Evaluation of Penalized-Likelihood Estimation Reconstruction on a Digital Time-of-Flight PET/CT Scanner for 18 F-FDG Whole-Body Examinations

Open access
Lindström, Elin, Sundin, Anders, Trampal, Carlos, Lindsjö, Lars, Ilan, Ezgi, Danfors, Torsten, Antoni, Gunnar, Sörensen, Jens +1 more·2018·Journal of Nuclear Medicine

Abstract

The resolution and quantitative accuracy of PET are highly influenced by the reconstruction method. Penalized-likelihood estimation algorithms allow for fully convergent iterative reconstruction, generating a higher image contrast than ordered-subsets expectation maximization (OSEM) while limiting noise. In this study, a type of penalized reconstruction known as block-sequential regularized expectation maximization (BSREM) was compared with time-of-flight OSEM (TOF OSEM). Various strengths of noise penalization factor β were tested along with various acquisition durations and transaxial fields of view (FOVs) with the aim of evaluating the performance and clinical use of BSREM for 18F-FDG PET/CT, both quantitatively and in a qualitative visual evaluation. Methods: Eleven clinical whole-body 18F-FDG PET/CT examinations acquired on a digital TOF PET/CT scanner were included. The data were reconstructed using BSREM with point-spread function recovery and β-factors of 133, 267, 400, and 533—and using TOF OSEM with point-spread function—for various acquisition times per bed position and various FOVs. Noise level, signal-to-noise ratio (SNR), signal-to-background ratio (SBR), and SUV were analyzed. A masked evaluation of visual image quality, rating several aspects, was performed by 2 nuclear medicine physicians to complement the analysis. Results: The lowest levels of noise were reached with the highest β-factor, resulting in the highest SNR, which in turn resulted in the lowest SBR. A β-factor of 400 gave noise equivalent to TOF OSEM but produced a significant increase in SUVmax (11%), SNR (22%), and SBR (12%). BSREM with a β-factor of 533 at a decreased acquisition duration (2 min/bed position) was comparable to TOF OSEM at a full acquisition duration (3 min/bed position). Reconstructed FOV had an impact on BSREM outcome measures; SNR increased and SBR decreased when FOV was shifted from 70 to 50 cm. The evaluation of visual image quality resulted in similar scores for reconstructions, although a β-factor of 400 obtained the highest mean whereas a β-factor of 267 was ranked best in overall image quality, contrast, sharpness, and tumor detectability. Conclusion: In comparison with TOF OSEM, penalized BSREM reconstruction resulted in an increased tumor SUVmax and an improved SNR and SBR at a matched level of noise. BSREM allowed for a shorter acquisition than TOF OSEM, with equal image quality.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.2967/jnumed.117.200790

Cites

18 of 18 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Several studies on subjective image quality in whole-body [18F]FDG-PET with PL reconstruction found that image quality was highest at beta values of 450 (to 600) despite lower lesion contrast recovery or image sharpness at these beta values compared to lower values.

    “Several studies on subjective image quality in whole-body [18F]FDG-PET with PL reconstruction found that image quality was highest at beta values of 450 (to 600) despite lower lesion CR or “image sharpness” at these beta values compared to lower values [41,108,112,175,179].”

    From Influences on PET Quantification and Interpretation (2022) · cites this paper as [108]

    Split into 2 atoms — 0 confirmed, 1 partially supported, 1 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.

    Checked by a reasoning model — may be wrong.

    Experimental — accuracy not yet measured
  • Using a commercially available PL reconstruction (GE Q.Clear), phantom studies have shown that PL reconstruction can increase CR of standardized sphere inserts compared to conventional TOF and PSF reconstruction.

    “Using a commercially available PL reconstruction (General Electric [GE] Q.Clear), several phantom studies have shown that PL reconstruction can increase the CR of standardized sphere inserts compared to conventional TOF and PSF reconstruction [41,46,101,107,108].”

    From Influences on PET Quantification and Interpretation (2022) · cites this paper as [108]

    The Image Reconstruction and NEMA Phantom sections compare Q.Clear/BSREM with TOF OSEM using PSF compensation in a standardized sphere phantom. The phantom results report higher contrast-to-noise ratio (CNR) for BSREM at certain β-factors, but do not report an increase in contrast recovery (CR); those are different measures. No page numbers are available in our copy. (LLM, conf=0.85)

    Experimental — accuracy not yet measured
  • Inter-method SUV differences may no longer be significant at beta values of 300 to 600, which have been rated as optimal for visual reading of [18F]FDG-PET images by human readers.

    “However, this difference is directly dependent on the user-defined penalization factor beta during PL reconstruction, and at beta values of 300 to 600, which have been rated as optimal for visual reading of [18F]FDG-PET images by human readers [107,112,113], inter-method SUV differences may no longer be significant [41,108,111].”

    From Influences on PET Quantification and Interpretation (2022) · cites this paper as [108]

  • PL reconstruction has the potential to systematically improve SNR compared to OSEM-based algorithms.

    “Based on the potential of PL reconstruction to systematically improve SNR compared to OSEM-based algorithms [108,113,168]”

    From Influences on PET Quantification and Interpretation (2022) · cites this paper as [108]

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Lindström, E., Sundin, A., Trampal, C., Lindsjö, L., Ilan, E., Danfors, T., Antoni, G., Sörensen, J., & Lubberink, M.
Affiliations
  • Uppsala University (SE)
  • Uppsala University Hospital (SE)
Venue
Journal of Nuclear Medicine
Published
2018-02-15
Volume
59
Issue
7
Pages
1152-1158
ISSN
  • 0161-5505 (ISSN-L)
  • 2159-662X
OpenAlex
W2789214951
Cited by
99
References
18
Citations by year
2026: 8 · 2025: 12 · 2024: 11 · 2023: 13 · 2022: 14 · 2021: 18 · 2020: 12 · 2019: 10 · 2018: 1
FWCI
7.12
Type
journal article
Language
en
Open access
Locations
  • Journal of Nuclear Medicine (publishedVersion)
  • PubMed (publishedVersion)
Funders
  • Akademiska Sjukhuset
  • Uppsala Universitet
Topics
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
Keywords
penalized-likelihood reconstruction, 18F-FDG PET/CT, whole-body PET, signal-to-noise ratio, signal-to-background ratio, standardized uptake value, point spread function, acquisition time, image quality, tumor detectability
MeSH terms
Positron Emission Tomography Computed Tomography, Image Processing, Computer-Assisted, Time Factors, Likelihood Functions, Phantoms, Imaging, Fluorodeoxyglucose F18, Whole Body Imaging

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review