Quantitative comparison of PET performance—Siemens Biograph mCT and mMR
Open accessThe DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Integrated clinical whole-body PET/MR systems were introduced in 2010. In order to bring this technology into clinical usage, it is of great importance to compare the performance with the well-established PET/CT. The aim of this study was to evaluate PET performance, with focus on image quality, on Siemens Biograph mMR (PET/MR) and Siemens Biograph mCT (PET/CT). METHODS: A direct quantitative comparison of the performance characteristics between the mMR and mCT system was performed according to National Electrical Manufacturers Association (NEMA) NU 2-2007 protocol. Spatial resolution, sensitivity, count rate and image quality were evaluated. The evaluation was supplemented with additional standardized uptake value (SUV) measurements. RESULTS: The spatial resolution was similar for the two systems. Average sensitivity was higher for the mMR (13.3 kcps/MBq) compared to the mCT system (10.0 kcps/MBq). Peak noise equivalent count rate (NECR) was slightly higher for the mMR (196 kcps @ 24.4 kBq/mL) compared to the mCT (186 kcps @ 30.1 kBq/mL). Scatter fractions in the clinical activity concentration range yielded lower values for the mCT (34.9 %) compared to those for the mMR (37.0 %). Best image quality of the systems resulted in approximately the same mean hot sphere contrast and a difference of 19 percentage points (pp) in mean cold contrast, in favour of the mCT. In general, point spread function (PSF) increased hot contrast and time of flight (TOF) increased both hot and cold contrast. Highest hot contrast for the smallest sphere (10 mm) was achieved with the combination of TOF and PSF on the mCT. Lung residual error was higher for the mMR (22 %) than that for the mCT (17 %), with no effect of PSF. With TOF, lung residual error was reduced to 8 % (mCT). SUV was accurate for both systems, but PSF caused overestimations for the 13-, 17- and 22-mm spheres. CONCLUSIONS: Both systems proved good performance characteristics, and the PET image quality of the mMR was close to that of the mCT. Differences between the systems were mainly due to the TOF possibility on the mCT, which resulted in an overall better image quality, especially for the most challenging settings with higher background activity and small uptake volumes.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Focus on time-of-flight PET: the benefits of improved time resolution
Conti, M. (2011) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-010-1711-y · Cited by 335
Schäfer, J. F., Gatidis, S., Schmidt, H., Gückel, B., Bezrukov, I., Pfannenberg, C. A., Reimold, M., Ebinger, M., Fuchs, J., Claussen, C. D., et al. (2014) · Radiology · doi:10.1148/radiol.14131732 · Cited by 215
Qualitative and Quantitative Comparison of PET/CT and PET/MR Imaging in Clinical Practice
Al-Nabhani, K. Z., Syed, R., Michopoulou, S., Alkalbani, J., Afaq, A., Panagiotidis, E., O’Meara, C., Groves, A., Ell, P., & Bomanji, J. (2014) · Journal of Nuclear Medicine · doi:10.2967/jnumed.113.123547 · Cited by 101
Boellaard, R., Rausch, I., Beyer, T., Delso, G., Yaqub, M., Quick, H. H., & Sattler, B. (2015) · Medical Physics · doi:10.1118/1.4930962 · Cited by 62 · cited 2×
“[26] and Boellard et al [27] have demonstrated that CT-based AC is superior to MR-based AC for image quality evaluation of the NEMA phantom in integrated PET/MR.”
BackgroundSystematic Evaluation of Phantom Fluids for Simultaneous PET/MR Hybrid Imaging
Ziegler, S., Braun, H., Ritt, P., Hocke, C., Kuwert, T., & Quick, H. H. (2013) · Journal of Nuclear Medicine · doi:10.2967/jnumed.112.116376 · Cited by 50 · cited 1×
“fluids with high relative permittivity are imaged [25].”
Background
Cited by
Studies of a Next-Generation Silicon-Photomultiplier–Based Time-of-Flight PET/CT System
Hsu, D. F. C., Ilan, E., Peterson, W. T., Uribe, J., Lubberink, M., & Levin, C. S. (2017) · Journal of Nuclear Medicine · doi:10.2967/jnumed.117.189514 · Cited by 241
Feasibility of state of the art PET/CT systems performance harmonisation
Kaalep, A., Sera, T., Rijnsdorp, S., Yaqub, M., Talsma, A., Lodge, M. A., & Boellaard, R. (2018) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-018-3977-4 · Cited by 160
Performance evaluation of the next generation solid-state digital photon counting PET/CT system
Zhang, J., Maniawski, P., & Knopp, M. V. (2018) · EJNMMI Research · doi:10.1186/s13550-018-0448-7 · Cited by 123
“PET/CT model Discovery MI (4-ring) [51] Discovery 690 [43] Vereos (this work) Ingenuity TF [44] Gemini T [42] Biograph mCT flow [45] Biograph mCT [46, 47] Celesteion [48–50]”
Method
Metadata
- Authors
- Karlberg, A. M., Sæther, O., Eikenes, L., & Goa, P. E.
- Venue
- EJNMMI Physics
- Published
- 2016-02-25
- Volume
- 3
- Issue
- 1
- Pages
- 5-5
- ISSN
- 2197-7364 (ISSN-L)
- OpenAlex
- W2281358382
- Cited by
- 79
- References
- 5
- FWCI
- 7.10
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Quantitative comparison of PET performance—Siemens Biograph mCT and mMR. https://app.cochat.ai/papers/10.1186/s40658-016-0142-7
- The paper
- Karlberg, A. M., Sæther, O., Eikenes, L., & Goa, P. E. (2016). Quantitative comparison of PET performance—Siemens Biograph mCT and mMR. EJNMMI Physics. https://doi.org/10.1186/s40658-016-0142-7
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.