Impact of point spread function modelling and time of flight on FDG uptake measurements in lung lesions using alternative filtering strategies
Open accessThe DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
BACKGROUND: The use of maximum standardised uptake value (SUVmax) is commonplace in oncology positron emission tomography (PET). Point spread function (PSF) modelling and time-of-flight (TOF) reconstructions have a significant impact on SUVmax, presenting a challenge for centres with defined protocols for lesion classification based on SUVmax thresholds. This has perhaps led to the slow adoption of these reconstructions. This work evaluated the impact of PSF and/or TOF reconstructions on SUVmax, SUVpeak and total lesion glycolysis (TLG) under two different schemes of post-filtering. METHODS: Post-filters to match voxel variance or SUVmax were determined using a NEMA NU-2 phantom. Images from 68 consecutive lung cancer patients were reconstructed with the standard iterative algorithm along with TOF; PSF modelling - Siemens HD·PET (HD); and combined PSF modelling and TOF - Siemens ultraHD·PET (UHD) with the two post-filter sets. SUVmax, SUVpeak, TLG and signal-to-noise ratio of tumour relative to liver (SNR(T-L)) were measured in 74 lesions for each reconstruction. Relative differences in uptake measures were calculated, and the clinical impact of any changes was assessed using published guidelines and local practice. RESULTS: When matching voxel variance, SUVmax increased substantially (mean increase +32% and +49% for HD and UHD, respectively), potentially impacting outcome in the majority of patients. Increases in SUVpeak were less notable (mean increase +17% and +23% for HD and UHD, respectively). Increases with TOF alone were far less for both measures. Mean changes to TLG were <10% for all algorithms for either set of post-filters. SNR(T-L) were greater than ordered subset expectation maximisation (OSEM) in all reconstructions using both post-filtering sets. CONCLUSIONS: Matching image voxel variance with PSF and/or TOF reconstructions, particularly with PSF modelling and in small lesions, resulted in considerable increases in SUVmax, inhibiting the use of defined protocols for lesion classification based on SUVmax. However, reduced partial volume effects may increase lesion detectability. Matching SUVmax in phantoms translated well to patient studies for PSF reconstruction but less well with TOF, where a small positive bias was observed in patient images. Matching SUVmax significantly reduced voxel variance and potential variability of uptake measures. Finally, TLG may be less sensitive to reconstruction methods compared with either SUVmax or SUVpeak.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Fully 3-D PET reconstruction with system matrix derived from point source measurements
Panin, V. Y., Kehren, F., Michel, C., & Casey, M. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.876171 · Cited by 536
Pak, K., Cheon, G. J., Nam, H.-Y., Kim, S.-J., Kang, K. W., Chung, J.-K., Kim, E. E., & Lee, D. S. (2014) · Journal of Nuclear Medicine · doi:10.2967/jnumed.113.133801 · Cited by 301 · cited 2×
“Recently, TLG has also been shown to offer superior prognostic information than SUVmax [16-20].”
BackgroundMethodImpact of Time-of-Flight on PET Tumor Detection
Kadrmas, D. J., Casey, M. E., Conti, M., Jakoby, B. W., Lois, C., & Townsend, D. W. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.109.063016 · Cited by 277 · cited 3×
“The use of PSF modelling, with and without TOF, has been shown to improve signalto-noise ratio (SNR) [24-27] and lesion detectability [28-30] partly through decreasing voxel variance.”
BackgroundMethodImage-based point spread function implementation in a fully 3D OSEM reconstruction algorithm for PET
Rapisarda, E., Bettinardi, V, Thielemans, K., & Gilardi, M. C. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/14/012 · Cited by 142 · cited 1×
“However, the implementation of PSF modelling, both within projection space and image space, from different manufacturers and also academic institutions has been shown to produce Gibbs artefacts [21,31-35] (Nick Vennart, personal communication).”
BackgroundVariations in Clinical PET/CT Operations: Results of an International Survey of Active PET/CT Users
Beyer, T., Czernin, J., & Freudenberg, L. S. (2011) · Journal of Nuclear Medicine · doi:10.2967/jnumed.110.079624 · Cited by 136
Schaefferkoetter, J., Casey, M., Townsend, D., & El Fakhri, G. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/5/1465 · Cited by 85
Clinical evaluation of PET image reconstruction using a spatial resolution model
Andersen, F. L., Klausen, T. L., Loft, A., Beyer, T., & Holm, S. (2013) · European Journal of Radiology · doi:10.1016/j.ejrad.2012.11.015 · Cited by 62
Cited by
Rausch, I., Cal-González, J., Dapra, D., Gallowitsch, H. J., Lind, P., Beyer, T., & Minear, G. (2015) · EJNMMI Physics · doi:10.1186/s40658-015-0132-1 · Cited by 133
“Using a model numerical observer method to calculate localized receiver operator curves (LROC), the authors measured a reduction of approximately 16 % in the area under the ROC (AROC) when the acquisition time was reduced from 240 to 120 s per PET FOV from OSEM + PSF + TOF reconstructed data with a subsequently applie…”
BackgroundMethodResultInfluentialSher, A., Lacoeuille, F., Fosse, P., Vervueren, L., Cahouet-Vannier, A., Dabli, D., Bouchet, F., & Couturier, O. (2016) · EJNMMI Research · doi:10.1186/s13550-016-0177-8 · Cited by 111
“Furthermore, point spread function (PSF) reconstruction, also available in new generations of PET systems, is known not only to improve sensitivity but also to overestimate SUV [8, 9].”
BackgroundQuantitative implications of the updated EARL 2019 PET–CT performance standards
Kaalep, A., Burggraaff, C. N., Pieplenbosch, S., Verwer, E. E., Sera, T., Zijlstra, J., Hoekstra, O. S., Oprea-Lager, D. E., & Boellaard, R. (2019) · EJNMMI Physics · doi:10.1186/s40658-019-0257-8 · Cited by 79
“, who suggested that TLG may be less sensitive to reconstruction methods compared with either SUVmax or SUVpeak [33].”
MethodNew PET technologies – embracing progress and pushing the limits
Aide, N., Lasnon, C., Kesner, A., Levin, C. S., Buvat, I., Iagaru, A., Hermann, K., Badawi, R. D., Cherry, S. R., Bradley, K. M., et al. (2021) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-021-05390-4 · Cited by 73
Fuentes-Ocampo, F., López-Mora, D. A., Flotats, A., Paillahueque, G., Camacho, V., Duch, J., Fernández, A., Domènech, A., Estorch, M., & Carrió, I. (2019) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-018-4256-0 · Cited by 65
Metadata
- Authors
- Armstrong, I. S., Kelly, M. D., Williams, H. A., & Matthews, J. C.
- Venue
- EJNMMI Physics
- Published
- 2014-11-30
- Volume
- 1
- Issue
- 1
- Pages
- 99-99
- ISSN
- 2197-7364 (ISSN-L)
- OpenAlex
- W2145597967
- Cited by
- 84
- References
- 7
- FWCI
- 4.33
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Impact of point spread function modelling and time of flight on FDG uptake measurements in lung lesions using alternative filtering strategies. https://app.cochat.ai/papers/10.1186/s40658-014-0099-3
- The paper
- Armstrong, I. S., Kelly, M. D., Williams, H. A., & Matthews, J. C. (2014). Impact of point spread function modelling and time of flight on FDG uptake measurements in lung lesions using alternative filtering strategies. EJNMMI Physics. https://doi.org/10.1186/s40658-014-0099-3
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.