Standard MRI-based attenuation correction for PET/MRI phantoms: a novel concept using MRI-visible polymer

Open access
Rausch, Ivo, Valladares, Alejandra, Sundar, Lalith Kumar Shiyam, Beyer, Thomas, Hacker, Marcus, Meyerspeer, Martin, Unger, Ewald·2021·EJNMMI Physics

Abstract

Background: PET/MRI phantom studies are challenged by the need of phantom-specific attenuation templates to account for attenuation properties of the phantom material. We present a PET/MRI phantom built from MRI-visible material for which attenuation correction (AC) can be performed using the standard MRI-based AC. Methods: A water-fillable phantom was 3D-printed with a commercially available MRI-visible polymer. The phantom had a cylindrical shape and the fillable compartment consisted of a homogeneous region and a region containing solid rods of different diameters. The phantom was filled with a solution of water and [18F]FDG. A 30 min PET/MRI acquisition including the standard Dixon-based MR-AC method was performed. In addition, a CT scan of the phantom was acquired on a PET/CT system. From the Dixon in-phase, opposed-phase and fat images, a phantom-specific AC map (Phantom MR-AC) was produced by separating the phantom material from the water compartment using a thresholding-based method and assigning fixed attenuation coefficients to the individual compartments. The PET data was reconstructed using the Phantom MR-AC, the original Dixon MR-AC, and an MR-AC just containing the water compartment (NoWall-AC) to estimate the error of ignoring the phantom walls. CT-based AC was employed as the reference standard. Average %-differences in measured activity between the CT corrected PET and the PET corrected with the other AC methods were calculated. Results: The phantom housing and the liquid compartment were both visible and distinguishable from each other in the Dixon images and allowed the segmentation of a phantom-specific MR-based AC. Compared to the CT-AC PET, average differences in measured activity in the whole water compartment in the phantom of −0.3%, 9.4%, and −24.1% were found for Dixon phantom MR-AC, MR-AC, and NoWall-AC based PET, respectively. Average differences near the phantom wall in the homogeneous region were −0.3%, 6.6%, and −34.3%, respectively. Around the rods, activity differed from the CT-AC PET by 0.7%, 8.9%, and −45.5%, respectively. Conclusion: The presented phantom material is visible using standard MR sequences, and thus, supports the use of standard, phantom-independent MR measurements for MR-AC in PET/MRI phantom studies.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1186/s40658-021-00364-9

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Cross-calibration of PET/MRI scanners is generally hampered by water-filled phantoms, which can produce substantial artifacts in the MR-AC sequence with corresponding SUV errors.

    “It should also be noted that cross-calibration of PET/MRI scanners is generally hampered by water-filled phantoms, which are commonly used for SUV calibration in PET/CT and can produce substantial artifacts in the MR-AC sequence with corresponding SUV errors [63,64].”

    From Influences on PET Quantification and Interpretation (2022) · cites this paper as [64]

    The Background (page unavailable) says standard MR attenuation correction can miss phantom materials and thereby underestimate measured activity; the Discussion identifies signal variations and tissue-swap-like artifacts in Dixon images. But the Discussion also says the observed tissue-swap-like artifacts did not enter the phantom-specific MR-AC map. The paper does not establish that artifacts from water-filled phantoms cause substantial SUV errors during cross-calibration. (LLM, conf=0.82)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 38

  1. [1]Bailey DL Pichler BJ Gückel B Combined PET/MRI: global warming—summary report of the 6th International Workshop on PET/MRI, March 27–29, 2017, Tübingen, Germany Mol Imaging Biol 2018 20 1 4 20 10.1007/s11307-017-1123-5 28971346 PMC5775351 · DOI
  2. [2]Rausch I, Quick HHH, Cal-Gonzalez J, Sattler B, Boellaard R, Beyer T. Technical and instrumentational foundations of PET/MRI. Eur J Radiol. 2017;94. 10.1016/j.ejrad.2017.04.004. 28431784 · DOI
  3. [3]Cabello J Ziegler SI Advances in PET/MR instrumentation and image reconstruction Br J Radiol 2018 91 1081 20160363 10.1259/bjr.20160363 27376170 PMC5966194 · DOI
  4. [4]Blumhagen JO Ladebeck R Fenchel M Scheffler K MR-based field-of-view extension in MR/PET: B 0 homogenization using gradient enhancement (HUGE) Magn Reson Med 2013 70 4 1047 1057 10.1002/mrm.24555 23203976 · DOI
  5. [5]Wagenknecht G Kaiser H-JJ Mottaghy FM Herzog H MRI for attenuation correction in PET: methods and challenges Magn Reson Mater Physics Biol Med 2013 26 1 99 113 10.1007/s10334-012-0353-4 PMC3572388 23179594 · DOI
  6. [6]Teuho J, Torrado-Carvajal A, Herzog H, et al. Magnetic resonance-based attenuation correction and scatter correction in neurological positron emission tomography/magnetic resonance imaging—current status with emerging applications. Front Phys. 2020;7. 10.3389/fphy.2019.00243. · DOI
  7. [7]Ladefoged CN Law I Anazodo U A multi-centre evaluation of eleven clinically feasible brain PET/MRI attenuation correction techniques using a large cohort of patients Neuroimage. 2017 147 346 359 10.1016/j.neuroimage.2016.12.010 27988322 PMC6818242 · DOI
  8. [8]Martinez-Moller A Souvatzoglou M Delso G Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data J Nucl Med 2009 50 4 520 526 10.2967/jnumed.108.054726 19289430 · DOI
  9. [10]Izquierdo-Garcia D Catana C MR imaging–guided attenuation correction of PET data in PET/MR imaging PET Clin 2016 11 2 129 149 10.1016/j.cpet.2015.10.002 26952727 PMC5798860 · DOI
  10. [11]Seith F Schmidt H Gatidis S SUV-quantification of physiological lung tissue in an integrated PET/MR-system: impact of lung density and bone tissue PLoS One 2017 12 5 e0177856 10.1371/journal.pone.0177856 28562622 PMC5451041 · DOI
  11. [12]Keereman V, Mollet P, Fierens Y, Espana S, Vandenberghe S. Design of a realistic PET-CT-MRI phantom. In: 2011 IEEE Nuclear Science Symposium Conference Record. New York: IEEE; 2011, pp. 3173–7, 10.1109/NSSMIC.2011.6153651. · DOI
  12. [13]Harries J Jochimsen TH Scholz T A realistic phantom of the human head for PET-MRI EJNMMI Phys 2020 7 1 52 10.1186/s40658-020-00320-z 32757099 PMC7406590 · DOI
  13. [14]Steinmann A Stafford RJ Sawakuchi G Developing and characterizing MR/CT-visible materials used in QA phantoms for MRgRT systems Med Phys 2018 45 2 773 782 10.1002/mp.12700 29178486 PMC5807190 · DOI
  14. [15]Valladares A, Beyer T, Rausch I. Physical imaging phantoms for simulation of tumor heterogeneity in PET, CT, and MRI: an overview of existing designs. Med Phys. 2020. 10.1002/mp.14045. PMC7216968 31981214 · DOI
  15. [16]Keller SH Jakoby B Svalling S Kjaer A Højgaard L Klausen TL Cross-calibration of the Siemens mMR: easily acquired accurate PET phantom measurements, long-term stability and reproducibility EJNMMI Phys 2016 3 1 11 10.1186/s40658-016-0146-3 27387738 PMC4936986 · DOI
  16. [17]Mantlik F Hofmann M Werner MK The effect of patient positioning aids on PET quantification in PET/MR imaging Eur J Nucl Med Mol Imaging 2011 38 5 920 929 10.1007/s00259-010-1721-9 21308373 · DOI
  17. [18]Büther F Vrachimis A Becker A Stegger L Impact of MR-safe headphones on PET attenuation in combined PET/MRI scans EJNMMI Res 2016 6 1 20 10.1186/s13550-016-0178-7 26936769 PMC4775719 · DOI
  18. [19]Ferguson A McConathy J Su Y Hewing D Laforest R Attenuation effects of MR headphones during brain PET/MR studies J Nucl Med Technol 2014 42 2 93 100 10.2967/jnmt.113.131995 24556458 · DOI
  19. [20]Tellmann L Quick HH Bockisch A Herzog H Beyer T The effect of MR surface coils on PET quantification in whole-body PET/MR: results from a pseudo-PET/MR phantom study Med Phys 2011 38 5 2795 2805 10.1118/1.3582699 21776816 · DOI
  20. [21]Boellaard R Rausch I Beyer T Quality control for quantitative multicenter whole-body PET/MR studies: a NEMA image quality phantom study with three current PET/MR systems Med Phys 2015 42 10 5961 10.1118/1.4930962 26429271 · DOI · Paper page
  21. [22]Paulus DH Quick HH Hybrid positron emission tomography/magnetic resonance imaging Investig Radiol 2016 51 10 624 634 10.1097/RLI.0000000000000289 27175550 · DOI
  22. [23]Paulus DH Braun H Aklan B Quick HH Simultaneous PET/MR imaging: MR-based attenuation correction of local radiofrequency surface coils Med Phys 2012 39 7 4306 4315 10.1118/1.4729716 22830764 · DOI
  23. [24]Heußer T Rank CM Berker Y Freitag MT Kachelrieß M MLAA-based attenuation correction of flexible hardware components in hybrid PET/MR imaging EJNMMI Phys 2017 4 1 12 10.1186/s40658-017-0177-4 28251575 PMC5332322 · DOI
  24. [25]Weiger M Pruessmann KP Short-T2 MRI: principles and recent advances Prog Nucl Magn Reson Spectrosc 2019 114-115 237 270 10.1016/j.pnmrs.2019.07.001 31779882 · DOI
  25. [26]Rai R Manton D Jameson MG 3D printed phantoms mimicking cortical bone for the assessment of ultrashort echo time magnetic resonance imaging Med Phys 2018 45 2 758 766 10.1002/mp.12727 29237232 · DOI
  26. [27]Mitsouras D Lee TC Liacouras P Three-dimensional printing of MRI-visible phantoms and MR image-guided therapy simulation Magn Reson Med 2017 77 2 613 622 10.1002/mrm.26136 26864335 PMC5108690 · DOI
  27. [28]Rai R Wang YF Manton D Dong B Deshpande S Liney GP Development of multi-purpose 3D printed phantoms for MRI Phys Med Biol 2019 64 7 075010 10.1088/1361-6560/ab0b49 30818301 · DOI
  28. [31]Price RR Axel L Morgan T Quality assurance methods and phantoms for magnetic resonance imaging: report of AAPM nuclear magnetic resonance Task Group No. 1 Med Phys 1990 17 2 287 295 10.1118/1.596566 2333055 · DOI
  29. [33]Carney JPJ Townsend DW Rappoport V Bendriem B Method for transforming CT images for attenuation correction in PET/CT imaging Med Phys 2006 33 4 976 983 10.1118/1.2174132 16696474 · DOI
  30. [34]de Graaf RA In vivo NMR spectroscopy: principles and techniques 2007 2 Chichester Wiley
  31. [35]Roemer PB Edelstein WA Hayes CE Souza SP Mueller OM The NMR phased array Magn Reson Med 1990 16 2 192 225 10.1002/mrm.1910160203 2266841 · DOI
  32. [36]Van den Berg CAT van den Bergen B Van de Kamer JB SimultaneousB1+ homogenization and specific absorption rate hotspot suppression using a magnetic resonance phased array transmit coil Magn Reson Med 2007 57 3 577 586 10.1002/mrm.21149 17326185 · DOI
  33. [37]Garwood M Ke Y Symmetric pulses to induce arbitrary flip angles with compensation for rf inhomogeneity and resonance offsets J Magn Reson 1991 94 3 511 525 10.1016/0022-2364(91)90137-I · DOI
  34. [38]Bley TA Wieben O François CJ Brittain JH Reeder SB Fat and water magnetic resonance imaging J Magn Reson Imaging 2010 31 1 4 18 10.1002/jmri.21895 20027567 · DOI
  35. [39]Rausch I Rust P DiFranco MD Reproducibility of MRI Dixon-based attenuation correction in combined PET/MR with applications for lean body mass estimation J Nucl Med 2016 57 7 1096 1101 10.2967/jnumed.115.168294 26917707 · DOI
  36. [40]Ladefoged CN Hansen AE Keller SH Impact of incorrect tissue classification in Dixon-based MR-AC: fat-water tissue inversion EJNMMI Phys 2014 1 1 101 10.1186/s40658-014-0101-0 26501459 PMC4545462 · DOI
  37. [41]Reeder SB Pineda AR Wen Z Iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL): application with fast spin-echo imaging Magn Reson Med 2005 54 3 636 644 10.1002/mrm.20624 16092103 · DOI
  38. [42]Rausch I Rischka L Ladefoged CNCN PET/MRI for oncologic brain imaging: a comparison of standard MR-based attenuation corrections with a model-based approach for the Siemens mMR PET/MR system J Nucl Med 2017 58 9 1519 1525 10.2967/jnumed.116.186148 28254872 · DOI

Cited by

Sorted by citations

Authors
Rausch, I., Valladares, A., Sundar, L. K. S., Beyer, T., Hacker, M., Meyerspeer, M., & Unger, E.
Affiliations
Medical University of Vienna (AT)
Corresponding author
Rausch, Ivo
Venue
EJNMMI Physics
Published
2021-02-18
Volume
8
Issue
1
Pages
18-18
ISSN
2197-7364 (ISSN-L)
OpenAlex
W3130762018
Cited by
11
References
2
Citations by year
2025: 3 · 2023: 2 · 2022: 3 · 2021: 3
FWCI
1.06
Type
journal article
Language
en
Open access
Locations
  • EJNMMI Physics (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Funders
  • H2020 Marie Skłodowska-Curie Actions (No. 764458)
  • Austrian Science Fund (I3451-N32)
  • Österreichische Forschungsförderungsgesellschaft (858060)
Topics
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
  • Physical Sciences › Engineering › Biomedical Engineering › Advanced X-ray and CT Imaging
Keywords
PET/MRI, attenuation correction, Dixon MRI, 3D-printed phantom, 18F-FDG, CT attenuation correction

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review