Factors Affecting the Performance of 5 Cerebral Oximeters During Hypoxia in Healthy Volunteers
The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Cerebral oximetry is a noninvasive optical technology that measures frontal cortex blood hemoglobin-oxygen saturation. Commercially available cerebral oximeters have not been evaluated independently. Unlike pulse oximeters, there are currently no Food and Drug Administration standards for performance or accuracy. We tested the hypothesis that cerebral oximeters accurately measure a fixed ratio of the oxygen saturation in cerebral mixed venous and arterial blood. METHODS: We evaluated the performance of 5 commercially available cerebral oximeters: the EQUANOX® 7600 in 3- and 4-wavelength versions (Nonin Medical, Plymouth, MN), FORE-SIGHT® (Casmed, Branford, CT), INVOS® 5100C (Covidien, Boulder, CO), and the NIRO-200NX® (Hamamatsu Photonics, Hamamatsu City, Japan) during stable isocapnic hypoxia in volunteers. Twenty-three healthy adults (14 men, 9 women) had sensors placed on each side of the forehead. The subject's inspired oxygen (FIO2) was then changed to produce 6 steady-state arterial oxygen saturation (SaO2) levels between 100% and 70%, while end-tidal CO2 was maintained constant. At each plateau, simultaneous blood samples from the jugular bulb and radial artery were analyzed with a hemoximeter (OSM-3, Radiometer Medical A/S, Copenhagen, Denmark). Each cerebral oximeter's bias was calculated as the difference between the instrument's reading (cerebral saturation, ScO2) with the weighted saturation of venous and arterial blood (Sa/vO2), as specified by each manufacturer (INVOS: 25% arterial/75% venous; FORE-SIGHT, EQUANOX, and NIRO: 30% arterial/70% venous). RESULTS: Five hundred forty-two comparisons between paired blood samples and oximeter readings were analyzed. The pooled root mean square error was 8.06%, a value higher than for pulse oximeters, which is ±3% by Food and Drug Administration standards. The mean % bias ± SD (precision) and root mean square errors were: FORE-SIGHT 1.76 ± 3.92 and 4.28; INVOS 0.05 ± 9.72 and 9.69; NIRO-200NX -1.13 ± 9.64 and 9.68; EQUANOX-3 λ 2.48 ± 8.12 and 8.47; EQUANOX-4 λ 2.84 ± 6.27 and 6.86. The FORE-SIGHT, NIRO-200NX, and EQUANOX-3 λ had significantly more positive bias at lower SaO2. The amount of bias during hypoxia was reduced when the bias was calculated on the basis of difference between oximeter reading and the arterial and mixed venous saturation difference rather than the weighted average of blood saturation, indicating that differences in the ratio between arterial and venous blood volumes account for some of the positive bias at low saturation. Dark skin pigment tended to produce more negative bias in all instruments but bias was significantly larger than zero only for the FORE-SIGHT oximeter. Bias was significantly more negative in women for INVOS and EQUANOX devices but not for the FORE-SIGHT device. CONCLUSIONS: While responsive to desaturation, cerebral oximeters exhibited large variation in reading errors between subjects, with mean bias possibly related to variations in the ratio of arterial and venous blood in the sampling area of the brain. This ratio is probably not fixed, as assumed by the manufacturers, but dynamically changes with hypoxia. Better understanding these factors could improve the performance of cerebral oximeters and help establish saturation or blood flow thresholds for brain well-being.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Jöbsis, F. F. (1977) · Science · doi:10.1126/science.929199 · Cited by 3,899
Agreement Between Methods of Measurement with Multiple Observations Per Individual
Bland, J. M., & Altman, D. G. (2007) · Journal of Biopharmaceutical Statistics · doi:10.1080/10543400701329422 · Cited by 2,065
Effects of skin pigmentation on pulse oximeter accuracy at low saturation
Bickler, P. E., Feiner, J. R., & Severinghaus, J. W. (2005) · Anesthesiology · doi:10.1097/00000542-200504000-00004 · Cited by 427
Self-citationFeiner, J. R., Severinghaus, J. W., & Bickler, P. E. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000285988.35174.d9 · Cited by 400
Self-citation
Cited by
Holmer, A., Marotz, J., Wahl, P., Dau, M., & Kämmerer, P. W. (2018) · Biomedical Engineering / Biomedizinische Technik · doi:10.1515/bmt-2017-0155 · Cited by 195
Impact of skin tone on photoacoustic oximetry and tools to minimize bias
Mantri, Y., & Jokerst, J. V. (2022) · Biomedical Optics Express · doi:10.1364/boe.450224 · Cited by 75
Weber, F., & Scoones, G. P. (2019) · Pediatric Anesthesia · doi:10.1111/pan.13726 · Cited by 53
Metadata
- Authors
- Bickler, P. E., Feiner, J. R., & Rollins, M. D.
- Venue
- Anesthesia & Analgesia
- Published
- 2013-10
- Volume
- 117
- Issue
- 4
- Pages
- 813-823
- ISSN
- 0003-2999 (ISSN-L)
- OpenAlex
- W2322330265
- Cited by
- 222
- References
- 4
- FWCI
- 9.74
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Factors Affecting the Performance of 5 Cerebral Oximeters During Hypoxia in Healthy Volunteers. https://app.cochat.ai/papers/10.1213/ane.0b013e318297d763
- The paper
- Bickler, P. E., Feiner, J. R., & Rollins, M. D. (2013). Factors Affecting the Performance of 5 Cerebral Oximeters During Hypoxia in Healthy Volunteers. Anesthesia & Analgesia. https://doi.org/10.1213/ane.0b013e318297d763
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.