Dam, R. M., Heyland, D. K., Cook, D. J., Malhotra, A., Dhaliwal, R., Henderson, W. R., Chittock, D. R., Finfer, S., et al. (2009) · Canadian Medical Association Journal · doi:10.1503/cmaj.090206 · Cited by 1,096
Hypoglycemia and risk of death in critically ill patients
The NICE-SUGAR Study Investigators (2012) · New England Journal of Medicine · doi:10.1056/nejmoa1204942 · Cited by 1,009
Hypoglycemia and outcome in critically ill patients
Egi, M., Bellomo, R., Stachowski, E., French, C. J., Hart, G. K., Taori, G., Hegarty, C., & Bailey, M. (2010) · Mayo Clinic Proceedings · doi:10.4065/mcp.2009.0394 · Cited by 432 · cited 1×
“Because many studies have shown that even mild hypoglycemia is significantly associated with increase in mortality [39,40], accuracy of blood-glucose monitoring in the hypoglycemic range is important.”
BackgroundSelf-citationToward Understanding Tight Glycemic Control in the ICU
Marik, P. E., & Preiser, J.-C. (2010) · Chest · doi:10.1378/chest.09-1737 · Cited by 365
Tight glucose control in the intensive care unit: Are glucose meters up to the task?
Scott, M. G., Bruns, D. E., Boyd, J. C., & Sacks, D. B. (2009) · Clinical Chemistry · doi:10.1373/clinchem.2008.117291 · Cited by 162
The Accuracy of Point-of-Care Glucose Measurements
Rebel, A., Rice, M. A., & Fahy, B. G. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600228 · Cited by 144 · cited 1×
“Although several reviews focused on the accuracy of point of blood-glucose monitoring in critically ill patients [10,37,38], our review is the first systematic review for this issue.”
Background
Cited by
Surviving Sepsis Campaign: International guidelines for management of sepsis and septic shock: 2016.
Rhodes, A., Evans, L. E., Alhazzani, W., Levy, M. M., Antonelli, M., Ferrer, R., Kumar, A., Sevransky, J. E., Sprung, C. L., Nunnally, M. E., et al. (2017) · Intensive Care Medicine · doi:10.1007/s00134-017-4683-6 · Cited by 6,621
“org March 2017 • Volume 45 • Number 3 inaccurate, with frequent false elevations (479–481) over the range of glucose levels, but especially in the hypoglycemic and hyperglycemic ranges (482) and in shock patients (receiving vasopressors) (478, 480).”
BackgroundSurviving Sepsis Campaign: International guidelines for management of sepsis and septic shock: 2016.
Rhodes, A., Evans, L. E., Alhazzani, W., Levy, M. M., Antonelli, M., Ferrer, R., Kumar, A., Sevransky, J. E., Sprung, C. L., Nunnally, M. E., et al. (2017) · Critical Care Medicine · doi:10.1097/ccm.0000000000002255 · Cited by 3,652
Villena Gonzales, W., Mobashsher, A., & Abbosh, A. (2019) · Sensors · doi:10.3390/s19040800 · Cited by 772
Perioperative Hyperglycemia Management: An Update
Duggan, E. W., Carlson, K., & Umpierrez, G. E. (2017) · Anesthesiology · doi:10.1097/aln.0000000000001515 · Cited by 402
Continuous Glucose Monitoring Systems: A Review
Vashist, S. (2013) · Diagnostics · doi:10.3390/diagnostics3040385 · Cited by 261
DuBois, J. A., Slingerland, R. J., Fokkert, M., Roman, A., Tran, N. K., Clarke, W., Sartori, D. A., Palmieri, T. L., Malic, A., Lyon, M. E., et al. (2017) · Critical Care Medicine · doi:10.1097/ccm.0000000000002252 · Cited by 59
Vihonen, H., Kuisma, M., & Nurmi, J. (2018) · Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine · doi:10.1186/s13049-018-0480-7 · Cited by 11
How can point-of-care technologies support in-hospital diabetes care?
Misra, S., Avari, P., Lumb, A., Flanagan, D., Choudhary, P., Rayman, G., & Dhatariya, K. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221137360 · Cited by 9
Metadata
- Authors
- Inoue, S., Egi, M., Kotani, J., & Morita, K.
- Venue
- Critical Care
- Published
- 2013-03-18
- Volume
- 17
- Issue
- 2
- Pages
- R48-R48
- ISSN
- 1364-8535 (ISSN-L)
- OpenAlex
- W2153014880
- Cited by
- 144
- References
- 8
- FWCI
- 2.08
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.